# BASE IT (K134022)

_Spident Co., Ltd. · EJK · Jun 11, 2014 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K134022

## Device Facts

- **Applicant:** Spident Co., Ltd.
- **Product Code:** [EJK](/productcode/EJK.md)
- **Decision Date:** Jun 11, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3250
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Base it is a light-cured, radiopaque dental liner and base material that Contains calcium hydroxyl apatite in a urethane dimethacrylate oligomer. Base it is used to line cavity preparations before restoration.

## Device Story

Base It is a light-cured, radiopaque dental liner and base material; composed of calcium hydroxyl apatite in a urethane dimethacrylate oligomer. Used by dental professionals to line cavity preparations prior to placement of restorative materials (adhesives, composites). Applied to the cavity floor/walls; light-cured to harden. Provides a protective barrier; radiopaque for radiographic visibility. Benefits include protection of the pulp and support for restorative materials.

## Clinical Evidence

Bench testing only. Performance characteristics evaluated include flexural strength, depth of cure, compressive strength, radio-opacity, water sorption, and solubility. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity, sensitization, irritation/intracutaneous reactivity, subacute toxicity, and genotoxicity; all tests passed.

## Technological Characteristics

Light-cured, radiopaque dental liner. Composition: calcium hydroxyl apatite in urethane dimethacrylate oligomer. Complies with ISO 9917-2. Biocompatible per ISO 10993-1. Form factor: dental liner/base material.

## Regulatory Identification

A calcium hydroxide cavity liner is a device material intended to be applied to the interior of a prepared cavity before insertion of restorative material, such as amalgam, to protect the pulp of a tooth.

## Predicate Devices

- LIME-LITE ([K953079](/device/K953079.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

-k134022

Image /page/0/Picture/1 description: The image shows the word "SPIDENT" in a stylized, bold font. The letters are black and are enclosed within a black rectangular border. The top and bottom lines of the rectangle are interrupted, creating a unique visual effect.

# SPIDENT Co., Ltd.

JUN 1 1 2014

203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82(32)819-4570 Fax : +82(32)819-4572

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).

Date: November. 22, 2013

1. Company and Correspondent making the submission:

|         | Company                                                                                        |
|---------|------------------------------------------------------------------------------------------------|
| Name    | SPIDENT Co., Ltd.                                                                              |
| Address | 203 & 312, Korea Industrial Complex, 722,<br>Gojan-Dong, Namdong-Gu, Incheon, Korea<br>405-821 |
| Phone   | +82(32)819-4570                                                                                |
| Fax     | +82(32)819-4572                                                                                |
| Contact | J. M. Ahn/President                                                                            |

### 2. Device:

Proprietary Name - Base it Common Name - Calcium hydroxide cavity liner. Classification Name - liner, cavity, calcium hydroxide

#### 3. Predicate Device:

LIME-LITE, K953079 (LIMELIGHT)

- 4. Classifications Names & Citations: EJK, 872.3250
#### 5. Description:

Base it is light-cured Cavity Liner. It is radiopaque cavity liner and base material specially formulated for use with adhesives, composites and conventional restorative materials. It has urethane dimethacrylate base and is similar in chemistry to dental composites.

Base it is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. It is substantially equivalent in design, function and intended use to the predicate devices.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "SPIDENT" in bold, sans-serif font. The word is enclosed in a rectangular box, with the top and bottom lines of the box broken at the letter "I". The letters are black, and the background is white.

# SPIDENT Co., Ltd.

203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82(32)819-4570 Fax : +82(32)819-4572

### 6. Indication for use:

Base it is a light-cured, radiopaque dental liner and base material that Contains calcium hydroxyl apatite in a urethane dimethacrylate oligomer. Base it is used to line cavity preparations before restoration.

### 7. Performance

Testing performed including Flexural strength, Depth of cure, Compressive strength, Radio-opacity, Water sorption, Solubility, Radio-opacity use which demonstrated that the Base it is equivalent to the predicates in specifications and performance characteristics.

Therefore, it was considered that the subject device was as effective as and performs as good as the predicate device.

In conclusion, it can be said that the effectiveness and performance of the subject device are substantially equivalent to those of the predicate device.

#### 8. Biocompatibility

Biocompatibility testing confirmed that the device meets the applicable requirements of the FDA Blue Book Memorandum #95-1 entitled Use of International Standards ISO 10993 Biological Evaluation of Medical Devices Part -1: Evaluation and Testing, and is biocompatible. Accordingly, it was considered that the subject device was p substantially equivalent in safety to the predicate device.

| No. | Testing Biocompatibility                | Results |
|-----|-----------------------------------------|---------|
| 1   | Cytotoxicity                            | Pass    |
| 2   | Sensitization                           | Pass    |
| 3   | Irritation or Intracutaneous reactivity | Pass    |
| 4   | Subacute toxicity                       | Pass    |
| 5   | Genotoxocity                            | Pass    |

#### 9. Review:

The Base it has the similar device characteristics as the predicate device, the LIME-LITE; intended use, material, chemical composition, design and use concept are similar.

The Base it has the similar mechanical properties as the predicate device: Flexural Strength. Depth of cure, Compressive Strength, and Solubility,

The Base it has been subjected to extensive safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the US regulations and ISO 9917-2.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with the word "SPIDENT" in a stylized font. The letters "SPI" are enclosed in a square box with rounded corners, while the letters "DENT" are placed to the right of the box, with the top line of the box extending over the letters "DE". The overall design is simple and modern, with a focus on the company name.

# SPIDENT Co., Ltd.

203 & 312, Korea Industrial Complex, 722, Gojan-Dong, Namdong-Gu, Incheon, Korea 405-821 Tel: +82(32)819-4570 Fax : +82(32)819-4572

Based on the comparison of intended use and technical features, the Base it is substantially equivalent to the predicate devices.

- 10. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, FDA's "Guidance for the Preparation of Premarket notifications for Dental Composite" and based on the information provided in this premarket notification SPIDENT Co., Ltd. concludes that the Base it is safe and effective and substantially equivalent to predicate devices as described herein.

- 11. SPIDENT Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA.
END

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with flowing lines, representing movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem. The logo is simple, using a single color, and is designed to be easily recognizable.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 11, 2014

SPIDENT Company, Limited C/O Ms. Lena Pak SPIDENT USA, Incorporated 2115 Linwood Avenue 50 Floor, Suite 521 Fort Lee, NJ 07024

Re: K134022

Trade/Device Name: Base It Regulation Number: 21 CFR 872.3250 Regulation Name: Calcium Hydroxide Cavity Liner Regulatory Class: II Product Code: EJK Dated: March 5, 2014 Received: March 14, 2014

Dear Ms. Pak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

:

. .

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{4}------------------------------------------------

Page 2 - Ms. Pak

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

· .

Sincerely yours,

Mary S. Runner -S . . . . . . . . . . . . . . . . . .

Erin I. Keith. M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number K __ K134022

Device Name: Base it

Indication for use:

Base it is a light-cured, radiopaque dental liner and base material that Contains calcium bydroxyl apatite in a urethane dimethacrylate oligomer. Base it is used to line cavity preparations before restoration.

.

Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21CFR801 Subpart D)

AND/OR Over-The-Counter Use _ (Per 21CFR807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

:

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheena A. Green -S 2014.06.11 11:21:30 -04'00'

0 0 0 0 0 2 3

---

**Source:** [https://fda-staging.innolitics.com/device/K134022](https://fda-staging.innolitics.com/device/K134022)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).