NICHD multi-center contraceptive study (historical control)
Historical control data from a previously conducted multi-center study was used to compare the contraceptive effectiveness and safety of the Caya diaphragm against the predicate ORTHO ALL-FLEX diaphragm.
The Caya® contoured diaphragm is indicated for the prevention of pregnancy in women who elect to use diaphragms as a method of contraception. It is recommended for use with a contraceptive gel.
Device Story
Caya contoured diaphragm is a single-sized, reusable, silicone elastomer cup with a nylon spring-encapsulated rim; designed to cover cervix and act as mechanical barrier to sperm. Device features grip dimples and finger removal dome to assist user insertion/removal. Patient folds device by squeezing rim sides to insert into vagina; spring creates seal against vaginal wall. Device serves as repository for contraceptive gel. Used by women for contraception; self-operated. Provides non-hormonal pregnancy prevention. Clinical effectiveness and safety profile comparable to traditional multi-sized diaphragms.
Clinical Evidence
Pivotal study (n=450 couples) compared single-sized SILCS (Caya) diaphragm with contraceptive gels (BufferGel or N-9) against historical control (ORTHO ALL-FLEX). Primary endpoints: pregnancy probability, safety, and ease of use. 6-month typical-use cumulative pregnancy probability was 10.4 per 100 women (95% CI 6.9-14.0). 6-cycle perfect-use probability was 7.9 per 100 women (95% CI 1.7-14.0). Results showed non-inferiority to predicate device. Safety profile similar to predicate.
Technological Characteristics
Medical-grade silicone elastomer cup; nylon spring-encapsulated rim. Anatomically contoured shape; single-size (fits 60-85 mm range). Mechanical barrier principle. Conforms to ISO 8009:2004(E). Reusable. No software or electronic components.
Indications for Use
Indicated for prevention of pregnancy in women electing to use diaphragms as contraception. Recommended for use with contraceptive gel.
Regulatory Classification
Identification
A contraceptive diaphragm is a closely fitting membrane placed between the posterior aspect of the pubic bone and the posterior vaginal fornix. The device covers the cervix completely and is used with a spermicide to prevent pregnancy. This generic type of device may include an introducer.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26, 2014
Kessel medintin GmbH Martin Kessel CEO Kelsterbacher Str. 28. 64546 Moerfelden-Walldorf, Frankfurt Germany
Re: K140305
> Trade/Device Name: Caya® contoured diaphragm Regulation Number: 21 CFR§ 884.5350 Regulation Name: Contraceptive diaphragm and accessories Regulatory Class: II Product Code: HDW Dated: July 24, 2014 Received: July 28, 2014
Dear Martin Kessel,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 4. Indications for Use Statement
#### 510(k) Number (if known):
Device Name: Caya® contoured diaphragm
#### Indications for Use:
The Caya® contoured diaphragm is indicated for the prevention of pregnancy in women who elect to use diaphragms as a method of contraception. It is recommended for use with a contraceptive gel.
Prescription Use _XXX_ Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) Summary-Revised
Submitter Name
Kessel medintim GmbH
#### Submitter Address
Kelsterbacher Str. 28, 64546, Moerfelden-Walldorf, Frankfurt, Germany
#### Establishment Registration Number
Kessel medintim GmbH will register its establishment and list the Caya " contoured diaphragm prior to initial importation of any devices intended for commercial distribution within the United States.
Phone Number +49 6105 203 728
Fax Number +49 6105 455 901
Contact Person Martin Kessel
Date Prepared February 5, 2014 (revised July 23, 2014)
Device Trade Name Caya® contoured diaphragm
Common Name Diaphragm
### Classification Name, Number, and Product Code
Classification Name: Contraceptive diaphragm and accessories (Class II)
Classification Number: 21 CFR §884.5350
Product Code: HDW
#### Predicate Devices ORTHO® ALL-FLEX® Diaphragm (silicone), Johnson & Johnson Produtos Profissionais Ltda. – A division of Johnson & Johnson, 510(k) number K080040.
#### Device Description and Statement of Intended Use
Description: The Caya® contoured diaphragm is a contoured, flexible, single-sized, silicone elastomer cup for covering the cervix that is circumscribed by a rim encapsulating a nylon spring. The Caya contoured diaphragm has an anatomically shaped spring that has been designed to fit women representing a range of traditional diaphragm sizes.
