BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) QX AMPLIFIED DNA ASSAY
K140446 · Becton, Dickinson & CO · MKZ · May 20, 2014 · Microbiology
Device Facts
Record ID
K140446
Device Name
BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) QX AMPLIFIED DNA ASSAY
Applicant
Becton, Dickinson & CO
Product Code
MKZ · Microbiology
Decision Date
May 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3120
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140446 · May 20, 2014
BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) QX AMPLIFIED DNA ASSAY
Becton, Dickinson & CO
Retrospective clinical specimens (swabs and urine) from OB/GYN, STD, and family planning clinics
Retrospective clinical specimens were used to assemble blinded comparison panels to evaluate the clinical performance (PPA/NPA) of the BD ProbeTec CTQ Assay on the new BD Viper LT System compared to the predicate BD Viper System.
653 female and 170 male subjects attending OB/GYN, STD, and family planning clinics; Sample Size: 617 compliant female and 167 compliant male subjects; Number of Sites: 4
BD Viper System in extracted mode
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)
Indications for Use
The BD ProbeTec Chlamydia trachomatis (CT) Q® Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper™ LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.
Device Story
The BD ProbeTec CTQ Assay is an in vitro diagnostic test for qualitative detection of Chlamydia trachomatis DNA. It utilizes Strand Displacement Amplification (SDA) technology. The BD Viper LT System is a bench-top instrument that automates nucleic acid extraction, amplification, and detection. Input samples include clinician-collected swabs, patient-collected vaginal swabs, urine, and liquid-based cytology specimens. The system rehydrates dried reagents in Priming and Amplification microwells, performs thermal incubation, and monitors fluorescence via a reader. An automated algorithm processes CT-specific and Extraction Control (EC) signals to determine results. The device is intended for use in clinical laboratory settings by trained personnel. Output is a qualitative result (positive/negative/EC failure) used by clinicians to aid in diagnosing chlamydial urogenital disease, facilitating timely patient treatment.
Clinical Evidence
No clinical diagnostic study was performed; the submission relies on a clinical comparison study between the subject and predicate instruments. 823 subjects were enrolled to create a panel of 660 positive and 951 negative specimens across various matrices. Results showed an overall Positive Percent Agreement (PPA) of 97.9% and Negative Percent Agreement (NPA) of 97.7% between the BD Viper LT and the BD Viper System. Analytical studies included reproducibility (n=96 per level), LoD determination (Probit analysis), carry-over (n=630 per diluent), and environmental stress testing.
Technological Characteristics
SDA technology using BsoBI restriction enzyme and Bst DNA polymerase. Detection via fluorescently-labeled detector probe (fluorophore/quencher). Instrument: BD Viper LT (bench-top). Connectivity: Standalone. Reagents: Dried in disposable microwells. Software: Automated algorithm for signal processing and result reporting.
Indications for Use
Indicated for asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease. Specimens include clinician-collected female endocervical and male urethral swabs, patient-collected vaginal swabs, male/female urine (neat and UPT), and gynecological specimens in BD SurePath or PreservCyt media.
Regulatory Classification
Identification
Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Chlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).
Predicate Devices
BD ProbeTec™ Chlamydia trachomatis (CT) Q® Amplified DNA Assay on the BD Viper System (K081824)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K140446
B. Purpose for Submission:
The purpose of this submission is to migrate the BD ProbeTec™ Chlamydia trachomatis (CT) Qˢ Amplified DNA Assay (“BD ProbeTec CTQ Assay”) from the previously cleared (K081824) BD Viper System operating in extracted mode (“BD Viper”) to the new BD Viper LT System.
C. Measurand:
Chlamydia trachomatis (CT) DNA
D. Type of Test:
Strand displacement nucleic acid amplification (SDA) assay
E. Applicant:
Becton, Dickinson and Company
F. Proprietary and Established Names:
BD ProbeTec™ Chlamydia trachomatis (CT) Qˢ Amplified DNA Assay
BD ProbeTec CTQ Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3120, Chlamydia serological reagents
2. Classification:
Class I
3. Product code:
MKZ
4. Panel:
Microbiology (83)
{1}
H. Intended Use:
1. Intended use(s):
The BD ProbeTec Chlamydia trachomatis (CT) Q® Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper™ LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
Prescription only
4. Special instrument requirements:
The BD ProbeTec CTQ Assay uses SDA technology on the BD Viper System in Extracted Mode or the BD Viper LT System, which extracts, amplifies, and detects the target DNA.
I. Device Description:
The BD ProbeTec CTQ Assay uses SDA technology for the detection of CT DNA. The BD Viper LT System is a table-top instrument that is designed to be fully contained on a standard laboratory bench-top. The system performs automated extraction of nucleic acids from multiple specimen types in addition to amplification and detection of target nucleic acid sequences when utilized with legally marketed in vitro diagnostic assays.
