The eNeura Therapeutics SpringTMS® is indicated for the acute treatment of pain associated with migraine headache with aura.
Device Story
SpringTMS is a portable, hand-held transcranial magnetic stimulator (sTMS) for acute migraine treatment. Device delivers brief single pulse of magnetic energy (0.9 Tesla) to occipital cortex to induce electrical current, aiming to stop or lessen migraine effects. Intended for patient self-administration in home or office settings via physician prescription. Device features a SIM chip-based authorization system to control treatment duration; LCD display communicates status. Powered by internal rechargeable lithium-ion battery. Output affects clinical decision-making by providing non-invasive, drug-free acute pain relief for migraineurs.
Clinical Evidence
Bench testing only. Performance testing included magnetic pulse characteristics, magnetic field mapping, 5 Gauss line location, software verification/validation, and electromagnetic compatibility/electrical safety (IEC 60601-1, IEC 60601-1-2). Results demonstrated substantial equivalence to the predicate device.
Technological Characteristics
Hand-held portable stimulator; polycarbonate case with integral coil. Operating principle: time-varying magnetic field (0.9 Tesla peak @ 180 μs) inducing electrical current in occipital cortex. Power: internal rechargeable lithium-ion battery. Connectivity: SIM chip for use authorization. Software: verification/validation per SRS. Standards: IEC 60601-1:2005, IEC 60601-1-2:2007.
Indications for Use
Indicated for acute treatment of pain associated with migraine headache with aura in patients prescribed the device.
Regulatory Classification
Identification
A transcranial magnetic stimulator device for headache is a device that delivers brief duration, rapidly alternating, or pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electrical currents for the treatment of headache.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety.
(2) Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
(3) The elements of the device that contact the patient must be assessed to be biocompatible.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
(5) Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
(6) The physician and patient labeling must include the following:
(i) A summary of the clinical performance testing, including any adverse events and complications.
(ii) The intended use population in terms of the types of headaches appropriate for use with the device.
(iii) Information on how to report adverse events and device malfunctions.
(iv) A diagram or picture depicting the proper placement of the device on the user.
1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
2. Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
3. The elements of the device that contact the patient must be assessed to be biocompatible.
4. Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
5. Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
6. The physician and patient labeling must include the following:
a. A summary of the clinical performance testing, including any adverse events and complications.
b. The intended use population in terms of the types of headaches appropriate for use with the device.
c. Information on how to report adverse events and device malfunctions.
d. A diagram or picture depicting the proper placement of the device on the user.
Predicate Devices
eNeura Therapeutics® Cerena™ Transcranial Magnetic Stimulator (K130556)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Notification K140094
#### GENERAL INFORMATION
#### Applicant:
eNeura Therapeutics, LLC 240 North Wolfe Road Sunnyvale, CA 94085 U.S.A. Phone: 408-245-6500 FAX: 408-245-6424
#### Contact Person:
Larry Getlin Regulatory Consultant for eNeura Therapeutics, LLC 2690 Pheasant Road Orono, MN 55331 U.S.A. Phone: 612-850-8144
## DEVICINFORMATION
Trade Name: SpringTMS®
#### Generic/Common Name:
Transcranial magnetic stimulator for the treatment of migraine headache 21 CFR§882.5808
# Classification:
Class II
#### Product Code: OKP
#### PREDICATE DEVICE(S)
eNeura Therapeutics® Cerena™ Transcranial Magnetic Stimulator (K130556)
#### INTENDED USE
The eNeura Therapeutics SpringTMS® is indicated for the acute treatment of pain associated with migraine headache with aura.
## PRODUCT DESCRIPTION
The SpringTMS® is a portable, hand-held device that delivers a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a
{1}------------------------------------------------
portion of the brain called the occipital cortex to stop or lessen the effects of migraine headaches. Since a single pulse of magnetic stimulation is emitted, this method of stimulation is called single pulse transcranial magnetic stimulation or sTMS. The SpringTMS is indicated for the acute treatment of pain associated with migraine headache with aura. The device is designed for patient use where treatments are selfadministered and can be delivered in a variety of settings including the home or office. The device is intended for prescription use only.
## TECHNOLOGICAL CHARACTERISTICS
The technological characteristics of the SpringTMS are substantially equivalent to the predicate device. Table 1 lists the technological characteristics of the SpringTMS and the predicate device and provides the rationale to support a determination of substantial equivalence. Any differences in the technological characteristics of the device do not affect the safety and effectiveness of the device.
