K134051 · Transonic Systems, Inc. · DXG · Jan 29, 2015 · Cardiovascular
Device Facts
Record ID
K134051
Device Name
TRANSONIC ELSA SYSTEM
Applicant
Transonic Systems, Inc.
Product Code
DXG · Cardiovascular
Decision Date
Jan 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1435
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of the ELSA system is to provide clinically relevant data to healthcare providers treating neonatal, pediatric, adult patients with arterial and venous lines for routine monitoring of diagnostic parameters: Delivered Flow; Recirculation: Oxygenator Blood Volume and other associated hemodynamic parameters.
Device Story
The Transonic ELSA system is a diagnostic monitor using transit time ultrasound indicator dilution technology to assess hemodynamic status during extracorporeal life support. It utilizes clamp-on sensors placed on extracorporeal circuit tubing to measure delivered flow, recirculation, and oxygenator blood volume (OXBV). The system consists of a monitor (HCE101), sensor pairs or single-head sensors, a fluid bag warmer, and a data transfer module. Used in ICUs or ORs by clinicians, the device displays measurements via a touchscreen, providing trend plots and user instructions. By quantifying hemodynamic parameters and oxygenator blood volume, the system assists healthcare providers in evaluating the efficacy of extracorporeal life support procedures and managing patient care.
Clinical Evidence
Bench testing only. Safety testing conducted per IEC 60601-1 and electromagnetic compatibility standards. Performance validated via physical testing and visual examination against established release specifications and Quality Control Test Procedures.
Technological Characteristics
Transit time ultrasound indicator dilution principle. Components include HCE101 monitor, HnFX series sensors, HFW1000 fluid bag warmer, and HCED01 data transfer module. Features touchscreen interface with trend plotting. Complies with IEC 60601-1 safety standards.
Indications for Use
Indicated for neonatal, pediatric, and adult patients undergoing extracorporeal life support procedures requiring routine monitoring of hemodynamic parameters.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
TRANSONIC COstatus Cardiac Output System (K113821)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol depicts a stylized caduceus with three human profiles incorporated into the design, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2015
Transonic Systems Inc. Naveen Thuramalla VP, Engineering & Clinical Studies 34 Dutch Mill Road Ithaca, NY 14850
Re: K134051
Trade/Device Name: Transonic ELSA System Regulation Number: 21 CFR 870.1435 Regulation Name: Computer, Diagnostic, Pre-programmed, Single-function Regulatory Class: II Product Code: DXG Dated: December 26, 2014 Received: December 29, 2014
Dear Naveen Thuramalla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
M.A. Hillebrand
for Bram Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K134051
#### Device Name
ELSA (Extracorporeal Life Support Assurance) System
#### Indications for Use (Describe)
The intended use of the ELSA system is to provide clinically relevant data to healthcare providers treating neonatal, pediatric, adult patients with arterial and venous lines for routine monitoring of diagnostic parameters: Delivered Flow; Recirculation: Oxygenator Blood Volume and other associated hemodynamic parameters.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Submitter's Name & Address: | Transonic Systems Inc<br>34 Dutch Mill Road,<br>Ithaca, NY 14850 |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person & Telephone: | Naveen Thuramalla<br>607-257-5300 (*326) |
| Date Summary Prepared: | January 28, 2015 |
| Device Name: | Classification Name: Computer, Diagnostic, Pre-<br>programmed, Single-function, 21 CFR 870.1435.<br>Product Class and Code: DXG and Class II<br>Classification Panel: Cardiovascular<br>Common/Usual Name: Extracorporeal, Diagnostic<br>monitor<br>Proprietary Name: Transonic ELSA System |
| Predicate Devices: | K960817: HD01-SERIES<br>TRANSONIC HEMODIALYSIS MONITOR<br><br>K113821: TRANSONIC COstatus Cardiac Output<br>System |
#### Device Description:
In accordance with section 510(k) for the Federal Food, Drug, and Cosmetic Act, Transonic Systems Inc. intends to introduce into interstate commerce the Transonic ELSA system, which is an apparatus based on transit time ultrasound indicator dilution techniques that provides clinically relevant data to healthcare providers treating patients undergoing Extracorporeal Life Support Procedures. The clinically relevant data (such as delivered flow, recirculation, oxygenator blood volume and other related hemodynamic parameters), shall indicate the efficacy of such procedures and quantify the patient's hemodynamic status. These patients could be in the intensive care units (ICU), operating room (OR) or other such environments.
### Components:
ELSA system consists of the following components:
{4}------------------------------------------------
| Model/Part # | Description |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| HCE101 | ELSA monitor/meter |
| HnFX | - Sensor pair configuration (H6FX; H7FX; H9FX.<br>Other sizes such as H5FX are possible<br>depending on the size of the tubing used byclinicians with extracorporeal circuits).<br>- Single head sensor configuration could also be<br>used with ELSA system. |
| HFW1000 | Fluid Bag Warmer |
| HCED01 | Data Transfer Module |
#### Substantial Equivalence:
The Transonic ELSA System for use with patients undergoing extracorporeal life support procedures is similar to the HD01- Series Transonic Hemodialysis Monitor (K960817) which is used in the diagnostic assessment of delivered blood flow, and recirculation. Both systems use clamp-on sensors placed on standard extracorporeal circuits to obtain diagnostic measurements.
Oxygenator blood volume (OXBV) measurement by ELSA system is similar to Central blood volume (CBV) measurement by Transonic COstatus Cardiac Output System (K113821). Both represent the volume between point of injection and point of detection of the indicator. The two systems are also similar in terms of touch screen device, onscreen user instructions, and measurement trend plots on the screen.
Both HD01 and COstatus devices are manufactured and marketed by Transonic Systems Inc. All three systems (ELSA, HD01 and COstatus) are based on transit time ultrasound dilution principles.
The difference between COstatus system and the proposed ELSA system is that the ELSA system would allow use of the system with patients being treated on extracorporeal life support procedures while COstatus system could be used with patients with in-situ arterial and venous catheters. COstatus measured CBV is the volume between injection site and recording site and includes the volume of blood in the heart, lungs and large vessels. Whereas, ELSA measured OXBV is the volume between injection site and recording site and includes the volume of blood in the oxygenator and any tubing in between the sensors.
The difference between ELSA system and the HD01 is that the ELSA system would be used with patients undergoing extracorporeal life support procedures while HD01 would be used with patients undergoing dialysis procedures.
{5}------------------------------------------------
These differences do not raise any new issues of safety or effectiveness regarding the use of Transonic ELSA System.
# Bench Testing:
The ELSA system (HCE101) is deemed to be safe and effective based on the safety testing conducted in accordance with the IEC 60601-1 standard and the electromagnetic compatibility test report.
In addition, bench testing was conducted by Transonic Systems Inc. and the validation report can be found in Section 18 of this 510(k) submission. Prior to shipment, the finished product will be tested and must meet all required release specifications before distribution. The array of testing required for release includes, but are not limited to: physical testing and visual examination (in-process and finished product). The physical testing is defined by Quality Control Test Procedure documents. These tests are established testing procedures that ensure the product's performance parameters conform to the product design specifications. The testing instruction records for each of the individually required procedures are approved, released, distributed, and revised in accordance with document control cGMP's.
### Conclusion:
Based on the information provided above, we conclude that the Transonic ELSA system is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.