The Transonic HCM102 system is intended for diagnostic assessment of cardiovascular status and circulatory variables in patients aged 1 month and older. This includes COstatus measurements by indicator dilution for quantifying patient's cardiac and volume status as well as continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation, systemic vascular resistance and related index parameters) by continuous arterial pressure waveform analysis. In this system, cardiac output (CO) value from dilution is used to calibrate the pressure wave form and assess continuous cardiac output. In addition the system also measures heart rate, systolic, diastolic pressures, mean arterial pressure and indexed parameters.
Device Story
System performs diagnostic assessment of cardiovascular status; inputs include ultrasound dilution indicator data and arterial pressure waveforms. Device uses intermittent ultrasound dilution cardiac output (CO) measurements to calibrate continuous arterial pressure waveform analysis. Outputs include continuous CO, stroke volume variation, pulse pressure variation, systemic vascular resistance, heart rate, and blood pressures. Used in ICU/OR by clinicians. System connects to patient via extracorporeal AV tubing set between standard arterial and central venous lines; utilizes standard pressure transducers. Provides real-time hemodynamic monitoring to assist clinical decision-making regarding patient volume and cardiac status.
Clinical Evidence
Bench and animal testing conducted. Safety testing performed per IEC 60601-1 and electromagnetic compatibility standards. Finished products undergo physical testing and visual examination per Quality Control Test Procedures to ensure conformance to design specifications. No human clinical trial data reported.
Technological Characteristics
Extracorporeal hemodynamic monitor; utilizes ultrasound dilution and arterial pressure waveform analysis. Components include monitor, flow/dilution sensors, AV loop tubing sets, and pressure transducers. Connectivity via standard in-situ arterial/venous lines. Safety tested to IEC 60601-1. Software-based processing for calibration and hemodynamic parameter derivation.
Indications for Use
Indicated for diagnostic assessment of cardiovascular status and circulatory variables in patients aged 1 month and older in ICU or OR settings.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 16, 2014
Transonic Systems, Inc. Naveen Thuramalla VP, Engineering & Clinical Studies 34 Dutch Mill Rd Ithaca, New York 14850
Re: K140017
Trade/Device Name: Transonic HCM102 System Regulation Number: 21 CFR 870.1435 Regulation Name: Single-Function, Preprogrammed Diagnostic Computer Regulatory Class: Class II Product Code: DXG Dated: June 17, 2014 Received: June 18, 2014
Dear Naveen Thuramalla,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mude Jellm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Transonic. The word "transonic" is written in blue, with the words "THE MEASURE OF" written in smaller font below the "tran" portion of the word. To the right of that text, the words "BETTER RESULTS" are written in smaller font below the "sonic" portion of the word. In the middle of the word, there is a gray symbol that looks like sound waves.
# Indications for Use
510(k) Number (if known): New Submission
Device Name: Transonic HCM102 System
Indications for Use: (Prescription Device)
The Transonic HCM102 system is intended for diagnostic assessment of cardiovascular status and circulatory variables in patients aged 1 month and older. This includes COstatus measurements by indicator dilution for quantifying patient's cardiac and volume status as well as continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation, systemic vascular resistance and related index parameters) by continuous arterial pressure waveform analysis. In this system, cardiac output (CO) value from dilution is used to calibrate the pressure wave form and assess continuous cardiac output. In addition the system also measures heart rate, systolic, diastolic pressures, mean arterial pressure and indexed parameters.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "transonic" in blue, with the words "THE MEASURE OF" in gray below the left side of the word and "BETTER RESULTS." in gray below the right side of the word. In the middle of the word, there is a gray symbol that looks like sound waves.
Transonic Systems Inc. 34 Dutch Mill Rd., Ithaca, NY 14850 p: 800.353.3569 f: 607.257.7256
# 510(k) SUMMARY
## Summary of Safety & Effectiveness
| Submitter's Name & Address: | Transonic Systems Inc<br>34 Dutch Mill Road,<br>Ithaca, NY 14850 |
|-----------------------------|------------------------------------------------------------------|
|-----------------------------|------------------------------------------------------------------|
- Contact Person & Telephone: Naveen Thuramalla 607-257-5300 (*326)
- Date Summary Prepared: Dec 30, 2013
Classification Name: Computer, Diagnostic, Device Name: Preprogrammed, Single-function, 21 CFR 870.1435. Product Class and Code: DXG and Class II Classification Panel: Cardiovascular Common/Usual Name: Extracorporeal, Diagnostic monitor Proprietary Name: Transonic HCM102 System
- K060898: PULSION PICCO PLUS, MODEL 8100. Predicate Devices:
K023960: LIDCOPLUS HEMODYNAMIC MONITOR SYSTEM, MODEL HM 70.
#### Device Description:
In accordance with section 510(k) for the Federal Food, Drug, and Cosmetic Act, Transonic Systems Inc. intends to introduce into interstate commerce the Transonic HCM102 system, which is an apparatus based on ultrasound dilution and arterial pressure waveform analysis to provide diagnostic assessment of cardiovascular status including COstatus (HCM101) measurements by indicator dilution for quantifying patient's cardiac and volume status as well as continuous cardiac output and associated hemodynamic parameters (such as stroke volume variation; pulse pressure variation, systemic vascular resistance and related index parameters) by continuous arterial pressure waveform analysis. In this system, cardiac output is first measured by COstatus (HCM101) system and this is then automatically transferred to calibrate the
{4}------------------------------------------------
arterial pressure waveform to obtain continuous cardiac output (when that mode is selected). Thus cardiac output is determined both intermittently through ultrasound technique and continuously through arterial pressure waveform analysis. In addition the system also measures heart rate, systolic, and diastolic pressures and derives mean arterial pressure. These patients could be in the intensive care units (ICU), operating room (OR) or other such environments.
