K140099 · Seedings Life Science Ventures, LLC · EZL · Mar 4, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K140099
Device Name
LIGHTHOUSE URINARY CATHETER
Applicant
Seedings Life Science Ventures, LLC
Product Code
EZL · Gastroenterology, Urology
Decision Date
Mar 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lighthouse™ Urinary Catheter is intended for bladder/urinary tract drainage during prostate biopsy and brachytherapy procedures where transrectal ultrasound (TRUS) imaging is used to visualize the prostate. The catheter additionally enhances identification of the course of the posterior aspect of the prostatic urethra during TRUS-guided prostate biopsy and brachytherapy procedures by creating an acoustic interface (i.e., air) between the catheter and the urethral wall via its urethral identification bladder.
Device Story
Lighthouse™ Urinary Catheter is a 3-lumen Foley catheter designed for use during prostate biopsy and brachytherapy. It features a standard Foley balloon lumen, a urine drainage lumen, and a third lumen connected to a Urethral Identification Bladder (UIB). The UIB is a 4-cm bladder along the shaft that holds 0.1cc–1.0cc of air. When inserted, the air-filled UIB creates an acoustic interface between the catheter and the urethral wall, providing an echogenic marker visible under TRUS imaging. This allows clinicians to identify the posterior aspect of the prostatic urethra during procedures. The device is used in clinical settings by physicians. It benefits patients by improving visualization of the urethra, potentially reducing complications associated with prostate biopsy and brachytherapy procedures.
Clinical Evidence
Bench testing only. Evidence includes sterilization and shelf-life validation, biocompatibility testing, and device performance specifications. Performance testing included visual inspection, dimensional analysis, adherence to ASTM F623-99 standards, and echogenicity testing in a phantom model to verify the acoustic interface functionality.
Indicated for patients undergoing prostate biopsy or brachytherapy procedures requiring bladder/urinary tract drainage and visualization of the prostatic urethra via transrectal ultrasound (TRUS).
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
MAR 0 4 2014 .
← ۶۰
## 510(k) SUMMARY
| K140099<br>pg 1 of 3 |
|------------------------------------------------------------------------------------------|
| seedlings |
| Life • Science • Ventures • LLC<br>230 East 15th Street, Suite 1-A<br>New York, NY 10003 |
| 510(k) Owner: | Seedlings Life Science Ventures |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Address: | 230 East 15th Street, Suite 1-A, New York, NY, 10003 |
| Phone: | 954.336.0777 |
| FAX: | 866-277-3269 |
| Contact: | Ken Solovay, Chief Operating Officer |
| Email: | ken@seedlings.net |
| Date Prepared: | March 3, 2014 |
| Device Trade name: | Lighthouse™ Urinary Catheter |
| Common name: | Urinary Catheter |
| Classification name: | Catheter, Retention Type, Balloon |
| Classification Product Code: | EZL, FGI |
| Regulation Number: | 21 CFR 876.5130 |
| Predicate Devices: | |
| (1) Lighthouse Urinary Catheter<br>Manufacturer:<br>FDA 510(k) number:<br>Classification Code: | Seedlings Life Science Ventures, LLC<br>K131655<br>78 EZL |
| (2) BeamCath, Urological Catheter Patient Positioning Device<br>Manufacturer: | Beampoint AB |
| FDA 510(k) number: | K042110 |
| Classification Code: | 78 EZL |
.
:
·
:
/
{1}------------------------------------------------
| | K140099 |
|--|-----------|
| | pg 2 of 3 |
seedlings
Life · Science · Ventures · LLC
230 East 15th Street, Suite 1-A
New York, NY 10003
### 510(k) SUMMARY
- The Lighthouse™ Urinary Catheter is a 3-lumen Foley catheter. In addition to a System lumen for inflating the Foley balloon and a urine drainage lumen, the Lighthouse Description has a third lumen for filling 0.1cc - 1cc of air in a 4-cm length bladder along the shaft of the catheter. The Urethral Identification Bladder (UIB) has a maximum outer diameter of 24F (i.e. smaller than other urinary catheters that meet ASTM F623-99) when filled with a maximum 1cc volume described in the instructions for use (IFU) and provides echogenicity when viewed by a Transrectal Ultrasound (TRUS). The non-inflated UIB contains a narrow air space between the catheter shaft and the urethra, which provides an echogenic marker under TRUS and when necessary can be increased by adding 0.1cc - 1.0cc of volume via syringe.
- The Lighthouse™ Urinary Catheter is intended for bladder/urinary tract drainage Intended Use during prostate biopsy and brachytherapy procedures where transrectal ultrasound (TRUS) imaging is used to visualize the prostate. The catheter additionally enhances identification of the course of the posterior aspect of the prostatic urethra during TRUS-guided prostate biopsy and brachytherapy procedures by creating an acoustic interface (i.e., air) between the catheter and the urethral wall via its urethral identification bladder.
