← Product Code [HQF](/productcode/HQF) · K140362

# KATALYST REVOLVER LASER PROBES, KATALYST REVOLVER ILLUMINATED LASER PROBES, KATALYST REVOLVER ILLUMINATED PROBES (K140362)

_Katalyst Surgical, LLC · HQF · Sep 15, 2014 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140362

## Device Facts

- **Applicant:** Katalyst Surgical, LLC
- **Product Code:** [HQF](/productcode/HQF.md)
- **Decision Date:** Sep 15, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4390
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

Revolver Laser Probes, Revolver Illuminated Laser Probes & Revolver Illuminated Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments. The Revolver Laser Probes can only be used with a medical laser at operating wavelengths of 500nm to 900mm. The Revolver assemblies containing illumination should only be used with the light wavelength range of 425nm to 700nm.

## Device Story

Device consists of fiber optic cables, reusable handles, and reusable laser adapters for vitreoretinal surgery. Inputs: laser energy (500-900nm) and light for illumination (425-700nm). Operation: surgeon manipulates handle to insert metal tubing into surgical site; fiber optics transmit laser for photocoagulation and light for visualization. Output: endo-ocular laser delivery and illumination. Used in OR by ophthalmologists. Benefits: enables precise tissue coagulation and visualization during vitreoretinal procedures. Reusable handle/adapter components paired with single-use sterile fiber optics.

## Clinical Evidence

Bench testing only. Performance verified via optical power ratings, light transmission (lumens), and biocompatibility per ISO 10993. Sterilization validated per ISO 11135-1. Shelf-life and packaging integrity established per ASTM F1980-09, F1929-04, and F88-09. Optical radiation safety tested per ISO 15752 and ISO 15004-2.

## Technological Characteristics

Materials: glass (silica/silica) and plastic (acrylic) fibers, metal tubing. Sensing/Actuation: passive optical transmission. Energy: laser (500-900nm) and illumination (425-700nm). Dimensions: 8-10 ft length; 100-750 micron fiber core diameters; 20g-27g gauge sizes. Connectivity: standalone accessory to ophthalmic laser/light sources. Sterilization: ethylene oxide (ISO 11135-1).

## Regulatory Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

## Predicate Devices

- Katalyst Laser Probe and Illuminated Laser Probe ([K121187](/device/K121187.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are arranged in a way that creates a sense of depth and unity. Encircling the design is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2014

Katalyst Surgical, LLC Ms. Meryl Koch Ouality Assurance and Regulatory Affairs Manager 754 Goddard Ave. Chesterfield, MO 63005

Re: K140362

Trade/Device Name: Katalyst Revolver Laser Probes, Revolver Illuminated Laser Probes & Revolver Illuminated Probes

Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: HQF, HQB, MPA Dated: August 8, 2014 Received: August 15, 2014

Dear Ms. Koch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K140362

Device Name

Katalyst Revolver Laser Probes, Revolver Illuminated Laser Probes & Revolver Illuminated Probes

Indications for Use (Describe)

Revolver Laser Probes, Revolver Illuminated Laser Probes & Revolver Illuminated Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments.

The Revolver Laser Probes can only be used with a medical laser at operating wavelengths of 500nm to 900mm. The Revolver assemblies containing illumination should only be used with the light wavelength range of 425nm to 700nm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K140362)

| Manufacturer:      | Katalyst Surgical, LLC<br>754 Goddard Avenue<br>Chesterfield, MO 63005<br>636-399-0109 (phone)<br>636-787-0603 (fax)                             |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:           | Meryl Koch<br>Quality Assurance and Regulatory Affairs Manager<br>636-536-5950 (phone)<br>636-787-0603 (fax)<br>m.koch@katalystsurgical.com      |
| Date Prepared:     | February 9th, 2013                                                                                                                               |
| Date Updated:      | September 10th, 2014                                                                                                                             |
| Device Trade Name: | Revolver Laser Probes, Revolver Illuminated Laser Probes<br>& Revolver Illuminated Probes                                                        |
| Common Name:       | (1) Ophthalmic photocoagulator, (2) Endoscope and<br>accessories, and (3) Ophthalmic laser (accessory<br>for)                                    |
| Classification:    | (1) 21 CFR 886.4690; Ophthalmic photocoagulator, (2) 21<br>CFR 876.1500; Endoscope and accessories, and (3) 21 CFR<br>886.4390; Ophthalmic laser |
| Class: II          |                                                                                                                                                  |
| Product Code:      | HOF. HOB. and MPA                                                                                                                                |

# Indications For Use:

Revolver Laser Probes, Revolver Illuminated Laser Probes, and Revolver Illuminated Probes:

For use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments.

