K133963 · Adhezion Biomedical, LLC · MPN · Sep 9, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133963
Device Name
SURGISEAL TWIST (TM) TOPICAL SKIN ADHESIVE
Applicant
Adhezion Biomedical, LLC
Product Code
MPN · General, Plastic Surgery
Decision Date
Sep 9, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SURGISEAL Topical Skin Adhesive is intended for topical applications only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, simple, thoroughly cleansed, trauma induced lacerations. SURGISEAL may be used in conjunction with, but not in place of, deep dermal sutures.
Device Story
SURGISEAL Twist is a sterile, single-use topical skin adhesive containing 2-octyl cyanoacrylate and D & C Violet #2 colorant. The device consists of a plastic ampoule within an applicator sleeve and tip. Used by clinicians to hold skin edges of surgical incisions or lacerations together; the adhesive polymerizes in minutes upon application to form a physical barrier to microbial penetration. The device is applied manually by the clinician to the wound area. The current version features a modified applicator tip dimension compared to the predicate. Benefits include wound closure and potential microbial barrier protection. No clinical studies were conducted to correlate barrier properties with infection reduction.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included flexibility, linear and surface coverage, wound closure strength, and force expression. Biocompatibility testing was leveraged from the predicate device (K130474) per ISO 10993.
Technological Characteristics
Monomeric 2-octyl cyanoacrylate formulation with D & C Violet #2. Single-use plastic applicator. Sterilization: Gamma irradiation (ISO 11137-2:2006) for filled ampoule; Ethylene oxide (ISO 11135-1 and 2:2008) for finished bulk applicator. Low viscosity for layered application.
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive surgery punctures) and simple, cleansed, trauma-induced lacerations. Not for use in place of deep dermal sutures.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2014
Adhezion Biomedical, LLC Ms. Caridad Smith Director of Regulatory Affairs and Quality Assurance One Meridian Boulevard, Suite 1B02 Wyomissing, Pennsylvania 19610
Re: K133963
Trade/Device Name: SURGISEAL Twist" Topical Skin Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue Adhesive Regulatory Class: Class II Product Code: MPN Dated: August 13, 2014 Received: August 14, 2014
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133963
Device Name SURGISEAL Twist™ Topical Skin Adhesive
#### Indications for Use (Describe)
SURGISEAL Topical Skin Adhesive is intended for topical applications only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, simple, thoroughly cleansed, trauma induced lacerations.
SURGISEAL may be used in conjunction with, but not in place of, deep dermal sutures
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Adhezion BIOMEDICAL. The logo consists of a blue abstract design on the left, resembling curved lines or waves. To the right of the design is the word "Adhezion" in a dark blue, sans-serif font. Below "Adhezion" is the word "BIOMEDICAL" in a smaller, lighter blue font.
# 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92 is below:
| Submitter Information: | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Adhezion Biomedical, LLC |
| Address | One Meridian Boulevard<br>Suite 1B02<br>Wyomissing, PA 19610 |
| Phone Number | (828) 572-4201 |
| Fax Number | (828) 726-7117 |
| Establishment Registration | 3006385287 |
| Name of contact person(s) | Caridad Smith – Director of Regulatory Affairs/Quality Assurance<br>Tracy Lofgren – Regulatory Consultant |
| Date prepared | August 27, 2014 |
| Name of Device: | |
| Trade or proprietary names | SURGISEAL Twist™ Topical Skin Adhesive |
| Common or usual name | Device, Tissue Adhesive for Topical Approximation |
| Classification name | Class II |
| Classification Panel | General and Plastic Surgery Devices |
| Regulation | Class II, under 21 CFR 878.4010 |
| Product Code(s) | MPN |
| Legally Marketed device(s) to which equivalence is claimed | SURGISEAL Stylus Twist™ Topical Skin Adhesive (K130474) |
| Reason for 510(k) submission | Device Modification – Dimensional Specification Change to Tip |