J. Substantial Equivalence Information:
1. Predicate device name(s):
BD ProbeTec™ Chlamydia trachomatis (CT) Q® Amplified DNA Assay on the BD Viper System
2. Predicate 510(k) number(s):
K081824
3. Comparison with predicate:
{2}
| Item | Predicate Device: BD ProbeTec CTQ Assay on the BD Viper System – K081824 | Subject Device: BD ProbeTec CTQ Assay on the BD Viper LT System – K140446 |
| --- | --- | --- |
| Intended Use | The BD ProbeTec CT Q^{x} Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease. | The BD ProbeTec Chlamydia trachomatis (CT) Q^{x} Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper™ LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease. |
| Assay Results | Qualitative | Same as predicate |
| Technology | Strand displacement amplification (SDA) | Same as predicate |
| Instrument | BD Viper | BD Viper LT |
| Specimen Types | Key: liquid based cytology (LBC) urine preservative transport media (UPT) | Key: liquid based cytology (LBC) urine preservative transport media (UPT) |
{3}
| Item | Predicate Device: BD ProbeTec CTQ Assay on the BD Viper System – K081824 | Subject Device: BD ProbeTec CTQ Assay on the BD Viper LT System – K140446 |
| --- | --- | --- |
| | Female specimens: Endocervical swab Patient-collected vaginal swab Neat urine UPT urine Male Specimens: Urethral swab Neat urine UPT urine | Female specimens: Endocervical swab Patient-collected vaginal swab Neat urine UPT urine LBC specimens collected in SurePath preservative fluid LBC specimens collected in PreservCyt Solution Male Specimens: Urethral swab Neat urine UPT urine |
| Priming Microwell | | |
| Primers | CT cryptic plasmid | Same as predicate |
| Detector | Linear Detector Flourescein (fluorophore) Dabcyl (quencher) | Same as predicate |
| Nucleotides | 4 of 4 nucleotides required for SDA | Same as predicate |
| Non-specific reagents and cofactors | Buffering components, magnesium ions, salt, and stabilizing reagents | Same as predicate |
| Amplification Microwell | | |
| Restriction enzyme | BsoBI restriction enzyme | Same as predicate |
| Polymerase | Bst DNA polymerase | Same as predicate |
| Nucleotides | 0 of 4 nucleotides required for SDA | Same as predicate |
| Non-specific reagents and cofactors | Buffering components, magnesium ions, salt, and stabilizing reagents | Same as predicate |
# K. Standard/Guidance Documents Referenced (if applicable):
1. CLSI. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition. Sections 5 and 7 - Protocols for Evaluation of the Limit of Blank & Limit of Detection and Verification of Detection
{4}
Capability Claims. CLSI document EP 17-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2012.
2. CLSI. Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods; Approved Guideline. Sections 5, 6, and 8 – Specimen Collection, Specimen Transport and Storage, & Sample Storage. CLSI document MM13-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2005.
3. CLSI. Molecular Diagnostic Methods for Infectious Diseases; Approved Guideline – Second Edition. Sections 10 and 13 – Establishment and Evaluation of Performance Characteristics of Molecular Diagnostic Tests & Controlling False-Positive Nucleic Acid Target Amplification Reactions. CLSI document MM3-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2006.
4. FDA. Assay Migration Studies for in vitro Diagnostic Devices. Guidance for Industry and FDA Staff; 2013.
## L. Test Principle:
The BD ProbeTec CTQ Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides, and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper LT System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of CT DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescence Units; Max RFU) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.
In addition to the fluorescent probe used to detect amplified CT target DNA, a second fluorescently-labeled oligonucleotide is incorporated into each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the CT-specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is re-hydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper LT System and an automated algorithm is applied to both the EC and CT-specific signals to report specimen results as positive, negative, or EC failure.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
## Reproducibility
To create the panel for the reproducibility study, CT (serovar H) and Neisseria gonorrhoeae
{5}
(GC; ATCC strain 19424) were spiked into clinical matrix in their respective diluents: vaginal swab matrix in Swab Diluent for the BD ProbeTec CT/GC Amplified DNA Assays ("Q^x Swab Diluent"), female urine in Urine Preservative Transport media for the BD ProbeTec Q^x Amplified DNA Assays ("UPT"), or liquid based cytology specimens (LBC) in PreservCyt LBC media diluted into BD LBC Specimen Dilution Tubes.
