← Product Code [MUE](/productcode/MUE) · K133972

# ASPIRE CRISTALLE (K133972)

_Fujifilm Medical Systems U.S.A, Inc. · MUE · Mar 25, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133972

## Device Facts

- **Applicant:** Fujifilm Medical Systems U.S.A, Inc.
- **Product Code:** [MUE](/productcode/MUE.md)
- **Decision Date:** Mar 25, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1715
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

## Device Story

ASPIRE Cristalle is an integrated full-field digital mammography (FFDM) system; utilizes amorphous selenium (a-Se) detector with TFT readout to capture x-ray energy; converts x-rays to digital images. System includes gantry, x-ray generator, and Acquisition Workstation (AWS). AWS features Windows 7 PC, 5-megapixel portrait monitor, and hub for signal transmission. Technologist operates system to perform patient imaging; provides powered compression and three AEC modes. AWS enables automated post-processing, technologist preview, QC, and transmission to printers/PACS. Radiologists view images on diagnostic workstations to screen/diagnose breast cancer. Benefits include digital image capture replacing traditional screen-film methods.

## Clinical Evidence

Clinical image attribute review conducted by independent mammographic radiologists. Six image sets of screening and diagnostic cases reviewed across all exposure modes. Results confirmed ASPIRE Cristalle provides sufficiently acceptable image quality for mammographic use. Non-clinical testing included Sensitometric Response, Spatial Resolution, Noise Analysis, DQE, Dynamic Range, AEC performance, and ACR MAP/Contrast Detail phantom testing.

## Technological Characteristics

FFDM system; amorphous selenium (a-Se) detector with TFT readout; powered compression; three AEC modes. AWS uses Windows 7 OS, 5MP portrait monitor. Connectivity via hub to control cabinet and exposure stand. Software-based automated post-processing. Complies with FFDM 510(k) guidance for imaging performance (MTF, DQE, CNR).

## Regulatory Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

## Predicate Devices

- FUJIFILM Aspire HD Plus (FDR MS-2500) ([K121674](/device/K121674.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

**MAR 2 5 2014**

114 K133 972
Page 1 of 2

19 WEST AVE 203-324-2000 203-353-0926

# 510(k) Summary

#### Date Prepared: December 18, 2013

#### Submitter's Information

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (203) 602-3576 Facsimile: (203) 602-3785 Contact: Peter Altman

# Device Name and Classification:

| Product Name:         | ASPIRE Cristalle                            |
|-----------------------|---------------------------------------------|
| Model Number:         | FDR MS-3500                                 |
| Classification Name:  | Full-Field Digital Mammography X-ray System |
| Classification Panel: | Radiology                                   |
| CFR Section:          | 21 CFR 892.1715                             |
| Device Class:         | Class II                                    |
| Product Code:         | MUG                                         |

#### Substantial Equivalence/Predicate Device:

FUJIFILM Aspire HD Plus (FDR MS-2500) (K121674)

The predicate and proposed devices have the same Indication For Use (IFU). Both devices generate digital mammographic images that are intended for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.

Both systems employ amorphous selenium digital x-ray detectors integrated into the gantry (stand) based x-ray systems. The Cristalle uses hexagonal pixels as compared to the square pixels of the HD Plus. The x-ray stands of the ASPIRE Cristalle and Aspire HD Plus are the same and the generators are extremely similar. The technological characteristics of the devices are similar as demonstrated by the comparison of imaging characteristics such as MTF, Noise Analysis, DOE, CNR, Phantom testing, etc. measured during non-clinical testing. Non-clinical testing was conducted in accordance with the FFDM 510(k) Guidance document covering Sensitometric Response, Spatial Resolution, Noise Analysis, Signal-to-Noise Ratio Transfer - DQE, Dynamic Range, Image Erasure and Fading, Repeated Exposure Test, AEC Performance, ACR MAP Phantom Testing, Contrast Detail Phantom Testing, Patient Radiation Dose Testing, and Breast Compression system Testing.

A clinical image attribute review was conducted by independent mammographic radiologists in accordance with the FFDM 510(k) Guidance document. The mammographic attributes of six (6) image sets of screening and diagnostic cases were reviewed for all exposure modes concluding that the ASPIRE Cristalle provides sufficiently acceptable image quality for mammographic use.

{1}------------------------------------------------

# FUJKFILM

#### FUJIFILM MEDICAL SYSTEMS USA,

19 WEST AVENUE TAMFORD, CT 06902 203-324-2000 TOLL-FREE: 800-431-1850 203-353-0926

## Description of the Device:

The ASPIRE Cristalle is an integrated FFDM system combining an X-ray system made by Fujifilm with Fujifilm's a-Se detector and Acquisition Workstation (AWS). The ASPIRE Cristalle creates digital mammography images by direct capture of x-ray energy using the a-Se detector. The detector is a Fujifilm design utilizing an a-Se photo-conversion layer with TFT Readout circuitry to acquire image data and transfer images to the AWS for automated post processing, technologist preview and QC, and subsequent transmission to hard copy printers, diagnostic workstations and archiving systems. The ASPIRE Cristalle provides powered compression and three AEC modes.

The ASPIRE Cristalle Acquisition Workstation (FDR 3000A WS) includes an off the shelf personal computer, the application software, Windows 7 Operating System, a 5megapixel portrait type monitor, and a hub. The hub transmits signals between the personal computer and control cabinet, and between the personal computer and exposure stand.

The AWS display primarily consists of three windows:

- Patient Information Input window .
- Exposure Menu Selection window .
- . Study window.

The user may switch between these windows depending on the operation being performed. The X-ray control panel, which controls and observes the exposure stand, is always displayed in the lower part of each window. This allows setting the exposure conditions and confirming the radiation conditions on a single view.

# Intended Use:

The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates fullfield digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

#### Safety Information:

The ASPIRE Cristalle introduces no new safety or efficacy issues other than those already identified with the predicate device. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and is consistent with the level of concern indicated in the "Class II Special Controls Guidance Document: Full-Field Digital Mammography System" document issued on: November 5, 2010.

#### Conclusion:

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject devices to be as safe and effective as the predicate device based upon the clinical and non-clinical data summarized above.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2014

Fujifilm Medical Systems USA, Inc. % Mr. Peter Altman Regulatory Consultant 419 West Avenue STAMFORD CT 06902

Re: K133972

Trade/Device Name: ASPIRE Cristalle Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: 11 Product Code: MUE Dated: December 23, 2013 Received: December 26, 2013

Dear Mr. Altman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

# Page 2 - Mr. Peter Altman

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.

Sincerely yours,

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

| K133972                                                                                                                                    |  |
|--------------------------------------------------------------------------------------------------------------------------------------------|--|
| and and the county of the county of the first of the first of the first of the first of the first of the first of the first<br>Device Name |  |
| ASPIRE Cristalle                                                                                                                           |  |

Indications for Use (Describe)

The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

[ ] Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (9/13)

Page 1 of 2

PSC Publishing Servers (1) 0)) 441-6140

ASPIRE Cristalle 510(k)

Section 4-Page 2

---

**Source:** [https://fda-staging.innolitics.com/device/K133972](https://fda-staging.innolitics.com/device/K133972)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
