30.0 INCH COLOR LCD MONITOR CCL650I2 (CL30650, MD302C6)
Applicant
Jvckenwood Corporation
Product Code
LLZ · Radiology
Decision Date
Feb 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
6M pixel 30.0 inch Color LCD Monitor, CCL650i2 (CL30650, MD302C6) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is not meant to be used for digital mammography.
Device Story
6M 30.0-inch color LCD monitor; displays medical images for diagnostic review by trained medical practitioners. Inputs via DVI-D or DisplayPort; supports multiple display modes including 6M (3280 x 2048) and dual 3M configurations. Features LED backlight for luminance stability; integrated Front Sensor for automatic grayscale calibration; Ambient Light Sensor for environmental monitoring. Used in clinical settings; operated by physicians/radiologists. Output provides high-resolution visual representation of medical imaging data; facilitates clinical decision-making by enabling accurate image interpretation. Benefits include high-resolution display, consistent luminance via LED technology, and automated maintenance/calibration via Medivisor Nx software.
Clinical Evidence
Bench testing only. Performance validated against safety and EMC standards. Image quality characteristics (luminance uniformity, chromaticity, pixel defects) compared to predicate device. Calibration accuracy verified using internal Front Sensor versus external sensors.
Technological Characteristics
30.0-inch color LCD; 3280 x 2048 resolution; LED backlight; DVI/DisplayPort connectivity. Integrated Front Sensor and Ambient Light Sensor. Complies with ANSI/AAMI ES60601-1, CAN/CSA C22.2 No.60601-1, FCC Class B, VCCI-B. Luminance uniformity <30% (AAPM-TG18 4.4); Chromaticity Delta (u', v') ≤ 0.01 (AAPM-TG18 4.8.4).
Indications for Use
Indicated for display and viewing of medical images for diagnosis by trained medical practitioners. Not for use in digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
21.3 inch (54 cm) Color LCD Monitor CCL358i2 (CL21358) (K133185)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 9 2014
# TOTOKU
K140214
Page 1 of 5
## 510(k) SUMMARY
| Submitted Information: | JVC KENWOOD CORPORATION<br>3-12, MORIYA-CHO, KANAGAWA-KU,<br>YOKOHAMA-SHI, KANAGAWA, 221-0022 JAPAN | | | | |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Contact Person: | Tsukasa Tashiro, General Manager<br>Email: tashiro.tsukasa@jvckenwood.com<br>Tel: +81.258.24.6611<br>Fax: +81.258.24.6617 | | | | |
| Date Prepared: | January 20, 2014 | | | | |
| Device Name: | 6M 30.0 inch Color LCD Monitor CCL650i2 (CL30650, MD302C6) | | | | |
| Common Name: | CCL650i2, CL30650, MD302C6 | | | | |
| Classification Name: | Class II<br>(Part 892 Radiology Devices<br>Sec. 892.2050 Picture Archiving and Communication System) | | | | |
| Predicate Device: | 21.3 inch (54 cm) Color LCD Monitor CCL358i2 (CL21358)<br>(K133185) | | | | |
| Device Description: | CCL650i2 (CL30650, MD302C6) is a 6M 30.0-inch Color LCD<br>monitor whose display resolution is 3280 x 2048 supporting DVI<br>(digital visual interface) and DisplayPort. | | | | |
| Intended Use: | 6M pixel 30.0 inch Color LCD Monitor, CCL650i2 (CL30650,<br>MD302C6) is intended to be used in displaying and viewing<br>medical images for diagnosis by trained medical practitioners. It is<br>not meant to be used for digital mammography. | | | | |
| Substantial Equivalence: | CCL650i2 (CL30650, MD302C6) realizes 2 (two) 3M screens of the<br>predicate device CCL358i2 (CL21358) (K133185) in one device.<br>The characteristics of CCL650i2 (CL30650, MD302C6) achieves<br>more than that of 2 (two) 3M monitors, CL358i2 (CL21258)<br>(K133185). The both devices do not share common components. | | | | |
### JVC KENWOOD Corporation
Professional & Healthcare Division 3-12, Moriya-cho, Kanagawa-ku, Yokohama-shi, Kanagawa, 221-0022 Japan
{1}------------------------------------------------
### Technical Specification
1. Luminance uniformity: [SPEC] Less than 30% based on AAPM-TG18 4.4. Refer to actual Luminance uniformity data 2. Pixel Defects / Fault [SPEC] Class II or more. ISO13406-2 3. Artifacts • pixel defects/faults (ISO13406-2): · artifacts: miscellaneous including ringing, ghosting, image sticking 4. Chromaticity Measurement of 5%, 50%, 95% Level
[SPEC] Refer to actual data. 5. Chromaticity [SPEC] Delta (u', v') ≤ 0.01 measured at 80% Lmax based on AAPM-TG18 4.8.4 Refer to actual chromaticity data.
