K140342 · Mindways Software, Inc. · KGI · Aug 29, 2014 · Radiology
Device Facts
Record ID
K140342
Device Name
QCT PRO ASYNCHRONOUS CALIBRATION MODULE, CLINIQCT
Applicant
Mindways Software, Inc.
Product Code
KGI · Radiology
Decision Date
Aug 29, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1170
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140342 · Aug 29, 2014
QCT PRO ASYNCHRONOUS CALIBRATION MODULE, CLINIQCT
Mindways Software, Inc.
Retrospective clinical CT scans
Retrospective analysis of existing clinical CT scans was used to evaluate the performance of the CliniQCT asynchronous calibration module compared to the predicate QCT Pro device.
Retrospective analysis; Bone mineral density; Clinical CT scans
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective data analysis; Retrospective analysis
Patients who underwent CT scans for other clinical purposes
QCT Pro (predicate device)
Bone mineral content (BMC) and bone mineral density (BMD) estimation accuracy and precision
Indications for Use
The QCT Pro Asynchronous Calibration Module is intended to provide an alternative method for calibrating QCT data sets that are intended for analysis with QCT Pro Spine and Hip application modules. Installation of the QCT Pro Asynchronous Calibration Module does not alter the clinical indications for use of the QCT Pro Spine and Hip application modules. It does, however, provide an additional means of obtaining calibration data that does not require the simultaneous scanning of a patient with a CT calibration standard as is required when using the QCT Pro Spine and Hip application modules without the installation of the QCT Pro Asynchronous Calibration Module. Thus the QCT Pro Asynchronous Calibration Module facilitates: (1) retrospective assessment of bone density from CT scans acquired for other purposes, (2) assessment of bone density in conjunction with another medically appropriate procedure involving CT scans of the anatomical regions where estimating bone density is prescribed, and (3) assessment of bone density without a phantom as an independent measurement procedure.
Device Story
CliniQCT (QCT Pro Asynchronous Calibration Module) extends existing QCT Pro bone densitometry software; enables BMD/BMC estimation from CT images without requiring a calibration phantom to be scanned simultaneously with the patient. Inputs: patient CT images and separate phantom calibration scans acquired on the same scanner. Operation: phantom scans (acquired before or after patient scan) provide calibration data for QCT analysis. Output: BMD/BMC values for spine and hip. Used in clinical settings by physicians to assess bone density; facilitates retrospective analysis of existing CT scans or combined-purpose imaging. Benefits: reduces patient radiation exposure by eliminating need for dedicated QCT scans; allows opportunistic bone density screening from diagnostic CTs.
Clinical Evidence
Performance evaluated via phantom testing, retrospective data analysis, and prospective patient studies. Results demonstrated non-inferiority in precision and accuracy relative to synchronous calibration predicate devices. Accuracy is reported as unbiased. In vitro precision: 1.4 mg/cm3 (volume) and 0.007 g/cm2 (area). In vivo precision: spine up to 1%, total hip 1.1%, femoral neck 1.2%.
Technological Characteristics
Software-based bone densitometer; utilizes Quantitative Computed Tomography (QCT) principles. Calibrates using aqueous K2HPO4 standard. Operates on standard medical CT scanner images. Asynchronous calibration allows phantom scans to be acquired separately from patient scans. Connectivity: integrates with existing QCT Pro software modules.
Indications for Use
Indicated for patients requiring bone mineral density (BMD) and bone mineral content (BMC) assessment of the spine and proximal femur via Quantitative Computed Tomography (QCT).
