K140155 · Cardinal Health200, LLC · FXX · Jun 20, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140155
Device Name
SECURE-GARD SURGICAL AND PROCEDURE FACE MASK
Applicant
Cardinal Health200, LLC
Product Code
FXX · General, Plastic Surgery
Decision Date
Jun 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Cardinal Health Secure-Gard® surgical and procedure masks are intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Secure-Gard® surgical and procedure masks are single use, disposable devices provided non-sterile.
Device Story
Secure-Gard surgical and procedure masks are four-layer, single-use, disposable face masks. Construction includes inner nonwoven cellulose, two middle layers of nonwoven polyolefin melt blown and polyester/polyethylene blend media, and an outer nonwoven polyolefin spunbond layer. Layers are sonically welded; masks feature malleable nosepieces for fit. Variants include ties or earloops, with optional integrated eye shields or anti-fog foam strips. Used in clinical settings by healthcare personnel to provide a barrier against microorganisms, blood, body fluids, and airborne particulates. Masks are worn over the nose and mouth to reduce exposure risk for both patient and wearer.
Clinical Evidence
Bench testing only. Performance evaluated against ASTM F2100-11 Level 3 standards. Results: Synthetic blood penetration (160 mmHg) passed; BFE 99.4-99.7%; PFE (0.1 micron) 98.6-99.1%; Delta P 3.1-3.5 mmH2O/cm²; Flammability Class 1. All 32/32 samples passed.
Technological Characteristics
Four-layer construction: nonwoven cellulose (inner), nonwoven polyolefin melt blown (media), nonwoven polyester/polyethylene blend (media), polyolefin spunbond (outer). Sonic welding. Malleable nosepiece. Ties or polyester-spandex earloops. Tested to ASTM F2100-11 Level 3. Biocompatibility per ISO 10993. Non-sterile.
Indications for Use
Indicated for operating room personnel and general healthcare workers to protect patients and wearers against transfer of microorganisms, blood, body fluids, and airborne particulates. Single-use, disposable, non-sterile.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
JUN 2 0 2014
K140155
Image /page/0/Picture/2 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The company name, "CardinalHealth", is written in a simple, sans-serif font.
1500 Waukegan Road
Waukegan, IL 60085
www.cardinalhealth.com
510(k) SUMMARY Secure-Gard® Surgical and Procedure Masks
Manufacturer:
Cardinal Health 200, LLC 1500 Waukegan Road Waukegan, IL 60085
> 1500 Waukegan Road Waukegan, IL 60085
Lavenia Ford
(847) 887-3323
(847) 887-2461
Surgical Mask
FXX
Regulatory Affairs Contact:
Telephone Number:
Fax Number:
Date summary Prepared: May 21, 2014
Trade Name:
Secure-Gard® Surgical Mask
Secure-Gard® Procedure Mask
Regulation Number/Device Class: Class II per 21 CFR § 878.4040
Regulation Name: Surgical Apparel
Common Name:
Procedure Mask
Product Code:
Predicate Device:
K111402 -Kimberly-Clark, KC300 Face Mask(s)
{1}------------------------------------------------
### Description
The Cardinal Health Secure-Gard® surgical and procedure masks are identified by Regulation 21 CFR 878.4040 with product code FXX.
The Cardinal Health Secure-Gard® surgical and procedure masks are all the same four-layer masks constructed of 1 layer of nonwoven cellulose (inner facing), 1 layer of nonwoven polvolefin melt blown (media). 1 laver of nonwoven polyester/polvethylene blends (media), and 1 layer polyolefin spunbond material (outer facing). The four layers of the mask body are collated and sonically welded around the edges to enclose the filter media. The mask is provided with nonwoven polyolefin ties or polyester-spandex earloops (both attached by ultrasonic welding). A malleable nosepiece is placed within the binding for comfort and individualized fit around the wearer's nose. The surgical masks will be provided with or without an eye shield. Cardinal Health surgical and procedure masks are a single use, disposable device provided non-sterile
#### Indications for Use
Cardinal Health Secure-Gard® surgical and procedure masks are intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Secure-Gard® surgical and procedure masks are single use, disposable devices provided non-sterile.
This submission covers 7 catalog numbers of Secure-Gard® surgical and procedure masks models, see Table 1 below. Each model is the same four-laver mask constructed of 1 laver of nonwoven cellulose (inner facing), 1 layer of nonwoven polyolefin melt blown (media), 1 layer of nonwoven polyester/polyethylene blends (media), and 1 layer polyolefin spunbond material (outer facing) and has been tested according to ASTM 2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks.
