K140331 · Orthalign, Inc. · OLO · Jun 10, 2014 · Neurology
Device Facts
Record ID
K140331
Device Name
ORTHALIGN PLUS SYSTEM
Applicant
Orthalign, Inc.
Product Code
OLO · Neurology
Decision Date
Jun 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The OrthAlign Plus System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior. Example orthopedic surgical procedures include but are not limited to: Total Knee Arthroplasty . Total Hip Arthroplasty: Anterior/Posterior
Device Story
OrthAlign Plus is a non-invasive, computer-assisted surgical navigation system for knee and hip arthroplasty. It uses a palm-sized computer module and reference sensor to track angular and positional changes in a triaxial format. During surgery, the physician registers anatomical landmarks; the device calculates mechanical axes and displays instrument positioning relative to these axes on a screen. In knee procedures, it assists with femoral/tibial mechanical axis alignment and cutting block orientation. In hip procedures, it assists with acetabular shell inclination/anteversion and measures intraoperative changes in leg length and offset. The system provides real-time graphical and numerical feedback, allowing surgeons to adjust implant placement intraoperatively. This improves surgical accuracy compared to manual estimation, potentially reducing leg length discrepancies and improving implant alignment, which may lead to better patient outcomes.
Clinical Evidence
Bench testing included software verification, hardware validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-1), and biocompatibility (ISO 10993-1). Simulated use testing was conducted in a cadaver model (n=5 hips, 18 data points) to validate accuracy of leg length and offset measurements against radiographic gold standards. Results confirmed system accuracy and performance.
Technological Characteristics
System includes single-use computer module and reusable reference sensor/jigs. Utilizes triaxial sensor technology to convert spatial inputs into graphical/numerical displays. Connectivity is standalone. Biocompatibility per ISO 10993-1. Sterilization validated for reusable components. Software performs spatial coordinate calculations and automated measurement of anatomical axes.
Indications for Use
Indicated for patients undergoing total knee or total hip arthroplasty procedures requiring stereotactic orthopedic surgical navigation to assist in determining reference alignment axes and positioning of implants. Aids in controlling leg length and offset discrepancies during posterior approach total hip arthroplasty.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
### K140331
#### 3. 510(K) SUMMARY
DATE
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).
OrthAlign, Inc. APPLICANT 120 Columbia Suite 500 Aliso Viejo, CA 92656 Tel: (949) 715-2424 Fax: (949) 831-9500 David Vancelette
May 15, 2014
OFFICIAL CORRESPONDENT
OrthAlign. Inc. 120 Columbia. Suite 500 Aliso Viejo, CA 92656 dvancelette@orthalign.com Tel: (858) 692-0335 Fax: (949) 831-9500
OrthAlign Plus® System TRADE NAME COMMON NAME Stereotaxic Instrument
DEVICE Class II, 21 CFR §882.4560 CLASSIFICATION
PRODUCT CODES OLO: Orthopedic Stereotaxic Instrument
OrthAlign Plus® System (K130387) PREDICATE Navitrack System - S&N Image Free Hip (K041369) DEVICES DASH Hip (K110021)
Traditional 510(k), K140331. The subject device is a SUBMISSION TYPE modification to the previously cleared OrthAlign Plus® System (K130387).
#### SUBSTANTIALLY EQUIVALENT TO:
The OrthAlign Plus System is substantially equivalent to the previously cleared OrthAlign Plus® System (K130387), Navitrack System - S&N Image Free Hip (K041369) and DASH Hip (K110021).
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The OrthAlign Plus® System is an innovative non-invasive computer assisted surgical
Page 1 of 5
Response to Questions i
# JUN 1 0 2014
L-1514
{1}------------------------------------------------
navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format.
The current standard of care for knee arthroplasty procedures has the physician estimating these changes either by visual observation and mechanical guides with tactile feedback or with the assistance of computer assisted surgery devices.
The current standard of care for hip arthroplasty procedures has the physician estimating angular orientation of the acetabular shell either by visual observation and mechanical guides or with the assistance of computer assisted surgery devices.