Four levels were created for each matrix/diluent type: negative, high negative (C₂₀-C₈₀), low positive (1.5X LoD), and moderate positive (3X LoD). Reproducibility for GC can be found in K140448. The reproducibility panel was comprised of twelve panel members (3 clinical matrices x 4 levels = 12 panel members). Testing was conducted at 2 external sites (each had one BD Viper LT) and one internal site (one BD Viper LT and three BD Viper Systems). One lot of extraction and amplification reagents and one lot of the Control Set for the BD ProbeTec CT/GC Q^x Amplified DNA Assays ("BD CT/GC Q^x Control Set") were included in this study.
For the BD Viper LT System, there were two operators per site testing one run per day on one instrument over a total of 8 days. Each panel member was tested in duplicate on three BD Viper LT Systems over eight days by two operators (2 x 3 x 8 x 2 = 96 tests per level). The reproducibility the BD ProbeTec CTQ Assay on the BD Viper is shown in Table 1 below.
Table 1 - Reproducibility of the BD ProbeTec CTQ Assay on the BD Viper LT System
| | | | | | Within Run | | Between Run within Day | | Between Day within Site | | Between Site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Specimen Type | Panel | % Expected Results# | 95% CI | Mean RFU | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PreservCyt LBC | Negative | 99.0% (95/96) | (94.3 - 99.8%) | 14.6 | 104.7 | 718.8 | 0.0 | 0.0 | 7.2 | 49.1 | 6.9 | 47.4 | 105.2 | 722.0 |
| | High Negative | 41.7% (40/96) | (32.3 - 51.7%) | 565.0 | 533.2 | 94.4 | 226.5 | 40.1 | 0.0 | 0.0 | 0.0 | 0.0 | 579.3 | 102.5 |
| | Low Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1435.3 | 372.3 | 25.9 | 0.0 | 0.0 | 85.2 | 5.9 | 86.4 | 6.0 | 391.6 | 27.3 |
| | Moderate Positive | 98.9% (93/94)* | (94.2 - 99.8%) | 1752.4 | 241.0 | 13.8 | 0.0 | 0.0 | 36.1 | 2.1 | 0.0 | 0.0 | 243.7 | 13.9 |
| Vaginal Swab | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 22.1 | 28.6 | 129.3 | 8.7 | 39.5 | 0.0 | 0.0 | 5.0 | 22.5 | 30.3 | 137.0 |
| | High Negative | 24.0% (23/96) | (16.5 - 33.4%) | 974.0 | 716.7 | 73.6 | 0.0 | 0.0 | 0.0 | 0.0 | 105.1 | 10.8 | 724.4 | 74.4 |
| | Low Positive | 95.8% (91/95)* | (89.7 - 98.4%) | 1603.8 | 535.8 | 33.4 | 115.0 | 7.2 | 0.0 | 0.0 | 0.0 | 0.0 | 548.0 | 34.2 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1780.6 | 320.3 | 18.0 | 0.0 | 0.0 | 59.6 | 3.3 | 52.2 | 2.9 | 330.0 | 18.5 |
| Female UPT | Negative | 99.0% (95/96) | (94.3 - 99.8%) | 21.5 | 171.4 | 797.4 | 0.0 | 0.0 | 0.0 | 0.0 | 5.8 | 27.0 | 171.5 | 797.9 |
| | High Negative | 32.3% (31/96) | (23.8 - 42.2%) | 678.3 | 578.6 | 85.3 | 276.0 | 40.7 | 0.0 | 0.0 | 0.0 | 0.0 | 641.1 | 94.5 |
| | Low Positive | 99.0% (95/96) | (94.3 - 99.8%) | 1549.4 | 432.5 | 27.9 | 0.0 | 0.0 | 85.9 | 5.5 | 0.0 | 0.0 | 441.0 | 28.5 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1747.0 | 223.9 | 12.8 | 133.5 | 7.6 | 0.0 | 0.0 | 22.5 | 1.3 | 261.6 | 15.0 |
*There were two moderate positive LBC samples and one low positive swab sample which resulted in an extraction transfer error and therefore no valid results were available for analysis.
The results for negative panel members were calculated according to an expected result of 'negative for CT.' All other panel members were calculated according to an expected result of 'positive for CT.'
For the BD Viper System, there were two operators testing three runs per day on each of the 3 instruments (6 runs total per day) over a total of 4 days. Each panel member was tested in duplicate on three BD Viper Systems over four days by two operators (2 x 6 x 4 x 2 = 96 tests
{6}
per level). The reproducibility the BD ProbeTec CTQ Assay on the BD Viper is shown in Table 2 below.
