Optimus CMF System is implantable bone plates and bone screws for maxillofacial and mandible surgery procedures including: 1. Fractures 2. Osteotomies 3. Reconstructive procedures 4. Revision procedures where other treatments or devices have failed.
Device Story
Optimus CMF System consists of implantable bone plates and screws for maxillofacial and mandible surgery. Plates (0.4-2.6mm thick) and screws (1.3-2.7mm diameter; 3.0-20.0mm length) are used by surgeons to stabilize bone segments during fracture repair, osteotomies, or reconstructive procedures. Devices are non-sterile and require sterilization before use. Surgeons select appropriate plate shapes and screw sizes based on anatomical needs. The system provides mechanical fixation to support bone healing and structural restoration.
Clinical Evidence
Bench testing only. Performance validated via 4-Point Bending (ASTM F382-99), Torsion and Axial Pullout Strength (ASTM F543-13e1), and biocompatibility testing (ISO 10993-5, -10, -3, -6). Sterilization validated per ANSI/AAMI ST79 and ISO 17665-1.
Technological Characteristics
Plates: Unalloyed Titanium (ASTM F67). Screws: Ti-6Al-4V ELI titanium alloy (ASTM F136). Surface treatment: Anodizing. Mechanical fixation via bone plates and screws. Non-sterile, single-use. Various shapes (straight, Y, X, L, Z, etc.) and dimensions.
Indications for Use
Indicated for patients requiring maxillofacial and mandible surgery for fractures, osteotomies, reconstructive procedures, or revision procedures following failed treatments.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
LeForte System Bone Plate & Screw by Jeil Medical Corporation (K112457)
Synthes Matrix Mandibule Plate and Screw System by Synthes Inc. (K113567)
Biomet Microfixation Facial Plating System by Biomet Microfixation (K121589)
Synthes (USA) Midfacial System by Synthes (USA) (K953806)
Synthes 1.3mm Self-Drilling Screw by Synthes (USA) (K983485)
Leibinger IMF Screw by Howmedical Leibinger, Inc. (K963030)
Synthes (USA) Craniofacial Plates by Synthes (USA) (K040272)
Leforte System Bone Plate by Jeil Medical Coporation (K091679)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stylized to resemble a bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2015
Osteonic Co., Ltd C/O Ms. Priscilla Chung Official Correspondent LK Consulting Group USA, Inc. 2651 E. Chapman Ave., Ste. 110 Fullerton, CA 92831
Re: K140037
Trade/Device Name: Optimus CMF System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY, DZL Dated: December 9, 2014 Received: December 12, 2014
## Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) k140037
### Device Name OPTIMUS CMF SYSTEM
### Indications for Use (Describe)
Optimus CMF System is implantable bone screws for maxillofacial and mandible surgery procedures including: 1. Fractures
- 2. Osteotomies
- 3. Reconstructive procedures
- 4. Revision procedures where other treatments or devices have failed.
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
̄ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
Date:
#### 1. Applicant / Submitter:
OSTEONIC Co., Ltd. 505-3Ho, Digital-ro 29-gil, Guro-gu, Seoul, Republic of Korea 152-779 Tel: +82-2-6082-8885
#### 2. Submission Correspondent
Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110. Fullerton CA 92831 Email: juhee.c@lkconsultinggroup.com
#### 3. Device:
- · Proprietary Name OPTIMUS CMF SYSTEM
- Common Name Dental Bone Plate & Screw System
- · Classification Name Dental Bone Plate
- Screw, fixation, intraosseous
#### 4. Predicate Device:
- · LeForte System Bone Plate & Screw by Jeil Medical Corporation (K112457)
- · Synthes Matrix Mandibule Plate and Screw System by Synthes Inc. ( K113567)
- · Biomet Microfixation Facial Plating System by Biomet Microfixation (K121589)
- = Synthes (USA) Midfacial System by Synthes (USA) (K953806)
- = Synthes 1.3mm Self-Drilling Screw by Synthes (USA) (K983485)
- Leibinger IMF Screw by Howmedical Leibinger, Inc. (K963030)
- = Synthes (USA) Craniofacial Plates by Synthes (USA) (K040272)
- · Leforte System Bone Plate by Jeil Medical Coporation (K091679)
#### 5. Product Code & Regulation Number:
- · JEY (21CFR872.4760)
- · DZL (21CFR872.4880)
#### 6. Description:
The OPTIMUS CMF SYSTEM comprises plates and screws. The plate thickness sizes range from 0.4 to 2.6mm. They are made of unalloyed Titanium (ASTM F67) and anodized in
{4}------------------------------------------------
four colors (silver, blue, green, and gold). The diameters of the screws range from 1.3 to 2.7mm, while lengths range from 3.0 to 20.0mm. They are made of Ti-6A1-4V ELI titanium alloy (ASTM F136) and anodized in six colors (light blue, silver, purple, blue, gold and green). The plates and screws are single use only, non-sterile products. They must be sterilized before use.
