← Product Code [FMF](/productcode/FMF) · K140194

# NAMIC RCS SYRINGE (K140194)

_Navilyst Medical, Inc. · FMF · Apr 14, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K140194

## Device Facts

- **Applicant:** Navilyst Medical, Inc.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Apr 14, 2014
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The NAMIC RCS is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media and saline solutions during an angiographic procedure.

## Device Story

NAMIC RCS is a piston syringe used for intra-arterial or intra-venous administration of radiographic contrast media and saline solutions during angiographic procedures. Operated by clinicians in a hospital setting, the device facilitates the delivery of fluids required for imaging. It functions as a manual delivery tool, ensuring controlled administration of contrast agents to enhance diagnostic visualization. The device benefits patients by enabling accurate fluid delivery during angiography.

## Clinical Evidence

Bench testing only. Performance evaluation conducted in accordance with FDA guidance for piston syringes (April 1993), ISO 7886-1:1997 (sterile hypodermic syringes), and ISO 594-2:1998 (luer lock fittings).

## Technological Characteristics

Piston syringe; manual operation. Materials, design, and components are consistent with predicate devices. Features 6% (Luer) taper lock fittings per ISO 594-2:1998. Designed for single use. Sterilization method not specified.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- [K113198](/device/K113198.md)
- [K875196](/device/K875196.md)
- [K873955](/device/K873955.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Navilyst Medical. Inc. NAMIC RCS. Abbreviated 510(k) January 24, 2014

A. Sponsor

#### K140194

#### 510(K) SUMMARY FOR THE NAMIC RCS

## APR 1 4 2014

Date prepared: January 24, 2014

- Navilyst Medical, Inc 26 Forest Street Marlborough, MA 01752 B. Contact Brandon M. Brackett OR Wanda Carpinella Specialist, Global Regulatory Affairs Director of Global Regulatory Affairs 508-658-7984 508-658-7929 C. Device Name Trade Name NAMIC RCS Common/Usual name: Piston Syringe Classification Name: Syringe, Piston (21CFR§880.5860. Class II) Classification Panel: General Hospital D. Predicate Device(s) Common/Usual name: Piston Syringe Classification Name Syringe, Piston - 21CFR\$880.5860. Class H Classification Panel: General Hospital
#### E. Device Description

Premarket Notification

#### Intended Use

The NAMIC RCS is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media and saline solutions during an angiographic procedure.

### F. Summary of Similarities and Differences in Technological Characteristics and Performance

The proposed NAMIC RCS syringes incorporate similar materials, design, components, technological characteristics, and intended us as the predicate syringes.

K113198, K875196, K873955

#### G. Performance Data

The NAMIC RCS is substantially equivalent to the predicate devices based on comparison of technological characteristics and the results of non-clinical tests which included the performance evaluation conducted in accordance with the following FDA guidance documents, international standards, and testing which included:

- . FDA's "Guidance on the Content of Premarket Notification 510(k) Submissions for Piston Syringes dated April 1993"
- . ISO 7886-1:1997 "Sterile Hypodernic Syringes for Single Use -- Par1 1: Syringes for Manual Use'
- . ISO 594-2:1998 "Conical Fittings 6% (Luer) Taper for Syringes, Needles, Certain Other Medical Equipment - Part 2: Lock Fittings"

#### H. Conclusion

The results of the non-clinical testing and a comparison of similarities and differences demonstrate that the proposed and predicate devices are substantially equivalent.

{1}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three arms or lines extending upwards. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the left side of the logo.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 14, 2014

Navilyst Medical, Inc. Brandon M. Brackett Global Regulatory Affairs Specialist 26 Forest Street Marlborough, MA 01752

Re: K140194

Trade/Device Name: NAMIC RCS Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: January 24, 2014 Received: January 27, 2014

Dear Mr. Brackett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical

{2}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Erin I. Keith -S

Erin 1. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K140194

Device Name NAMIC RCS

Indications for Use (Describe)

The NAMIC RCS is intended to be used for the intra-venous administration of radiographic contrast media and saline solutions during an angiographic procedure.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Digitally signed by Richard C. Chapman Date: 2014.04.14 08:59:45 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K140194](https://fda-staging.innolitics.com/device/K140194)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
