Insuflow Synergy™ Port (5 mm) and Insuflow® Synergy™ XL Port (8 mm, 10mm and 12mm) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
Device Story
Gas conditioning device for minimally invasive surgery; attaches to CO2 insufflator. System comprises sterile, disposable single-use component (filter, heater/humidifier, tubing, path-of-entry access port) and reusable control module. CO2 gas flows through filter, heater/humidifier, and dual-lumen access port into surgical cavity. Access port includes duckbill/tool seals for instrument entry and separate lumen for gas delivery. Obturator features transparent optical window for endoscopic visualization during insertion. Used by surgeons in OR to maintain surgical cavity insufflation while providing conditioned (heated/humidified) gas to reduce potential tissue desiccation/hypothermia. Output is physical delivery of conditioned gas and instrument access channel.
Clinical Evidence
Bench testing only. Performance testing included flow/pressure characterization and insertion/removal testing for dual-lumen models. Seal leak integrity and gas conditioning (temperature/humidity) data leveraged from predicate devices.
Technological Characteristics
Dual-lumen access port; disposable filter, heater/humidifier, and tubing; reusable control module. Features transparent optical window on obturator for endoscopic visualization. Dual-lumen configuration with thinner inner lumen wall. Gas conditioning via integrated heating element and humidification media.
Indications for Use
Indicated for patients undergoing thoracic, abdominal, or gynecologic minimally invasive endoscopic surgical procedures requiring CO2 insufflation and instrument access.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
140263 age 1 of 3
#### 510(k) Summary 5.0
Insuflow® Synergy™ and Insuflow® Synergy™ XL Port
January 31, 2014 (revised February 27, 2014) Date Prepared:
- Lexion Medical, LLC Submitter: 545 Atwater Circle St. Paul, MN 55103 Telephone: 651-855-1447 Fax: 651-636-1671
Mr. Bernard Horwath Contact: Regulatory Affairs Consultant 4486 Timberline Ct Vadnais Heights, MN 55127 Telephone: 651- 231-1761
Proprietary Name Modified Device: Insuflow Synergy™ Port (5 mm) and Insuflow® Synergy™ XL Port (8 mm, 10mm and 12mm)
Common/Usual Name: Gas Conditioner Insufflator Device with integral path of entry device
Classification Name: Class II per regulations 884.1730, Product Code: HIF Class II per regulations 876.1500, Product Code: GCJ
Establishment Registration Number: 2135348
## Predicate Devices:
LEXION Medical believes that the modified Insuflow Synergy™ and Synergy™ XL Port devices are substantially equivalent to the following devices:
-
- nang oquare and the and the article of noon
Insuflow" Synergy™ XL Port, K.1206440 Synergy
-
- SurgiQuest AirSeal, K121336
## Description:
The Insuflow® Synergy™ and Synergy™ XL Port devices are gas conditioning devices that
attach to the outlet port of an insufflator or other regulated CO2 source and are design and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow® Synergy™ XL Port consist of a
sterile, disposable single use device with a filter, heater/humidifier, tubing set, and a path of entry access port device. A reusable control module houses the control and safety circuits for the system.
{1}------------------------------------------------
The technology for gas conditioning and path of entry for the modified devices are the same as e predicate Insuflow Synergy™ marketed devices. Regulated CO2 gas flows into the nsuflow Synergy™ and Synergy™ XL Port, through the in-line filter, continues along the ubing to enter the path of entry access device that contains the heating element and umidification media, and through the path of entry access device lumen for delivery into the atient's surgical cavity.
be modifications included in this Special 510fk) include the following:
- Adding an 8 mm size path of entry access device to the Insuflow® Synergy™ family. Sizes now available are 5 mm, 8 mm, 10 mm and 12 mm.
- All access devices will be in a dual lumen configuration with a thinner inner lumen wall design that results in a smaller overall outer lumen diameter.
e integral path of entry access device in the modified Insuflow® Synergy™ and Synergy™ XL ort is designed and constructed similarly to the predicate Insuflow® Synergy™ and Synergy™ KL Port devices but an additional 8 mm access device configuration is being added, just as the predicate EndoPath and SurgiQuest trocar devices. The modified Insuflow Synergy™ (5 mm) and Synergy™ XL Port devices (8 mm, 10 mm, and 12 mm) now feature a dual lumen configuration only. The predicate Insuflow® Synergy™ Port (5 mm) came in single and dual lumen configurations and the predicate Insuflow Synergy™ XL Port (10 mm and 12 mm) had nly a single-lumen working channel configuration for conditioned gas delivery and instrument entry. The modified dual lumen configuration has an inner lumen wall thickness thinner than in the predicate design which results in a smaller overall outer lumen diameter. The dual lumen access device has an inner lumen working channel with duckbill and tool seals for instrument ntry into the surgical cavity and delivers conditioned insufflation gas through the outer lumen channel.
