DISPOSABLE ADJUSTABLE PEEP VALVE MR CONDITIONAL USE
Applicant
Galemed Corp.
Product Code
BYE · Anesthesiology
Decision Date
Apr 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5965
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Single patient use adjustable positive end expiratory pressure (PEEP) valves for use in hospital, transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions. For use with neonates to adults. Environment of Use: Hospital, extended care facilities, emergency medical services, and patient transport. May be used in a Magnetic Resonance (MR) environment, not to exceed a 3.0 Tesla static magnetic field.
Device Story
Disposable adjustable PEEP valve; spring-actuated mechanism; elevates lung pressure above atmospheric at end of exhalation. Used in hospital, transport, emergency, and post-hospital care; operated by clinicians. Device connects to breathing circuits or manual resuscitators. Modified from predicate to include non-magnetic materials (push rod, springs) for MR conditional use up to 3.0 Tesla. Provides mechanical pressure regulation; no electronic or software components. Benefits patient by maintaining positive end-expiratory pressure during constant or intermittent gas flow.
Clinical Evidence
Bench testing only. Performance data included leakage testing, pressure/resistance accuracy, drop testing, environmental stress testing, and age testing. MR conditional status verified per ASTM F2052-02 in a 3 Tesla suite, confirming deflection < 45 degrees.
Technological Characteristics
Spring-actuated mechanical valve; 22/30 mm connections. Materials: non-magnetic components (rod and springs) for MR compatibility. MR conditional up to 3.0 Tesla static magnetic field. No software, electronics, or energy source. Disposable, single-patient use.
Indications for Use
Indicated for neonates to adults requiring elevated end expiratory pressure therapy in hospital, transport, emergency, or post-hospital settings. Contraindicated in individuals not requiring elevated end expiratory pressure therapy.
Regulatory Classification
Identification
A positive end expiratory pressure (PEEP) breathing attachment is a device attached to a ventilator that is used to elevate pressure in a patient's lungs above atmospheric pressure at the end of exhalation.
{0}------------------------------------------------
## K133957
# APR 1 4 2014
## 510(k) Summary Page 1 of 4 12-Feb-14
Galemed Corp. Amoy Export Processing Zone, 39 Section 3, Haijing East Road Xiamen, Fujian Province, China
| Official Contact: | Enhe Liu, QA Director |
|----------------------------|-------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | Disposable Adjustable PEEP Valve |
| Common/Usual Name: | PEEP Valve |
| Classification Name: | BYE - Attachment, breathing, positive end expiratory<br>pressure<br>21 CFR 868.5965 |
| Predicate Devices: | K082092 - Galemed PEEP valve<br>K081266 - Mercury Medical - PEEP Valve MR |
## Device Description
The proposed Galemed Disposable Adjustable PEEP valves are identical to their predicate PEEP Valves, K.082092, except we have replaced some components that were magnetic with nonmagnetic material in order to meet the requirements for MR conditional environments of use. Otherwise the PEEP valves are identical.
Galemed disposable adjustable PEEP valves elevate the pressure in a patient's lungs above atmospheric pressure at the end of exhalation.
The Galemed disposable adjustable PEEP valve for use in MR environments is identical to our predicate, K082092, disposable PEEP valve, except we have replaced the materials in 2 components, a push rod and the springs to materials which are less magnetic. Other than these 2 changes, the device is identical to the predicate. These changes / modifications have been evaluated and demonstrated that they do not alter nor change the performance or safety profile of the new device as compared to the predicates.
### Modifications to Proposed Device compared to Predicate
The proposed Galemed Disposable Adjustable PEEP valves are identical to their predicate PEEP Valves, K.082092, except we have replaced some components that were magnetic with nonmagnetic material in order to meet the requirements for MR conditional environment of use. Otherwise the PEEP valves are identical.
## Indications for Use
Single patient use adjustable positive end expiratory pressure (PEEP) valves for use in hospital, transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions. For use with neonates to adults.
{1}------------------------------------------------
510(k) Summary Page 2 of 4 12-Feb-14
## Environment of Use
Hospital, extended care facilities, emergency medical services, and patient transport
May be used in a Magnetic Resonance (MR) environment, not to exceed a 3.0 Tesla static magnetic field.