<sup>1</sup> Caya is a registered trademark of Kessel medintim GmbH of Frankfurt, Germany.
<sup>2</sup> The Caya contoured diaphragm was known as the SILCS diaphragm throughout design, development, and evaluation.
{4}------------------------------------------------
As with the predicate device, the Caya contoured diaphragm is inserted by squeezing the rim at the sides to fold the diaphragm into a narrow profile. The Caya® contoured diaphragm has grip dimples along the sides of the rim that provide a tactile cue as to where to squeeze the diaphragm during insertion. The single-size diaphragm was developed and tested as the SILCS diaphragm.
Statement of Intended Use: The Caya® contoured diaphragm is indicated for the prevention of pregnancy in women who elect to use diaphragms as a method of contraception. It is recommended for use with a contraceptive gel.
#### Summary of Technological Characteristics
When inserted into the vagina, the Caya® contoured diaphragm functions as a mechanical barrier that prevents sperm from entering the cervical canal. The spring within the perimeter of the device causes the device to create a seal against the vaginal wall, covering the cervix and preventing sperm from entering the cervical canal. The silicone cup also serves as a repository for contraceptive gel. See Table 5.1 (below) for a comparison of the Caya" contoured diaphragm to its predicate device, the ORTHO® ALL-FLEX® Diaphragm.
#### Performance Data
The subject device has passed biocompatibility testing in accordance with ISO 10993, performance testing in accordance with ISO 8009, and cleaning validation. Specifically the device has been subjected to and passed the following biocompatibility and performance tests:
#### Biocompatability testing
- . Cytotoxicity Growth Inhibition Test
- . Vaginal Irritation Study
- Delayed Type Hypersensitivity
- Genotoxicity, bacterial reverse mutation study ●
- Inductively Coupled Plasma Investigations ●
- Gas Chromato-graphic and FTIR Fingerprint Investigations ●
- ISO intracutaneous study ●
- Mouse peripheral blood microcutaneous study ●
- Genotoxicity, mouse lymphoma ●
- 12-week muscle implant ●
- Subchronic Systemic Toxicity
#### Performance testing
- Visual inspection
- Weight ●
- Kneel-down bending ●
- Membrane tensile strength and elongation ●
- Shore A hardness
- Membrane thickness
- Chemical resistance ●
- Compression Test ●
- Twist Test ●
{5}------------------------------------------------
#### Clinical Testing
In a Phase I post coital study of barrier effectiveness, the SILCS diaphragm3 used with Nonoynol-9 (N-9) spermicide reduced the average number of progressively motile sperm per high powered field from a baseline of 12.5 to 0, confirming that the SILCS diaphragm would likely perform well in an effectiveness study.
The Phase II/II contraceptive effectiveness study-conducted at six sites in the United States-recruited 450 couples to participate in the study. Couples were randomized to two groups: 300 to SILCS diaphragm used with BufferGel (BG-an investigational lactic-acid based contraceptive gel) and 150 to SILCS diaphragm with N-9 spermicide gel. Healthy, sexually active female volunteers 18–40 years old, at risk for pregnancy, and desiring contraception but at low risk for HIV/STIs, were considered for enrollment. Eligible volunteers were not pregnant, had normal menstrual cycles, were not actively desiring pregnancy and were willing to accept an unknown risk of pregnancy and engage in at least 4 acts of vaginal intercourse per cycle. Participants were followed for at least 190 days and 6 menstrual cycles, and were seen at enrollment and after menstrual cycles 1. 3. and 6. Study outcomes included pregnancy probability. safety, acceptability, diaphragm fit, and ease of use. Results on effectiveness and safety were compared to an historical control group who used the ORTHO® ALL-FLEX® Diaphragm with these gels.
The historical control for the SILCS diaphragm pivotal study was a multi-center contraceptive study conducted by the National Institute for Child Health and Human Development (NICHD). That study demonstrated that BG used with an ORTHO® ALL-FLEX® Diaphragm worked about as well as N-9 with that diaphragm (6-month typical-use cumulative probability of pregnancy rates of 10.1 per 100 women (95% CI 7.1-13.1) and 12.3 (95% CI 7.7-16.9), respectively. * The SILCS pivotal study was based on that previous study which was then used in the SILCS historical control analysis to compare the single-size SILCS diaphragm with the ORTHO® ALL-FLEX® Diaphragm on contraceptive effectiveness and safety.