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Mindways QCT Pro Bone Mineral Density Analysis Software (K894854)
Mindways QCT Pro CTXA Hip Bone Mineral Densitometer (K002113)
QCT Pro CTXA Hip Extended Reference Data (K030330)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or a cape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2014
Mindways Software, Inc. % Keenan Brown, Ph. D. President/ Director of Research and Development 3001 S Lamar Blvd. Suite 302 AUSTIN, TX 78704
Re: K140342
Trade/Device Name: OCT Pro Asynchronous Calibration Module, CliniOCT Regulation Number: 21 CFR 892.1170 Regulation Name: Bone Mineral Densitometer Regulatory Class: II Product Code: KGI Dated: July 25, 2014 Received: July 31, 2014
Dear Dr. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140342
#### Device Name
QCT Pro Asynchronous Calibration Module, CliniQCT
### Indications for Use (Describe)
The QCT Pro Asynchronous Calibration Module is intended to provide an alternative method for calibrating QCT data sets that are intended for analysis with QCT Pro Spine and Hip application modules. Installation of the QCT Pro Asynchronous Calibration Module does not alter the clinical indications for use of the QCT Pro Spine and Hip application modules. It does, however, provide an additional means of obtaining calibration data that does not require the simultaneous scanning of a patient with a CT calibration standard as is required when using the QCT Pro Spine and Hip application modules without the installation of the QCT Pro Asynchronous Calibration Module. Thus the QCT Pro Asynchronous Calibration Module facilitates: (1) retrospective assessment of bone density from CT scans acquired for other purposes, (2) assessment of bone density in conjunction with another medically appropriate procedure involving CT scans of the anatomical regions where estimating bone density is prescribed, and (3) assessment of bone density without a phantom as an independent measurement procedure.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Mindways Software, Inc. 3001 S Lamar Blvd, Suite 302 Austin, TX 78704, USA
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92(c).
| Contact Information: | J. Keenan Brown, Ph.D. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| | President and Director of Research and Development |
| | Mindways Software, Inc |
| | 3001 S Lamar Blvd, Ste 302 |
| | Austin, TX 78704 |
| | Phone:512-912-0871 |
| | Fax: 512-912-0871 |
| | Email: keenan@qct.com |
| Date: | July 24, 2014 |
| Model Number: | CLINIQCT |
| Device/Trade Name: | QCT Pro Asynchronous Calibration Module, CliniQCT, QCT Pro<br>CliniQCT |
| Common/Usual Name: | Bone Mineral Densitometer |
| Classification Name: | Bone Densitometer, 21 CFR 892.1170, Class II |
| Predicate Devices: | K894854: Mindways QCT Pro Bone Mineral Density Analysis<br>Software<br>Intended Use: Estimate bone mineral density within the spine. |
{4}------------------------------------------------
# K002113: Mindways QCT Pro CTXA Hip Bone Mineral Densitometer Intended Use: Estimate bone mineral density at the proximal femur, T-Score calculation, aid in determining fracture risk.
K030330: OCT Pro CTXA Hip Extended Reference Data Intended Use: Reference data for Z-Score calculation between ages 20 and 80.
### Device Description
The OCT Pro Asynchronous Calibration Module is intended to extend the capabilities of OCT Pro bone mineral densitometer products (K894854. K002113. K030330) currently marketed by Mindways to the measurement of bone mineral content (BMC) and bone mineral density (BMD) from patient-specific CT images acquired without the simultaneous use of a CT calibration phantom. The QCT Pro bone mineral densitometer products distributed with or supplemented with the QCT Pro Asynchronous Calibration Module will be marketed under the name QCT Pro CliniQCT or more simply CliniQCT. Mineral calibration with the asynchronous calibration method is obtained from phantom measurements acquired on the same CT scanner, operated in a substantially similar mode used to acquire a patient CT scan. The phantom measurements used for calibration purposes may be acquired before or after a patient CT scan to which they are to be applied. The asynchronous calibration mode facilitates: (1) retrospective assessment of bone density from CT scans acquired for other purposes, (2) assessment of bone density in conjunction with another medically appropriate procedure involving CT scans of the anatomical regions where estimating bone density is prescribed, and (3) assessment of bone density without a phantom as an independent measurement procedure. Retrospective bone density analysis provides bone density information without exposing a patient to additional ionizing radiation while combining CT studies results in an overall reduction in patient exposure to ionizing radiation relative to the expected radiation dose associated with performing the studies separately.
## Intended Use
The QCT Pro Asynchronous Calibration Module is intended to provide an alternative method for calibrating QCT data sets that are intended for analysis with currently marketed QCT Pro Spine and Hip application modules. Installation of the OCT Pro Asynchronous Calibration Module does not alter the intended use of the QCT Pro Spine and Hip application modules, however, it facilitates an additional means of providing calibration data to QCT Pro that does not require the
{5}------------------------------------------------
simultaneous scanning of a patient with a CT calibration standard as is required when using the QCT Pro Spine and Hip application modules without the installation of the QCT Pro Asynchronous Calibration Module. Calibration data necessary for use with the QCT Pro Asynchronous Calibration Module is obtained from a standard OCT Pro QA phantom CT scan performed on the same CT scanner used to acquire patient CT images intended for analysis within OCT Pro operating in conjunction with the OCT Pro Asynchronous Calibration Module. The phantom CT scans used to obtain calibration data may be obtained either prior to or subsequent to patient CT scans.
