U2 FEMORAL COMPONENT, CR, CEMENTED

K140073 · United Orthopedic Corp. · JWH · Jun 16, 2014 · Orthopedic

Device Facts

Record IDK140073
Device NameU2 FEMORAL COMPONENT, CR, CEMENTED
ApplicantUnited Orthopedic Corp.
Product CodeJWH · Orthopedic
Decision DateJun 16, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee cannot be satisfactorily balanced and stabilized at the time of surgery. For cemented femoral components, patellar components, tibial baseplate components and tibial inserts components: This device is a single use implant and intended for cemented use only. For porous coated femoral component: This device is a single use implant and intended for cementless use only.

Device Story

U2 Femoral Component, CR, Cemented is a metallic implant for total knee arthroplasty. Device replaces damaged femoral condyles to restore joint function and reduce pain. Manufactured from cast Co-Cr-Mo alloy; available in seven sizes (#1-#7). Implanted by orthopedic surgeons in clinical/OR settings using bone cement for fixation. Device serves as a mechanical replacement component within the U2 Total Knee System. Benefits include improved joint stability and mobility for patients with severe knee pathology.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including range of motion, contact area/pressure on femorotibial and femoropatellar joints, subluxation testing, and fatigue compression testing.

Technological Characteristics

Material: Cast Co-Cr-Mo alloy (ASTM F75). Configuration: Seven sizes (#1-#7), left and right. Fixation: Cemented. Design: CR (Cruciate Retaining) type. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients requiring knee arthroplasty for pain relief or improved function due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, pseudogout, or posttraumatic joint loss. Applicable for salvage of failed surgical attempts. Contraindicated for non-cemented use in components specified for cemented use.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN | 6 2014 9 U2 Femoral Component, CR, Cemented 510(k) Summary # 510(k) Summary of Safety and Effectiveness | Submitter Information | | |---------------------------------------|-------------------------------------------------------------------------------------------------------------------------| | Name | United Orthopedic Corporation | | Address | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan | | Phone Number | +886-3-5773351 ext. 2212 | | Fax Number | +886-3-577156 | | Name of Contact Person | Fang-Yuan Ho<br>Regulation and Document Management | | Date prepared | January 10, 2014 | | Name of Device | | | Trade Name | U2 Femoral Component, CR, Cemented, #7 | | Common Name | Femoral Component | | Classification Name and<br>Regulation | Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis under 21CFR §<br>888. 3560 | | Device Class | Class II | | Classification Panel | Orthopaedics | | Product Code | JWH | | Predicate Device | 1. “United” U2 Total Knee System (K051640).<br>2. “Smith & Nephew” Genesis II Total Knee System (K951987) | ### Device Description: U2 Cemented Femoral Component-CR Type is manufactured from cast Co-Cr-Mo alloy confirming to ASTM F75. It is available in seven proportional sizes (#1~ #7, AP/ML ranging from 52mm/ 56mm to 76mm/ 80mm) and is provided in left and right configurations. The femoral component is the same as that of cleared U2 Total Knee System (K051640), except adding size #7. Fixation of the femoral component to the femur {1}------------------------------------------------ ## U2 Femoral Component, CR, Cemented is achieved using bone cement. #### Intended Use: The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee cannot be satisfactorily balanced and stabilized at the time of surgery. For cemented femoral components, patellar components, tibial baseplate components and tibial inserts components: This device is a single use implant and intended for cemented use only. For porous coated femoral component: This device is a single use implant and intended for cementless use only. #### Comparison to Predicate Device: U2 Cemented Femoral Component-CR Type has the same materials, design, manufacturing and sterilization method as the femoral component of "United" U2 Total Knee System (K051640), except adding size #7. Besides, the basic design; size distribution, intended use of the subjected device and the method of fixation are similar with the femoral component of "Smith & Nephew" Genesis II Total Knee System (K951987). #### Performance Data: ### ● Non-clinical Performance This 510(k) submission was prepared in accordance with the Agency's, " Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial {2}------------------------------------------------ #### ਹੈ U2 Femoral Component, CR, Cemented 510(k) Summary Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA". Tests as follows were conducted to evaluate the safety and effectiveness of the subjected device, and the test results demonstrated that this device is safe and effective. - Range of Motion a. - Contact Area and Contact Pressure on Femorotibial Joint b. - Contact Area and Contact Pressure on Femoropatellar Joint c. - Subluxation of Femoroltibial Joint d. - e. Fatigue Compression Test of Femoral Component ## ● Clinical Performance Data/Information None provided as a basis for substantial equivalence. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines forming its body and wings. The symbol is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. The logo is black and white. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002 #### June 16, 2014 United Orthopedic Corporation Fang-Yuan Ho Regulatory Affairs Manager Number 57, Park Avenue 2, Science Park Ksinchu 300 Taiwan Re: K140073 Trade/Device Name: U2 Femoral Component, CR, Cemented Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 24, 2014 Received: March 25, 2014 Dear Fang-Yuan Ho: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {4}------------------------------------------------ #### Page 2 - Fang-Yuan Ho (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. ## Lori A. Wiggins - for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indication for Use ### 510 (k) Number (if known): K140073 #### Device Name: U2 Femoral Component, CR, Cemented #### Indications for Use: The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee cannot be satisfactorily balanced and stabilized at the time of surgery. For cemented femoral components, patellar components, tibial baseplate components and tibial inserts components: This device is a single use implant and intended for cemented use only. For porous coated femoral component: This device is a single use implant and intended for cementless use only. Prescription Use x AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **Casey L. Hanley, Ph.D.** Division of Orthopedic Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...