K140213 · Dongguan Microview Medical Technology Co., Ltd. · HGH · Mar 25, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K140213
Device Name
DISPOSABLE ENDOSCOPIC CANNULA
Applicant
Dongguan Microview Medical Technology Co., Ltd.
Product Code
HGH · Obstetrics/Gynecology
Decision Date
Mar 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Endoscopic Cannula is indicated for rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy. It has an additional feature of visualization.
Device Story
Disposable Endoscopic Cannula facilitates transcervical aspiration of uterine cavity during first trimester pregnancy; features integrated LED light source and CMOS camera for visualization. Device comprises dual-channel polycarbonate cannula; one channel for aspiration, one for image signal transmission via cable to external Endoscopic Workstation. Operator uses workstation display to locate embryo tissues; then applies vacuum source to suction orifice to remove tissues. Used in clinical setting by physician. Visualization feature aids in locating target tissue; provides real-time guidance for aspiration procedure; potentially improves procedural accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-5 and ISO 10993-10. Electrical safety and EMC testing per IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Optical performance testing per ISO 8600-1, 3, 4, and 5. Mechanical testing included tensile strength, vacuum integrity, and vacuum performance.
Technological Characteristics
Materials: Polycarbonate (patient-contacting). Sensing: CMOS camera, LED light source. Energy: Electrical (external workstation). Dimensions: 170mm length, 5-9mm outer diameter. Connectivity: Wired cable to Endoscopic Workstation. Sterilization: Sterile, single-patient use. Standards: ISO 10993, IEC 60601, ISO 8600.
Indications for Use
Indicated for rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy.
Regulatory Classification
Identification
A vacuum abortion system is a device designed to aspirate transcervically the products of conception or menstruation from the uterus by using a cannula connected to a suction source. This device is used for pregnancy termination or menstrual regulation. This type of device may include aspiration cannula, vacuum source, and vacuum controller.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized eagle or bird, composed of three interconnected human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2015
Dongguan Microview Medical Technology Co. % Long Yang Company Representative Shenzhen Hlongmed Biotech Co., Ltd 15-08 East Block, Yihai Plaza, Chuangye Road North Nanshan District, Shenzhen P.R. China, 518054
Re: K140213
Trade/Device Name: Disposable Endoscopic Cannula Regulation Number: 21 CFR 884.5070 Regulation Name: Vacuum abortion system Regulatory Class: II Product Code: HGH Dated: February 10, 2015 Received: February 20, 2015
Dear Long Yang,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
### 510(k) Number (if known)
K140213
### Device Name
Disposable Endoscopic Cannula, model: RL1205, RL1206, RL1207, RL1208, RL1209
### Indications for Use (Describe)
The Disposable Endoscopic Cannula is indicated for rapid transcervical aspiration of the first trimester of pregnancy. It has an additional feature of visualization.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
_ | Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(K) number is: Date of Preparation: March 24, 2015
# 1. Submitter information
Manufacturer Name: DONGGUAN MICROVIEW MEDICAL TECHNOLOGY CO., LTD. Address: Galaxy Industrial Area, Qingxi, Dongguan, Guangdong, China Tel: +86-769-87738870-862 Fax: +86-769-87895932
# 2. Contact person
Long Yang (COO.) Shenzhen Hlongmed Biotech Co., Ltd. R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District,Shenzhen, P. R. China Tel: +86-755-86664986 Fax: +86-755-86664933 E-mail: yanglong@hlongmed.com
# 3. Device Information
Trade Name: Disposable Endoscopic Cannula Model: RL1205, RL1206, RL1207, RL1208, RL1209 Common Name: Uterine Cannulae Regulatory Class: II
| Classification Name | Product Codes | Regulation Number |
|------------------------|---------------|-------------------|
| Vacuum Abortion System | HGH | 21 CFR 884.5070 |
{5}------------------------------------------------
# 4. Predicative Device
| Item | Predicate Device |
|---------------------------|------------------------|
| Device Name | Rigid Uterine Cannulae |
| Common Name | Uterine Cannulae |
| Manufacturer | Doranne Frano |
| Classification regulation | 21CFR 884.5070 |
| Classification and | Class II |
| Product code | HGH |
| 510(k) number | K093508 |
# 5. Indications for Use
The Disposable Endoscopic Cannula is indicated for rapid transcervical aspiration of the uterine cavity during the first trimester of pregnancy. It has an additional feature of visualization.