## Summary of Technological Characteristics and Comparison with Predicate Devices
QCT Pro when operated with the QCT Pro Asynchronous Calibration Module, collectively marketed as CliniQCT. provides estimates of bone mineral content (BMC) and bone mineral density (BMD) values similar to those obtained from the predicate OCT Pro bone densitometer devices (K894854. K002113. K030330) for regions of interest in the spine and proximal femur. With the exception of the use of an asynchronous calibration method, the new device has the same technological characteristics as its predicate QCT Pro devices that calibrate bone mineral density with reference to a CT calibration phantom scanned simultaneously with a patient. BMD estimates derived with the asynchronous calibration method are suitable for comparison to the same reference data sets as are used with the QCT Pro predicate devices. Such reference data comparisons provide an aid to the physician in identifying patients with abnormally high or low bone mineral density.
| Device Characteristic | CliniQCT<br>Installed and Used in<br>Conjunction with<br>QCT Pro | QCT Pro<br>(K894854, K002113,<br>K030330) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device provides estimates of bone mineral content (BMC). | ✓ | ✓ |
| Device provides estimates of bone mineral density (BMD). | ✓ | ✓ |
| BMC and BMD estimates are provided for the spine and hip. | ✓ | ✓ |
| BMC and BMD estimates are derived using the principals of<br>Quantitative Computed Tomography (QCT) | ✓ | ✓ |
| BMC and BMD estimates are provided standardized to an<br>aqueous K2HPO4 bone density standard. | ✓ | ✓ |
| | | |
| BMC and BMD estimates are derived relative to a phantom<br>standard scanned:<br><br>with the patient<br><br>separate from the patient, either before or after patient<br>scan | both scan modes<br>accessible with<br>installation of<br>CliniQCT<br><br><br> | <br> |
| Precision, ( <i>in vitro</i> ) relative to calibration reference standard. | Non-inferior to QCT<br>Pro | Volume Density: 1.4<br>mg/cm3 (0.7% at a<br>nominal volume density<br>of 200 mg/cm3)<br><br>Area Density: 0.007<br>g/cm2 (0.7% at a<br>nominal area density of<br>1.0 g/cm2) |
| Precision, ( <i>in vivo</i> ) patient. | Non-inferior to QCT<br>Pro | Spine: Up to 1%<br>Total Hip ROI:<br>0.011 g/cm2 (1.1% at<br>1.0 g/cm2)<br><br>Femoral Neck:<br>0.012 g/cm2 (1.2% at<br>1.0 g/cm2) |
| Accuracy, relative to calibration reference standard | Unbiased | Unbiased |
| Radiation dose associated with examination is within currently<br>accepted guidelines regarding radiation health risks associated<br>with common radiologic diagnostic procedures.<br>Note, <i>QCT Pro</i> does not directly control the delivery of<br>radiation to a patient, but rather makes use of images acquired<br>from medical CT scanners approved for diagnostic, medical x-<br>ray, CT imaging purposes. | | |
| Device can be used to provide retrospective measurements<br>from CT scans acquired for other purposes without exposing<br>the patient to additional ionizing radiation beyond that required<br>for the other purposes. | | |
| Device can be used to provide BMD and BMC measurements<br>from CT scans acquired for multiple purposes (e.g., dual use of<br>images acquired for CT colonography and bone density<br>screening) without exposing the patient to additional ionizing<br>radiation beyond that required for the other, non-QCT,<br>purpose(s). | | |
{6}------------------------------------------------
{7}------------------------------------------------
## Device Testing
The design and performance of CliniQCT installed with QCT Pro was evaluated and compared to QCT Pro using retrospective data analysis, phantom testing and prospective patient studies. Software validation was performed using standard techniques.
### Summary
Based on the similarities in analytical methods, calibration standards and performance testing, it is concluded that CliniQCT used in conjunction with QCT Pro is substantially equivalent to predicate devices currently marketed under the Federal Food, Drug and Cosmetic Act. No new safety or effectiveness issues are raised by CliniQCT.
Signature
J. Keenan Brown, PhD Director of Research and Development Mindways Software, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.