← Product Code [DRG](/productcode/DRG) · K133997

# LG SMARTHEALTH (K133997)

_Lg Electronics · DRG · Nov 21, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133997

## Device Facts

- **Applicant:** Lg Electronics
- **Product Code:** [DRG](/productcode/DRG.md)
- **Decision Date:** Nov 21, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The LG SmartHealth is intended to collect vital sign measurements from physiological measurement devices intended for use in the home. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers (or nurse). Patients can also engage in video conferences with caregivers (or nurse). The LG SmartHealth is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. The LG SmartHealth is contraindicated for patients requiring direct-medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.

## Device Story

LG SmartHealth is an Android-based smartphone application for remote patient monitoring. It collects vital sign data from external Bluetooth-enabled medical devices (glucometers, blood pressure cuffs, weight scales, body fat readers, activity monitors). The app fetches readings, stores them locally, and transmits encrypted data (SHA256) to a server. Patients use the app to review measurements, receive educational/motivational content, and conduct video conferences with caregivers/nurses. The system is not interpretive, does not provide real-time data, and is not for diagnosis. It is intended for home use by patients capable of self-management. Caregivers use server-side software to review patient data and set thresholds. The device facilitates communication between patients and healthcare providers, supporting non-time-critical care management.

## Clinical Evidence

No clinical testing performed. Substantial equivalence supported by bench testing, software verification/validation, and a usability study confirming human factors, ease of use, and comprehension by participants and nurses.

## Technological Characteristics

Android-based software application (OS 2.3.3+). Connectivity via Bluetooth (2.402-2.480 GHz) to external medical sensors. Data transmission to server via internet. Security features include SHA256 encryption and MAC address filtering. Operates on commercially available Android smartphones/tablets.

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- Intel Health Guide Express ([K103276](/device/K103276.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 21, 2014

LG Electronics c/o Ms. Diane Sudduth Senior Consultant, RA/Q/A 816 Congress Avenue, Suite 1400 Austin, TX 78701

Re: K133997

Trade/Device Name: LG SmartHealth Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Code: DRG Dated: November 19, 2014 Received: November 20, 2014

Dear Ms. Sudduth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Ms. Diane Sudduth

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K133997

Device Name

LG SmartHealth

#### Indications for Use (Describe)

The LG SmartHealth is intended to collect vital sign measurement devices intensurement devices intended for use in the home . Patients can review the stored vital sign measurement information and motivational content from caregivers(or nurse). Patients can also engage in video conferences with caregivers(or nurse).

The LG SmartHealth is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required.

The LG SmartHealth is contraindicated for patients requiring direct-medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

for

## LG SmartHealth

#### 1. Submission Sponsor

LG Electronics 38 Baumoe-ro, Seocho-gu Seoul, 137-724 Korea Phone: +82-2-526-4527 Fax: +82-2-526-4370 Contact: Chang Hee Lee, Chief Research Engineer

#### 2. Submission Correspondent

Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (561) 305-5075 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Dr. Diane Sudduth, Senior Consultant, QA Email: project.management@emergogroup.com

#### 3. Date Prepared

December 21, 2013

#### 4. Device Identification

| Trade/Proprietary Name:    | LG SmartHealth                                                     |
|----------------------------|--------------------------------------------------------------------|
| Common/Usual Name:         | Remote Patient Monitoring System                                   |
| Classification Name:       | Transmitters and Receivers, Physiological Signal<br>Radiofrequency |
| Classification Regulation: | 870.2910                                                           |
| Product Code:              | Product Code DRG                                                   |
| Device Class:              | Class II                                                           |
| Classification Panel:      | Cardiovascular                                                     |

#### 5. Legally Marketed Predicate Device(s)

Intel Health Guide Express (K103276)

{5}------------------------------------------------

## 6. Device Description

The LG SmartHealth application is software planned to operate on the Android operating system (OS) that can be loaded as an application or "app" on a commercially available smartphone. LG intended the software to work on the Android OS, as well as to operate on smartphones as well. The scalability and differences between many smartphones today are not too great to allow this flexibility and option.

