The single-use disposable, non-sterile Koala TOCO is intended for use in conjunction with standard fetal monitors for the evaluation of relative external uterine activity during antepartum and intrapartum periods in a clinical setting. The Koala IUP/TOCO Reusable Cable is also for use with the Koala Intrauterine Pressure Catheter for intrapartum, intrauterine pressure monitoring. The Koala TOCO and Koala IUP/TOCO Reusable Cable are not intended for home monitoring of pre-term labor.
Device Story
The Koala TOCO is a single-use, pneumatic tocodynamometer that detects maternal abdominal changes during uterine contractions. It features an elastic membrane over a guard-ring that traps air; contraction-induced pressure changes are transmitted via a low-volume air channel to a pressure transducer housed within the Koala IUP/TOCO Reusable Cable. The system is used in clinical settings by healthcare providers in conjunction with standard fetal monitors. The reusable cable also supports intrauterine pressure monitoring when connected to the Koala Intrauterine Pressure Catheter. Output is displayed on compatible fetal monitors (Philips/Corometrics), allowing clinicians to assess contraction frequency and duration to guide labor management. The device benefits patients by providing non-invasive monitoring of uterine activity.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing, including predicate comparison, simulated use, hardware testing, electrical performance, biocompatibility (ISO 10993-1), and IUPC compatibility testing.
Technological Characteristics
Thermoplastic polyurethane housing and cable; pneumatic sensing diaphragm; Wheatstone resistive bridge pressure transducer; 10-foot cable; non-sterile; single-use TOCO; reusable cable; compatible with standard elastic belts; non-waterproof.
Indications for Use
Indicated for pregnant patients, especially during labor, for evaluation of external uterine activity during antepartum and intrapartum periods in clinical settings. Contraindicated for home monitoring of pre-term labor.
Regulatory Classification
Identification
An external uterine contraction monitor (i.e., the tokodynamometer) is a device used to monitor the progress of labor. It measures the duration, frequency, and relative pressure of uterine contractions with a transducer strapped to the maternal abdomen. This generic type of device may include an external pressure transducer, support straps, and other patient and equipment supports.
{0}------------------------------------------------
# MAY 2 9 2014
# 510(K) SUMMARY FOR THE CLINICAL INNOVATIONS, LLC Koala TOCO and Koala IUP/TOCO Reusable Cable
(per 21CFR 807.92 and http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm)
# 1. SUBMITTER/510(K) HOLDER
Clinical Innovations, LLC 747 West 4170 South Murray, UT 84123 Phone: 1-801-268-8200 Internet: www.clinicalinnovations.com Establishment Registration No: 1722684
Contact: Tom Haueter Contact's Phone: 1-801-268-8200 Contact's Fax: 1-801-263-7373 Contact's Email: T.Haueter@clinicalinnovations.com
## 2. DEVICE NAME
| Proprietary Name: | Koala TOCO and Koala IUP/TOCO Reusable Cable |
|------------------------|------------------------------------------------------|
| Common/Usual Name: | Tocodynamometer and cable |
| Classification Name: | external uterine contraction monitor and accessories |
| Classification Panel: | Obstetrics/Gynecology |
| Device Class: | Class II |
| Classification Number: | 21 CFR 884.2720 |
| Product Code: | HFM, HFN |
# 3. PREDICATE DEVICE
The following device is a legally marketed device to which equivalence is being claimed:
- . FeatherLite™ Tocodynamometer (Ventrex, Inc., K013477)
# 4. DEVICE DESCRIPTION
The Koala TOCO is a single-use tocodynamometer that is a transducer pressuresensing device that can detect the changes in a mother's abdomen as her uterus tightens during a contraction. The device detects how often contractions occur and the length of each.
The Koala TOCO device is a pneumatic tocodynamometer that comprises a guardring with a thin elastic membrane stretched across a shallow depression in the center
{1}------------------------------------------------
of the guard-ring. The elastic membrane traps a small volume of air in the depression. The air volume beneath the membrane of the Koala TOCO is connected via a low volume air channel to a pressure transducer in the Koala IUP/TOCO Reusable Cable. The operational response of the Koala TOCO is substantially equivalent to that of the standard guard-ring TOCO. This requires that the Koala TOCO have essentially the same physical shape, guard-ring area and pressure sensing area of the predicate devices. The pressure sensing area will present resistance to applied pressure. In other words, the displacement of air in the elastic membrane mimics the compression of the strain gauge of the guard-ring TOCO.
The Koala IUP/TOCO Reusable Cable can be used for intrauterine pressure monitoring with the Koala Intrauterine Pressure Catheter. The Koala TOCO and the Koala IUP/TOCO Reusable Cable are compatible with Philips and Corometrics monitors.
#### INDICATIONS FOR USE / INTENDED USE 5.
The single-use disposable, non-sterile Koala TOCO is intended for use in conjunction with standard fetal monitors for the evaluation of relative external uterine activity during antepartum and intrapartum periods in a clinical setting.
The Koala IUP/TOCO Reusable Cable is also for use with the Koala Intrauterine Pressure Catheter for intrapartum, intrauterine pressure monitoring.
