MEDISYSTEMS ONESITE DUAL LUMEN NEEDLE WITH MASTERGUARD ANTI-STICK NEEDLE PROTECTOR, MEDISYSTEMS ONESITE DUAL LUMEN BUTTO

K134019 · Nxstage Medical, Inc. · FIE · Sep 11, 2014 · Gastroenterology, Urology

Device Facts

Record IDK134019
Device NameMEDISYSTEMS ONESITE DUAL LUMEN NEEDLE WITH MASTERGUARD ANTI-STICK NEEDLE PROTECTOR, MEDISYSTEMS ONESITE DUAL LUMEN BUTTO
ApplicantNxstage Medical, Inc.
Product CodeFIE · Gastroenterology, Urology
Decision DateSep 11, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medisystems Dual Lumen Needle is indicated for use as a vascular access device for dialysis procedures. The Medisystems Dual Lumen Buttonhole Needle is indicated fo use as a vascular access device for dialysis procedures using a constant-site or "buttonhole" method of needle insertion.

Device Story

Vascular access device for dialysis procedures; consists of dual lumen needle or dual lumen buttonhole needle. Used by clinicians to establish blood access. Dual lumen design allows for simultaneous blood withdrawal and return through a single needle site. Buttonhole variant facilitates consistent, repeated access at the same site. Devices include safety features (MasterGuard Anti-Stick Needle Protector or SteriPick). Used in clinical dialysis settings. Provides reliable access to circulatory system for extracorporeal blood treatment; enables efficient dialysis therapy.

Clinical Evidence

Bench testing only. Performance, verification, and validation testing were conducted to characterize device performance and confirm that predetermined acceptance criteria were met.

Technological Characteristics

Dual lumen needle design for vascular access. Includes safety mechanisms (MasterGuard Anti-Stick Needle Protector, SteriPick). Configuration and materials are similar to predicate devices. Sterilization performed at Kawasumi Laboratories Co. (Thailand) Ltd.

Indications for Use

Indicated for patients requiring vascular access for dialysis procedures. The Dual Lumen Needle is for standard insertion; the Dual Lumen Buttonhole Needle is for constant-site or "buttonhole" insertion methods.

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting community and connection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 11, 2014 NxStage Medical, Inc. Laura F. Plath Regulatory Affairs Manager 350 Merrimack Street Lawrence, MA 01843 Re: K134019 > Trade/Device Name: Dual Lumen needle/Dual Lumen Buttonhole needle Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: FIE Dated: July 30, 2014 Received: July 31, 2014 Dear Laura F. Plath, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known): not known at the time of submission | Device Name: | Medisystems Dual Lumen Needle | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use: | The Medisystems Dual Lumen Needle is indicated for use as a vascular access device for dialysis procedures. | | Device Name: | Medisystems Dual Lumen Buttonhole Needle | | Indications for Use: | The Medisystems Dual Lumen Buttonhole Needle is indicated fo use as a vascular access device for dialysis procedures using a constant-site or "buttonhole" method of needle insertion. | | Prescription Use | X | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {3}------------------------------------------------ This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k) summary has been provided in conformance with 21 CFR §807.92 #### Date: December 27, 2013 | A. Submitter's Information:<br>Name: | NxStage Medical, Inc. | |---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | 350 Merrimack Street<br>Lawrence, MA 01843<br>United States | | FDA Establishment<br>Owner/Operator Number: | 9045797 | | Contact Person: | Laura F. Plath<br>Regulatory Affairs Manager | | Phone: | (978) 332-5973 | | Fax: | (978) 687-4750 | | Manufacturing and<br>Sterilization Site: | Kawasumi Laboratories Co. (Thailand) Ltd.<br>48 Moo 8 Ratchasima-Chokchai Rd.,<br>Tambon Tha-ang, Amphur<br>Chokchai, Nakhorn Rachasima, Thailand | | FDA Establishment<br>Registration Number: | 9615908 | | B. Device Name:<br>Trade/Proprietary Name: | Medisystems onesite Dual Lumen Needle with<br>MasterGuard® Anti-Stick Needle Protector<br>Medisystems onesite Dual Lumen Buttonhole<br>Needle with SteriPick® | | Device: | Needle, fistula | | Regulation Description: | Blood access device and accessories. | | Regulation Medical<br>Specialty: | Gastroenterology/Urology Devices | | Review Panel: | Gastroenterology/Urology | | Product Code: | FIE | | Submission Type: | Traditional 510(k) | {4}------------------------------------------------ #### NxStage Medical, Inc. NxStage® Dual Lumen Needle and Dual Lumen Buttonhole Needle 510(k) Premarket Notification Submission Regulation Number: 876.5540 Device Class: = # C. Substantial Equivalence: These two proposed devices are substantially equivalent to two predicate devices: Medisystems Buttonhole Needles cleared under K990803 and the Cordis Corp. Bionics Bi-Flo double lumen needle cleared under K801355. The proposed Dual Lumen Needle and Dual Lumen Buttonhole Needle are substantially equivalent to these two cleared predicates. # D. Device Description/Indications for Use: The proposed devices are to be used as a vascular access device for dialysis procedures using either a standard needle insertion technique (for Dual Lumen Needle) or a constant-site or "buttonhole" method of insertion (for Dual Lumen Buttonhole Needle). ## Indications for use: | Dual Lumen Needle: | The Medisystems Dual Lumen Needle is indicated for use<br>as a vascular access device for dialysis procedures. | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Dual Lumen<br>Buttonhole Needle: | The Medisystems Dual Lumen Buttonhole Needle is<br>indicated for use as a vascular access device for dialysis<br>procedures using a constant-site or "buttonhole" method<br>of needle insertion. | ## E. Technological Characteristics: The proposed devices have the same technological characteristics and are similar in design and configuration as compared to the predicate devices. # F. Summary of Non-Clinical Test/Performance Testing - Bench NxStage believes that the information and data provided in this submission clearly describes the proposed devices and demonstrates that the devices are adequately designed for the labeled indications for use. Performance, verification and validation testing was conducted to characterize performance of the proposed devices and the predetermined acceptance criteria was met. Results of this testing have documented that the proposed devices are substantially equivalent to the predicate devices and are suitable for the labeled indications for use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...