K140504 · Vascular Pathways, Inc. · FOZ · Jul 17, 2014 · General Hospital
Device Facts
Record ID
K140504
Device Name
ACCUCATH BC INTRAVASCULAR CATHETER SYSTEM
Applicant
Vascular Pathways, Inc.
Product Code
FOZ · General Hospital
Decision Date
Jul 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The AccuCath BC Intravascular Catheter System is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath for use with low pressure power injectors having a maximum pressure setting of 300 psi and maximum flow rate of 6mL second.
Device Story
Single-use, sterile intravascular catheter system (18, 20, 22 Gauge) for vascular access. Features include echogenic needle for ultrasound visualization, integrated blood control septum in hub to prevent blood flow upon needle retraction, and safety mechanism to shield needle post-placement. Device includes protective cover, luer lock fitting, flashback chamber, guide wire, slider, spring, and release button. Used by clinicians in clinical settings for fluid administration, blood sampling, or pressure monitoring. Blood control septum automatically activates upon needle removal, restricting flow until secure luer connection is established. Power injection capability supports contrast media delivery. Benefits include reduced blood exposure for clinicians and improved needle visualization during insertion.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1), catheter performance (ISO 10555-1/5), luer fitting integrity (ISO 594-1/2), radiopacity (ASTM F640), packaging validation (ISO 11607, ASTM F2096, ASTM F88, ASTM F1980), and sharps injury prevention (Guidance No. 934).
Technological Characteristics
Over-the-needle peripheral catheter; echogenic needle; integrated blood control septum; luer lock fitting. Materials evaluated per ISO 10993-1. Sterilization via ethylene oxide (AAMI TIR16). Standards: ISO 10555-1/5, ISO 594-1/2, ASTM F640, ISO 11607-1/2, ASTM D4169, ASTM F2096, ASTM F88, ASTM F1980.
Indications for Use
Indicated for patients requiring vascular access for blood sampling, blood pressure monitoring, or intravenous fluid administration. Suitable for use with low-pressure power injectors (max 300 psi, 6 mL/sec).
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Becton Dickinson BD INSYTE AUTOGUARD BC device (K110443)
Ethicon Endo-Surgery PROTECTIV* ACUVANCE IV Catheter System (K012128)
Vascular Pathways, Inc., Rapid Intravascular Catheter Start System (K112347)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 8. 510(k) Summary
In accordance with 21 CFR 807.87(h), a 510(k) Summary (21 CFR 807.92) follows:
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
510(k) Number:
# Applicant Information:
| Date Prepared: | February 22, 2014 |
|----------------|-------------------------------------------|
| Name: | Vascular Pathways, Inc. |
| Address: | 1847 Trade Center Way<br>Naples, FL 34109 |
| Phone: | 239-254-0391 |
Contact Person: Scott Pease, VP of RA and QA
1847 Trade Center Way Naples, Fl 34109 Cell - 678-689-8010 e-mail: spease@vascularpathways.com
### Device Information:
| Device Trade Name: | AccuCath BC Intravascular Catheter System |
|---------------------------|-------------------------------------------|
| Common Name: | Intravascular Catheter |
| Classification Name(s): | Intravascular Catheter. |
| Product Code/ Regulation: | FOZ / 21 CFR 880.5200 |
| Classification: | Class II |
### Predicate Devices:
Becton Dickinson BD INSYTE AUTOGUARD BC device (K110443), Ethicon Endo-Surgery PROTECTIV* ACUVANCE IV Catheter System (K012128) Vascular Pathways, Inc., Rapid Intravascular Catheter Start System (K112347)
### Device Description:
The AccuCath BC Intravascular Catheter System has a usable length catheter of 2.25 inches in 18, 20 and 22 Gauge sizes. The devices are single use, sterile intravascular catheters designed to be inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This longer AccuCath BC catheter is identical to the AccuCath with the addition of needle echogenicity and a septum with plunger integrated into the catheter hub. The AccuCath BC catheter's hub has a built in blood control septum. The blood control feature is a single-use septum that automatically activates to stop the blood flow in the catheter hub when the needle is removed from the catheter during initial insertion by the clinician. Blood flow from the catheter hub will be restricted immediately after the needle retraction until a secure luer connection is made. The flow path is permanently opened once a secure luer connection has been made. Similar to the original AccuCath system, the device is provided with a safety mechanism which allows the needle to be shielded following placement of the catheter. All devices have the basic structure of a protective cover, a catheter with a bier lock fitting, an echogenic needle connected to a flashback chamber, a safety container, a guide wire within the lumen of the needle which is connected to a slider and a spring and release button.