| Catalog # | Trade Name | Design Features | Color |
|-----------|-----------------------------|------------------------------------------------------------|-----------------|
| AT74535 | Secure-Gard® surgical mask | Anti-fog foam strip<br>and ties | Teal and Purple |
| AT74635 | Secure-Gard® surgical mask | Anti-fog foam strip,<br>eye shield, and ties | Teal and Purple |
| AT74635-1 | Secure-Gard® surgical mask | Anti-fog foam strip,<br>anti-glare eye shield,<br>and ties | Teal and Purple |
| AT744235 | Secure-Gard® surgical mask | Light weight anti-fog<br>strip and ties | Teal and Purple |
| AT744335 | Secure-Gard® surgical mask | Light weight anti-fog<br>strip with eye shield<br>and ties | Teal and Purple |
| AT74531 | Secure-Gard® procedure mask | Earloops | Teal and Purple |
| AT74631 | Secure-Gard® procedure mask | Eye shield and<br>earloops | Teal and Purple |
#### Table 1: Product Description and Catalog Number
{2}------------------------------------------------
## Device and Predicate Device Technical Characteristics
The Cardinal Health Secure-Gard® surgical and procedure face masks are substantially equivalent to the predicate device Kimberly-Clark KC300 of K111402 in intended use and principles of operation see Table 3 below. The Cardinal Health Secure-Gard® surgical and procedure masks and predicate masks meet ASTM F2100-11 as shown below in Table 11.
| Element of Comparison | Kimberly-Clark (KC300) | Cardinal Health Secure-Gard® Surgical and Procedure Masks |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Kimberly Clark KC300 Face Mask(s) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Kimberly Clark KC300 face mask(s) is a single use, disposable device(s), provided non-sterile. | <b>Secure-Gard Surgical Mask</b><br>Cardinal Health Secure-Gard® surgical masks are intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Secure-Gard® surgical masks are single use, disposable devices provided non-sterile<br><br><b>Secure-Gard Procedure Mask</b><br>Cardinal Health Secure-Gard® procedure masks are intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Secure-Gard® procedure masks are single use, disposable devices provided non-sterile |
| Material Composition | 4 layer mask made from nonwoven polyester blends and polypropylene materials | <b>Secure-Gard Surgical Mask</b><br>4 layer mask made from 1 layer of nonwoven cellulose (inner facing), 1 layer of nonwoven polyolefin melt blown (media), 1 layer of nonwoven polyester/polyethylene blends (media), and 1 layer polyolefin spunbond material (outer facing)<br><br><b>Secure-Gard Procedure Mask</b><br>4 layer mask made from 1 layer of nonwoven cellulose (inner facing), 1 layer of nonwoven polyolefin melt blown (media) 1 |
| Table 3: Comparison of Predicate device and Cardinal Health Secure-Gard® Surgical and | | | | |
|---------------------------------------------------------------------------------------|--|--|--|--|
| Procedure masks | | | | |
{3}------------------------------------------------
| | | layer of nonwoven<br>polyester/polyethylene blends<br>(media), and 1 layer polyolefin<br>spunbond material (outer facing) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Specifications and Dimensions | KC300 Surgical Mask<br>7.0"x3.7" | Secure-Gard Surgical Mask<br>AT74535 - 7"x4" |
| | KC300 Procedure Mask<br>6.9"x3.8" | AT74635 - 7 ¼"x4 1/8"<br>AT74635-1 - 7 ¼"x4 1/8"<br>AT744235 - 7"x4"<br>AT744335 - 7"x4" |
| | | Secure-Gard Procedure Mask<br>AT74531 - 7"x3 3/8"<br>AT74631 - 7"x3 3/8" |
| | | See Appendix I for mask<br>specifications |
| Mask Style | Pleated | Secure-Gard Surgical Mask<br>Pleated |
| | | Secure-Gard Procedure Mask<br>Pleated |
| Design Features | KC300 Surgical Mask<br>Fluidshield Fog-Free | Secure-Gard Surgical Mask<br>AT74535 - Anti-fog foam strip<br>and Ties<br>AT74635 - Anti-fog foam strip,<br>Eye shield, and Ties<br>AT74635-1 - Anti-fog foam strip,<br>Anti-glare eye shield, and Ties<br>AT744235 - Lightweight anti-fog<br>strip and Ties<br>AT744335 - Lightweight anti-fog<br>strip, Eye shield, and Ties |
| | KC300 Procedure Mask<br>Fluidshield Fog-Free | Secure-Gard Procedure Mask<br>AT74531 - Earloops<br>AT74631 - Eye shield and<br>Earloops |
| Physical Testing | Predicate device was tested<br>according to ASTM F2100-11<br>Level 3 in previous 510(k)<br>submission K111402. | Secure-Gard Surgical Mask<br>Device was tested in accordance<br>with ASTM F2100-11, and meets<br>Level 3 requirements |
| | | Secure-Gard Procedure Mask<br>Device was tested in accordance<br>with ASTM F2100-11, and meets<br>Level 3 requirements |
| | | See Appendix G for all test data |
| Biocompatibility | Predicate device was tested<br>according ISO 10993 | Secure-Gard Surgical Mask<br>Device was tested in accordance<br>to ISO 10993 and passed |
-
'
. .