In the SILCS pivotal study, 35 study pregnancies were reported which yielded 6-month Kaplan-Meier cumulative typical-use pregnancy probabilities per 100 women (with 95% confidence intervals) of 10.4 (6.9, 14.0), 9.6 (5.5, 13.6) and 12.5 (5.4, 19.5) for all SILCS with BG and SILCS with N-9, respectively. The rate for all SILCS users was non-inferior to the rate for all users of the ORTHO® ALL-FLEX® Diaphragm, using data from the historical control groups. The observed 6-month cumulative typical-use pregnancy probability was 10.4 per 100 women (95% CI: 6.9, 14.0). The observed 6-cycle cumulative perfect-use probability was 7.9 per 100 women (95% CI: 1.7. 14.0). Extrapolated to 12 months these estimates are for typical use: 17.8 per 100 women (95% CI: 12, 23.6) and for perfect use: 14.0 per 100 women (95% CI: 3.0, 23.6) (See Table 1 (below)).
<sup>3</sup> The Caya® contoured diaphragm was known as the SILCS diaphragm throughout design, development, and evaluation.
<sup>4</sup> Barnhart KT, Rosenberg MJ, MacKay HT, et al. Contraceptive efficacy of a novel spermicidal microbicide used with a diaphragm: a randomized controlled trial. Obstet Gynecol. 2007;110(3):577-586.
{6}------------------------------------------------
| | Estimated number of<br>pregnancies per 100 women<br>during time period | 95% confidence interval |
|--------------|------------------------------------------------------------------------|-------------------------|
| 6 months | | |
| Typical use | 10.4 | 6.9, 14.0 |
| Perfect use* | 7.9 | 1.7, 14.0 |
| | | |
| 12 months | | |
| Typical use | 17.8 | 12.0, 23.6 |
| Perfect use | 14.0 | 3.0, 23.6 |
Table. 5.1. Typical and Perfect Use Pregnancy Probabilities at 6- and 12-months
*6-cycle
#### Conclusion
Differences in technological characteristics between the Caya® contoured diaphragm and the predicate device raise no new safety concerns, and the devices have intended use. Results from the pivotal study show that the single-sized Caya diaphragm used with a contraceptive gel provided similar use effectiveness as results from a previously implemented study that evaluated the predicate device (ORTHO® ALL-FLEX® Diaphragm) used with the same contraceptive gels. The single-sized Caya diaphragm has a similar clinical safety profile as the predicate device. The above information demonstrates that the single-sized Caya diaphragm is substantially equivalent to the predicate device.
{7}------------------------------------------------
| Attribute | Subject device | Predicate device |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device information | | |
| Trade/Device name | Caya® contoured diaphragm | ORTHO® ALL-FLEX®<br>Diaphragm |
| 510(k) number | K140305 | K080040 |
| Manufacturer | Kessel medintim GmbH | Johnson & Johnson Produtos<br>Profissionais Ltda. – A division<br>of Johnson & Johnson |
| Classification/Product code | 21 CFR §884.5350 / HDW | 21 CFR §884.5350 / HDW |
| Similarities | | |
| Indication for use | The Caya® contoured diaphragm is<br>indicated for the prevention of<br>pregnancy in women who elect to<br>use diaphragms as a method of<br>contraception. It is recommended<br>for use with a contraceptive gel. | The ORTHO® ALL-FLEX®<br>Diaphragm, in conjunction with<br>an appropriate spermicide, is<br>indicated for the prevention of<br>pregnancy in women who elect<br>to use diaphragms as a method<br>of contraception. |
| Mode of action | Mechanical contraceptive barrier | Same |
| Reusable | Yes | Same |
| Material (barrier) | Medical-grade silicone | Same |
| Standard of conformity | ISO 8009:2004(E) | Same |
| Differences | | |
| Rim shape | Anatomically contoured (arcing<br>spring) in one size (60 to 85 mm) | Circular (arcing spring) in four<br>sizes (65, 70, 75, and 80 mm) |
| Spring stiffness (relative) | Soft | Stiff |
| Grip dimples | Present | Not present |
| Finger removal dome | Present | Not present |
| Material: spring | Nylon | Metal |
| Material: color (of silicone) | Violet | Buff |
Table 5.2. Summary comparison of technological characteristics between the single-sized Caya contoured diaphragm and its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.