Table 2 – Reproducibility of the BD ProbeTec CTQ Assay on the BD Viper System
| | | | | | Within Run | | Between Run within Day | | Between Day within Instrument | | Between Instrument | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Specimen Type | Panel | % Expected Results# | 95% CI | Mean RFU | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| PreservCyt LBC | Negative | 99.0% (95/96) | (94.3 - 99.8%) | 14.1 | 135.6 | 962.7 | 0.0 | 0.0 | 1.1 | 7.8 | 4.2 | 29.9 | 135.6 | 963.2 |
| | High Negative | 44.8% (43/96) | (35.2 - 54.7%) | 723.3 | 771.3 | 106.6 | 0.0 | 0.0 | 211.8 | 29.3 | 0.0 | 0.0 | 799.8 | 110.6 |
| | Low Positive | 95.8% (92/96) | (89.8 - 98.4%) | 1542.5 | 530.9 | 34.4 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 530.9 | 34.4 |
| | Moderate Positive | 99.0% (95/96) | (94.3 - 99.8%) | 1951.3 | 269.2 | 13.8 | 0.0 | 0.0 | 0.0 | 0.0 | 112.7 | 5.8 | 291.8 | 15.0 |
| Vaginal Swab | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 1.3 | 5.6 | 437.0 | 2.9 | 224.2 | 0.0 | 0.0 | 1.7 | 133.2 | 6.5 | 508.9 |
| | High Negative | 30.2% (29/96) | (21.9 - 40.0%) | 832.6 | 759.7 | 91.2 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 759.7 | 91.2 |
| | Low Positive | 97.9% (94/96) | (92.7 - 99.4%) | 1660.9 | 445.5 | 26.8 | 0.0 | 0.0 | 0.0 | 0.0 | 54.6 | 3.3 | 448.8 | 27.0 |
| | Moderate Positive | 99.0% (95/96) | (94.3 - 99.8%) | 1677.3 | 413.3 | 24.6 | 0.0 | 0.0 | 105.1 | 6.3 | 71.1 | 4.2 | 432.3 | 25.8 |
| Female UPT | Negative | 100.0% (96/96) | (96.2 - 100.0%) | 0.6 | 2.5 | 407.6 | 0.0 | 0.0 | 0.0 | 0.0 | 0.6 | 92.3 | 2.6 | 417.9 |
| | High Negative | 29.2% (28/96) | (21.0 - 38.9%) | 973.5 | 795.9 | 81.8 | 201.9 | 20.7 | 0.0 | 0.0 | 0.0 | 0.0 | 821.1 | 84.3 |
| | Low Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1780.6 | 394.4 | 22.1 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 394.4 | 22.1 |
| | Moderate Positive | 100.0% (96/96) | (96.2 - 100.0%) | 1925.5 | 225.7 | 11.7 | 0.0 | 0.0 | 49.3 | 2.6 | 34.7 | 1.8 | 233.6 | 12.1 |
# The results for negative panel members were calculated according to an expected result of 'negative for CT.' All other panel members were calculated according to an expected result of 'positive for CT.'
The data presented in Tables 1 and 2 demonstrate good reproducibility for the BD ProbeTec CTQ Assay on the BD Viper LT and the BD Viper System. On both systems, the assay shows $\geq$ 95% detection of CT at low positive/close to the detection limit and moderate positive levels. For the high negative samples, detection was >5% and <95% positive on both systems. Taken together, this data indicates that the reproducibility for the BD Viper LT and the BD Viper System is similar.
b. Linearity/assay reportable range: N/A
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# Post Pre-Warm Specimen Stability
The existing workflow previously cleared for the BD ProbeTec CTQ Assay on the BD Viper includes a pre-warm step for swab and urine samples. The workflow proposed in K140446 for the BD Viper LT System includes a pre-warm step for LBC specimens in SurePath or PreservCyt media so all sample types (urine, swabs, and LBC specimens) can be tested on the same instrument run. For this reason, a post pre-warm specimen stability study was conducted to support the storage and transport stability claims for LBC specimens in SurePath or PreservCyt LBC media.
{7}
Mini-pools were created by diluting CT and GC negative LBC specimens in either SurePath or PreservCyt LBC media at a ratio of 1:3.4 mL (12 mini-pools were created for each LBC media). Each mini-pool was split equally. Half were spiked with CT serovar H and GC strain ATCC 19424 (at 90 EB/mL and 300 cells/mL, respectively) and half were not spiked (negative samples). Results for GC can be found in K140448. From these split mini-pools, 2.2 mL aliquots were removed and dispensed into BD LBC Specimen Dilution Tubes (15 samples per group; 30 samples total/mini-pool). Fourteen of these tubes were pre-warmed in the BD Viper LT heat block at 114°C for 15 min, one tube was read at "baseline" (no warming).