#### 7. Indication for use:
Optimus CMF System is implantable bone plates and bone screws for maxillofacial and mandible surgery procedures including:
- 1. Fractures
- 2. Osteotomies
- 3. Reconstructive procedures
- 4. Revision procedures where other treatments or devices have failed.
#### 8. Non-Clinical Testing:
The following tests were performed to validate the performance and the safety of the subject device and the test results supported substantial equivalence to the predicate devices.
- י 4-Point Bending test in accordance with ASTM F382-99(2008)e1
- י Torsion and Axial Pullout Strength Test in accordance with ASTM F543-13e1
- Cytotoxicity test in accordance with ISO 10993-5 י
- ' Sensitization test in accordance with ISO 10993-10
- י Genotoxicity test in accordance with ISO 10993-3
- 트 Implantation test in accordance with ISO 10993-6
- י Sterilization validation test in accordance with ANSI/AAMI ST79, ISO 17665-1. ISO 11737-1, and USP 30-NF25 <61> Microbial Limited Test.
#### 9. Substantial Equivalence:
The subject and predicate device(s) share the same intended use, primary design and equivalent material of manufacture. Any minor differences do not impact device performance as compared to the predicate devices and demonstrate that the Optimus CMF System is substantially equivalent to the predicate devices. The performance test results support that the subject device performs as well as the predicate devices.
{5}------------------------------------------------
| | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Predicate Device 5 | Predicate Device 6 | Predicate Device 7 | Predicate Device 8 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | OSTEONIC<br>Co., Ltd. | Jeil Medical<br>corporation | Synthes | Biomet<br>Microfixation | Synthes | Synthes | Howmedica<br>Leibinger Inc. | Synthes | Jeil Medical<br>Coporation |
| Device<br>Name | OPTIMUS<br>CMF<br>SYSTEM | LeForte System<br>Bone Plate &<br>Screw | MatrixMANDI<br>BULE Plate<br>and Screw<br>System | Biomet<br>Microfixation<br>Facial Plating<br>System | SYNTHES<br>(USA)<br>MIDFACIAL<br>SYSTEM | Synthes<br>1.3mm Self-<br>Drilling<br>Screw | Leibinger IMF<br>Sscerw | SYNTHES<br>(USA)<br>CRANIOFACI<br>AL PLATES | LEFORTE<br>SYSTEM<br>BONE PLATE |
| 510(K) # | N/A | K112457 | K113567 | K121589 | K953806 | K983485 | K963030 | K040272 | K091679 |
| Class | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| Product<br>Code | JEY, DZL | JEY, DZL | JEY | JEY | JEY | DZL | DZE | JEY | JEY |
| Intended<br>Use | Optimus CMF<br>System is<br>implantable<br>bone plates<br>and bone<br>screws for<br>maxillofacial<br>and mandible<br>surgery<br>procedures<br>including:<br>1. Fractures<br>2. Osteotomies<br>3.<br>Reconstructive<br>procedures<br>4. Revision<br>procedures<br>where other<br>treatments or<br>devices have<br>failed. | This device is<br>intended for use<br>in selective<br>trauma of the<br>mid-face,<br>reconstruction<br>procedures and<br>selective<br>orthognathic<br>surgery of the<br>maxilla and chin | The<br>SynthesMatrix<br>MANDI BULE<br>Plate and<br>Screw System<br>is intended for<br>oral,<br>maxillofacial<br>surgery:<br>• Trauma<br>• Reconstructi-<br>ve surgery<br>• Orthognatic<br>surgery<br>(surgical<br>correction of<br>deformities) | These devices are<br>implantable bone<br>plates and bone<br>screws for facial<br>procedures<br>including:<br>1. Fractures<br>2. Osteotomies<br>3. Reconstructive<br>procedures<br>4. Revision<br>procedures where<br>other treatments<br>or devices have<br>failed | Intended for use<br>in selective<br>trauma of the<br>midface and<br>craniofacial<br>skeleton;<br>craniofacial<br>surgery;<br>reconstructive<br>procedures and<br>selective<br>orthognathic<br>surgery of the<br>maxilla and<br>chin. | Synthes<br>1.3mm Self-<br>Drilling<br>Screws are<br>intended for<br>selective<br>trauma of the<br>midface and<br>craniofacial<br>skeleton:<br>craniofacial<br>surgery;<br>reconstructive<br>procedures;<br>and selective<br>orthognathic<br>surgery of the<br>maxilla and<br>chin. | Intended for<br>use in<br>temporary<br>maxillomandib<br>ular fixation to<br>provide indirect<br>stabilization of<br>the maxilla,<br>mandible or<br>both. | Intended for<br>use in selective<br>trauma of the<br>midface and<br>craniofacial<br>skeleton;<br>craniofacial<br>surgery;<br>reconstructive<br>procedures and<br>selective<br>orthognathic<br>surgery of the<br>maxilla and<br>chin. | This device is<br>intended for<br>use in selective<br>trauma of the<br>mid-face and<br>craniofacial<br>skeleton<br>craniofacial<br>surgery<br>reconstruction<br>procedures and<br>selective<br>orthognathic<br>surgery of the<br>maxilla and<br>chin |