As in the predicates, the obturator for the proposed Insuflow® Synergy™ (5 mm) and Synergy™ (8 mm, 10 mm, and 12 mm) Port devices has a transparent optical window at the distal end, ich when used with an endoscope, provides visibility of individual tissue layers during insertion.
#### ndications for Use:
he modified Insuflow Synergy™ Port (5 mm) and the Insuflow® Synergy™ XL Port (8 mm, 10 nm, and 12 mm) have the same intended use as the Insuflow Synergy™ and Synergy™ 3 Port predicate devices. The intended use of the modified device, as described in the labeling, has ot changed as a result of the modifications.
ow Synergy™ Port (5 mm) and Insuflow® Synergy™ XL Port (8 mm, 10mm and 12mm) applications in thoracic, abdominal and gynecologic minimally invasive endoscopic urgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
{2}------------------------------------------------
#### Technological characteristics, comparison to predicate device.
The modified Insuflow Synergy™ (5 mm) and Synergy™ XL Port (8 mm, 10 mm and 12 mm) utilize the same gas conditioning technology and path of entry access port technology as the predicate Insuflow® Synergy™ Port (5 mm) cleared under 510(k) K120640 and the Insuflow® Synergy™ XL Port (10 mm and 12 mm) cleared under 510(k) K132204. All of the modified devices perform the same gas conditioning function as the predicates with minor design configuration changes. The new Insuflow Synergy™ XL Port (8 mm) path of entry access device is technically equivalent to the predicate Insuflow® Synergy™ and Synergy™ XL Port devices but is configured for 8 mm surgical tools, just as the predicate EndoPath and SurgiQuest trocar devices. The modified Insuflow Synergy™ and Synergy™ XL Port devices are configured in the dual lumen configuration while the predicate Insuflow Synergy™ Port (5 mm) is configured in single and dual lumen configurations and the predicate Insuflow Synergy™ XL Port (10 mm and 12 mm) devices have only a single lumen configuration. All materials in the modified Insuflow Synergy™ devices are the same as the predicate Insuflow® Synergy™ devices.
The obturators for the modified Insuflow Synergy™ and Synergy™ XL Port use the same technology and materials as the predicates with a transparent optical window at the distal end, which when used with an endoscope, provides visibility of individual tissue layers during insertion.
#### Discussion of performance testing.
The risk analysis and resulting performance testing have been conducted to assure that the modified Insuflow® Synergy™ and Insuflow® Synergy™ XL Port devices perform in accordance with specifications and applicable standards. All of the modified devices utilize the same exact design/parts as the predicates for gas conditioning and sealing. As a result, the existing testing for gas temperature and humidity characterization and seal leak integrity testing previously submitted applies directly to these devices. Since the lumen designs were changing, flow/pressure performance and insertion/removal testing were repeated and successfully completed for the modified Insuflow® Synergy™ (5 mm) and Synergy™ XL (8 mm,10 mm and 12 mm) Port dual lumen models. As a result, the risk analysis and applicable testing were successfully conducted to confirm that the Insuflow" Synergy™ and Synergy™ XL dual lumen configuration changes do not substantially affect their performance.
#### Conclusion:
Based on a comparison to the predicate devices and information provided, it is the conclusion of LEXION Medical that the modified Insuflow Synergy™ Port (5 mm) and the Insuflow Synergy™ XL Port (8 mm, 10 mm and 12 mm) devices are substantially equivalent to devices already on the market being used for these applications (cleared by the 510(k) process) and present no new concerns about safety and effectiveness.
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 5, 2014
LEXION Medical, LLC % Bernard Horwath Regulatory Consultant HRG 4486 Timberline Court Vadnais Heights, MN 55127
Re: K140263
Trade/Device Name: Insuflow® Synergy™ and Insuflow® Synergy™ XL Port Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF, GCJ Dated: January 31, 2014 Received: February 3, 2014
Dear Bernard Horwath.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device fabeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
#### Page 2 - Bernard Horwath
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D., Director Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K140263
## Insuflow Synergy™ and Insuflow Synergy™ XL Port:
Indications for Use:
Insuflow Synergy™ Port (5 mm) and Insylow® Synergy™ XL Port (8 mm, 10mm and 12mm) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Office of Device Evaluation (ODE)
Herbert P. Lerner -S 2014.03.05 15:54:19 -05'00'
Lexion Medical Insuflow® Synergy™ and Synergy™ XL Port Special 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.