{2}------------------------------------------------
| | CCL35812 (CL21358) | CCL650i2 (CL30650, MD302C6) |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K133185 | Unknown |
| Display Area | Horizontal: 433.152mm, Vertical: 324.864mm | Horizontal: 645.504mm, Vertical: 403.0464mm |
| Input Signal | DVI-D Digital Video Signal, DisplayPort | DVI-D Digital Video Signal, DisplayPort |
| Maximum Display | 1536 x 2048 dots | 3280 x 2048 dots |
| Pixel Pitch | 0.2115 x 0.2115mm | 0.197 x 0.197mm |
| | DVI | DVI/DisplayPort |
| Scanning Frequency | 46.6KHz, Vertical: 30Hz (Landscape)<br>61.9KHz, Vertical: 30Hz (Portrait)<br>93.1KHz, Vertical: 60Hz (Landscape)<br>123.9KHz, Vertical: 60Hz (Portrait)<br>DisplayPort<br>47.4KHz, Vertical: 30Hz (Landscape)<br>63.2KHz, Vertical: 30Hz (Portrait)<br>94.8KHz, Vertical: 60Hz (Landscape)<br>126.3KHz, Vertical: 60Hz (Portrait) | 126.5KHz, Vertical: 60Hz (3M) *1<br>123.8KHz, Vertical: 60Hz (3M PLUS) *2<br>63.2KHz, Vertical: 30Hz (6M) *3<br>61.9KHz, Vertical: 30Hz (6M PLUS) *4<br><br>*1 3M=1536 x 2048<br>*2 3M PLUS=1640 x 2048<br>*3 6M=3072 x 2048<br>*4 6M PLUS=3280 x 2048 |
| Maximum Luminance | 410 cd/m² DICOM calibrated<br>800 cd/m² typ.as LCD component | 410 cd/m² DICOM calibrated<br>800 cd/m² typ.as LCD component |
| Luminance Calibration<br>(Optional) | Software: Medivisor Nx<br>Calibration Sensor: Chroma5 (X-Rite) | Software: Medivisor Nx<br>Calibration Sensor: Chroma5 (X-Rite) |
| Contrast Ratio | 1400:1 | 1000:1 |
| Serial Communication | USB: upstream port (x 1), downstream port (x 2) | USB: upstream port (x 1), downstream port (x 2) |
| Safety Standards | ANSI/AAMI ES60601-1, CAN/CSA C22.2 No.60601-1, FCC<br>(Class B), MDD/CE, VCCI-B (Class B) | ANSI/AAMI ES60601-1, CAN/CSA C22.2 No.60601-1, FCC<br>(Class B), MDD/CE, VCCI-B (Class B) |
| Weight & Dimension | Net: 12kg<br>474(w) x 468.4 - 529.9(H) x 220(D) mm (Landscape)<br>367(w) x 521.9 - 583.4(H) x 220(D) mm (Portrait)<br>Packed: 15.0kg<br>470(w) x 670(H) x 340(D) mm | Net: 15.5kg<br>695.6(w) x 512.1 - 573.6(H) x 243.5(D) mm<br>Packed: 20.3kg<br>791(w) x 690(H) x 344(D) mm |
| Power Supply | 100-240V AC 50/60Hz | 100-240V AC 50/60Hz |
# ubstantial Equivalence Compariso
{3}------------------------------------------------
CCL650i2 (CL30650, MD302C6) can be considered to have equivalent display performances to those of the predicate device CCL358) (K133185) due to the following reasons:
- a. The maximum display size (3280 x 2048) is twice or more larger than that of the predicate device (1536 x 2048) and the active area size (645.504mm (H) x 403.0464mm (V)) is larger than that of the predicate device (433.152mm (H) x 324.864mm (V)). The proposed 6M device realizes two 3M screens in one device.
- b. The DICOM calibrated luminance values of the both devices are the same (410 cd/m² and the typical maximum luminance values are also same (800 cd/m²) between the both devices. The high luminance to be maintained constantly was realized by the employment of LED backlight deteriorating more slowly than conventional CCFL backlights.