#### Components:
Transonic HCM102 system consists of the following components. The below components are essentially COstatus system components with the addition of pressure transducer and connecting cable.
| Model/Part # | Description |
|--------------|----------------------------------------------------------------------|
| HCM102 | Monitor (HCM101 monitor with the addition of pressure<br>capability) |
| HCM101 | Cardiac output monitor/meter |
| HC2TP | Flow/dilution sensor pair with pump cable and pressure |
| | transducer cables. |
| HCP01 | AV Loop Pump |
| ADT2005D | Arteriovenous (AV) Loop - Tubing Set |
| ADT2005E | Arteriovenous (AV) Loop - Tubing Set |
| ADT2006E | Arteriovenous (AV) Loop - Tubing Set |
| ADT1020 | Extension Set/ Sensor Adapter Tubing |
| HCS3011 | AV Loop kit with ADT2005D AV Loop |
| HCS3021 | AV Loop kit with ADT2005E AV Loop |
| HCS3022 | AV Loop kit with ADT2006E AV Loop |
| HCS3002 | Sensor Adapter Tubing Pack with ADT1020 |
| HCS40XX | MX950 TranStar Disposable Transducer<br>(K061573)<br>or |
| | equivalent. |
| HCR01 | Printer |
| HFW1000 | Fluid Bag Warner |
| HCED02 | Data Transfer Module |
| HCR01 | Printer |
### Substantial Equivalence:
The Transonic HCM102 System for use with patients to provide diagnostic assessment of cardiovascular status including intermittent and continuous cardiac output and associated hemodynamic parameters (such as total end diastolic blood volume, central blood volume, active circulation volume, stroke volume variation; pulse pressure variation, etc) is similar to the Pulsion Continuous Pulse Contour Cardiac Output (PiCCO Plus) System (K060898) and LiDCO plus monitor (K023960).
{5}------------------------------------------------
Proposed device and predicate devices first measure cardiac output and other related hemodynamic parameters intermittently and then use the intermittent Cardiac output to calibrate the arterial blood pressure waveform to measure continuous cardiac output. Arterial blood pressure waveform is also analyzed to obtain other related hemodynamic parameters such as stroke volume variation and pulse pressure variation. All three systems are extracorporeal in nature and use both indicator dilution as well as arterial pressure waveform approaches. All three would require access to both arterial and central venous catheters to obtain indicator dilution and arterial pressure waveform measurements.
The difference between Pulsion PiCCO Plus system and the proposed HCM102 system is that the Pulsion system requires use of a dedicated special arterial catheter insertion into the patient, which is then connected to the Pulsion monitor for both types of measurements. However, Transonic HCM102 system works off an extracorporal AV tubing set connected between standard in-situ arterial and central venous lines (like COstatus system cleared by FDA K113821) for indicator dilution measurements and requires connecting a standard pressure transducer (such as MX950 or equivalent cleared by FDA K942377) in-line with the hospital pressure transducer for arterial pressure waveform measurements.
Pulsion system uses thermodilution and hence requires injection of cold or room temperature isotonic saline whereas Transonic HCM102 system uses ultrasound dilution and hence requires injection of body temperature isotonic saline.
The difference between LiDCPlus hemodynamic monitor and the proposed HCM102 system is that the LiDCO plus monitor uses lithium dilution, which requires injection of lithium, is used to obtain intermittent dilution cardiac output measurement and other related parameters. Transonic HCM102 system works off an extracorporal AV tubing set, which is connected between standard in-situ arterial and central venous lines (like COstatus system cleared by FDA K113821), and uses body temperature isotonic saline to obtain intermittent dilution cardiac output measurement and other related parameters.
For continuous arterial pressure waveform measurements, LiDCO Plus monitor uses the analogue arterial blood pressure trace that is slaved from the conventional blood pressure monitor to produce continuous arterial pressure waveform measurements while in case of Transonic HCM102 system a standard pressure transducer (such as MX950 or equivalent cleared by FDA K942377) is connected in-line with the hospital pressure transducer for continuous arterial pressure waveform measurements.
These differences do not raise any new issues of safety or effectiveness regarding the use of Transonic HCM102 System.
{6}------------------------------------------------
### Safety and Effectiveness:
The HCM102 is deemed to be safe and effective based on the safety testing conducted in accordance with the IEC 60601-1 standard and the electromagnetic compatibility test report.
In addition, bench and animal testing was conducted by Transonic Systems Inc. and the test reports can be found in Section 19 of this 510(k) submission. Prior to shipment, the finished product will be tested and must meet all required release specifications before distribution. The array of testing required for release includes, but are not limited to; physical testing and visual examination (in-process and finished product). The physical testing is defined by Quality Control Test Procedure documents. These tests are established testing procedures that ensure the product's performance parameters conform to the product design specifications. The testing instruction records for each of the individually required procedures are approved, released, distributed, and revised in accordance with document control cGMP's.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.