After the pre-warm step, the seven tubes were stored at the following conditions: two tubes at 2-8°C for 7 days, two tubes at 30°C for 7 days, and two tubes at -20°C for 90 days (one tube was available for testing and one was available for retesting, if required). At each storage condition/time point, one negative and one positive tube from each of the 12 mini-pools was removed and tested in a dual dispensing mode, which generated two replicate test results (1 tube x 2 replicates x 12 mini-pools = 24 measurements/storage condition/time point for each LBC media type). Four lots of extraction reagents, 2-3 lots of amplification reagents, three lots of the BD CT/GC Q⁴ Control Set, and three BD Viper LT Systems were included in this study. Results are summarized in Tables 2 and 3 below.
Table 3 - PreservCyt Specimens Post Pre-Warm Stability Results
| Temperature | Timepoint | Positives | Negatives |
| --- | --- | --- | --- |
| N/A | Baseline | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass) |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass)* |
*Data generated by Out of Specification (OOS) testing after contaminated environment was deemed clean.
Table 4 - SurePath Specimens Post Pre-Warm Stability Results
| Temperature | Timepoint | Positives | Negatives |
| --- | --- | --- | --- |
| N/A | Baseline | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 3 | 24/24 (Pass) | 0/24 (Pass) |
| 2-8 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| 30 °C | Day 7 | 24/24 (Pass) | 0/24 (Pass) |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass) |
| -20 °C | Day 90 | 24/24 (Pass) | 0/24 (Pass) |
The data indicates that, for all conditions and time points tested, the percent positive rate for prewarmed CT positive and percent negative rate for CT negative LBC samples was 100% for samples stored in either PreservCyt or SurePath LBC media.
{8}
9
# Run Control Failure Rate
This analysis was conducted to determine the failure rate for the BD CT/GC Q⁺ Control Set across four different studies. One positive and negative control result was recorded for each run for each of the following:
Clinical agreement study = 56 runs
Carry-over study (in K140446, referred to as “contamination validation study”) = 63 runs
LoD/LoB validation studies = 71 runs
Reproducibility study = 51 runs
A total of 241 results were collected for each control type. There were no false negatives found in the positive control samples assessed (0/241 x 100 = 0% failure rate). There were 2 false positives found in the negative control samples assessed (2/241 x 100 = 0.83% failure rate).
# Environmental Study
The environmental study was conducted to assess the performance of the BD ProbeTec CTQ Assay on the BD Viper LT when the instrument is exposed to variations in environmental temperature and humidity. One BD Viper LT was placed in an environmental chamber and equilibrated to either 1) 18°C with 20% humidity (low temperature, low humidity) or 2) 27°C with 85% humidity (high temperature, high humidity). Three lots of positive and negative controls from the BD CT/GC Q⁺ Control Set were reconstituted with Viper QC Wash Buffer and run as samples. For the negative controls, 30 “samples” were prepared from each lot with one run per lot. An extraction control (EC) was included in each of the negative controls to show that the negative result was valid (30 negative samples/ECs x 3 lots = 90 measurements). For the positive controls, 90 “samples” were prepared from each lot with 3 runs/lot (90 positive samples/lot x 3 lots = 270 measurements).
Each rack was loaded with all positive or all negative controls (30/rack). These were logged in as “samples.” On each rack, one pair of positive and negative controls were run normally (e.g., reconstituted, extracted, and amplified by the BD Viper LT). This study required 12 runs total with one lot of extraction and amplification reagents for testing with the assay. The results are summarized in Table 5 below.
Table 5 - Environmental System Testing Summary
| Conditions/Parameter | BD ProbeTec CTQ Assay Percent Correct (N) | | | Pass/Fail ≥97% |
| --- | --- | --- | --- | --- |
| | EC | Negative | Positive | |
| 1: 18°C / 20% RH | 100% (90/90) | 100% (90/90) | 100% (270/270) | Pass |
| 2: 27°C / 85% RH | 100% (90/90) | 100% (90/90) | 100% (270/270) | Pass |
There were no false negatives or false positives detected in this study, indicating that the performance of the BD ProbeTec CTQ Assay on the BD Viper LT is not affected by the temperature and humidity tested in this study.
{9}
# d. Detection limit:
# Limit of Detection (LoD)
The following clinical matrices were tested in the LoD studies: vaginal swabs in $\mathbf{Q}^{\mathrm{x}}$ Swab Diluent, urine in UPT, and SurePath or PreservCyt LBC specimens in LBC Specimen Dilution Tubes. For all LoD studies, CT serovar H (ATCC VR879; strain UW-43/Cx) and GC (ATCC strain 19424) were paired for testing and CT serovar D (ATCC VR885; strain UW-3Cx) and GC (ATCC strain 49226) were paired for testing. Results of the LoD studies for GC can be found in K140448.
# i. LoD Detection - BD Viper and BD Viper LT
For each matrix, four mini-pools were prepared and then used to create LoD detection panels (one panel was created for each of the four mini-pools). Each LoD detection panel contained six analyte levels for CT and GC (CT concentrations in Table 6 below; GC results can be found in K140448). There were six replicates created for the six analyte levels (4 mini-pools x 6 analyte levels x 6 replicates for each of the analyte levels = 144 total tubes tested per matrix).