### Contraindications
Contraindicated in individuals not requiring elevated end expiratory pressure therapy
## Predicate Device Comparison:
The proposed disposable adjustable PEEP valves are identical to our predicate Galemed PEEP valves, K082092, except for a change in materials for the rod and the springs to less magnetic materials.
Comparative Table 1 discusses the major features of the proposed device and the legally marketed predicate device.
The Galemed Disposable Adjustable PEEP Valve is viewed as substantially equivalent to the predicate device based upon the following:
#### Indications -
- Indicated as a Single patient use adjustable positive end expiratory pressure (PEEP) . valve for use in hospital, transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions.
Discussion - The indications for use are identical for the proposed and predicate device, Galemed PEEP Valve, K082092.
- The addition of the indications for use in a Magnetic Resonance (MR) environment, not . to exceed a 3.0 Tesla static magnetic field is new but testing has demonstrated that the device meets the recognized standard as a MR conditional device.
Discussion - The indications for use as MR Conditional is identical for the proposed and predicate device, Mercury Medical, PEEP Valve, K081266.
#### Technology -
- The technology of the use as a, spring actuated valve, which when placed into a circuit or . manual resuscitator, provides positive end expiratory pressure for the patient is identical to the predicates.
Discussion - There are no differences in technology between the proposed device and the predicates - Galemed PEEP Valve, K082092 and Mercury Medical, PEEP Valve, K081266.
#### Materials -
- The materials are identical to predicate with the exception of a change in the spring . material.
{2}------------------------------------------------
## 510(k) Summary Page 3 of 4 12-Feb-14
- According to ISO 10993-1 the proposed device would be considered External . communicating, Tissue / mucosal, limited duration (< 24 hr)
- . Stainless steel is well understood and no addition biocompatibility testing is required for this specific application of the material.
Discussion: We are utilizing identical materials to our predicate - Galemed PEEP Valve, K082092. The differences in the spring material or the rod do not affect performance nor raise any new safety concerns.
## Environment of Use -
- Hospital, extended care facilities, emergency medical services, and patient transport and . MR suites to 3 Tesla.
Discussion - The environment of use are identical between the proposed device and the predicates - Galemed PEEP Valve, K082092 and Mercury Medical, PEEP Valve, K081266.
## Non-clinical Testing -
We have performed the following performance data and testing:
- MR conditional in accordance with ASTM F2052-02 in a 3 Tesla MR suite .
- According to ASTM F2052-02 the proposed device should not deflect more than o 45 degree, the proposed device did not deflect this far and thus met the requirements.
- . Leakage
- The device was pressurized and leakage measured before and after age testing. o It met the pass /fail criteria
- Pressure / Resistance Accuracy .
- The device was tested for repeatability and accuracy after being subjected to a o number of conditions; it met its performance specifications criteria.
- . Drop test
- The device was dropped and tested for leakage and pressure / resistance. It met o the pass /fail criteria.
- Environmental Conditions .
- The device was subjected to high and low temperature conditions and then O tested, it met its performance specifications criteria.
- . Transportation
- The device was subjected to transportation testing and then tested it met its O performance specifications criteria.
- . Age testing
- The device was age testing and then evaluated it met its performance o specifications criteria.
Discussion - The performance testing when compared to the predicates - Galemed PEEP Valve, K082092 and Mercury Medical, PEEP Valve, K081266 (for MR conditional environments) support the conclusion of substantial equivalence.
## Substantial Equivalence Conclusion -
The sponsor believes that the testing and comparative data presented in the 510(k) supports our claim of substantially equivalence to the identified predicate devices, listed above.