The current standard of care for measuring the change in leg length and offset during total hip arthroplasty has the physician estimating the position of the femoral bony cut, targeting a location determined from preoperative templating, and/or using mechanical guides or anatomic comparisons to the contralateral leg.
The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.
In knee arthroplasty procedures, the device assists the surgeon in:
- Establishing the mechanical axis of the femur, determining the varus/valgus . angle and the flexion/extension angle of the cutting block relative to the femur.
- Establishing the mechanical axis of the tibia, determining the varus/valgus . angle and the posterior slope angle of the cutting block relative to the tibia.
In hip arthroplasty procedures, the device assists the surgeon in:
- . Establishing the orientation of the anterior pelvic plane and determining the inclination angle and the anteversion angle of the shell impactor relative to the anterior pelvic plane.
- Measuring the intraoperative change in leg length and offset (posterior . approach).
The OrthAlign Plus® System comprises a single use computer module and reusable instrumentation.
The OrthAlign Plus® System is usable for a total knee arthroplasty or total hip arthroplasty procedure. The System includes two optional configurations: the KneeAlign® 3 System usable for total knee arthroplasty only, and the HipAlign® System usable for total hip arthroplasty only. The OrthAlign Plus® System includes the singleuse OrthAlign Plus® Unit, a KneeAlign® 3 Instrument Set and a HipAlign® Instrument
Page 2 of 5
{2}------------------------------------------------
Set. The optional configurations include a modified version of the OrthAlign Plus® Unit and only one of the Instrument Sets. Indications for Use for each optional configuration are limited to the applicable orthopedic procedure.
### INDICATIONS FOR USE:
The OrthAlign Plus® System has the same indications for use as the previously cleared OrthAlign Plus® System (K130387). Additional functionality has been added to the predicate device to enable the measurement of intraoperative change in leg length and offset. Also, Indications for Use are common to the Navitrack System- S&N Image Free Hip (K041369) and DASH Hip (K110021). Thus, the Indications for Use are as follows:
## OrthAlign Plus® System:
The OrthAlign Plus System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.
Example orthopedic surgical procedures include but are not limited to:
- Total Knee Arthroplasty .
- Total Hip Arthroplasty: Anterior/Posterior .
### KneeAlign® 3 System:
The KneeAlign® 3 System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The KneeAlign® System facilitates the accurate positioning of implants, relative to these alignment axes.
Example orthopedic surgical procedures include but are not limited to:
- Total Knee Arthroplasty .
### HipAlign® System:
The HipAlign® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The HipAlign® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.
f
Example orthopedic surgical procedures include but are not limited to:
/
- Total Hip Arthroplasty: Anterior/Posterior ●
Page 3 of 5
{3}------------------------------------------------
### TECHNICAL CHARACTERISTICS:
The OrthAlign Plus System comprises a single use computer module, a reusable reference sensor, a reusable femoral jig, a reusable tibial jig, a reusable posterior hip jig and a reusable anterior hip jig. The device utilizes algorithms to convert sensor outputs into spatial coordinates, providing graphical and numerical representation of instruments and anatomy on the user display screen.
The optional KneeAlign® 3 and HipAlign® system configurations also comprise the single use computer module, reusable reference sensor and applicable reusable jigs. They utilize the same algorithms, sensor conversions, graphical and numerical representations and surgical techniques as the OrthAlign Plus® System.
The terms "OrthAlign Plus® System" and "OrthAlign Plus® Unit" are used henceforth in this submission to refer to all three system configurations and all three single-use computer modules, respectively, except where all three configurations are cited and differentiated.
#### PERFORMANCE DATA:
Device performance testing confirms that the OrthAlign Plus® System can be used according to its intended use. The OrthAlign Plus® System has been verified and validated according to OrthAlign's procedures for product design and development. Performance testing included:
- Software verification and validation to ensure the integrity of the code and . functionality and reliability of the software in various use sequences.
- System hardware verification/validation testing to ensure the electronics . hardware meets its mechanical requirements.
- Electrical safety testing to IEC 60601-1:2005 + Corr. 1 (2006) + Corr. 2 . (2007) + AM1 (2012) or IEC 60601-1:2012.