The Revolver Laser Probes can only be used with a medical laser at operating wavelengths of 500nm to 900nm. The Revolver assemblies containing illumination should be used with the light wavelength range of 425nm to 700nm.

# Device Description:

The Katalyst Revolver Laser Probes are cables made out of one fiber optic (Single-Use), one laser adapter(Reusable), and one handle (Reusable) for surgeon manipulation, metal tubing extending from the handle which penetrates into the surgical site and protective sheath over the fiber.

{4}------------------------------------------------

The fiber for laser transmission is made out of glass, and is restricted for use within the wavelength range of 500nm to 900nm.

The fiber for illumination transmission is made out of either glass or plastic.

The illumination fiber (glass or plastic) is between 100 and 750 microns in size (measure based on the core diameter). The laser fiber (glass) is between 150 and 300 microns in size (measure based on the core diameter).

The tubing is provided in four different variations are referred to as straight, Curved, Flex-Curved, Steerable and Articulating. The configuration is chosen based on the surgeon requirements. The total length of the device is 8-10 feet.

In case of laser and illumination functionalities provided by the same probe, the common protective sheath runs for 1-2 feet, while both branches run for the remainder of the 8-10 feet with each branch having its own protective sheath and each branch ending with its own adapter. The same tubing will then hold within its internal diameter the laser fiber and the illumination fiber.

The Revolver Laser Probes can only be used with a medical laser at an operating wavelength range of 500nm to 900nm. The Revolver Illumination Probes can only be used with the light wavelength range of 425nm to 700nm.

# Predicate Devices:

The Katalyst Revolver Laser Probe, Revolver Illuminated Laser Probe and Illuminated Probe were shown to be substantially equivalent to the previously cleared devices K121187-Katalyst Laser Probe and Illuminated Laser Probe.

# Performance Testing Summary:

The testing for The Revolver Probes was conducted in accordance to All testing was performed in accordance with Attachment C of FDA Guidance on the Content and Organization for a Medical Laser, specifically, the "Specifications to be used in Establishing the Substantial Equivalence" for accessories and other information, is summarized below:

The data presented in Table 6-1 is the output power ratings of illumination and laser surgical machines and the input power capacity of the optical fibers used in the Laser and Illuminated Laser Probes.

|  |  | Table 6-1: Power Specifications for Katalyst Laser and Illuminated Laser Probes |  |  |  |  |  |  |
|--|--|---------------------------------------------------------------------------------|--|--|--|--|--|--|
|  |  |                                                                                 |  |  |  |  |  |  |

| Optical Fiber<br>Type | Typical<br>Maximum<br>Machine<br>Output | Typical<br>Machine<br>Output<br>Setting | Optical Fiber<br>Maximum Input<br>Power Capacity |
|-----------------------|-----------------------------------------|-----------------------------------------|--------------------------------------------------|
| Illumination          | 28 mW                                   | 14 mW                                   | 350 W*                                           |

{5}------------------------------------------------

| Laser                   | 2.5 W | 300 mW | 314 kW |
|-------------------------|-------|--------|--------|
| *Plastic optical fiber. |       |        |        |

Table 6-1A: Power Specifications for Glass Fiber

| Optical Fiber<br>Type | Typical<br>Maximum<br>Machine<br>Output | Typical<br>Machine<br>Output<br>Setting | Optical Fiber<br>Maximum Input<br>Power Capacity |
|-----------------------|-----------------------------------------|-----------------------------------------|--------------------------------------------------|
| Glass                 | 2.5 W                                   | 300 mW                                  | 314 kW                                           |

#### Table 6-1B: Power Specifications for Plastic Fiber

| Optical Fiber<br>Type | Typical<br>Maximum<br>Machine<br>Output | Typical<br>Machine<br>Output<br>Setting | Optical Fiber<br>Maximum Input<br>Power Capacity |
|-----------------------|-----------------------------------------|-----------------------------------------|--------------------------------------------------|
| Plastic               | 28 mW                                   | 14 mW                                   | 350 W                                            |

Additionally, light transmission data was collected for the Katalyst Revolver Illuminated Laser Probes, Revolver Illuminated Probes and the predicate Illuminated Laser Probes.

