K140265 · Guangzhou Finecure Medical Equipment Co.,Ltd · KPI · Jan 23, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K140265
Device Name
ELECTRODE FOR URINARY INCONTINENCE
Applicant
Guangzhou Finecure Medical Equipment Co.,Ltd
Product Code
KPI · Gastroenterology, Urology
Decision Date
Jan 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Urinary Incontinence Probe, model PR-51 Vaginal Probe & PR-52 Anal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
Device Story
Urinary Incontinence Probe (models PR-51 Vaginal, PR-52 Anal) consists of lightweight cylinder with two independent conductive rings; used for pelvic floor muscle rehabilitation. Device provides electromyographic feedback or electrical stimulation to pelvic musculature. Operated in home or clinic settings by single user for up to one year. Probe connects to external stimulation/feedback devices via lead wires (2mm pigtail, 3.5mm female, or 2.35mm female terminals). Device is watertight for cleaning with soap and water; does not require sterilization. Output facilitates muscle strengthening and neuromuscular control, aiding treatment of urinary incontinence.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10.
Technological Characteristics
Cylindrical probe with two independent conductive rings; watertight design. Connectivity via 2mm pigtail, 3.5mm female, or 2.35mm female terminals. Materials evaluated per ISO 10993-1, 10993-5, and 10993-10. Non-sterile, reusable by single user. Passive electrode component; no internal software or active circuitry.
Indications for Use
Indicated for patients with urinary incontinence requiring pelvic floor muscle rehabilitation or neuromuscular control restoration via electromyographic feedback or electrical stimulation.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2015
Guangzhou Finecure Medical Equipment Co., Ltd % Field Fu Consultant Shenzhen Joyantech Consulting Co., Ltd. 4th Floor, Jinhui Building, Nanhai Blvd, Nanshan District, Shenzhen, 518000 China
Re: K140265
Trade/Device Name: Electrode for Urinary Incontinence Regulation Number: 21 CFR 876.5320 Regulation Name: Non-Implanted electrical continence device Regulatory Class: II Product Code: KPI, HIR Dated: December 10, 2014 Received: December 14, 2014
Dear Field Fu,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140265
#### Device Name Electrode for Urinary Incontinence
#### Indications for Use (Describe)
The Urinary Incontinence Probe, model PR-51 Vaginal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Product: Electrode for Urinary Incontinence
# SECTION 05 510(k) Summary
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
## 1 Administrative Information
| Date of Summary<br>prepared | Jan, 19, 2014 |
|--------------------------------------|---------------------------------------------------------------------|
| Manufacturer<br>information | Company title:<br>Guangzhou Finecure Medical Equipment Co.,<br>Ltd. |
| | Company address: |
| | F19, No.1 Kesheng Road, Baiyun Torch |
| | Building, No.1633 Beitai Road, Baiyun District |
| | Guangzhou, CHINA |
| | Phone: +86(020) 2802 6079 |
| | Fax: +86(020) 2802 6059 |
| | Contact Person: Chiping Ma |
| | E-mail: Finecure_export@163.com |
| Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd. |
| | 4th Floor, Jinhui Building, Nanhai BLVD, |
| | Nanshan District, Shenzhen, Guangdong, |
| | China. |
| | Contact person: Mr. Field.Fu |
| Image: Joyantech logo | E-Mail: cefda13485@163.com |
| | QQ: 670312758 |
| | Website: www.cefda.com |
| Establishment registration<br>number | No. |
## 2 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|------------------------------------|
| Trade Name: | Electrode for Urinary Incontinence |
| Model: | PR-51, PR-52 |
{5}------------------------------------------------
Guangzhou Finecure Medical Equipment Co., LTD
SEC_005:001_510K Summary
Version:A/0
Product: Electrode for Urinary Incontinence
| Classification name: | Stimulator, electrical, non-implantable, for<br>incontinence;<br>Perineometer. |
|----------------------|--------------------------------------------------------------------------------|
| Review Panel: | Gastroenterology/Urology;<br>Obstetrics/Gynecology. |
| Product Code: | KPI; HIR |
| Device Class: | II |
| Regulation Number: | 876.5320;884.1425 |
### 3 Predicate Device Information
| Sponsor: | VERYWAY MEDICAL INSTRUMENT CO., LTD |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device: | Everyway Incontinence Stimulation Electrode<br>Model: PR-02/02A, PR-03103A, PR-04104A, PR-10A,<br>PR-h1A, PR-14A for Life-Care VaginalProbe & PR-<br>06/06A, PR-1 2A, PR-1 3/13A for Life-Care Anal Probe. |
| 510(K) Number: | K122194 |
#### 4 Device Description
The Urinary Incontinence Probe, model PR-51 Vaginal Probe & PR-52 Anal Probe are the light weight cylinder consisting of two independent conductive rings that are paired and isolated, physically and electrically. The cylinder is shaped with a handle for comfort positioning in vaginal/anal canal for incontinent treatment as above mentioned and easy for removing after treatment. It is watertight to allow for washing with soap and water between uses. The probe is designed for repeated intermittent use in home or clinic for up to one year by a single user. It does not require sterilization, but does required washing for reuse according to the validated cleaning method as recommended in user manual.
To fit with different type of device connection, the lead wire of device was provided with three different type of wire connection terminals, 2mm pigtail, 3.5mm femelle and 2.35 mm femelle. Either of the PR-51 Vaginal Probe & PR-52 Anal Probe can be connected to any one of three connection terminals.
### 5 Intended Use
The Urinary Incontinence Probe, model PR-51 Vaginal Probe & PR-52 Anal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
{6}------------------------------------------------
Version:A/0
Product: Electrode for Urinary Incontinence
## 6 Indication for Use
The Urinary Incontinence Probe. model PR-51 Vaginal Probe & PR-52 Anal Probe are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
# 7 Technological characteristics of the proposed device compared to the predicate device
The proposed device and the Predicate device have the same intended use, design principle, and similar material composition. The differences do not exert adverse effect on the proposed device. The proposed device is substantially equivalent to the predicate devices.
## 8 Brief discussion of the nonclinical tests
Electrode for Urinary Incontinence conforms to the following standards:
- & ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process;
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for � In Vitro cytotoxicity;
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests � for irritation and skin sensitization.
# 9 Brief discussion of clinical tests
Not applicable.
# 10 Other information (such as required by FDA guidance)
No other information.
# 11 Conclusions
The subject device---Electrode for Urinary Incontinence PR-51, PR-52 are substantially equivalent to Everyway Incontinence Stimulation Electrode whose 510(k) number is K122194.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.