| Characteristic | Lighthouse™<br>(Seedlings) | Lighthouse™<br>(Seedlings) | BeamCath™<br>(Beampoint) |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K140099 | K131655 | K042110 |
| Intended Use | The Lighthouse™ Urinary<br>Catheter is intended for<br>bladder/urinary tract drainage<br>during prostate biopsy and<br>brachytherapy procedures<br>where transrectal ultrasound<br>(TRUS) imaging is used to<br>visualize the prostate. The<br>catheter additionally enhances<br>identification of the course of<br>the posterior aspect of the<br>prostatic urethra during TRUS-<br>guided prostate biopsy and<br>brachytherapy procedures by<br>creating an acoustic interface<br>(i.e., air) between the catheter<br>and the urethral wall via its<br>urethral identification bladder. | The Lighthouse™ Urinary<br>Catheter is intended for<br>bladder/urinary tract<br>drainage during prostate<br>brachytherapy procedures<br>where transrectal<br>ultrasound (TRUS) imaging<br>is used to visualize the<br>prostate. The catheter<br>additionally enhances<br>identification of the course<br>of the posterior aspect of<br>the prostatic urethra<br>during TRUS-guided<br>brachytherapy by creating<br>an acoustic interface (i.e.,<br>air) between the catheter<br>and the urethral wall via its<br>urethral identification<br>bladder. | The BeamCath is a<br>positioning device used<br>with external beam<br>conformal radiation<br>therapy for visualizing<br>the prostate position<br>during treatment of<br>prostate cancer. The<br>BeamCath is introduced<br>into the urethra as a<br>sterile urological<br>catheter and is retained<br>in place by inflating the<br>balloon tip. Its<br>radiopaque markers aid<br>in radiographic<br>visualization of the<br>prostate position for<br>planning, simulation<br>and treatment of<br>prostate cancer using<br>dose escalation<br>radiotherapy |
The Lighthouse is substantially equivalent to the following predicate devices: the Lighthouse Urinary Catheter (Seedlings Life Science Ventures) with its original
{2}------------------------------------------------
# 510(k) SUMMARY
Image /page/2/Picture/1 description: The image shows the logo for Seedlings Life Science Ventures LLC. The logo includes the word "seedlings" in a stylized font, with the words "Life, Science, Ventures, LLC" underneath. The address "230 East 15th Street, Suite 1-A, New York, NY 10003" is printed below the logo.
| Characteristic | Lighthouse™<br>(Seedlings) | Lighthouse™<br>(Seedlings) | BeamCath<br>(Beampoint) |
|----------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| 510(k) number | K140099 | K131655S | K042110 |
| Design | Urological catheter with Foley<br>balloon & echogenic 24Fr air-<br>filled bladder | Urological catheter with<br>Foley balloon & echogenic<br>24Fr air-filled bladder | Urological catheter<br>with Foley balloon &<br>tantalum radiopaque<br>markers |
| Materials | Biocompatible silicone | Biocompatible silicone | PVC shaft & latex<br>balloon |
| Sterile, Patient Use | Sterile, Single-patient use | Sterile, Single-patient use | Sterile, Single-patient<br>use |
| Sizes | 16Fr | 16Fr | 14Fr |
#### Performance The following bench tests have been performed to verify the Lighthouse performance specifications, with results demonstrating substantial equivalence:
- 1.) Sterilization and Shelf Life
- 2.) Biocompatibility
- 3.) Device Performance Specifications
### Data
- a. Visual Inspection
- Dimensional Analysis ف
- C. ASTM F623-99
- d. Echogenicity Testing in a Phantom Model
Determination Based on the comparison above, the urine drainage and enhanced visualization of Substantial under external imaging is equivalent to the predicate devices. The small risk of Equivalence using a urinary catheter during prostate biopsy is overwhelmingly outweighed by the benefits of using the Lighthouse Catheter. The risk of urethral complication from prostate biopsy without using the Lighthouse Catheter is much greater than the risk of a complication due to use of the Lighthouse Catheter and, as such, the device does not raise new questions of safety and effectiveness.
Conclusion We believe the Lighthouse Catheter is substantially equivalent based on the indications for use and the performance testing results.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. The text is in all capital letters.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2014
Seedlings Life Science Ventures Ken Solovay Chief Operating Officer 230 East 15th Street, Suite 1-A New York, NY 10003
K140099 Re:
> Trade/Device Name: Lighthouse™ Urinary Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL, FGI Dated: January 15, 2014 Received: January 16, 2014
Dear Ken Solovay,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ken Solovay
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/6 description: The image shows the text "Benjamin Fisher-S". The text is written in a simple, sans-serif font. The letters are black and the background is white. The letters "FDA" are in the middle of the text, and they are larger than the other letters.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(k): K140099 (To be assigned)
Device Name: Lighthouse™ Urinary Catheter
Indications for Use: The Lighthouse™ Urinary Catheter is intended for bladder/urinary tract drainage during prostate biopsy and brachytherapy procedures where transrectal ultrasound (TRUS) imaging is used to visualize the prostate. The catheter additionally enhances identification of the course of the posterior aspect of the prostatic urethra during TRUS-guided prostate biopsy and brachytherapy procedures by creating an acoustic interface (i.e., air) between the catheter and the urethral wall via its urethral identification bladder.
Prescription Use __ Yes (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ No (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANDTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/9 description: The image shows the text "Benjamin R. Fisher -S 2014.03.04 11:59:24 -05'00'". The text appears to be a name, followed by a date and time stamp. The name is "Benjamin R. Fisher", and the date and time stamp is "2014.03.04 11:59:24 -05'00'". The text is black and is on a white background.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.