- c. The LED backlight is employed with both CCL650i2 (CL30650, MD302C6) and the predicate device CCL358i2 (CL21358) (K133185) because it is mercury-free, consumes less power and deteriorates slowly. We have not recognized any adverse effects of the LED backlight on the quality of displayed images. Refer to "Technical Data" where several image quality characteristics of the proposed device are compared with those of the predicate device.
- d. The both devices display images in accordance with DICOM GSDF by default utilizing the factory calibrated display mode stored in lookup tables inside of them.
- e. Both devices support Digital Visual Interface (DVI) and DisplayPort.
- Pixel pitch of CCL650i2 (CL30650, MD302C6) 0.197mm x 0.197mm is smaller than that f. of the oredicate device CCL358i2 (CL21358) (K133185) 0.2115mm x 0.2115mm. The proposed device can display image more finely than the predicate device.
- g. CCL650i2 (CL30650, MD302C6) can display in 6 modes: 1) Full DVI (6M) 3280 x 2048 (30Hz) 2) Input 1: DVI (3M) 1640 x 2048 (60Hz), Input 2: DVI (3M) 1640 x 2048 (60Hz) 3) Input 1: DVI (3M) 1640 x 2048 (60Hz), Input 2: DP (3M) 1640 x 2048 (60Hz) 4) Input 1: DP (3M) 1640 x 2048 (60Hz), Input 2: DVI (3M) 1640 x 2048 (60Hz) 5) Input 1: DP (3M) 1640 x 2048 (60Hz), Input 2: DP (3M) 1640 x 2048 (60Hz) 6) Full DP (6M) 3280 x 2048 (30Hz) This unique characteristics is only for the proposed device.
As for the maintenance, the same QC software is used for both devices. Both devices have Front Sensor to stabilize the luminance.
As for built-in sensors, both devices have 2 (two) kinds of common sensors, Front Sensor and Ambient Light Sensor. Front Sensor is related to the maintenance or calibration and Ambient Light Sensor is used to measure the ambient light by Ix. Front sensor enables automatic grayscale calibration by measuring the luminance at the screen surface. Without Front sensor, the grayscale calibration process requires human intervention and the use of and external sensor. The accuracy data of the calibration with external sensors and that with Front Sensor is included in section 9 "Verification" in "Application".
{4}------------------------------------------------
K140214
Page 5 of 5
The overall design of the CCL650i2 (CL30650, MD302C6) has been validated in accordance with internationally recognized Safety and EMC standards by third-party certifiers. Besides, JVC KENWOOD Corporation performed a range of system and performance tests to ensure that the CCL650i2 (CL30650, MD302C6) performs in accordance with its specifications. None of the tests revealed behaviors inconsistent with the expected performance.
### Conclusion
The 6M pixel Color LCD Monitor, CCL650i2 (CL30650, MD302C6) is substantially equivalent to 2 (two) of the predicate 3M device with respect to technical characteristics, application and intended use. The specifications of the primary component employed by the proposed device are not the same to those of the predicate device, but the proposed device has been accepted by the same Safety Standards of the predicate device and the differences have been independently validated. Any differences between the devices do not affect safety or effectiveness. Therefore, the proposed device CCL650i2 (CL30650, MD302C6) were determined to be substantially equivalent to the predicate device CCL358i2 (CL21358) (K133185).
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993.0002
February 19. 2014
JVC KENWOOD Corporation % Mr. Tsukasa Tashiro General Manager 3-12. Moriya-Cho. Kanagawa-Ku Yokohama-Shi. Kanagawa 221-0022 JAPAN
Re: K140214
Trade/Device Name: 6M 30.0 inch Color I.CD Monitor CCL650i2 (C1.30650, MD302C6) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ. Dated: January 21, 2014 Received: January 28, 2014
Dear Mr. Tashiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2—Mr. Tashiro
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Salctv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sm.h.7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement on last page. |
| 510(k) Number (if known) | K140214 |
|--------------------------|---------|
| Not known | |
| Device Name |
|-------------------------------|
| CCI.650i2 (CI.30650, MD302C6) |
### Indications for Use (Describe)
6M pixel 30.0 inch Color LCD Monitor CCL650i2 (CL30650. MD302C6) is intended to be used in displaying and viewing medical images for diagnosis by trained Medical practitioners. It is not meant to be used in digital mammography.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY | |
|---------------------------|--|
| Health (CDRH) (Signature) | |
Concurrence of Center for Devices and Radiological Health (C
Sm.h.7)
FORM FDA 3881 (1/14)
Page of of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.