The results show that the Katalyst Illuminated Probes will perform in the same manner and

|                                         | Light Output in Lumens<br>(average) | % Difference |
|-----------------------------------------|-------------------------------------|--------------|
| Katalyst Surgical Probes                | 26.3                                | 0.38%        |
| Katalyst Revolver<br>Illuminated Probes | 26.2                                |              |

efficacy as the predicate, thus causing no extra risk to the patient and/or the user.

# Biocompatibility testing:

Biocompatibility testing was performed on the probes per ISO 10993. Testing was performed by Toxikon, Inc. The testing was conducted on the materials that are in contact with the patient, being the fibers (glass and plastic), and metal tubing, simulating the actual device. The testing successfully determined the Katalyst Revolver Probes to be biocompatible.

# Substantial Equivalence:

Bench testing performed on this device and compared to the predicate indicates that the Katalyst Revolver Laser Probes, Revolver Illuminated Laser Probes and Revolver Illuminated Probes are substantially equivalent to predicate devices. Bench testing of the Katalyst Revolver Probes was performed in accordance with FDA Guidance on the Content and Organization for a Medical Laser. Biocompatibility testing was performed per ISO 10993. Sterilization development, validation, and control were performed in accordance with ISO 11135-1. Product shelf-life was established in accordance with ASTM Standard #F

{6}------------------------------------------------

1980-09 (2011) and packaging integrity was established in accordance with ASTM F 1929-04 and ASTM F881F88M-09. Optical radiation safety testing was performed in accordance with ISO 15752 and ISO 15004-2.

| FDA File Reference No.                           | 510(k) No. K021696 for Laser | 510(k) No. K050807 for Illumination |
|--------------------------------------------------|------------------------------|-------------------------------------|
| TECHNOLOGICAL CHARACTERISTICS                    | Comparison Result            | Comparison Result                   |
| Indications for Use                              | Identical                    | Identical                           |
| Target Population                                | Identical                    | Identical                           |
| Design                                           | Similar                      | Similar                             |
| Optical Output                                   | Identical                    | Identical                           |
| Materials                                        | Identical                    | Identical                           |
| Performance                                      | Identical                    | Identical                           |
| Sterility                                        | Identical                    | Identical                           |
| Biocompatibility                                 | Identical                    | Identical                           |
| Anatomical Sites                                 | Identical                    | Identical                           |
| Human Factors                                    | Identical                    | Identical                           |
| Energy Used and/or Delivered                     | Similar                      | Similar                             |
| Compatibility with Environment and Other Devices | Identical                    | Identical                           |
| Where Used                                       | Identical                    | Identical                           |
| Standards Met                                    | Identical                    | Identical                           |
| Electrical Safety                                | Identical (not applicable)   | Identical (not applicable)          |
| Thermal Safety                                   | Identical (not applicable)   | Identical (not applicable)          |
| Radiation Safety                                 | Identical (not applicable)   | Identical (not applicable)          |

#### SUMMARY OF EQUIVALENCE

The technological characteristics are identical to the predicate device, in terms of the indication for use, target population, optical output, materials, performance, sterility, biocompatibility, anatomical sites, human factors, compatibility with environment and other devices. The electrical, thermal and radiation safety is identical, because it's not applicable to any of the devices.

The Revolver Probes design is similar to the predicate devices. The Revolver Probes comes four different gauges (20g, 23g, 25g, 27g) and four different tip configuration (i.e., straight, curved, flex-curved, and steerable). The tip configuration is straight or curved for the predicate devices. The difference in tip variations of the Katalyst Revolver Probes and the straight or curved tips of the predicate devices does not affect the safety or effectiveness of the Katalyst Revolver Probes when used as indicated. A surgeon may use any variation of the tip, straight. curved, flex-curved and steerable: in an identical manner; (2) to perform an identical surgical procedure; and (3) to achieve the same surgical result.

The predicate devices come in 20g, 23g, and 25ga in laser probes, and 20ga in Illuminated laser probes. Katalyst Revolver Probes come in 20ga, 23ga, and 27ga, for laser, illumination and laser illumination probes. The addition of the various gauge sizes does not affect the safety and effectiveness of the device. All the micron sizes are within the range of the pervious cleared predicate device.

