WONDFO MULTI-DRUG URINE TEST CUP, PANEL
Device Facts
| Record ID | K133968 |
|---|---|
| Device Name | WONDFO MULTI-DRUG URINE TEST CUP, PANEL |
| Applicant | Guangzhou Wondfo Biotech Co., Ltd. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | Feb 25, 2014 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Indications for Use
Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Amphetamine, of Methylenedioxymethamphetamine, Morphine, Phencyclidine, Oxycodone, Buprenorphine, Methadone and Notriptyline in human urine at the cutoff concentrations of: [Table of cut-offs]. Configuration of the Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. The test may vield positive results for the prescription drugs buprenorphine, oxazepam, oxycodone, and secobarbital when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.
Device Story
Device modification combines previously cleared single-analyte drug-of-abuse (DOA) tests into multi-test device; detects 2 to 13 drugs simultaneously in human urine. Input: urine sample; principle: lateral flow immunoassay; output: qualitative visual result (presence/absence of drug). Used at point-of-care sites by healthcare professionals. Output assists clinicians in assessing patient drug compliance.
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control procedures to confirm performance characteristics of the multi-test format compared to single-analyte predicates.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Form factors: test cup and test panel (dip card). Multi-analyte configuration (2-13 drugs). In vitro diagnostic use. No electronic components or software.
Indications for Use
Indicated for qualitative, simultaneous detection of drugs of abuse (Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, MDMA, Morphine, Phencyclidine, Oxycodone, Buprenorphine, Methadone, Notriptyline) in human urine. Intended for OTC and prescription use. Not intended to distinguish between prescription use and abuse.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Wondfo single DOA Tests (K112071)