Table 6 - CT Levels (in EB/mL) Tested in LoD Detection Study for Each Matrix Type
| Level | Vaginal Swab in Qx Swab Diluent | | UPT Urine | SurePath LBC Specimen in LBC Diluent | PreservCyt LBC Specimen in LBC Diluent |
| --- | --- | --- | --- | --- | --- |
| | Initial | Repeat* | | | |
| 0 | 0 | 0 | 0 | 0 | 0 |
| 1 | 3 | 5 | 1 | 3 | 3 |
| 2 | 7 | 10 | 3 | 7 | 7 |
| 3 | 15 | 20 | 7 | 15 | 15 |
| 4 | 30 | 40 | 15 | 30 | 30 |
| 5 | 60 | 60 | 30 | 60 | 60 |
*Repeat of CT serovar H vaginal swab LoD with new CT target levels and increased replicates due to variable positive rates and poor reproducibility at low target levels thus impacting the model fit and the accuracy of the LoD estimates. Target levels chosen are representative of the claimed LoD per K091824.
The LoD detection study was conducted in one day for each matrix type by two operators running samples on 3 BD Viper and 3 BD Viper LT Systems. Table 7 demonstrates the setup of a rack on one instrument. Each rack was run twice on the same instrument (4 replicates x 6 levels x 2 runs/rack = 48 measurements/rack).
Table 7 - Sample Rack Set up for LoD Detection Study
| | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | |
| | | Level 5 | Level 4 | Level 3 | Level 2 | Level 1 | null |
| | | MP 1 | MP 1 | MP 1 | MP 1 | MP 1 | MP 1 |
| | | Level 5 | Level 4 | Level 3 | Level 2 | Level 1 | null |
| | | MP 2 | MP 2 | MP 2 | MP 2 | MP 2 | MP 2 |
{10}
Three different extraction and amplification reagent lots and three lots of BD CT/GC Qx Control Set were used for the LoD detection study
A target titration plot was created by graphing the percent positive versus the concentration of CT EBs/mL in Log10. Probit analysis was used to estimate the 95% LoD concentration for the BD Viper LT and BD Viper System. Results of the LoD detection study are shown in the Table 8 below.
Table 8 - LoD for BD Viper LT and BD Viper Systems
| Specimen Matrices | Serovar | BD Viper LT LoD (95% CI) | BD Viper LoD (95% CI) | LoD Ratio (95% CI) |
| --- | --- | --- | --- | --- |
| Vaginal Swabs in Qx Swab Diluent | CT Serovar H | 47 (32, 69) | 45 (31, 67) | 1.0 (0.9, 1.1) |
| | CT Serovar D | 53 (33, 85) | 48 (30, 78) | 1.0 (0.9, 1.2) |
| UPT Urine | CT Serovar H | 39 (28, 55) | 27 (19, 37) | 1.1 (1.0, 1.2) |
| | CT Serovar D | 16 (11, 22) | 12 (9, 17) | 1.1 (1.0, 1.3) |
| SurePath LBC Sample in LBC Diluent | CT Serovar H | 17 (12, 27) | 26 (17, 40) | 0.9 (0.7, 1.0) |
| | CT Serovar D | 16 (13, 21) | 14 (11, 18) | 1.0 (0.9, 1.2) |
| PreservCyt LBC Sample in LBC Diluent | CT Serovar H | 14 (11, 18) | 15 (12, 20) | 1.0 (0.9, 1.1) |
| | CT Serovar D | 11 (8, 15) | 15 (11, 20) | 0.9 (0.7, 1.0) |
The data shown above is acceptable as it meets the pre-determined acceptance criteria that the LoD ratio between the BD Viper LT and BD Viper Systems must be 0.8-1.2 (BD Viper LT/BD Viper).
# ii. LoD Confirmation - BD Viper LT
For each matrix, one pool was created and then split into two aliquots of equal volume. One aliquot was spiked with CT serovar H and the other aliquot was spiked with CT serovar D. The LoD concentrations from the LoD determination study (see table 8 above) were used for spiking levels. Twenty replicates were generated from each aliquot. All samples were tested using one lot of reagents and one BD Viper LT. The results of the LoD confirmation are shown in Table 9 below.
