K140358 · Dallen Medical, Inc. · JDR · Jun 2, 2014 · Orthopedic
Device Facts
Record ID
K140358
Device Name
COMPRESSYN STAPLE
Applicant
Dallen Medical, Inc.
Product Code
JDR · Orthopedic
Decision Date
Jun 2, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Compressyn™ Staple is intended for: 1) hand and foot bone fragment and osteotomy fixation and joint arthrodesis, 2) fixation of proximal tibial metaphysis osteotomy, 3) adjunctive fixation of small bone fragments. These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremity; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in the flat bones such as the pelvis, scapula and sternum.
Device Story
Compressyn™ Staple is a sterile, single-use bone fixation system; consists of stainless steel staple preloaded on a handheld mechanical delivery device (ABS and polycarbonate). Used by surgeons for bone fragment stabilization, osteotomy fixation, and joint arthrodesis. Procedure involves pre-drilling holes in tissue segments; staple is deployed via delivery device to provide compression. Barbs on staple prevent back-out. System includes drill bit and drill guide. Provides stabilized fixation to facilitate bone healing.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Stainless steel staple; handheld delivery device constructed of ABS and polycarbonate. Mechanical delivery system. Provided sterile in disposable tray. No software or electronic components.
Indications for Use
Indicated for bone fragment/osteotomy fixation and joint arthrodesis in hand/foot, proximal tibial metaphysis, and adjunctive fixation of small bone fragments in long bones (femur, fibula, tibia, humerus, ulna, radius, clavicle, ribs) and flat bones (pelvis, scapula, sternum).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K140358
510(k) Summary
llen
MEDICAL
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### GENERAL INFORMATION
| APPLICANT: | Dallen Medical, Inc.<br>1046 Calle Recodo, Suite G<br>San Clemente, CA 92673<br>(949) 218-0030<br>(949) 218-0040 Fax |
|------------------------|----------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Al Memmolo<br>Chief Operating Officer |
| DATE PREPARED: | February 11, 2014 |
| DEVICE DESCRIPTION: | |
| TRADE NAME: | Compressyn™ Staple |
| GENERIC/COMMON NAME | Fixation Staple |
| CLASSIFICATION NAME: | Single/multiple component metallic bone fixation appliances<br>and accessories, CFR 888.3030 (code JDR) |
| DEVICE CLASSIFICATION: | Class II |
| PREDICATE DEVICES: | Dallen Medical, Inc. Compressyn™ Staple (K122871)<br>Dallen Medical, Inc. Compressyn™ Staple (K132875) |
#### Product Description:
The Compressyn™ Staple is a fully integrated system that consists of a stainless steel staple preloaded on a handheld ABS and polycarbonate mechanical delivery device. The staple is offered in several sizes with barbs to prevent back out. It is a compression staple fixation device that is placed into two tissue segments using pre-drilled holes to provide stabilized fixation. The Compressyn™ Staple is provided sterile in a disposable tray with a drill bit and drill guide.
{1}------------------------------------------------
## 510(k) Summary
Image /page/1/Picture/4 description: The image shows the logo for Dallen Medical. The word "dallen" is written in a simple, sans-serif font, with the "d" being significantly taller than the other letters. Below "dallen" is the word "MEDICAL" in a blocky, sans-serif font, with each letter being capitalized. The overall design is clean and modern.
## Indications for Use:
The Compressvn™ Staple is intended for: 1) hand and foot bone fragment and osteotomy fixation and joint arthrodesis, 2) fixation of proximal tibial metaphysis osteotomy, 3) adjunctive fixation of small bone fragments. These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremity; the humerus. ulna or radius in the upper extremities; the clavicle and ribs; and in the flat bones such as the pelvis, scapula and sternum.
#### Technical Characteristics and Performance Data:
The Compressyn™ Staple is identical to the predicate and therefore has the same physical and technical characteristics. The preloaded mechanical delivery device was tested for impact force, staple delivery, compressive force and bond strength, and was shown to be equivalent to the predicate.