- Electromagnetic compatibility testing to IEC 60601-1-1: 2007. .
- Instrumentation cleaning, sterilization and shipping validations for the . specified processes.
- Navigation device sterilization, packaging, shelf life, environmental . conditions and shipping validations for the specified ranges of conditions involved in each process. (Summary data for the identical predicate device is referenced for some validations.)
- Reference Sensor verification/validation testing to ensure the unit meets . requirements for the use environment and stresses from mechanical interactions, cleaning and the sterilization processes.
- System components biocompatibility assessment per ISO 10993-1 (2009). .
- Customer requirements validation with an advising surgeon to validate the . system meets design input requirements for its functions in a simulated use
{4}------------------------------------------------
environment.
- System accuracy testing: bench testing with mechanical fixtures and foam . models to verify leg length and offset and updated scale reader measurement accuracy.
- Simulated use testing in cadaver to validate system accuracy vs. the gold . standard of radiographic measurement in a simulated use environment (see discussion below).
For simulated use testing, a prospective cadaver validation was done in a simulated operating room environment with a surgeon conducting the procedures.
Measurement of the intraoperative change in leg length and offset was validated for the posterior approach with 18 data points (on 5 hips), using radiographic evaluation of the leg length and offset changes.
This testing regime demonstrates that the subject device is as safe, as effective, and performs as well as or better than the predicate devices. This testing regime demonstrates that the subject device is substantially equivalent to the legally marketed predicate devices. for its intended use in facilitating the accurate measurement of intraoperative change in leg length and offset in THA, posterior approach.
The information provided by OrthAlign in this 510(k) application confirms that the OrthAlign Plus® System is substantially equivalent to predicate devices such as the OrthAlign Plus® System (K130387), Navitrack System- S&N Image Free Hip (K041369) and DASH Hip (K110021).
#### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
A technological comparison and bench and cadaver testing demonstrate the substantial equivalence of the OrthAlign Plus® System to the predicate devices.
The subject device is identical to the predicate OrthAlign Plus® System (K130387), with the following exceptions:
- The subject device allows for the measurement of intraoperative changes in . leg length and joint offset.
- The subject device automates the user input of the axial position of the probe. .
The subject device is substantially equivalent to the predicate device DASH Hip (K) 10021) regarding the function of measurement of intraoperative changes in leg length and offset.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2014
OrthAlign, Incorporated Mr. David Vancelette Director. Ouality Assurance and Regulatory Affairs 120 Columbia, Suite 500 Aliso Viejo, California 92656
Re: K140331
Trade/Device Name: OrthAlign Plus® System, KneeAlign® 3 System, HipAlign® System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: May 15, 2014 Received: May 16, 2014
Dear Mr. Vancelette:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
### Page 2 - Mr. David Vancelette
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
· : Sincerely yours,
# Lori A. Wiggins
#### for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### 2. INDICATIONS FOR USE STATEMENT
### 2.1. ORTHALIGN PLUS® SYSTEM
510(k) Number (if known): K140331
Device Name: OrthAlign Plus® System
Indications for Use:
The OrthAlign Plus System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.
Example orthopedic surgical procedures include but are not limited to:
- Total Knee Arthroplasty .
- . Total Hip Arthroplasty: Anterior/Posterior
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Device
Page 1 of 3
Response to Questions K140331
{8}------------------------------------------------
# 2.2. KNEEALIGN® 3 SYSTEM
510(k) Number (if known): K140331
Device Name: KneeAlign® 3 System
Indications for Use:
The KneeAlign® 3 System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The KneeAlign® 3 System facilitates the accurate positioning of implants, relative to these alignment axes.
Example orthopedic surgical procedures include but are not limited to:
- Total Knee Arthroplasty .
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Casey L. Hanley, Ph.D.**
Division of Ortho
Page 2 of 3
{9}------------------------------------------------
# 2.3. HIPALIGN® SYSTEM
510(k) Number (if known): K140331
Device Name: HipAlign® System
Indications for Use:
The HipAlign® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The HipAlign® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.
: .
Example orthopedic surgical procedures include but are not limited to:
- Total Hip Arthroplasty: Anterior / Posterior .
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.