The ultrasonic scaler is intended to: 1. Remove supra and sub gingival calculus deposits and stain from the teeth; 2. Carry out periodontal pocket lavage with simultaneous ultrasonic tip movement; 3. Clean and irrigate root canals.
Device Story
Ultrasonic scaler P7 uses piezoelectric technology to generate high-frequency vibrations for dental procedures. Device consists of a control unit, handpiece with cord, and detachable scaling tips. Connects to external water supply for irrigation during operation. Operated by dental professionals in clinical settings. Piezoelectric transducer converts electric power into mechanical vibrations (28kHz ± 3kHz) at the tip to remove calculus, stains, and perform lavage or root canal irrigation. Water delivery through handpiece provides cooling and cleaning. Output allows clinicians to perform periodontal and endodontic maintenance, improving oral hygiene and patient dental health.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), output characteristics (IEC 61205), biocompatibility (ISO 10993-5, ISO 10993-10), and steam sterilization validation (AAMI/ANSI ST79).
Technological Characteristics
Piezoelectric ultrasonic scaler. Handpiece shell: aluminum alloy. Frequency: 28kHz ± 3kHz. Components: control unit, handpiece, cord, detachable tips. External water supply connection. Sterilization: moist heat (steam) per AAMI/ANSI ST79. Standards: IEC 60601-1, IEC 60601-1-2, IEC 61205, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for dental patients requiring removal of supra and subgingival calculus and stains, periodontal pocket lavage, and root canal cleaning/irrigation.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three faces overlapping to symbolize the department's focus on health and well-being. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2015
Nanning Baolai Medical Instruments Co., Ltd. c/o Mr. Mike Gu Osmunda Medical Device Service Group Level 7, Jin Gui Business Center 982 Cunyun Road, Baiyun District Guangzhou, Guangdong 510420 CHINA
Re: K140233
Trade/Device Name: Ultrasonic scaler P7 Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic scaler Regulatory Class: II Product Code: ELC Dated: February 28, 2014 Received: March 3, 2015
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Mike Gu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140233
Device Name Ultrasonic scaler P 7
#### Indications for Use (Describe)
The ultrasonic scaler is intended to:
- 1. Remove supra and sub gingival calculus deposits and stain from the teeth;
- 2. Carry out periodontal pocket lavage with simultaneous ultrasonic tip movement;
- 3. Clean and irrigate root canals.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for BAOLAI MEDICAL. The logo features the company name in blue, with a stylized "b" above it. The "b" is made up of several parallel lines that curve around to form the shape of the letter. The logo is set against a white background with a blue border.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Submitter: | Nanning Baolai Medical Instruments Co., Ltd.,<br>7th Floor, General Building, Hetai Sci. &Tech. Park, No.9 Gaoxin<br>4th RD., Nanning ,Guangxi, 530007, China |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Mike Gu<br>Regulatory Affairs Manager<br>Osmunda Medical Device Consulting Co., Ltd.<br>(+86)-020-62321333<br>(+86)-020-86330253 |
| Secondary Contact Person: | Wanqing Bi<br>Deputy Manager<br>Nanning Baolai Medical Instrument Co., Ltd.<br>0086-771-3815998<br>0086-771-3217883 |
| Device: | |
| Trade Name: | Ultrasonic scaler P7 |
| Common/Usual Name: | Ultrasonic scaler |
| Classification Names: | Ultrasonic scaler |
| Product Code: | ELC |
| RegulationNumber | 21 CFR 872.4850 |
| Date prepared | April 1th, 2015 |
| Number of submission | K140233 |
| Predicate Device(s): | Guilin Woodpecker USD-L ultrasonic scaler, K053555 |
| Device Description: | Ultrasonic scaler employs piezoelectric technology to produce a<br>steady power. It is composed of a control unit housing a<br>generator that produces piezo-electric vibrations. The working<br>instrument for the ultrasonic scaler is the handpiece, which is<br>connected to the control unit via a handpiece cord. A scaling tip<br>specific to a particular procedure is attached to the end of the<br>handpiece. The ultrasonic scaler connects to an external water<br>supply and delivers water via connected handpiece. |
| Intended Use: | The ultrasonic scaler is intended to:<br>1. Remove supra and sub gingival calculus deposits and<br>stain from the teeth;<br>2. Carry out periodontal pocket lavage with<br>simultaneous ultrasonic tip movement;<br>3. Clean and irrigate root canals. |
| Technology: | The ultrasonic scaler is powered by the piezoelectric technology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for BAOLAI MEDICAL. The logo features a stylized letter "b" made of concentric circles. The text "BAOLAI MEDICAL" is written in a bold, sans-serif font below the letter "b". The background is a gradient of light blue to white.
which is also employed by the predicate device. The electric power is transduced into high-frequency vibration via an ultrasonic transducer.
The working instrument for both the proposed and predicate devices is the handpiece, which is connected to the control unit via a handpiece cord. Both the proposed and predicate ultrasonic scalers connect to an external water supply and deliver water via the handpiece.
Determination of Substantial The ultrasonic scaler and its predicate device : Equivalence:
- have the same intended use;
- both employ piezoelectric technology and produce ultrasonic vibrations at a frequency of 28kHz ± 3kHz;
- Have similar main components, i.e. control unit, handpiece cord, handpiece and detachable scaling tips.
Compared to the predicate device, most of their specifications are identical except that the handpiece of the proposed ultrasonic scaler has a metal shell made from aluminum alloy rather than the plastic material used by the predicate device. The difference does not affect the safety of the proposed device as a biocompatibility testing aiming to the handpiece metal shell was conducted and result pass.
Summary of of Non-Clinical Electrical safety tests according to IEC 60601-1: 2005+CORR. 1 Tests: (2006)+CORR. 2 (2007) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;
> Electromagnetic Compatibility Test according to IEC 60601-1-2:2007: General requirement for basic safety and essential performance standard: compatibility - Requirements and tests;
> Bench testing according to IEC 61205:1993 Ultrasonics-Dental descaler systems-Measurement and declaration of the output characteristics.
> Biocompatibility testing according to ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2006 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for BAOLAI MEDICAL. The logo features a stylized letter "b" made of several concentric lines. Below the "b" is the text "BAOLAI MEDICAL" in a bold, sans-serif font. The text and the "b" are in a blue color, and the background is white.
Moist sterilization validation according to AAMI / ANSI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, (Consolidated Text) Comprehensive Guide To Steam Sterilization And Sterility Assurance In Health Care Facilities
Summary of Clinical Tests: The subject of this premarket submission, ultrasonic scaler, did not require clinical studies to support substantial equivalence. Conclusion: Based on the similarities in intended use, principles of operation, design rationale, test results, and performance, Nanning Baolai Medical Instrument Co., Ltd. considers the ultrasonic scaler to be substantially equivalent to the predicate device USD-L ultrasonic scaler.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.