TOPAZ EZ MICRODEBRIDER COBLATION WAND WITH INTEGRATED FINGER SWITCH
K140521 · ArthroCare Corporation · GEI · Jun 10, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140521
Device Name
TOPAZ EZ MICRODEBRIDER COBLATION WAND WITH INTEGRATED FINGER SWITCH
Applicant
ArthroCare Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 10, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch is indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: Procedure: Fasciotomy, Body Structure: Foot; Procedure: Synovectomy, Body Structure: Foot; Procedure: Tendonotomy, Body Structure: Knee, Wrist, Elbow, Ankle, Shoulder, Foot; Procedure: Rotator Cuff Tendonotomy, Body Structure: Shoulder; Procedure: Capsulotomy, Body Structure: Foot.
Device Story
The TOPAZ® EZ Microdebrider Coblation® Wand is a bipolar, high-frequency electrosurgical device used in arthroscopic and orthopedic procedures. It consists of a tungsten electrode tip, alumina ceramic spacer, stainless steel shaft, saline irrigation tubing, and a handle with an integrated finger switch. The device connects to an ArthroCare Quantum Controller. During operation, the device delivers radiofrequency energy to saline to create a plasma field, enabling precise soft tissue ablation, resection, and hemostasis. It is operated by a physician in a clinical or surgical setting. The integrated finger switch allows the surgeon to activate the wand directly. By providing controlled tissue removal and coagulation, the device assists in procedures such as fasciotomy, synovectomy, and tendonotomy, potentially reducing surgical time and improving procedural outcomes.
Clinical Evidence
No clinical data are included in this submission. Substantial equivalence is supported by non-clinical performance testing, including visual and dimensional inspection, Hi-Pot testing, electrical impedance, saline flow rate, ablation set points, finger switch actuation, shaft stiffness, axial compression force, biocompatibility (cytotoxicity, sensitization, maximization), and electrical safety testing per IEC 60601-2-2:2009.
Indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures of the foot, knee, wrist, elbow, and shoulder.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
TOPAZ® Microdebrider Coblation® Wand with Integrated Finger Switch (K080282)
Submission Summary (Full Text)
{0}------------------------------------------------
JUN | 0 2014
# K140521
# 510(k) Summary
# ArthroCare® Corporation TOPAZ® EZ Microdebrider Coblation® Wand With Integrated Finger Switch
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
# General Information
| Submitter Name: | ArthroCare Corporation |
|-----------------|-------------------------------------------------------------------------------------------------------|
| Address | 7000 West William Cannon Drive<br>Austin, TX 78735 |
| Contact Person: | Mitchell A. Dhority<br>Vice President, Regulatory Affairs<br>Phone: 512-358-5995<br>Fax: 512-895-1489 |
| Date Prepared: | February 27, 2014 |
## Device Name
| Proprietary Name: | TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch |
|--------------------|-----------------------------------------------------------------------|
| Regulation Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulatory Class: | I I |
| Product Code: | GEI |
| Regulation Number: | 21 CFR 878.4400 |
#### Predicate Device
TOPAZ® Microdebrider Coblation® Wand with Integrated Finger Switch included in: ArthroCare Topaz Wand, K080282 (cleared February 15, 2008)
#### Description
The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch (Topaz EZ IFS) is a bipolar, sterile, high frequency electrosurgical device, which consists of a distal electrode tip composed of tungsten, an alumina ceramic spacer, stainless steel shaft, saline irrigation tubing and a molded handle with an integrated single finger switch used to activate the Wand. The handle connects proximally to the radiofrequency Controller via an electrical cable. The Topaz EZ IFS is only compatible with the ArthroCare Quantum Controllers.
{1}------------------------------------------------
## .. ......... Intended Use/Indications for Use
The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch is indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:
| Procedure | Body Structure as Described Below |
|--------------------------|-------------------------------------------|
| Fasciotomy | Foot |
| Synovectomy | Foot |
| Tendonotomy | Knee, Wrist, Elbow, Ankle, Shoulder, Foot |
| Rotator Cuff Tendonotomy | Shoulder |
| Capsulotomy | Foot |
## Summary of Technological Characteristics
This Special 510(k) proposes modifications to the materials, dimensional specifications, and performance specifications of the ArthroCare Topaz Wand cleared under K080282. The indications for use, fundamental scientific technology, principle of operation, and sterilization methodology remain the same as in the previously cleared 510(k).