| Material | Plate -<br>Unalloyed<br>Titanium<br>Screw-<br>Titanium Alloy | Plate -Unalloyed<br>Titanium<br>Screw- Titanium<br>Alloy | Unalloyed<br>Titanium | Plate - Unalloyed<br>Titanium<br>Screw- Titanium<br>Alloy | Unalloyed<br>Titanium | Titanium<br>Alloy | Titanium Alloy | Unalloyed<br>Titanium | Unalloyed<br>Titanium |
| Surface | Anodizing | Plate: Anodizing<br>Screw: N/A | Anodizing | Plate: N/A<br>Screw: Anodizing | Anodizing | Anodizing | - | Anodizing | Plate:<br>Anodizing |
| Anodizing<br>color | Light blue,<br>silver, purple, | Silver, blue,<br>green and gold | Silver, blue,<br>green and gold | Silver, magenta,<br>blue, green and | Silver, blue,<br>green and gold | Silver, blue,<br>green and gold | - | Silver, blue,<br>green and gold | Silver, blue,<br>green and gold |
| | blue, gold and<br>green | | | gold | | gold | | | |
| Single Use | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Sterile | Non sterile | Non sterile | Sterile or non-<br>sterile | Non sterile | Non sterile | Non sterile | Non sterile | Non sterile | Non sterile |
| Shape and<br>Dimension<br>of plate | Plate has<br>various shapes,<br>length and<br>thickness.<br>Straight, Y, D-<br>Y, X, L, Z,<br>Square,<br>Matrix,<br>Orbital, Chin<br>and<br>Reconstruction<br>types with<br>various<br>lengths. The<br>range of<br>plate's profile<br>is from 0.4 to<br>2.6mm<br>thickness and<br>4 colors<br>(silver, blue,<br>green and<br>gold). | Length<br>5.2~223.5mm<br>Thickness<br>0.2~2.5mm | The system<br>consists of a<br>variety of<br>plates offered<br>in multiple<br>shape and sizes<br>and a variety of<br>screws offered<br>in multiple<br>diameters and<br>lengths to meet<br>the anatomical<br>needs of the<br>patient. | The plates that<br>are the subject of<br>this 510(k)<br>submission<br>include variations<br>of straight, angle,<br>curved, L-shape,<br>T-shape, double<br>T-shape, Z-shape,<br>X-shape, Y-shape,<br>double Y-shape,<br>H-shape, triangle,<br>square, rectangle,<br>matrix, mesh,<br>orbital floor,<br>LeFort, and chin<br>options with<br>various lengths<br>and thickness.<br>Plates are offered<br>flat or pre-bent. | The plate size is<br>available in a<br>variety of<br>configurations<br>to accommodate<br>various fracture<br>sites. | - | - | The synthes<br>double<br>adaption plate<br>can be cut or<br>trimmed to the<br>desired length<br>to<br>accommodate<br>various<br>fractures and<br>meet the<br>anatomical<br>need of the<br>patient. The<br>plate has a low<br>profile head,<br>uses 2.0mm<br>self-tapping or<br>self-drilling<br>bone screw<br>and 2.4mm<br>emergency<br>screw. | The system<br>consists of a<br>variety of<br>plates offered<br>in multiple<br>shape and<br>sizes. |
| Screw<br>Sizes | The diameter<br>of screw is<br>from 1.3 to<br>2.7mm in<br>lengths of 3.0<br>to 20.0mm and<br>6 colors (light<br>blue, silver,<br>purple, blue,<br>gold and<br>green). | Outer(head)<br>diameter<br>1.2~2.65mm<br>Inner diameter<br>0.7~1.6mm<br>Length<br>4.0~18.0mm | | Screw range in<br>diameters of<br>1.0mm to 2.3mm<br>and lengths from<br>2.0mm to<br>29.0mm | Screw size is 1.3<br>or 1.5mm. | Available in<br>lengths<br>ranging from<br>4~6mm | 2.0mm in<br>diameter and<br>ranges from<br>10.5 to 8.5mm<br>in total<br>length(6~14m<br>m in thread<br>length) | - | |
{6}------------------------------------------------
{7}------------------------------------------------
#### 10. Conclusions:
The subject and predicate device(s) share the same intended use, primary design and equivalent material of manufacture. The information provided in this submission including various test results support that the subject device is substantially equivalent to the predicate devices in the market.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.