# 6. Device Description
The Disposable Endoscopic Cannula is a sterile, single patient use device composed of transparent cover, suction orifice, dual-function cannula, handle, suction port, connection port, and data cable. The patient contacting components of this device are transparent cover, suction orifice and dual function cannula that are manufactured with polycarbonate.
The transparent cover contains a LED light source and CMOS camera at the distal end for visualization of the cervical canal and uterine cavity. This device has two channels, one for transcervical aspiration of the uterine cavity and the other for transmitting the image signal through a cable connected to the Endoscopic Workstation. The images captured by the camera are transmitted to the endoscopic workstation through the cable for display. When the embryo tissues are located, the operator sucks out the embryo tissues through the suction orifice connected to a vacuum source.
The Disposable Endoscopic Cannula is provided with five sizes based on outer cannula diameter: 5mm (RL1205 model), 6mm (RL1206 model), 7mm (RL1207 model), 8mm (RL1208 model), and 9mm (RL1209 model) (outer diameter). All models have a length of 170mm.
# 7. Summary of non-clinical data
Biocompatibility: The biocompatibility data from cytotoxicity, irritation and
{6}------------------------------------------------
sensitization testing demonstrated that the Disposable Endoscopic Cannula is in compliance with requirements in ISO 10993-5 and ISO 10993-10.
Electrical Safety and Electromagnetic Compatibility: Electrical Safety and Electromagnetic Compatibility data demonstrates that the device is in compliance with IEC 60601-1, IEC 60601-2-18 and IEC 60601-1-2.
Bench testing: Optical performance testing has demonstrated that this device is in compliance with ISO 8600-1, ISO 8600-3, ISO 8600-4, and ISO 8600-5. In addition, this device has been evaluated for tensile strength, vacuum integrity, and vacuum performance
| Parameter | Subject device (K140213) | Predicate device (K093508) |
|-----------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Device name | Disposable Endoscopic Cannula | Rigid Uterine Cannulae |
| Intended use | Same as the predicate, with<br>additional feature of visualization | For rapid transcervical aspiration of<br>the uterine cavity during the first<br>trimester of pregnancy |
| Design | Curved in shape two channels<br>One channel for aspiration<br>The other channel for data<br>transmission via cable | Straight or curved in shape<br>One channel for aspiration |
| Working length | Maximal 170 mm | Maximal 190 mm |
| Outer diameter | 5-9 mm | 6-12 mm |
| Material | Polycarbonate | Plastic (Styrenic copolymer resin) |
| Built in camera | Yes | No |
| Light source | Yes | No |
# 8. Comparison to Predicate Device
Substantial equivalent discussion:
- The Disposable Endoscopic Cannula and the predicate device have the same . intended use - uterine aspiration.
- The Disposable Endoscopic Cannula and the predicate device have the same . fundamental technological designs (shape, suction channel, etc.) and comparable dimensions.
- The Disposable Endoscopic Cannula is different from the predicate device in that . it has a light source, raising a safety concern. The difference does not raise new
{7}------------------------------------------------
types question because light source is common in other uterine devices (e.g., hysteroscopes). The difference can be assessed by accepted electrical safety, EMC, thermal safety and photoviological safety testing methods. The electrical safety, EMC, thermal safety and photobiological safety testing showed that the Disposable Endoscopic Cannula is safe.
- . Unlike the predicate device, the Disposable Endoscopic Cannula has a camera that can capture the image of uterine cavity. The imaging function of this device is supported by optical performance data including image resolution, distortion, depth of view, field of view, direction of view, etc.
- The Disposable Endoscopic Cannula uses different material, raising safety . concerns. There is no new type of question, because biocompatibility is a common question for patient-contacting medical devices. The biocompatibility testing showed that the Disposable Endoscopic Cannula is safe.
- To ensure mechanical performance, the Disposable Endoscopic Cannula was . evaluated for tensile strength, suction integrity, and suction performance. The results demonstrated that the Disposable Endoscopic Cannula is safe and effective.
In conclusion, the Disposable Endoscopic Cannula is substantially equivalent to the predicate device in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.