The LG SmartHealth application connects to commercially available wireless medical devices that are commonly used by patients in a home-care setting. These "sensor devices" are FDA cleared or FDA registered (Exempt) devices that can communicate with the LG SmartHealth application software loaded on the smartphone using Bluetooth connectivity. To avoid wireless interference and confusion on the part of the patient, serial readings are performed (i.e. only one Bluetooth sensor used at a time) and MAC address filtering is used for the various medical devices. The LG SmartHealth application plans to support the following sensor available medical devices, some initially, and others as the technology becomes available:

- -Glucose meters or glucometers
- -Blood pressure cuffs
- -Weight scales
- -Body fat readers
- -Activity monitors

On each screen of the LG SmartHealth application software a help button is present, which when pressed provides step-by-step based help to guide a patient through an interaction.

The medical device takes a reading depending on the frequency previously established and the LG SmartHealth application software fetches the information for the reading automatically. The LG SmartHealth application software stores and displays the information on the smartphone and transmits the information to the server. The LG SmartHealth application encrypts the information in preparation for transmit using SHA256 bit encryption, complying with HIPAA requirements.

#### 7. Indication for Use Statement

The LG SmartHealth is intended to collect vital sign measurements from physiological measurement devices intended for use in the home. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers (or nurse). Patients can also engage in video conferences with caregivers (or nurse).

The LG SmartHealth is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required.

The LG SmartHealth is contraindicated for patients requiring direct-medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing

{6}------------------------------------------------

its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.

## 8. Substantial Equivalence Discussion

The comparison chart below provides evidence to facilitate the substantial equivalence determination between LG SmartHealth to the predicate device (K103276) with respect to intended use, technological characteristics and principles of operation.

| Features and<br>Specifications       | Intel® Health Guide Express<br>(K103276)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LG SmartHealth<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Information               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Intended Use/ Indications for<br>Use | The Intel® Health Guide Express is intended<br>to collect vital sign measurements from<br>physiological measurement devices<br>intended for use in the home.<br><br>Patients can review the stored vital sign<br>measurement information and receive<br>educational and motivational content from<br>caregivers. Patients can also engage in<br>video conferences with caregivers and<br>answer the caregivers' questions by<br>participating in surveys.<br><br>The Intel® Health Care Management Suite<br>allows the caregiver to review patient data<br>and initiate video conferencing with<br>patients, or select and send educational and<br>motivational content to patients.<br><br>The Intel® Health Guide Express is not<br>interpretive, nor is it intended for diagnosis<br>or as a substitute for medical care, and it is<br>not intended to provide real time data. It is<br>made available to patients when time-<br>critical care is not required.<br><br>The Intel® Health Guide Express is<br>contraindicated for patients requiring direct-<br>medical supervision or emergency<br>intervention. It is intended for patients who<br>are willing and capable of managing its use.<br>Clinical judgment and experience by a<br>caregiver are required to check and<br>interpret the information delivered. | The LG SmartHealth is intended to<br>collect vital sign measurements<br>from physiological measurement<br>devices intended for use in the<br>home.<br><br>Patients can review the stored vital<br>sign measurement information and<br>receive educational and<br>motivational content from<br>caregivers (or nurse).<br>Patients can also engage in video<br>conferences with caregivers (or<br>nurse).<br><br>The LG SmartHealth is not<br>interpretive, nor is it intended for<br>diagnosis or as a substitute for<br>medical care, and it is not intended<br>to provide real time data. It is<br>made available to patients when<br>time-critical care is not required.<br><br>The LG SmartHealth is<br>contraindicated for patients<br>requiring direct-medical<br>supervision or emergency<br>intervention. It is intended for<br>patients who are willing and<br>capable of managing its use.<br>Clinical judgment and experience<br>by a caregiver are required to<br>check and interpret the<br>information delivered. |
| OTC and/or Rx                        | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