{1}------------------------------------------------
# Premarket Notification 510(k) Summary - Continued
# Indications for Use:
The AccuCath BC Intravascular Catheter System is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath BC is suitable for use with low pressure power injectors having a maximum pressure setting of 300 psi and maximum flow rate of 6mL/second.
# Comparison to Predicate Devices:
The AccuCath BC Intravascular Catheter System has the same intended use, mechanism of action, and basic design attributes as the BD INSYTE AUTOGUARD BC device (K110443), the Ethicon PROTECTIV* ACUVANCE IV Catheter System (K012128) and the Vascular Pathways, Inc., Rapid Intravascular Catheter Start System (K112347) (the predicate devices). It has the same indications for use as the BD INSYTE AUTOGUARD BC device – adding power injection to the previously cleared Rapid Intravascular Catheter Start System indication consistent with the Ethicon PROTECTIV* ACUV ANCE IV Catheter System (K012128). Performance data demonstrates substantially equivalence in terms of safety and effectiveness to the predicate devices.
# Performance Testing:
Biocompatibility and functional bench testing performed by Vascular Pathways, Inc. demonstrate the substantial equivalence, in terms of the safety and effectiveness, of the AccuCath BC Intravascular Catheter System to the predicate devices cited. In vitro bench testing included an assessment of all design input requirements and confirmation that the output of the design process met all design input requirements as follows:
| 1. | ISO 10993-1:2009, Biological Evaluation of Medical Devices- Part I: Evaluation and Testing |
|-----|------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | ISO 10555-1: 1995 - Sterile, Single Use intravascular Catheters - Part 1: General Requirements |
| 3. | ISO 10555-5: 2012 - Sterile, Single Use Intravascular Catheters - Part 5: Over-Needle Peripheral Catheters |
| 4. | ISO 594-1:1986 Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain Other Equipment - Part 1: General Requirements |
| 5. | ISO 594-2:1998 Conical Fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock Fittings |
| 6. | ASTM F640:2012 - Std Test Methods for Determining Radiopacity for Medical Use |
| 7. | ISO 11607-1:2006 - Packaging for terminally sterilized medical devices - Part 1: Requirements for materials... |
| 8. | ISO 11607-2:2006 - Packaging for terminally sterilized medical devices - Part 2: Validation requirements.. |
| 9. | ASTM D4169-09, Std Practice for Performance Testing of Shipping Containers and Systems. |
| 10. | ASTM F2096-11, Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test). |
| 11. | ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials. |
| 12. | ASTM F1980-07 (2011) Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. |
| 13. | Guidance Document No. 934 - Medical Devices with sharps injury prevention features (August 9, 2005) |
| 14. | AAMI TIR16:2000 - Process Development and Performance Qualification for Ethylene Oxide Sterilization -<br>Microbiological aspects |
{2}------------------------------------------------
# Summary:
Based upon the device description and test data provided in this submission the AccuCath BC Intravascular Catheter System is substantially equivalent to the predicate devices cited.
.
ﻴ
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MID 20993-0002
July 17, 2014
Vascular Pathways, Inc. Mr. Scott Pease Vice President of Regulatory Affairs and Quality Assurance 1847 Trade Center Way Naples, FL 34109
Re: K140504
Trade/Device Name: AccuCath BC Intravascular Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: June 10, 2014, 2014 Received: June 16, 2014, 2014
Dear Mr. Pease:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Mr. Pease
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
. ആപ്പ Sheth, M. Tojashri Purohit-Sheth, M.D.
Tolashri Sheth, M.D. Cluical Deputy Director Tejashri Purofitit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K140504
### Device Name
AccuCath BC Intravascular Catheter System
#### Indications for Use (Describe)
The AccuCath BC Intravascular Catheter System is inserted into a patient's vascular system to sample blood, monitor blood pressure, or administer fluids intravenously. This device may be used with consideration given to adequacy of vascular anatomy, appropriateness of the solution being infused, and duration of therapy. The AccuCath for use with low pressure power injectors having a maximum pressure setting of 300 psi and maximum flow rate of 6mL second.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Digitally signed by Richard C. Chapman -S Date: 2014.07.16 16:29:43 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.