O2 Nasal cannula with CO2 monitoring is intended for use in patients who require supplemental oxygen delivery and CO2 monitoring.
Device Story
Device is a nasal cannula for simultaneous supplemental oxygen delivery and CO2 gas sampling. Input: oxygen source and patient exhaled breath. Output: delivered oxygen to patient; sampled CO2 gas to external capnograph. Used in hospital/emergency settings by clinicians. Device consists of co-extruded O2/CO2 tubing, cannula prongs, luer connectors, tube slide, Y-piece, and optional hydrophilic filter. Connects to external capnograph for CO2 monitoring. Benefits patient by enabling continuous respiratory monitoring during oxygen therapy. No software or electronic components; purely mechanical device.
Clinical Evidence
Bench testing only. Performance verified against BS EN 13544-2:2002+A1:2009 and ISO 594-1/ISO 594-2 standards. Tests included visual inspection, nipple strength, tubing flow resistance, tensile strength, kinking resistance, dimensional inspection, gauging, and leakage/separation force testing. All samples passed.
Indicated for adult patients requiring supplemental oxygen delivery and CO2 monitoring.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the word "flex" in a stylized font. The "f" is designed with two curved lines, and the rest of the letters are in a bold, sans-serif typeface. Below the word "flex" are the words "TOTAL QUALITY" in a smaller, sans-serif font.
# 510(k) Summary
Trade name
C.F.R Section Product Code:
## Manufacturer:
O2 Nasal cannula with CO2 monitorin
21 CFR 868.1400. Carbon Dioxide ga CCK
Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom
Christopher Watkins Quality Regulatory & Technical Direct Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom
00 44 1443 471 593
March 2014
Nasal cannula with CO2 monitoring
Class 2
Unomedical - Mac-Safe (was Hospita sampling/oxygen delivery cannula") w marketing by 510(k) No K915228.
02 Nasal cannula with CO2 monitorin that is intended to situate under the p prongs of the cannula inserted into pa oxygen is administered and CO2 mor The O2 Nasal cannula with CO2 mon administaring pronos tuto slide co-e
Telephone:
Date Summary Prepared:
Regulatory Affairs Contact:
Common Name:
Classification:
Predicate Device
Description:
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a stylized logo. The logo features a stylized letter 'f' with curved lines, followed by the letters 'le'. Below the letters 'le' is the word 'TOTAL'. The logo appears to be for a company or organization, but without additional context, the specific nature of the company is unclear.
The O2 Nasal cannula with CO2 mon extruded tubes
# Part Numbers:
| | Nasal cannula with CO2 monitoring |
|--------------|----------------------------------------------------------|
| 020-10-123U | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - MALE PIGT |
| 020-10-124U | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - FEMALE PI |
| 020-10-125U | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - MALE LUE |
| 020-10-125AU | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - MALE LUE |
| 020-10-126U | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - FEMALE L |
| 020-10-126AU | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - FEMALE L |
| 020-10-127U | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - MALE LUE |
| 020-10-128U | O2 NASAL CANNULA WITH CO2 MONITORING - ADULT - FEMALE L |
Intended Use:
02 Nasal cannula with CO2 monitoring is patients who require supplemental oxyge monitoring
Substantial Equivalence
02 Nasal cannula with CO2 monitorir use as predicate device:
- Unomedical Mac-Safe (was Hospita . sampling/oxygen delivery cannula") w marketing by 510(k) No K915228.
Both O2 delivering/CO2 monitoring de use devices.
Neither device is life supporting or life
Neither device uses software/ and are
Both O2 delivering/CO2 monitoring de sterile in individual poly bags.
Both O2 delivering/CO2 monitoring de came intended use in the same intent
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with a stylized letter "f" in a bold, black font. The letter "f" is composed of two parallel lines that curve and extend upwards. To the right of the "f", the letters "le" are visible, also in a bold, black font. Below the letters, the word "TOTAL" is written in a smaller, sans-serif font.
| | FlexicareO2 Nasal cannula with CO2<br>monitoring<br>K: Unknown | Unomedical - M<br>CO2 gas sampling<br>was c |
|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Conforms to: | Yes | |
| BS EN 13544-2:<br>2002 +A1:2009<br>Annex A4<br>And<br>BS EN 13544-2:<br>2002 +A1:2009<br>Annex A5 | | |
| Components | Co extruded O2/CO2 tubing<br>luer connector (male/female)<br>Cannula delivery/cannula prongs<br>O2 Connector<br>Tube slide<br>Y-piece<br>Venturi barrel<br>Hydrophilic filter (option) | Co ex<br>luer c<br>Cannula |
| Materials | PVC<br>Polypropylene<br>ABS | |
| Assembly<br>method | Bonded - Solvent adhesive | Bon |
| Target<br>population | Adult | |
| Home use | No | |
| Hospital/<br>emergency use | Yes | |
| Connectable to<br>Capnograph? | Yes | |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with a stylized letter "f" that appears to be part of a word. The letters are thick and bold, with a unique design. Below the letter, there is the word "TOT" in smaller, sans-serif font.