## Table 5A – Comparison of Characteristics

{7}------------------------------------------------

| User Characteristics                                                                        |                                                                                                                              |                                                                                                                                    |
|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Intended User                                                                               | Patient in Home Setting                                                                                                      | Same                                                                                                                               |
| Device                                                                                      |                                                                                                                              |                                                                                                                                    |
| Components                                                                                  | Web based software application Data Transfer Tool (Integrated)                                                               | Smart phone app based software application Data Transfer Tool (Integrated)                                                         |
| User Interface Device                                                                       | Computer Device (i.e. desktop, laptop)                                                                                       | Android Device(i.e. Smart phone, tablet)                                                                                           |
| Connectivity to Device                                                                      | Bluetooth or USB wired                                                                                                       | Bluetooth                                                                                                                          |
| Ability to Update Device Settings                                                           | Not available                                                                                                                | Automatic deploy                                                                                                                   |
| Connectivity to Web-based Services                                                          | Login Secured by SSL Encryption                                                                                              | Same                                                                                                                               |
| Operating System                                                                            | Microsoft Windows 7 (32-bit versions)                                                                                        | Android 2.3.3 or higher                                                                                                            |
| Supported Device Types                                                                      | Medical devices designed for home use:<br>Blood glucose meter<br>Weight scale<br>Blood pressure<br>Pulse Oximeter<br>FEV/PEF | Medical devices designed for home use:<br>Blood glucose meter<br>Weight scale<br>Blood pressure<br>Temperature<br>Activity Monitor |
| Software Delivery                                                                           | Web download for the uploader only                                                                                           | Same                                                                                                                               |
| Software Features                                                                           |                                                                                                                              |                                                                                                                                    |
| Patient Management                                                                          | Manage patient accounts, filter/sort/search patients, print reports.                                                         | The server connected to the mobile device manages patient accounts, filter/sort/search patients                                    |
| Implementation method of collecting data from sensors                                       | Bluetooth or USB wired                                                                                                       | Bluetooth                                                                                                                          |
| Communication frequency                                                                     | Bluetooth: 2.402 to 2.480 GHz                                                                                                | Same                                                                                                                               |
| Max data throughput under worst case conditions                                             | Multiply read data can be stored in the medical device and they are resent to the server after recovery of network condition | Same                                                                                                                               |
| Use of<br>thresholds/algorithms for<br>determining how<br>thresholds are set and<br>changed | Thresholds set by healthcare<br>professional in server software                                                              | Same                                                                                                                               |
| Communication method<br>with server                                                         | Connected through internet                                                                                                   | Same                                                                                                                               |

{8}------------------------------------------------

The LG SmartHealth is a medical device and has similar indications for use statement as the predicate device. The differences in the indications statement between LG SmartHealth and Intel Health Guide Express do not alter the intended use – both devices have essentially the same intended use.

The device also has similar technological characteristics as the predicate device. Both devices have the same data collection software functionality, communication method of patient device with central server, types of sensors which can be interfaced to the patient device, implementation method of collecting data from sensors, sensor software, connectivity, and communication protocol. Since the comparison of the descriptive characteristics of the proposed and predicate devices may not be sufficiently precise to assure equivalence, performance data are provided. The results of the performance testing demonstrate substantial equivalence.

## 9. Non-Clinical Performance Data

The device's software development, verification, and validation have been carried out in accordance with FDA guidelines. The software was tested against the established Software Design Specifications for each of the test plans to assure the device performs as intended. The device Hazard analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria of each module and interaction of processes. The LG SmartHealth device passed all testing and supports the claims of substantial equivalence and safe operation.

Validation activities included a usability study of the LG SmartHealth under actual use. The study demonstrated:

- Comprehension of the study nurses and participants with the LG SmartHealth,
- Appropriate human factors related to the LG SmartHealth, and ●
- Ease of use of the LG SmartHealth.

#### 10. Clinical Performance Data

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The verification and validation testing of the device's software was found to be acceptable and supports the claims of substantial equivalence.

{9}------------------------------------------------

## 11. Statement of Substantial Equivalence

The LG SmartHealth has the same or similar intended use as the predicate device and that any technological differences between the LG SmartHealth software and the predicate device do not raise any questions regarding LG SmartHealth')'s safety and effectiveness.

The information provided in this submission supports the substantial equivalence to the predicate device and that the system is safe and effective for the users/operators.

---

**Source:** [https://fda-staging.innolitics.com/device/K133997](https://fda-staging.innolitics.com/device/K133997)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
