CARESTREAM Image Suite is an image management system whose intended use is to receive, process, review, display, print and archive images and data from all imaging modalities. Tablet Viewer Software for Image Suite is used for patient management by clinicians in order to access and display patient data, medical reports, and medical images for diagnosis from different modalities including CR, DR, CT, MR, and US. Tablet Viewer Software for Image Suite provides wireless and portable access to medical images for remote reading or referral purposes from web browsers including usage with validated mobile devices. This device is not intended to replace the full Mini-PACS and should be used only when there is no access to the full Mini-PACS Web Viewer. This excludes mammography applications in the United States.
Device Story
Tablet Viewer Software for Image Suite is an optional extension to the Image Suite Mini-PACS. It enables clinicians to access patient images, reports, and data on an iPad 2 via wireless network. The software utilizes HTML5 to allow browser-enabled mobile devices to function as remote viewers. It provides a GUI for basic image manipulation, including zoom, pan, windowing, basic measurements, and cine. The system is intended for use in small clinics for remote reading or referral purposes when the full Mini-PACS Web Viewer is inaccessible. It does not perform primary diagnostic image processing but facilitates clinical review and decision-making by providing portable access to imaging data.
Clinical Evidence
Clinical assessment performed on the Apple iPad 2 to validate the display of medical images for diagnostic purposes. Bench testing included evaluation of luminance response, viewing angles, resolution, noise, reflectivity, and device settings. Functional QA testing confirmed software performance. No specific sensitivity/specificity metrics were reported; the device was validated for its intended use as a remote viewing platform.
Technological Characteristics
Software-based PACS extension using HTML5 for browser-based access on mobile devices (iPad 2). DICOM compliant. Features include zoom, pan, windowing, measurements, and cine. Validated for display performance (luminance, resolution, reflectivity).
Indications for Use
Indicated for clinicians to access and display patient data, medical reports, and medical images for diagnosis from CR, DR, CT, MR, and US modalities. Used for remote reading or referral via web browsers on validated mobile devices. Not for mammography applications in the US. Not intended to replace full Mini-PACS.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# APR 0 1 2014
K140271
Page 1 of 3
Carestream
Carestream Health Inc. 150 Verona Street Rochester, NY 14608
# 510(k) Summary Pursuant to 21 CFR 807.92
| 510{k) Owner Name: | Carestream Health, Inc. |
|------------------------|-------------------------------------------------------------------------------------------------------------|
| 510{k} Owner Address: | 150 Verona Street<br>Rochester, New York 14608 |
| 510(k) Owner Phone: | 585-627-8790 |
| 510(k) Owner Fax: | 585-323-7643 |
| Contact Name & Info: | Alexandra Chamberlain<br>Requlatory Affairs Manager<br>alexandra.chamberlain@carestream.com<br>585-627-8790 |
| Date Summary Prepared: | January 31, 2014 |
| Device Trade Name: | Tablet Viewer Software for Image Suite |
| Device Common Name: | PACS |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Name: | Picture Archiving and Communication System |
| Device Class: | Class II |
| Device Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Predicate Device: | Image Suite<br>Manufactured by Carestream Health, Inc.<br>510(k) Number - K100094 (March 11, 2010) |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Carestream" in a stylized font. The letters are outlined in black, giving them a distinct appearance. The font choice adds a unique visual element to the word, making it stand out.
Carestream Health Inc. 150 Verona Street Rochester, NY 14608
### Device Description:
CARESTREAM Tablet Viewer Software for Image Suite is an optional feature for Image Suite Mini-PACS users. The software technology uses HTML5 which allows a browserenabled mobile device to run the software application. The user is able to access patient images and study reports from an iPad 2 mobile device anywhere through a wireless network. Tablet Viewer Software for Image Suite has a simple GUI for viewing and includes some fundamental tools such as zoom, pan, windowing, basic measurements, cine, etc. Tablet Viewer Software for Image Suite functions as an extension to Image Suite.
CARESTREAM Image Suite is a stand-alone, self-contained radiographic imaging system designed to provide a low-cost platform to manage medical images, reports, patient/exam information and workflow in small clinics. The system performs capture, processing, review, archive, and printing of radiographic images as well as report writing and printing and is designed to run on a PC workstation. CARESTREAM Image Suite is designed to be simple and intuitive to both use and service.