#### Basis for Determination of Substantial Equivalence:
Upon reviewing the documentation provided in this submission and comparing intended use, principle of operation, performance data, and overall technological characteristics, the Compressyn™ Staple is determined by Dallen Medical to be substantially equivalent to existing legally marketed devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features the department's emblem, which is a stylized representation of a human figure with outstretched arms, symbolizing care and protection. The emblem is accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the emblem.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
June 2. 2014
Dallen Medical, Incorporated Mr. Al Memmolo Chief Operating Officer 1046 Calle Recodo. Suite G San Clemente, California 92673
Re: K140358
1140558
Trade/Device Name: Compressyn "M Staple Regulation Number: 21 CFR 888.3030 Regulation Number: 21 CFK 866.5050
Regulation Name: Single/Multiple Component Metallic Bonc Fixation Appliances and Accessories Regulatory Class: Class II Product Code: JDR Dated: April 24, 2014 Received: April 25, 2014
Dear Mr. Memmolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 3 IQ(x) premilated to equivalent (for the indications
referenced above and have determined the devices marked in interstate referenced above and have deemined the usedicate devices marketed in interstate
for use stated in the enclosure) to legally marketed previces marketed in interstate for use stated in the enclosure) to legily market accrises Device Amendments, or to
commerce prior to May 28, 1976. the ensections of the Medical Device Amends, Drug, commerce prior to May 28, 1976. the enactinent and of the Federal Food, Drug,
devices that have been reclassified in accordance with the provisions of the Federal (PMA) devices that have been reclassified in accordance will the proval application (PMA).
and Cosmetic Act (Act (Act (Act require approval of a premarket approval application of t and Cosmetic Act (Act) that do not require applied on the general controls of the Act. The Act. The Act. The Act. The Act. The Act. The Act. The Act. The You may, therefore, market the device, subject to the may begistration, listing of
general controls provisions of the Act include requirements for and general controls provisions of the Accurentiality tor manust misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability
adulteration. Please note: CDRH docs not evaluate information related in adulteration. Please note: CDRH does not evaluate information roads to 17.
warranties. We remind you, however, that device labeling must be truthful and not misleading.
lf your device is classified (see above) into either class II (Special Controls) or class III (PMA). it
ed be If your device is classified (see above) mio entire class if (spicent Gention) your device can be
may be subject to additional controls. Existing major regulations affective may be subject to additional controls. Exismig major regalations and one of the may
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may found in the Code of Federal Regulations. This 21, Parts 30
publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
ts of the Active in the since that your device complies with other requirement Please be advised that FDA's issuance of a substintial equirements of the Act
that FDA has made a determination that your device complies with other comply that FDA has made a determination that your develevels. When evenes. You must comply
or any Federal statutes and regulations administered by other Federal and listing (21 CFR or any Federal statues and regulations administered by oncertains and listing (2) CFR Part
with all the Act's requirements, including, but not limited to: registration and l with all the Act's requirements, including, but not millied to: registration and the one of the Access of the
{3}------------------------------------------------
## Page 2 - Mr. Al Memmolo
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the auvoise ovents) (21 CFR Part 820); and if applicable, the electronic product quality Systems (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your dovies on our lebening on Consumer Assistance at its toll-Comace the Division of S041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note highww.lda.g0v/McdicalDevices/1v/ress/100 by reference to premarket notification" (21 CFR Part the regulation entitled, "Misoranoing of receivents under the MDR regulation (21 CFR Part 803), please go to
CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/deficult.html/home ntlp.//www.ida.governmentics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general Information on your responsibility of the more of its toll-free number
Division of Small Manufacturers, International and Consumer Assistance at Division of Sinan Manazao, From or at its Internet address
(800) 638 204 l of (301) 776-7100 or at hs internet and of the street on the many of the street of the man
Sincerely yours,
# Lori A. Wiggins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K140358
Section 4
# Indications for Use Statement
Page I of 1
K140358 510(k) Number (if known):
Device Name: Compressyn™ Staple
Indications for Use:
The Compressyn™ Staple is intended for: 1) hand and foot bone fragment and osteotomy 3) I he Compressyn" Stapie is intention of proximal tibial metaphysis osted in long fixation and joint arthrodests, 2) fradion of proximal size be located in long bones
adjunctive fixation of small bone fragments. These fragments adjunctive fixation of small bolle naghiems. The something may as the pelvis in the such as the femur, Ilbula and tible in the flower extremity the halterials, and the pelvis, scapula and sternum.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) OR
Over-The-Counter-Use _________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Casey L. Hanley, Ph.D. |
|----------------------------------|
| ________________________________ |
| Division of Orthopedic Devices |
Page 1/1
Dallen Medical, Inc. Danell Neches Staple, Preloaded Sterile System Premarket Notification 510(k)
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.