K140441 · Promepla Sam · KNY · Jun 17, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K140441
Device Name
BI-FLEX URETERAL ACCESS SHEATH
Applicant
Promepla Sam
Product Code
KNY · Gastroenterology, Urology
Decision Date
Jun 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Indications for Use
The Bi-Flex Ureteral Access Sheath is intended to be a conduit for passage of endoscopes and other urological devices for the purpose of performing ureteroscopy procedures. The dual working lumen dilator with luer lock connections allows the user to insert guidewires and fluids.
Device Story
Bi-Flex Ureteral Access Sheath functions as conduit for endoscopes/urological instruments during ureteroscopy; consists of flexible, coil-reinforced sheath and semirigid dual-lumen dilator with tapered distal tip. Device inserted over guidewire; dilator removed, leaving sheath in place. Dual-lumen dilator features luer lock connections for guidewire/fluid insertion. Used in clinical urology settings by physicians. Radiopaque materials and hydrophilic coating facilitate navigation and placement. Benefits include safe, atraumatic access to urinary tract for diagnostic or therapeutic procedures.
Indicated for patients undergoing ureteroscopy procedures requiring a conduit for the passage of endoscopes and other urological instruments, and for the insertion of guidewires and fluids.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Submitter . Promepla SAM
# JUN 1 7 2014
K140441 Pg. 1 of 3 Bi-Flex Ureteral Access Sheath Traditional 510 (k)
# 510 (k) Summary
# A. Submitter Information:
| Submitter's Name: | PROMEPLA SAM | |
|--------------------------|-----------------------------------|--|
| Address: | 9 Avenue Prince Albert II | |
| | "LE COPORI" | |
| | MC 98000 MONACO (Principality of) | |
| Contact Person: | Mohamed Rekik | |
| Contact Person's Number: | (377) 97984233 | |
| Contact Person's Fax: | (377) 92056150 | |
| Date of Preparation: | January 23, 2014 | |
#### B. Device Name:
| Trade Name: | Bi-Flex Ureteral Access Sheath | |
|-------------------------|--------------------------------|--|
| Common Name: | Ureteral Access Sheath | |
| Classification Name(s): | Accessories, Catheter, G-U | |
| Produce Code: | KNY | |
| CFR Reference: | 21 CFR 876.5130 | |
#### C. Predicate Device Name:
Trade Name:
ROCAMED RocaUS Platinum (K120160)
#### D. Device Description:
The Bi-Flex Ureteral Access Sheath is designed to create a conduit for urological procedural instruments. The device consists of two components: a flexible, coil reinforced sheath and a semirigid dual lumen dilator catheter with tapered distal tip. Both components are radiopaque and have hydrophilic coating. This device is sold in two sizes, 10/12 and 12/14 FR, and two lengths, 35 and 45 cm.
#### E. Intended Use:
The Bi-Flex Ureteral Access Sheath is intended to be a conduit for passage of endoscopes and other urological devices for the purpose of performing ureteroscopy procedures. The dual working lumen dilator with luer lock connections allows the user to insert guidewires and fluids.
# F. Technological Characteristics Summary:
The Bi-Flex Ureteral Access Sheath is flexible coil reinforced sheath with hydrophilic coating. The device can be inserted by placing the dilator/sheath assembly over a guidewire, inserting it into the patient and unclipping the dilator from the sheath and removing the sheath in place. The sheath allows for safe passage of endoscopes, injection of fluids and other related instruments.
Table 1 provides a comparison summary of the technological characteristics of the Bi-Flex Ureteral Access Sheath versus the predicate devices.
Section 5
.
{1}------------------------------------------------
# Table 1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
#
| | Proposed Device | Predicate Device |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Bi-Flex Ureteral Access<br>Sheath | ROCAMED RocaUS<br>Platinum |
| 510(k) Number | | |
| Product Code, Regulation #,<br>Name | KNY<br>21 CFR 876.5130,<br>Urological catheter and<br>accessories. | KNY<br>21 CFR 876.5130,<br>Urological catheter and<br>accessories. |
| Manufacturer | Promepla SAM | Promepla SAM |
| Intended Use | The Bi-Flex Ureteral<br>Access Sheath is<br>intended to be a conduit<br>for passage of<br>endoscopes and other<br>urological devices for the<br>purpose of performing<br>ureteroscopy procedures.<br>The dual working lumen<br>dilator with luer lock<br>connections allows the<br>user to insert guidewires<br>and fluids. | The ROCAMED RocaUS<br>Platinum is intended to be<br>a conduit for passage of<br>endoscopes and other<br>urological devices for the<br>purpose of performing<br>ureteroscopy procedures.<br>The dual working lumen<br>dilator with luer lock<br>connections allows the<br>user to insert guidewires<br>and fluids. |
| Reuse Status | Disposable. For single<br>patient use only | Disposable. For single<br>patient use only |
| Sterile | Yes | Yes |
| Lumen | 2 | 2 |
| Dilator Material | LDPE+BaSO4 | LDPE+BaSO4 |
| Sheath Material | Pebax-SST-PTFE | Pebax-SST-PTFE |
| X-Ray Opaque | Yes | Yes |
| Coil Reinforced | Yes | Yes |
| Fr Size | 10/12,12/14 | 10/12,12/14 |
| Length | 35, 45 cm | 35 cm |
| Guide wire Compatibility | 0.032", 0.035" | 0.032", 0.035" |
| Atraumatic Tip | Yes | Yes |
| Tapered Dilator | Yes | Yes |
| Radiopaque Marks | Yes | Yes |
| Hydrophilic Coating | Yes | Yes |
| Injection of Contrast Media | Yes | Yes |
| Proximal End Funnel | Yes | Yes |
Section 5
{2}------------------------------------------------
# G. Performance Data:
Results of physical and functional testing support a determination of substantial equivalents for the Bi-Flex Ureteral Acces Sheath when compared to the predicate device.
The Bi-Flex Ureteral Access Sheath is substantially equivalent to devices currently market approved in terms of intended use, technology, principles of operation and materials.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a bird-like figure, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### June 17, 2014
Promepla SAM Mohamed Rekik Quality Manager 9 Avenue Albert II, "Le Copori" MC 98000 MONACO
K140441 Re:
Trade/Device Name: Bi-Flex Ureteral Access Sheath Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: KNY Dated: January 23, 2014 Received: February 21, 2014
Dear Mohamed Rekik,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use sured in the enorther. 776, the enactment date of the Medical Device Amendments, or to conninered province been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic rice, its (110) has the device, subject to the general controls provisions of the Act. The rou may, alores provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is blackling a controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r least of advised that I Drin based that your device complies with other requirements of the Act that I DA has made a continegulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mohamed Rekik
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
JoyceMDang-S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K140441
Device Name Bi-Flex Ureteral Access Sheath
Indications for Use (Describe)
The Bi-Fies: Ureteral Access Sheath is intended to be a conduit for passage of endoscopes and other urological devices for the purpose The DPF fest Offerent Needs onean is the dual working lumen dilator with luer lock connections allows the user to insert guidewires and fluids.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE FOR FOR FOR FOR FOR FOR FOR FOR FOR the first the rest Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
FORM FDA 3881 (1/14)
150 Publishing Services (301) 433-6740
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the I he burden instructions, search existing data sources, gather and maintain the data needed and complete ume to review that dolleding as a some of generate regarding this burden estimate or any other aspect and formation collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.