{7}------------------------------------------------

| Type                                                | Material               | Size<br>(microns) |
|-----------------------------------------------------|------------------------|-------------------|
| Laser                                               | Glass<br>SILICA/SILICA | 150-300           |
| Laser/Illumination<br>(plastic illumination fibers) | Glass<br>SILICA/SILICA | 150-300           |
|                                                     | Plastic<br>Acrylic     | 100-500           |
| Laser/Illumination<br>(glass illumination fibers)   | Glass<br>SILICA/SILICA | 100-300           |
|                                                     | Glass<br>SILICA/SILICA | 100-500           |

#### Table 3-1: Materials and Sizing of Optical Fibers

| Gauge | Laser Fiber   |              | Illumination Fiber |              | Laser illumination fiber |              |
|-------|---------------|--------------|--------------------|--------------|--------------------------|--------------|
|       | Core/         | Attenuation/ | Core/              | Attenuation/ | Core/                    | Attenuation/ |
|       | Cladding      | transmission | Cladding           | transmission | Cladding                 | transmission |
|       | Diameter      |              | Diameter           |              | Diameter                 |              |
| 20 ga | 200 µm ± 4 µm | See Below    | 486 µm ± 6%        | See Below    | 390 µm ± 6%              | See Below    |
|       | 220 µm ± 4 µm |              | 14 µm ± 6%         |              | 10 µm ± 6%               |              |
| 23 ga | 200 µm ± 4 µm | See Below    | 240 µm ± 9%        | See Below    | 240 µm ± 9%              | See Below    |
|       | 220 µm ± 4 µm |              | 10 µm ± 9%         |              | 10 µm ± 9%               |              |
| 25 ga | 150 µm ± 3 µm | See Below    | 240 µm ± 9%        | See Below    | 190 µm ± 9%              | See Below    |
|       | 165 µm ± 3 µm |              | 10 µm ± 9%         |              | 10 µm ± 9%               |              |
| 27 ga | 150 µm ± 3 µm | See Below    | 240 µm ± 9%        | See Below    | 140 µm ± 9%              | See Below    |
|       | 165 µm ± 3 µm |              | 10 µm ± 9%         |              | 10 µm ± 9%               |              |

Image /page/7/Figure/4 description: The image is a graph titled "Typical Attenuation for Glass Fiber". The x-axis is labeled "Wavelength, nm" and ranges from 150 to 1250. The left y-axis is labeled "UV Attenuation, dB/m" and ranges from 0.0 to 2.5, while the right y-axis is labeled "VIS-IR Attenuation, dB/km" and ranges from 0 to 250. There are two curves on the graph, one blue and one red, representing the attenuation of UV and VIS-IR light, respectively.

Attenuation for Plastic Fiber (Katalyst Fiber is Optical Grade)

{8}------------------------------------------------

Image /page/8/Figure/1 description: The image is a graph that shows the transmission loss in dB/m as a function of wavelength in nm. The x-axis represents the wavelength, ranging from 400 to 1100 nm, and is divided into three regions: Ultra Violet, Visible Band, and Infra-red. The y-axis represents the transmission loss, ranging from 0.1 to 30.0 dB/m. There are two curves on the graph, labeled "INDUSTRIAL" and "OPTICAL", which show the transmission loss for different materials.

Furthermore, the predicate devices are single-use devices. The revolver probes are assembled with single-use, and reusable components. Revolver probes comprise three components, handle, fiber and an adapter. The handle and adapter are reusable and are provided nonsterile. The replaceable fiber is a single-use component, and is provided sterile. This change from the predicate device does not have any impact on the safety and effectiveness of the device. The predicate devices and The Revolver Probes are intended as accessories to devices used to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam and to provide illumination during vitroretinal surgery. The Revolver Probes are substantially equivalent to the previously cleared devices.

#### Conclusion

The Katalyst Revolver Probes were shown to be substantially equivalent to previously cleared devices with respect to intended use, indications for use, technological characteristics, performance characteristics, and biocompatibility.

---

**Source:** [https://fda-staging.innolitics.com/device/K140362](https://fda-staging.innolitics.com/device/K140362)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
