K140461 · Dr. Mach GmbH & Co. KG · FSY · Apr 3, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140461
Device Name
MACH LED 2MC
Applicant
Dr. Mach GmbH & Co. KG
Product Code
FSY · General, Plastic Surgery
Decision Date
Apr 3, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The surgical light Mach LED 2MC is intended to illuminate the on the patient's body with a high intensity, shadow-free and "cold" light. The Mach LED 2MC lighting system is designed for illuminating an examination area and surgical field at the hospital and doctor's practice.
Device Story
Surgical lamp providing high-intensity, shadow-free, cold illumination for surgical and examination sites. Device comprises lamp housing, LED modules, optical/electrical/mechanical components, and sterilizable handle. Input: electrical power; Output: visible light. Operation: 84 LEDs (warm white, cold white, green, red) merged via computer-calculated optical system with light guides and facetted lenses. System mounted on ceiling suspension with horizontal/spring arms allowing 360-degree rotation and vertical movement. Controlled via keypad on lamp head or wall. Features include light field merging, intensity control, color temperature adjustment, and integrated laser pointer. Used in hospitals and clinics by medical staff. Accessories include camera modules, monitor yokes, instrument trays, and battery backup. Benefits include optimized visualization of surgical field.
Clinical Evidence
No clinical data is required for this device classification submission. Performance verified via bench testing against Medical Electrical Equipment standards and EN 61000 group requirements.
Technological Characteristics
LED-based surgical lamp. 84 LEDs (warm white, cold white, green, red). Computer-calculated optical system with light guides and facetted lenses. Ceiling-mounted suspension system. Keypad control. Accessories include camera modules, monitor yokes, and battery backup. Complies with Medical Electrical Equipment standards and EN 61000 series.
Indications for Use
Indicated for illumination of surgical fields and examination areas in hospitals and clinical practices.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2014
Dr. Mach GmbH & Company KG % Gudrun Busch, Ph.D. Dr. Busch Regulatory Strategy GmbH Fehmarnstrasse 31 Norderstedt, Schleswig-Holstein 22846 Germany
Re: K140461
Trade/Device Name: Mach LED 2MC Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: Class II Product Code: FSY Dated: February 27, 2014 Received: March 5, 2014
Dear Dr. Busch:
This letter corrects our substantially equivalent letter of April 3, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Dr. Gudrun Busch
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
## 510(k) Number (if known) K 1 4046 I
### Device Name Mach LED 2MC
### Indications for Use (Describe)
The surgical light Mach LED 2MC is intended to illuminate the on the patient's body with a high intensity, shadow-free and "cold" light.
Type of Use (Select one or both, as applicable)
[図] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
. :
{3}------------------------------------------------
K140461
Image /page/3/Picture/1 description: The image shows a logo for "Dr. Mach Medizinische Leuchten + Technik". The logo is in black and white and features a stylized font. There is a drawing of a medical examination light.
APR 0 9 2014
# 510(k) Summary for Mach LED 2MC
as required by section 807.92(c)
| Submitter | Dr. Mach GmbH & Co.KG |
|---------------------|----------------------------------------------------------------------------|
| | Flossmannstrasse 28<br>85560 Ebersberg<br>Germany |
| Contact Person | Rainer Adams<br>Technical Director |
| | Phone: +49 8092 209330<br>Fax: +49 8092 209350 |
| Preparation Date | February 14, 2014 |
| Trade Name | Mach LED 2MC |
| Common Name | Surgical Lamp |
| Classification Name | Surgical Lamp<br>Regulation 21 CFR 878.4580, Class II<br>Product Code: FSY |
| Predicate Device | Mach LED MC<br>Surgical lamp<br>510(k) No. K093010, April 1, 2010 |
Dr. Mach GmbH & Co KG Summary Mach LED 2MC
Page 1 of 3
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Dr. Mach Medzinische Leuchten + Technik. The logo features the name "Dr. Mach" in a bold, stylized font, with the words "Medzinische Leuchten" and "+ Technik" underneath in a smaller font. To the left of the text is a drawing of a surgical light.
# Device Description
The surgical light Mach LED 2MC is intended to illuminate the operation site on the patient's body with a high intensity, shadow free and "cold" light.
The Mach LED 2MC consists of lamp housing, LED modules, optical, electrical and mechanical components, one sterilizable handle sleeve as well as the cables.
One LED-module consists of 4 different-colored LEDs: warm white, cold white, green and red. The four different colors are merged inside the lamp head by a computercalculated optical system with light guide and facetted lenses.
The light system can be added to the ceiling mounted suspension system supporting the horizontal arms and spring arms. The horizontal arms can be rotated horizontally with 360°, the spring arms can be rotated horizontally with 360° and moved vertically with 50° downwards and 45° upwards.
The light system is operated by a keypad on the lamp head or, by special request of the customer, by a keypad on the wall.
The surgical light Mach LED 2MC will be marketed with merging of light fields, light intensity control, color temperature adjustment and integrated laser pointer.
Available accessories for the Mach LED 2MC lighting systems are as follows:
- Camera module
- Remote control of camera module .
- Remote control with network interface for camera module .
- Single monitor yoke for flat panel monitors .
- . Double monitor yoke for flat panel monitors
- . Instrument trays
- Trays for CRT monitors ●
- 24V DC battery backup support .
- Low profile wall control unit .
- Sterilizable handle sleeves .
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for "Dr. Mach Medizinische Leuchten + Technik". The logo is in black and white and features the text "Dr. Mach" in a stylized font. Below the name is the text "Medizinische Leuchten + Technik" in a smaller font. To the left of the text is a drawing of a medical examination light.
# Intended Use
The Mach LED 2MC lighting system is designed for illuminating an examination area and surgical field at the hospital and doctor's practice.
The intended use is identical to the intended use of the predicate device Mach LED MC, K093010.
## Technological characteristics Comparison
The Mach LED 2MC has the same technological characteristics (design, materials, chemical composition, energy source) as the predicate device. The Mach LED 2MC and the predicate device Mach LED MC are both based on the LED (light-emitting diode) technology.
For the new device, the number of LEDs was reduced. Therefore, dimensions of the light head diameter, the illumination depth, and the light intensity have been reduced accordingly.
| Technical Parameter | New Device | Predicate Devices | | |
|---------------------|--------------|-------------------|--------------|--|
| | Mach LED 2MC | Mach LED 3MC | Mach LED 5MC | |
| Number of LEDs | 84 | 112 | 160 | |
# Non-clinical data
Performance testing was conducted to verify that Mach LED 2MC meet the requirements for Medical Electrical Equipment as defined in the relevant standards for Medical electrical equipment and the EN 61000 group.
The accessories are identical to those used with the predicate device.
The risk analysis method used to assess the impact of the modifications was a Failure Modes and Effects Analysis (FMEA). The design verification test methods used are the same as those submitted for the predicate device submission.
# Clinical data
No clinical data is required for this device classification submission.
# Conclusions
The modifications incorporated into the Mach LED 2MC design arc minor (change in number of LEDs). Based on the information provided hercin it is concluded from the performed testing that the device is as safe, as effective, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.