| Colour/ size/<br>material | Devices displayed a high level of substantial equivalence. All<br>within 0.1mm of each other. ID's are classed as critical dir<br>resistance/rate through tub<br>All materials are the same, with any differences being min<br>e.g. colour/finish. | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| ID/ OD/ length | ID: 3.9mm | | |
| 02 tubing | OD: 5.0mm | | |
| | Length: 2100mm | | |
| CO2 tubing | ID: 1.4 | | |
| | OD: 2.7 | | |
| | Length: 2100mm | | |
Tests Performed:
| Test | Standard? / In-House? |
|----------------------------------------------------------------|-----------------------------------------|
| Visual inspection | In-House |
| Nipple dimensions (on venturi barrel<br>supplied with cannula) | |
| Strength of nipple(on venturi barrel<br>supplied with cannula) | BS EN13544-2: 2002 +A1:2009<br>Annex A5 |
| Tubing flow resistance | |
| O2 Connector to tubing tensile strength | |
| O2 Connector to nipple tensile strength | |
| Tube resistance to kinking | |
| Dimensional inspection of luer conical | ISO 594-1 / BS EN 20594-1:1994 |
| Dimensional inspection of luer conical | |
| Gauging tests on luer | |
| Liquid leakage from luer | |
| Air leakage from luer | |
| Luer separation force | ISO 594-2 (BS EN 1707:1997) |
| Luer unscrewing torque | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a stylized logo, possibly for a company or brand. The logo features a bold, stylized letter 'f' with a double-line design, giving it a unique and modern appearance. Below the 'f', there is a smaller word, 'TOTAL', in capital letters, suggesting that the logo might be associated with a company named 'Total' or a product line with that name.
All Samples passed the performance testing when tested against methods House test methods and relevant BS EN standards.
The results of this testing show that the O2 Nasal cannula with CO2 monite performance tests and performs at least as well as marketed predicate dev Unomedical.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a partial view of a logo or emblem. On the left side, there is a circular seal with text around the perimeter, including the words "HUMAN SERVICES" and "USA". Inside the seal, there is a stylized design that resembles an abstract representation of a human figure or a symbol related to health and human services. On the right side of the image, the word "DEPAR" is visible, suggesting that it is part of a larger word, likely "DEPARTMENT".
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6, 2014
Flexicare Medical Limited C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25TH Street NW Buffalo, MN 55313
Re: K140113
Trade/Device Name: 02 Nasal Cannula with CO2 Monitoring Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: February 18, 2014 Received: February 21, 2014
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence detern that FDA has made a determination that your device complies with other or any Federal statutes and regulations administered by other Federal agen comply with all the Act's requirements, including, but not limited to: regi CFR Part 807); labeling (21 CFR Part 801); medical device reporting (rep device-related adverse events) (21 CFR 803); good manufacturing practic forth in the quality systems (QS) regulation (21 CFR Part 820); and if app product radiation control provisions (Sections 531-542 of the Act); 21 CF
If you desire specific advice for your device on our labeling regulation (2 contact the Division of Small Manufacturers, International and Consumer free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.ht the regulation entitled, "Misbranding by reference to premarket notification 807.97). For questions regarding the reporting of adverse events under th CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under Division of Small Manufacturers, International and Consumer Assistance (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.h
Sincerely yours,
Image /page/6/Picture/5 description: The image shows a logo with the text "Tejashri Purohit-Sheth, M.D., Clinical Deputy Director, DAGRID". The logo is black and white and has a stylized design. The text is arranged in a way that it is easy to read. The logo is likely for a medical or clinical organization.
Erin I. Keith, M.S. Acting Director
{7}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on lest page.
510(k) Number (if known) K 1401 13
Device Name
..
O2 Nasal Cannula with CO2 monitoring
#### Indications for Use (Describe)
O2 Nasal Cannula with CO2 monitoring is intended for use in Adult patients who require supplemental oxygen delivery and CO2 monitoring
Type of Use (Select one or both, as epplicable)
. . ..
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/14 description: The image shows a combination of text and symbols. The text includes 'Todd D', '2014.03.06', and '-05'00'. There are also some illegible symbols and characters present in the image. The text and symbols are arranged in a somewhat disorganized manner.
FORM FDA 3881 (1/14)
Page Lot 1
es (104) 445-6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.