The Koala TOCO and Koala IUP/TOCO Reusable Cable are not intended for home monitoring of pre-term labor.
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE 6. DEVICE
A summary comparison of the technological characteristics, including design and materials, is provided in the table below:
| Parameter | Koala TOCO and Koala IUP/TOCO<br>Reusable Cable<br>(Clinical Innovations) | FeatherLite™ Tocodynamometer<br>(Ventrex Inc.) |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K140163 | K013477 |
| Indications for Use | For use in conjunction with standard fetal monitors for the evaluation of external uterine activity during antepartum and intrapartum periods in a clinical setting. | For use in conjunction with standard fetal monitors for the evaluation of external uterine activity during antepartum and intrapartum periods. It should only be used in a clinical setting. |
| Target Patient Population | Pregnant patients, especially during labor | Pregnant patients, especially during labor |
Clinical Innovations, LLC
{2}------------------------------------------------
| Parameter | Koala TOCO and Koala IUP/TOCO<br>Reusable Cable<br>(Clinical Innovations) | FeatherLite™ Tocodynamometer<br>(Ventrex Inc.) |
|-------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Water environment use | No | No |
| Anatomical sites | Surface of abdomen over fundal area of<br>the uterus | Surface of abdomen over fundal area of<br>the uterus |
| Size | Body diameter: 7.0cm<br>Body thickness: 0.8cm | Body diameter (sensing dome): 3.6cm<br>Body thickness: .10cm<br>Sensing dome thickness: .79cm |
| Weight (with cable) | 0.35 Ibs (159 g) | 0.11 lbs (51g) |
| Material | Housing material: thermoplastic<br>polyurethane<br>Cable: Thermoplastic polyurethane | Housing material: thermoplastic<br>Sensing nipple: flexible polymer<br>Cable: vinyl clad |
| Technology | Sensing diaphragm connected to a<br>pressure transducer centrally located in a<br>circular (guard ring) body | Sensing nipple attached to a strain gauge |
| Sensor | Wheatstone resistive bridge sensor | Wheatstone resistive bridge sensor |
| Location of electronics | Housed in the reusable cable | Housed in the TOCO device |
| Serviceable | No | No |
| Patient attachment | Elastics belt included.<br>Compatible with standard elastic belts. | Elastic belt included.<br>Compatible with standard elastic belts. |
| Cable length | 10 feet | 8 feet |
| Single Use/Disposable | Yes | Yes |
| Biocompatible | Yes | Yes |
| Sterile | Non-sterile | Non-sterile |
# 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Non-clinical testing performed to support the performance and safety claims for the Koala TOCO and the Koala IUP/TOCO Reusable Cable includes, but is not limited to, the following:
- Predicate comparison testing .
- Simulated use testing .
- Hardware testing .
- Electrical performance testing .
- Biocompatibility testing according to ISO 10993-1 .
- . Intrauterine pressure catheter (IUPC) compatibility testing
The evidence resulting from the non-clinical testing supports the substantial equivalence of the proposed and predicate devices.
{3}------------------------------------------------
# 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Not Applicable.
# 9. SUMMARY OF OTHER INFORMATION
This submission included proposed product labeling and packaging specifications.
# 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
The indications for use, principles of operation, and technological characteristics of the proposed Koala TOCO with the Koala IUP/TOCO Reusable Cable are substantially equivalent to the predicate device FeatherLite™ Tocodynamometer (subject of K013477). Differences between the proposed and predicate device are limited to minor differences in visual appearance, weight, sensing technology, and transducer location. These differences are minor and do not impact the safety and effectiveness of the device.
The safety, performance, and effectiveness of the Koala TOCO for its intended use are demonstrated by non-clinical testing. Based on the evidence provided, Clinical Innovations believes that the proposed Koala TOCO and Koala IUP/TOCO Reusable Cable are substantially equivalent to the predicate.
{4}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2014
Clinical Innovations, LLC % JoAnne L. Bronikowski Consultant Aptiv Solutions 62 Forest Street, Suite 300 Marlborough, MA 01752
Re: K140163
> Trade/Device Name: Koala TOCO and Koala IUP/TOCO Reusable Cable Regulation Number: 21 CFR§ 884.2720 Regulation Name: External uterine contraction monitor and accessories Regulatory Class: II Product Code: HFM, HFN Dated: April 29, 2014 Received: April 30, 2014
Dear JoAnne L. Bronikowski,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - JoAnne L. Bronikowski
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on fast page.
510(k) Number (if known) K140163
Device Name
Koala TOCO and Koala IUP/TOCO Reusable Cable
## Indications for Use (Describe)
The single-use disposable, non-sterile Koala TOCO is in conjunction with standard fetal monitors for the evaluation of relative external uterine activity during antepartum and intrapartum periods in a clinical setting.
The Kosla TUP/TOCO Reusable Cable is also for use with the Koala Intrauterine Pressure Catheter for intrapartum, intrauterine pressure monitoring.
The Koala TOCO and Koala IUP/TOCO Reusable Cable are not intended for home monitoring of pre-term labor.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
: : FOR FDA USE ONLY :: 、人民和人民族、据明确的人才和用品。
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# 2014
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.