CARESTREAM Image Suite connects with hardware including multiple radiographic image capture devices (CR and / or DR detectors) attached to a PC workstation with either a standard or a high-resolution monitor. CARESTREAM Image Suite is designed as a hardware-independent system and may be interfaced with verified and validated imaging modalities from both Carestream Health and 3rd party vendors, as well as Carestream Health PACS systems, and other 3rd party PACS systems. The Image Suite system can directly acquire an image from Carestream Health acquisition devices and is PC and monitor independent.
### Indications for Use:
CARESTREAM Image Suite is an image management system whose intended use is to receive, process, review, display, print and archive images and data from all imaging modalities.
Tablet Viewer Software for Image Suite is used for patient management by clinicians in order to access and display patient data, medical reports, and medical images for diagnosis from different modalities including CR. DR. CT, MR, and US.
Tablet Viewer Software for Image Suite provides wireless and portable access to medical images for remote reading or referral purposes from web browsers including usage with validated mobile devices. This device is not intended to replace the full Mini-PACS and should be used only when there is no access to the full Mini-PACS Web Viewer.
This excludes mammography applications in the United States.
દર્
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Carestream" in a simple, sans-serif font. The letters are outlined in black, giving them a slightly bold appearance. The word is presented in a straightforward, horizontal arrangement.
Carestream Health Inc. 150 Verona Street Rochester, NY 14608
## Technological Characteristics:
The Tablet Viewer Software for Image Suite is fully compatible with Image Suite and can be displayed on the iPad 2 mobile device.
The product is software, and has been extensively tested in accordance with General Principles of Software Validation; Final Guidance for Industry and FDA Staff. The product is DICOM Compliant. Testing included the following:
- . Bench testing on the Apple iPad 2 for luminance response, optimal viewing angles, resolution, noise, reflectivity, device and display settings, and exception handling
- A Clinical Assessment of Tablet Viewer Software for Image Suite on the Apple . iPad 2
- . Functional QA testing of the software
The technological characteristics of the new device are equivalent to the predicate device in that they use the same Image Suite PACS platform. The new type of display device is being introduced by the introduction of Tablet Viewer Software for Image Suite.
The Indications of the predicate and subject are similar, differing only by the added description related to the remote access from the validated mobile device and the list of supported modalities for the Tablet Viewer software.
### Substantial Equivalence:
Tablet Viewer Software for Image Suite is a new feature that can be used with Image Suite (predicate device, K100094). The software provides wireless and portable access to medical images for remote reading or referral purposes using an iPad 2 mobile Results of bench and clinical testing demonstrated that the Tablet Viewer device. Software for Image Suite is suitable as a platform to display patient data, medical reports, and medical images for diagnosis from different modalities including CR, DR, CT, MR, and US. No substantial differences that affect safety and efficacy were noted in comparison with the predicate device.
### Conclusion:
CARESTREAM Tablet Viewer Software for Image Suite is substantially equivalent to the CARESTREAM Image Suite predicate. The product has been validated and tested and no substantial differences that affect safety and efficacy were noted.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## April 1, 2014
Carestream Health, Inc. % Ms Alexandra Chamberlain Regulatory Affairs Manager l 50 Verona Street ROCHESTER NY 14608
Re: K140271
Trade/Device Name: Tablet Viewer Software for Image Suite Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 4, 2014 Received: February 5, 2014
Dear Ms Chamberlain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of general connell provision practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that i Drederal statutes and regulations administered by other Federal agencies. You must or any I odoral statution and including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Ms Chamberlain
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh-7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
# K140271
Device Name
CARESTREAM Tablet Viewer Software for Image Suite
### Indications for Use (Describe)
CARESTREAM Image Suite is an imagement system whose intended use is to receive, process. review. display, print and archive images and data from all imaging modalities,
Tablet Viewer Software for Image Suite is used for patient management by clinicians in order to access and display patient data. medical reports, and medical images for different modalities including CR. DR. CT. MR, and US.
Tablet Viewer Software for Image Suite provides wireless and portable images for remote reading or referral purposes from web browsers including usage with validated mobile device is not intended to replace the full Mini-PACS and should be used only when there is no access to the full Mini-PACS Web Viewer.
This excludes mammography applications in the United States.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Smh.7)
FORM FDA 3881 (9/13)
14 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4 1 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.