Table 9 - Results of LoD Confirmation Study
| Specimen Matrices | Serovar | Total # Tested | Total Positive | Total Negative | Percent Positive |
| --- | --- | --- | --- | --- | --- |
| Vaginal Swabs in Qx Swab Diluent | CT Serovar H | 20 | 19 | 1 | 95% |
| | CT Serovar D | 20 | 20 | 0 | 100% |
| UPT Urine | CT Serovar H | 20 | 19 | 1 | 95% |
{11}
| | CT Serovar D | 20 | 19 | 1 | 95% |
| --- | --- | --- | --- | --- | --- |
| SurePath LBC
Sample in LBC
Diluent | CT Serovar H | 20 | 20 | 0 | 100% |
| | CT Serovar D | 20 | 19 | 1 | 95% |
| PreservCyt LBC
Sample in LBC
Diluent | CT Serovar H | 20 | 20 | 0 | 100% |
| | CT Serovar D | 20 | 20 | 0 | 100% |
The results from the LoD confirmation study were found to be acceptable as they met the predetermined acceptance criteria of 95% positive agreement with the LoD levels from the LoD determination study.
## Limit of Blank (LoB)
Data from CT negative specimens (Level 0) from the analytical LoD studies described above were combined to calculate an overall LoB for the BD ProbeTec CTQ Assay on the BD Viper LT. Data from all specimen matrices were compared and the worst (or highest) LoB estimate among all the specimen matrices was selected as the overall LoB (68 MaxRFU). Because this value is below the assay threshold of 125 MaxRFU, this demonstrates that the assay has an acceptable separation between background fluorescence obtained from CT negative specimens and the assay cut-off for a positive result.
## e. Analytical specificity:
## Carry-Over/System Contamination
The carry-over study was conducted to look for the presence of contamination in negative specimens due to carry-over of DNA during nucleic acid extraction and PCR amplification in the BD Viper LT. Each run consisted of 15 high positive and 15 negative samples arranged in alternating positive or negative rows. To create high positive samples, 1 x 10⁵ EB/mL of CT was spiked into either one of two diluents (listed below). These diluents were selected because they represent two most challenging settings that can cause sample contamination.
1. 3 mL of Q⁸ Swab Diluent: This volume was selected to represent a maximum fill volume, thereby maximizing the amount of fluid adhering to the outside of the pipette tip.
2. PreservCyt LBC media diluted in BD LBC Specimen Dilution tube diluent (1.7 mL LBC diluent + 0.5 mL PreservCyt = 2.2 mL; "LBC Specimen Matrix"): This LBC media was selected because it has a high alcohol content and is therefore more likely to drip from the pipette tips during sample transfer due to low surface tension.
Each run consisted of 30 samples. Seven runs were conducted per instrument with 3 BD Viper LT Systems for a total of (30 samples/run x 7 runs x 3 instruments) 630 measurements for each diluent. Runs were conducted by three operators over a period of three days. Two lots of extraction reagents, one lot of amplification reagents, and two lots of the BD CT/GC Q⁸ Control Set were used in this study.
{12}
Table 10 - Carry Over on Viper LT
| BD Viper LT System | Qx Sample Diluent | | | LBC Specimen Matrix | | |
| --- | --- | --- | --- | --- | --- | --- |
| | n | Positive Results | Percent Positive | n | Positive Results | Percent Positive |
| 1 | 210 | 0 | 0.00% | 210 | 0 | 0.00% |
| 2 | 210 | 1 | 0.48% | 210 | 0 | 0.00% |
| 3 | 210 | 1 | 0.48% | 210 | 0 | 0.00% |
| Overall | 630 | 2 | 0.32% | 630 | 0 | 0.00% |
The overall rate of carry-over was $0.32\%$ (2/630) for CT spiked into the $\mathrm{Q}^{\mathrm{x}}$ Swab Diluent and $0.0\%$ (0/630) for CT spiked into PreservCyt media diluted in BD LBC Specimen Dilution tube diluent. These results are acceptable as they are comparable to the levels of carry-over cleared for the BD ProbeTec CTQ Assay on the BD Viper in K081824.
f. Assay cut-off:
The assay cut-off was established previously in K081824.
2. Comparison studies:
a. Method comparison with predicate device:
# Clinical Comparison Study
The results of the BD ProbeTec CTQ Assay run on the BD Viper LT were compared to the results obtained with the BD Viper System in a clinical comparison study. Clinical samples were collected between October 2012 and April 2013 from four geographically diverse clinical sites in North America (OB/GYN, sexually transmitted disease and family planning clinics). For each female subject, specimens were collected in the following order: (1) a first void urine specimen, (2) five patient-collected vaginal swab specimens, and (3) a SurePath liquid based cytology (LBC) specimen and a PreservCyt LBC specimen. The LBC specimen collection order was randomized throughout the study. For each male subject, a first void urine specimen was collected. Each urine specimen (male and female) was aliquoted into five UPTs (BD urine Preservative Transport for the $Q^{\mathrm{x}}$ Amplified DNA Assays).