The following table summarizes the technological differences between the subject and predicate devices and describes the rationale for the change.
| Parameter | Predicate Device:<br>Topaz Wand<br>(K080282) | Subject Device:<br>TOPAZ EZ IFS<br>Wand | Rationale for Change |
|---------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Spacer Material | Thermal-Set Silicone<br>Elastomer (Silicone) | Alumina | Facilitates consistency in<br>manufacturability |
| Shaft Length | 2.91 inches (74mm) | 1.57 inches (40mm) | Length shortened based on user<br>feedback (decreases the<br>working distance between the<br>user and the surgical site) |
| External Shaft<br>Insulation<br>Material | Nylon | Polyester | Material now used to insulate<br>the shaft as opposed to also<br>serving as the saline sheath |
| Saline Delivery<br>(irrigant<br>contacting<br>material) | Nylon and stainless<br>steel | Pebax and stainless<br>steel | Provides saline delivery close to<br>the tip of the wand and greater<br>visibility for the surgeon |
| Shaft OD | $0.032 \pm .005"$<br>(.812± .13mm) | $0.042 \pm .005"$ (1.067<br>±.13mm) | Increased stiffness of the shaft |
| Active Electrode<br>Insulation<br>Material | Polyimide | Pebax | Manufacturability |
| Shaft Y -<br>Connector and<br>Bushing | Separate components<br>comprised of PVC and<br>polycarbonate,<br>respectively | Single integrated<br>component<br>comprised of<br>polycarbonate | Cost savings and process<br>improvement. Adhesive was<br>increased to prevent the shaft<br>from sliding within Y-<br>connector. |
| Active/Return<br>Electrode<br>Insulator | N/A | Clear Pebax | Insulates electrodes; centers<br>electrode components during<br>manufacturing |
| Internal | N/A | Nylon | Holds cured adhesive to |
{2}------------------------------------------------
| Parameter | Predicate Device:<br>Topaz Wand<br>(K080282) | Subject Device:<br>TOPAZ EZ IFS<br>Wand | Rationale for Change |
|-------------------------|----------------------------------------------|-----------------------------------------|----------------------------------------------------------|
| Reinforcement<br>Tubing | | | facilitate bonding of electrodes<br>during manufacturing |
| Labeling | Topaz IFS | Topaz EZ IFS | Proprietary name change.<br>Instructions strengthened. |
#### Summary of Non-Clinical Performance Data
Non-clinical performance testing verifies that the Topaz EZ IFS performs as intended and that no additional risks or hazards have been identified as a result of the changes described in this Special 510(k). As such, results from this testing support the determination that the Topaz EZ IFS is substantially equivalent to the predicate device. The following table lists the non-clinical testing performed and the results obtained in support of substantial equivalence.
| Testing Type | Test Description | Result Supporting Substantial<br>Equivalence |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Verification Testing | • Visual Inspection<br>• Dimensional Inspection<br>• Hi-Pot Testing<br>• Electrical Impedance<br>• Saline Flow Rate<br>• Controller Ablation Set Points<br>• Finger Switch Actuation Testing<br>• Ablation Testing<br>• Shaft Stiffness<br>• Axial Compression Force Testing | Both the Topaz EZ IFS and<br>predicate device have substantially<br>equivalent testing specifications<br>and both performed within<br>acceptance criteria. These results<br>support that the Topaz EZ IFS and<br>predicate device are substantially<br>equivalent. |
| Biocompatibility<br>Testing | • Cytotoxicity<br>• Sensitization<br>• Maximization | The Topaz EZ IFS and the<br>predicate device are biocompatible.<br>These results support that the<br>Topaz EZ IFS and predicate device<br>are substantially equivalent. |
| Electrical Safety<br>Testing | Testing in accordance with IEC<br>60601-2-2:2009. | The Topaz EZ IFS and the<br>predicate met all acceptance<br>criteria in accordance with IEC<br>60601-2-2: 2009. These results<br>support that the Topaz EZ IFS and<br>predicate device are substantially<br>equivalent. |
## Clinical Data
No clinical data are included in this submission
#### Summary
The Topaz EZ IFS is substantially equivalent to the predicate Topaz Wand (K080282). Both the subject and predicate devices have the same indications for use and utilize the same fundamental scientific technology. All product performance testing demonstrates that the Topaz EZ IFS performs as intended and has acceptable mechanical properties when used in accordance with its labeling. The minor differences between the Topaz EZ IFS and the predicate device do not raise any new questions of safety or effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2014
ArthroCare Coporation Mr. Mitchell A. Dhority Vice President, Regulatory Affairs 7000 West William Cannon Drive Austin, Texas 78735
Re: K140521 Trade/Device Name: TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: May 13, 2014 Received: May 14, 2014
Dear Mr. Dhority:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
Image /page/3/Picture/10 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right.
{4}------------------------------------------------
Page 2 - Mr. Mitchell A. Dhority
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below,
510(k) Number (if known)
K140521
Device Name
TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch
T
Indications for Use (Describe)
The TOPAZ® EZ Microdebrider Coblation® Wand with Integrated Finger Switch is indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic procedures:
Body Structure Procedure Fasciotomy Foot Synovectomy Foot Knee, Wrist, Elbow, Ankle, Shoulder, Foot Tendonotomy Rotator Cuff Tendonotomy Shoulder Capsulotomy Foot
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
CONSTITUTION CONTROLLER COLLECTION FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR MONDENT OF CONSULTION
Concurrence of Center for Davices and Radiological Health (CDRH) (Signalure)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### ·DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Offica of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.