:
{4}------------------------------------------------
| acceptance criteria |
|-----------------------------------------------------------------------------------------------------------------|
| Secure-Gard Procedure Mask<br>Device was tested in accordance<br>to ISO 10993 and passed<br>acceptance criteria |
| Appendix H for all test data |
## Summary of Testing
The Cardinal Health Secure-Gard® surgical and procedure masks have been tested according to ASTM 2100-11 and standards which comprise ASTM F2100-11, Standard Specification for Performance of Materials Used in Medical Face Masks, see Table 11 below.
# Table 11: Summary of ASTM F2100-11 Testing - Cardinal Health Secure-Gard® Surgical and Procedure Masks compared to Predicate
| Test | Acceptance<br>Criteria per<br>ASTM F2100-<br>11 Level 3<br>(AQL = 4.0%) | Predicate<br>device test<br>results<br>K111402<br>(KC300) | Current device test results | |
|------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------|-----------------------------|-----------------------|
| | | | ASTM<br>F2100-11<br>Level 3 | Average |
| ASTM F1862<br>Synthetic<br>Blood | 160 mmHg | Pass | 32/32 Pass | Not Applicable |
| ASTM F2101<br>BFE | ≥ 98% | Pass | 32/32 Pass | 99.4 - 99.7% |
| ASTM F2299<br>PFE at 0.1<br>micron | ≥ 98% | Pass | 32/32 Pass | 98.6 - 99.1% |
| Mil-M-<br>36954C<br>Delta P | < 5.0<br>mmH₂0/cm² | Pass | 32/32 Pass | 3.1- 3.5<br>mmH₂0/cm² |
| CPSC 1610<br>Flammability | Class 1 | Pass | 32/32 Pass<br>Class 1 | Not Applicable |
All results of testing met ASTM F2100-11 Level 3 acceptance criteria. ASTM F2100-11 Level 3 was performed at an AQL of 4% (32 samples tested, accept on 3 failures and reject on 4 failures).
## Summary of Testing
Based on the results of the biocompatibility and physical performance testing the Secure-Gard® surgical and procedure masks are as safe and as effective for their intended use as the predicate. The Secure-Gard® surgical and procedure masks are substantially equivalent to the predicate device, in terms of general intended use performance testing, material composition, configurations/dimensions, and safety and effectiveness.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
June 20, 2014
Cardinal Health 200, LLC Ms. Lavenia Ford Manager, Regulatory Affairs 1430 Waukegan Road Waukegan, Illinois 60085
Re: K140155
Trade/Device Name: Secure-Gard® Surgical and Procedure Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Mask Regulatory Class: Class II Product Code: FXX Dated: May 21, 2014 Received: May 22, 2014
Dear Ms. Ford,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/5/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an abstract human figure with three lines representing the body, arms, and head. The graphic is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
{6}------------------------------------------------
Page 2 - Ms. Ford
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDeyices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Tejashri Purohit-Sheth Purohit Sheth M.D M.D. Cusical Deputy Director
GRODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140155
Device Name
Secure-Gard® surgical and procedure masks
Indications for Use (Describe)
Cardinal Health Secure-Gard® surgical and procedure masks are intended to be worn by operating room personnel and other general healthcare workers to protect both patients against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Secure-Gard® surgical and procedure masks are single use, disposable devices provided non-sterile.
This submission covers 7 catalog numbers of Secure-Gard® surgical and procedure masks models, see Table 1 below. Each model is the same four-layer mask constructed of I layer of nonwoven cellulose (inner facing), 1 layer of nonwoven polyolefin melt blown (media), 1 layer of nonwoven polyester/polyethylene blends (media), and 1 layer polyolefin spunbond material (outer facing) and has been tested according to ASTM F2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks
Table 1: Product Description and Catalog Number
| Catalog # | Trade Name | Design Features | Color |
|-----------|-----------------------------|------------------------------------------------------|-----------------|
| AT74535 | Secure-Gard® surgical mask | Anti-fog foam strip and ties | Teal and Purple |
| AT74635 | Secure-Gard® surgical mask | Anti-fog foam strip, eye shield, and ties | Teal and Purple |
| AT74635-I | Secure-Gard® surgical mask | Anti-fog foam strip, anti-glare eye shield, and ties | Teal and Purple |
| AT744235 | Secure-Gard® surgical mask | Light weight anti-fog strip and ties | Teal and Purple |
| AT744335 | Secure-Gard® surgical mask | Light weight anti-fog strip with eye shield and ties | Teal and Purple |
| AT74531 | Secure-Gard® procedure mask | Earloops | Teal and Purple |
| AT74631 | Secure-Gard® procedure mask | Eye shield and earloops | Teal and Purple |
Type of Use (Select one or both, as applicable)
L Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
A MARK LE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR THE CHILL Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sreekanth Gu
ly signed by Sreekanth Gutala -S
JS, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 9200300.100.1.1=2000540490. cn=Sreekanth Gutala -S 2044.06.19 21:59:43 -04'00'
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.