| Device Description | SURGISEAL Twist Topical Skin Adhesives are sterile, professional liquid<br>topical skin adhesives containing a monomeric (2-octyl cyanoacrylate)<br>formulation and the colorant D & C Violet #2. Each applicator is comprised of<br>a plastic ampoule container contained within an applicator sleeve with the<br>applicator tip. This Twist applicator tray is contained in an outer Tyvek pouch.<br>When SURGISEAL is applied to the skin, it polymerizes in minutes.<br><br><i>In vitro</i> studies have shown that SURGISEAL acts as a physical barrier to<br>microbial penetration as long as the adhesive film remains intact. Clinical<br>studies were not conducted to demonstrate microbial barrier properties and a<br>correlation between microbial barrier properties and a reduction in infection<br>have not been established. |
| Indications for use | SURGISEAL Twist Topical Skin Adhesive is intended for topical applications<br>only to hold closed easily approximated skin edges of wounds from surgical<br>incisions, including punctures from minimally invasive surgery, simple,<br>thoroughly cleansed, trauma induced lacerations.<br><br>SURGISEAL may be used in conjunction with, but not in place of, deep<br>dermal sutures. |
| Technological<br>Characteristics | The technological characteristics of SURGISEAL Stylus Twist and<br>SURGISEAL Twist Topical Skin Adhesives are equivalent in performance.<br><br>SURGISEAL Twist consists of a monomeric (2-octyl cyanoacrylate) liquid<br>adhesive formulation packaged in a single-use applicator. The device is a low<br>viscosity formulation to allow for varied layered applications of the adhesive to<br>the intended area and allow for either a single thick, continuous layer or two<br>thin layers of the adhesive to the wound area.<br><br>The smaller tip dimensions of the SURGISEAL Twist applicator is the only<br>difference between the subject and predicate devices. |
| Substantial<br>Equivalence | Biocompatibility:<br>The biocompatibility testing that was previously conducted to the currently<br>marketed devices, SURGISEAL Stylus Twist (K130474) per the International<br>Standard ISO-10993, “Biological Evaluation of Medical Devices, Part 1:<br>Evaluation and Testing”. The existing testing is deemed supportive of the<br>modified device. Based on the results from those studies, the subject device is<br>considered to be non-toxic, non-irritating, non-sensitizing and biocompatible<br>and no additional biocompatibility testing was necessary.<br><br>Performance Testing:<br>The performance testing of SURGISEAL Twist Topical Skin Adhesive<br>modified product is identical to the currently marketed product; therefore the<br>performance testing provided in the Premarket Notification K130474 is the<br>same. Additional bench testing was performed to support the dimensional tip<br>specification modification to the currently marketed product. Bench tests<br>included: Flexibility, Linear and Surface Coverage, Wound Closure Strength<br>and Force Expression.<br><br>In all cases, the dimensional tip specification modification SURGISEAL Twist<br>Topical Skin Adhesive met specifications and demonstrated equivalence to the<br>currently marketed device.<br><br>Sterilization and Shelf-Life<br>The sterilization process of the dimensional tip specification modification |
| device remains unchanged. All products within the SURGISEAL Twist | |
| Topical Skin Adhesive product family are sterilized in accordance with the | |
| order of operation of the assembly. First to be sterilized is the filled ampoule | |
| containing the adhesive, by gamma irradiation in accordance with ISO 11137- | |
| 2:2006. Then the finished bulk applicator in the secondary packaging is | |
| sterilized by ethylene oxide in accordance with ISO 11135-1and 2:2008. | |
| The proposed dimensional tip specification modification of SURGISEAL<br>Twist™ Topical Skin Adhesives does not impact the 24 month (2 year)<br>expiration date (shelf-life) of the currently marketing predicated devices. | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Adhezion Biomedical. The logo consists of a stylized graphic to the left of the company name. The graphic is made up of several curved lines in a light blue color. The company name is written in a dark blue font, with the word "Adhezion" on top and the word "BIOMEDICAL" underneath.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Adhezion Biomedical. The logo consists of a stylized graphic to the left of the company name. The graphic is made up of several curved lines in shades of blue. The company name "Adhezion" is in a dark blue sans-serif font, with the word "BIOMEDICAL" in a smaller, lighter blue font underneath.
Based on bench performance testing, the specification of the device, SURGISEAL Twist has demonstrated to be substantially equivalent to its currently marketed device from a safety and performance perspective, and has demonstrated that a dimensional tip modification to the product maintains approximation of wound edges.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.