A total of 823 subjects were enrolled (170 males and 653 females). Specimens were collected from subjects presenting with symptoms of CT and/or GC (symptomatic) and shipped to BD on cold packs for screening, aliquoting, and panel assembly. Following exclusion of non-compliant subjects and non-evaluable specimens, a total of 617 compliant female subjects and 167 compliant male subjects were available for the clinical comparison panel assembly.
To create the clinical comparison panel, for each subject, a UPT specimen was tested on the BD Viper. These results were used to confirm that the target number of CT positive panel members had been met. The clinical comparison panel consisted of randomly selected positive and negative specimens with four aliquots created from each specimen type. The number of
{13}
specimens selected for the clinical comparison panel was based on minimums suggested by FDA in previous correspondence (minimum of 40 CT positive and 40 CT negative). The positive and negative specimens were randomized within the panel and the testing sites were blinded to the results for each panel member. Panels were identical across all testing sites. Three external sites tested the panels on the BD Viper LT. An internal site tested the panels on the BD Viper System. Results from the two instruments were compared and expressed statistically as positive and negative percent agreement.
Table 11 - PPA and NPA for the BD ProbeTec CTQ Assay on the BD Viper LT System versus the BD Viper System
| | | | Positive Percent Agreement | | Negative Percent Agreement | |
| --- | --- | --- | --- | --- | --- | --- |
| Gender | Specimen Type | Site | Percent | 95% CI | Percent | 95% CI |
| Female | Vaginal Swab | A | 100.0% (46/46) | (92.3%, 100.0%) | 95.0% (57/60) | (86.3%, 98.3%) |
| | | B | 100.0% (46/46) | (92.3%, 100.0%) | 98.3% (59/60) | (91.1%, 99.7%) |
| | | C | 100.0% (46/46) | (92.3%, 100.0%) | 91.7% (55/60) | (81.9%, 96.4%) |
| | | Total | 100.0% (138/138) | NA | 95.0% (171/180) | (90.6%, 98.3%) |
| | Q* UPT | A | 97.9% (46/47) | (88.9%, 99.6%) | 98.3% (58/59) | (91.0%, 99.7%) |
| | | B | 97.9% (46/47) | (88.9%, 99.6%) | 100.0% (59/59) | (93.9%, 100.0%) |
| | | C | 95.7% (45/47) | (85.8%, 98.8%) | 98.3% (58/59) | (91.0%, 99.7%) |
| | | Total | 97.2% (137/141) | (92.2%, 100.0%) | 98.9% (175/177) | (96.6%, 100.0%) |
| | BD SurePath | A | 100.0% (39/39) | (91.0%, 100.0%) | 97.0% (65/67) | (89.8%, 99.2%) |
| | | B | 100.0% (39/39) | (91.0%, 100.0%) | 97.0% (65/67) | (89.8%, 99.2%) |
| | | C | 100.0% (39/39) | (91.0%, 100.0%) | 97.0% (65/67) | (89.8%, 99.2%) |
| | | Total | 100.0% (117/117) | NA | 97.0% (195/201) | (93.0%, 100.0%) |
| | PreservCyt | A | 97.5% (39/40) | (87.1%, 99.6%) | 97.0% (64/66) | (89.6%, 99.2%) |
| | | B | 95.0% (38/40) | (83.5%, 98.6%) | 98.5% (65/66) | (91.9%, 99.7%) |
| | | C | 95.0% (38/40) | (83.5%, 98.6%) | 98.5% (65/66) | (91.9%, 99.7%) |
| | | Total | 95.8% (115/120) | (89.2%, 100.0%) | 98.0% (194/198) | (96.0%, 99.5%) |
| Male | Q* UPT | A | 97.9% (47/48) | (89.1%, 99.6%) | 100.0% (65/65) | (94.4%, 100.0%) |
| | | B | 95.8% (46/48) | (86.0%, 98.8%) | 100.0% (65/65) | (94.4%, 100.0%) |
| | | C | 95.8% (46/48) | (86.0%, 98.8%) | 98.5% (64/65) | (91.8%, 99.7%) |
| | | Total | 96.5% (139/144) | (92.4%, 100.0%) | 99.5% (194/195) | (98.5%, 100.0%) |
| Total | All | Total | 97.9% (646/660) | (96.0%, 99.4%) | 97.7% (929/951) | (96.3%, 98.8%) |
{14}
*Denominator based on the results from the BD Viper System and numerator based on the results from the BD Viper LT System. For reproducibility by site, the 95% CI was calculated using the score method. For the total reproducibility, the 95% CI was calculated using the bootstrap approach (except when there was 100% agreement at all sites, then the 95% CI cannot be calculated).
NA = Not Applicable
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity and Specificity
N/A
4. Clinical cut-off:
N/A
5. Expected values/Reference range:
N/A
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.