1.5T ORBIT AND MANDIBLE ARRAY, 3.0T ORBIT AND MANDIBLE ARRAY
K133986 · Resonance Innovations, LLC · MOS · Jan 27, 2014 · Radiology
Device Facts
Record ID
K133986
Device Name
1.5T ORBIT AND MANDIBLE ARRAY, 3.0T ORBIT AND MANDIBLE ARRAY
Applicant
Resonance Innovations, LLC
Product Code
MOS · Radiology
Decision Date
Jan 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The family is to be used in conjunction with a GE 1.5T and GE 3.0T Magnetic Resonance Scanner, to produce diagnostic images of the orbits and surrounding anatomy, as well as mandible regions that can be interpreted by a trained physician. The intended use for the device family is to provide MRI antenna sets to facilitate targeting imaging of the orbit region and mandible region. Tissue of the brain and face with emphasis on the orbits, and of the mandible joint of the head.
Device Story
The Orbit and Mandible Array Family consists of receive-only MRI antenna sets (1.5T and 3.0T models) designed for targeted imaging of the orbit and mandible regions. The device housing is constructed from EVA foam covering a flexible circuit board. It functions as a passive coil, receiving signals stimulated by the MRI system's transmit coils; it does not generate its own power. Used in clinical settings by trained physicians to produce diagnostic images of the brain and face. The smaller loop design allows for closer proximity to the patient's face compared to general-purpose coils, facilitating higher-resolution performance in the target anatomy. The output is interpreted by a physician to aid in clinical diagnosis.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including sensitivity profiles, SNR analysis, hazard analysis, FMEA, blocking analysis, and heat tests. Imaging comparisons in sagittal, coronal, and transverse planes demonstrated equivalent performance to the predicate.
Technological Characteristics
Receive-only MRI coils. Materials: EVA foam housing, flexible circuit board. Biocompatible construction. No internal power source; stimulated by MRI system. Dimensions optimized for orbit/mandible anatomy. Decoupling circuitry included. Tested for dielectric withstand and liquid ingress.
Indications for Use
Indicated for patients requiring diagnostic MRI imaging of the orbit region, surrounding anatomy, and mandible joint. Used in conjunction with GE 1.5T and 3.0T scanners.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Orbit and Mandible Array Family
Resonance Innovations LL
| Section 8. 510(k) Summary | k133986 | JAN 27 2014 |
|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------|
| Submitter's Name: | Resonance Innovations LLC | |
| Submitter's Address: | 9840 South 140th St., Suite 8<br>Omaha, NE 68138 | |
| Submitter's Telephone: | 402-934-2650 | |
| Submitter's Contact: | Randall Jones, President | |
| Date 510(k) summary prepared: | January 22, 2014 | |
| Name of the device, including the trade or proprietary name if applicable, the common<br>or usual name and the classification name, if known: | | |
| Proprietary Name: | Orbit and Mandible Array Family |
|-----------------------|-------------------------------------|
| Common or Usual Name: | MRI coil(s) |
| Classification Name: | Coil, Magnetic Resonance, Specialty |
| Classification Code: | MOS |
| Predicate Device: | Pediatric Array Coil (K951649) |
Description of the Device
The devices are a 1.5T Orbit and Mandible Array and a 3.0T Orbit and Mandible Array. The designs and materials used to manufacture each coil are no different from standard MRI coil technology that has existed for years. The geometry of each coil housing is formed by an EVA foam cover around a flexible circuit board. The two devices in this family are identical except for the tuned frequency. For this reason, it is warranted to bundle these two devices into one 510(k) submission.
The following devices are bundled in this submission.
| Device Name | Model Number | Operational Field Strength<br>of MRI System |
|-------------------------------|--------------|---------------------------------------------|
| 1.5T Orbit and Mandible Array | 994GE1500 | 1.5T |
| 3.0T Orbit and Mandible Array | 992GE3000 | 3.0T |
## Indications for Use
The family is to be used in conjunction with a GE 1.5T and GE 3.0T Magnetic Resonance Scanner, to produce diagnostic images of the orbits and surrounding anatomy, as well as mandible regions that can be interpreted by a trained physician.
The intended use for the device family is to provide MRI antenna sets to facilitate targeting imaging of the orbit region and mandible region.
- Tissue of the brain and face with emphasis on the orbits, and of the mandible joint 0 of the head.
{1}------------------------------------------------
Orbit and Mandible Array Family
K133986
Resonance Innovations LLC Page 2 of 2
# Technological Characteristics
The differences between the predicate and the current submission are described, below,
- 1.) Design. The submission is for a smaller subset of anatomies of the larger predicate. itself having elements close to the patient's orbits. Very much similarly, this new submission of the Orbit & Mandible Family utilizes smaller dedicated loops closer to the patient's face.
- 2.) Material. The distinction between the predicate and Orbit & Mandible Family group is non-substantive in that they both have insulating qualities proven by dielectric withstand testing, and have passed liquid ingress testing.
- 3.) Chemical Composition. Both products, the predicate and the recent submission, have biocompatible construction as demonstrated by cytotoxicity testing and by their history of use in previously cleared devices.
- 4.) Encrgy Source. Both of these products are receive-only coils not generating their own power, but rather stimulated by the MRI system as an energy source.
# Non-clinical tests
The coils have similar dimension in the head area and have induced similar fields by the transmit coils that stimulate them. The predicate and current submission were compared by similar risk management efforts, as listed below, and determined to be substantially equivalent.
- Sensitivity Profile ●
- SNR Analysis .
- Hazard Analysis ●
- FMEA .
- Blocking Analysis .
- Heat Tests �
#### Clinical tests
Analyses in all 3-planes (sagittal, coronal and transverse) were run to show that the anatomies of the submitted and predicate coils have substantial equivalence; the predicate device images the complete brain including orbits, and the Orbit & Mandible Family submission simply dedicates to a more local, higher-resolution performance on the orbits and mandible.
#### Final Discussion
As described in the Design section. above, the submission is for smaller loops dedicated more closely to the patient's face, with the coils offering no energy source, and exhibiting strong decoupling: thus, the study determines that the submitted device is as safe, as effective, and performs as well or better than the legally-marketed device on the target anatomies.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27,2014
RESONANCE INNOVATIONS LLC RANDALL JONES PRESIDENT 9840 SOUTH 140TH ST., SUITE 8 OMAHA NE 68138
Re: KI33986
Trade/Device Name: Orbit and Mandible Array Family Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: December 31, 2013 Received: January 06, 2014
Dear Mr. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2-Mr. Jones
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mikara M. O'Hara
For
Janine Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
.
510(k) Number (if known): k133986
Device Name: Orbit and Mandible Array Family
Indications for Use:
The Indications for use are as follows:
- · The family is to be used in conjunction with a GE 1.5T and GE 3.0T Magnetic Resonance Scanner, to produce diagnostic images of the orbits and surrounding anatomy, as well as mandible regions that can be interpreted by a trained physician.
- · The intended use for the device family is to provide MRI antenna sets to facilitate targeting imaging of the orbit region and mandible region.
- o Tissue of the brain and face with emphasis on the orbits, and of the mandible joint of the head.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological IIcalth
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.