| Feature | SpringTMS® | Cerena™ Transcranial<br>Magnetic Stimulator | Substantial Equivalence Rationale |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K140094 | K130556 | New 510(k) submission |
| Operating<br>Principle | Induces electrical<br>current in region near<br>coil<br>Transcranial<br>Evoked response<br>Stimulation on the<br>occipital cortex | Induces electrical<br>current in region<br>near coil<br>Transcranial<br>Evoked response<br>Stimulation on the<br>occipital cortex | N/A (same) |
| Design | Time varying<br>magnetic field<br>Non-invasive | Time varying<br>magnetic field<br>Non-invasive | N/A (same) |
| Use<br>Authorization | The user must insert a<br>SIM chip to use the device<br>for a programmed<br>duration. The<br>programmed duration<br>corresponds to the<br>prescribed months of use.<br>The SIM chip is only<br>available under physician<br>prescription. | None | The addition of this Use<br>Authorization feature does not<br>affect the operating principle or<br>performance of the device and no<br>additional risks or hazards have<br>been identified related to this<br>change. |
| Display | LCD display | LED indicators | In both devices, the display serves<br>to communicate device status to the<br>patient and no additional risks or<br>hazards have been identified related<br>to this change. |
| Magnetic<br>Field | 0.9 Tesla Peak @ 180 μs<br>(total magnetic energy<br>140J) | 0.9 Tesla Peak @ 180 μs<br>(total magnetic energy<br>140J) | N/A (same) |
| Current | 4 mA/cm² induced at 1 cm | 4 mA/cm² induced at 1<br>cm | N/A (same) |
## Table 1: Summary of Technological Characteristics
{2}------------------------------------------------
| Electrical<br>Power | Internally powered with<br>rechargeable lithium ion<br>battery pack. Battery pack<br>charger mains input --<br>100-240V AC, 47/63 Hz,<br>output 12 V DC | Externally powered by<br>AC/DC power adapter.<br>Mains input --<br>100-240V AC, 50/60 Hz,<br>output 12 V DC | The SpringTMS and predicate<br>device utilize the same voltage and<br>power and both meet all criteria for<br>establishing electrical safety. |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Materials | Hand held portable<br>stimulator in<br>polycarbonate case<br>(integral coil) | Hand held portable<br>stimulator in<br>polycarbonate case<br>(integral coil) | N/A (same) |
| Where used | Home-use and where the<br>operator is | Home-use and where the<br>operator is | N/A (same) |
| Dimensions<br>and Weight | 9 in. (23 cm) long<br>5 in. (13 cm) wide<br>3 in. (8 cm) deep<br>3.8 lb. (1.7 kg) | 13 in. (33 cm) long<br>5 in. (13 cm) wide<br>5 in. (13 cm) deep<br>3.4 lb. (1.54 kg) | The form factor changes raise no<br>new issues of safety or efficacy. |
## SUBSTANTIAL EQUIVALENCE
The indications for use for the predicate device is identical to the proposed indications for use for the SpringTMS. The differences in the technological characteristics (Replacement of LED indicators with LCD screen, addition of the Use Authorization Feature, power supply type change and minor dimensional changes) do not raise any new issues of safety or effectiveness. Thus, the SpringTMS is substantially equivalent to the predicate device.
## TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
All necessary performance testing was conducted on the SpringTMS to support a determination of substantial equivalence to the predicate device. Table 2 lists the non-clinical performance testing conducted and the results supporting substantial equivalence.
| Testing Type | Test Description | Results Supporting Substantial Equivalence |
|---------------------------|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Bench Testing | Magnetic Pulse Characteristics vs. Time | Both devices have the same specification for magnetic pulse shape and both tested within specification. No new issues of safety or efficacy have been raised. The measured rate of change of the magnetic field is substantially equivalent. |
| | Magnetic Pulse Field Map | No new issues of safety or efficacy have been raised. The Magnetic Pulse Field Maps for the SpringTMS and the predicate device are substantially equivalent. |
| | Location of 5 Gauss Line | No new issues of safety or efficacy have been raised. The location of the 5 Gauss line for the SpringTMS and the predicate are substantially equivalent. |
## Table 2: Non-Clinical Performance Testing and Substantial Equivalence Support
{3}------------------------------------------------
| Software Verification<br>Validation Testing | SpringTMS Software Testing | SpringTMS Software met all<br>requirements of the SRS. No new<br>issues of safety or efficacy have been<br>raised. All safety and performance<br>specifications for the SpringTMS SRS<br>are substantially equivalent to those in<br>the SRS for the predicate. |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electromagnetic<br>Compatibility and<br>Electrical Safety | Testing in accordance with<br>the following standards:<br>• IEC 60601-1:2005<br>• IEC 60601-1-2:2007 | The SpringTMS and the predicate<br>device met all acceptance criteria. No<br>new issues of safety or efficacy have<br>been raised. Therefore the SpringTMS<br>is substantially equivalent to the<br>predicate. |
The collective results of performance testing demonstrate that the materials chosen, the manufacturing processes, and design of the SpringTMS meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the SpringTMS does not raise new questions of safety or effectiveness when compared to the predicate device.
## CONCLUSION
The SpringTMS is substantially equivalent to the predicate device. The indications for use is identical to that of the predicate device, and the product performance testing has demonstrated that the SpringTMS is as safe, as effective and performs in the same manner as the predicate device in terms of intended use, safety and technological characteristics, and patient populations. The differences include the device external design such as size and weight, display type, power supply type, and the authorization system for the device use; however, those changes do not raise any new safety or efficacy concerns. The information contained in this 510(k) premarket notification demonstrates the substantial equivalence of the SpringTMS to the predicate device.
#### SUMMARY
The SpringTMS is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WORE-COON Silver Spring, MD 20993-0002
May 21, 2014
eNeura Therapeutics, LLC c/o Larry Getlin 2690 Pheasant Road Orono, MN 55331
Re: K140094
Trade/Device Name: Spring TMS® Regulation Number: 21 CFR §882.5808 Regulation Name: Transcranial magnetic stimulator for the treatment of migraine headache Regulatory Class: Class II Product Code: OKP Dated: April 18, 2014 Received: April 21, 2014
Dear Mr. Getlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{5}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Carlos L. Pena -S
Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140094
Device Name eNeura Therapeutics Spring TMS
The eNeura Therapeutics® Spring TMS® is indicated for the acute treatment of pain associated with migraine headache with aura.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Carlos L. Pena -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the r inc baract to the concern existing data sources, gather and maintain the data needed and complete une to review the collections, but in the many comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.