{3}------------------------------------------------
| 510(k) Summary |
|----------------|
| Page 4 of 4 |
| 12-Feb-14 |
Table 1 - Comparative Table
| Table 1 - Comparative Table<br>Features | Predicate | Proposed |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Single patient use adjustable positive end expiratory<br>pressure (PEEP) valves for use in hospital, transport,<br>emergency, and post hospital care to evaluate end lung<br>pressure above atmospheric at the end of exhalation in<br>constant and intermittent gas flow conditions. | Single patient use adjustable positive end expiratory pressure<br>(PEEP) valves for use in hospital, transport, emergency, and post<br>hospital care to evaluate end lung pressure above atmospheric at<br>the end of exhalation in constant and intermittent gas flow<br>conditions. For use with neonates to adults. |
| MR Conditional | Mercury Medical PEEP Valve, K081266<br>May be used in a Magnetic Resonance (MR)<br>environment, not to exceed a 1.5 Tesla or less.<br>Spatial gradient of 450 gauss/cm or less. | Environment of Use: Hospital, extended care facilities,<br>emergency medical services, and patient transport. May be used<br>in a Magnetic Resonance (MR) environment, not to exceed a 3.0<br>Tesla static magnetic field.<br><br>May be used in a Magnetic Resonance (MR) environment, not to<br>exceed a 3.0 Tesla.<br>Spatial gradient of 720 gauss/cm or less. |
| Patient Use / Duration if use | Single patient use, disposable, < 24 hours | Single patient use, disposable, < 24 hours |
| Environment of Use | Hospital, extended care facilities, emergency medical<br>services, and patient transport | Hospital, extended care facilities, emergency medical services,<br>and patient transport |
| Patient Population | Patient population from infant (neonate) to adults | Patient population from neonates to adults |
| Contraindications | Contraindicated in individuals not requiring elevated end<br>expiratory pressure therapy | Contraindicated in individuals not requiring elevated end<br>expiratory pressure therapy |
| Features, Specifications, and Performance | | |
| Standard 22 / 30 mm connections | Yes | Yes |
| Pressure Range (cmH2O) | Adjustable - 2.5 to 10, 0 to 1, 5 to 20, and 0 to 20 | Adjustable - 2.5 to 10, 0 to 1, 5 to 20, and 0 to 20 |
| Principle of Operation | Spring actuated valve | Spring actuated valve |
| Placement | Placed into a circuit or attached to a manual resuscitator | Placed into a circuit or attached to a manual resuscitator |
| Leakage | After setting the sample to 5 cm H2O and pressurizing,<br>the pressure drop shall not be in excess of 2 cmH2O<br>within 5 seconds after reaching 5 cmH2O | After setting the sample to 5 cm H₂O and pressurizing, the<br>pressure drop shall not be in excess of 2 cmH2O within 5 seconds<br>after reaching 5 cmH2O |
| Pressure / Accuracy | At flow rates of 2, 3, 5, 10, and 15 lpm the resistance<br>value should when set to 5 cmH2O be 5 +2.5/-2 and when<br>set to 20 cmH.0 he 20 +2 5/2 | At flow rates of 2, 3, 5, 10, and 15 lpm the resistance value<br>should when set to 5 cmH2O be 5 +2.5/-2 and when set to 20<br>cmH.O he 20+25/-2 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure with three curved lines representing wings or feathers.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Silver Spring. MD 20993-0002
## April 14, 2014
GaleMed Corporation C/O Mr. Paul Dryden Amoy Export Processing Zone, 39 Section 3. Haijing East Road XIAMEN. FUJIAN. CHINA 361026
Re: K133957
Trade/Device Name: Disposable Adjustable PEEP Valve Regulation Number: 21 CFR 868.5965 Regulation Name: Breathing Attachment Positive End Expiratory Pressure Regulatorv Class: II Product Code: BYE Dated: February 12, 2014 Received: February 14, 2014
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can.be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Digitally sloned by Richard Date: 2014 04.15 13:17:29
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known) K133957
Device Name
Disposable Adjustable PEEP Valve
#### Indications for Use (Describe)
Single patient use adjustable positive end expiratory pressure (PEEP) valves for use in hospital. transport, emergency, and post hospital care to evaluate end lung pressure at the end of exhalation in constant and internittent gas flow conditions.
For use with neonates to adults.
Environment of Use: Hospital, extended care facilities, emergency medical services, and patient transport
May be used in a Magnetic Resonance (MR) environment, not to exceed a 3.0 Tesla static magnetic field.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY 1 1 - 1 -. " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - " - ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ ាន់ជាជួយដែល .. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Anya C. Harry -S 2014.04.13 10:24:51 -04'00'
FORM FDA 3881 (9/13)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.