The Irix-A Lumbar Integrated Fusion System is a stand-alone intervertebral body fusion device intended for use in patients with degenerative disc disease (DDD) at one (1) or two (2) contiguous levels of the lumbosacral spine (L2-S1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. In addition, these patients may have had a previous non-fusion spinal surgery at the involved level(s) and may have had up to a Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Irix-A System is intended to be used with autogenous bone graft material, and is to be used with three titanium alloy bone screws included as part of the system.
Device Story
Irix-A Lumbar Integrated Fusion System is a stand-alone intervertebral body fusion device used in anterior discectomy procedures to restore biomechanical height and facilitate spinal fusion. The device is a box-shaped implant with superior/inferior teeth, manufactured from titanium alloy (Ti6Al4V) and/or PEEK-Optima LT1, with optional plasma-sprayed commercially pure titanium coating. It is secured via three titanium alloy bone screws. Used in clinical settings by surgeons; the device provides structural support for autogenous bone graft material. Mechanical performance is validated through static/dynamic compression, compression-shear, and expulsion testing. Benefits include stabilization of the lumbosacral spine in patients with degenerative disc disease.
Clinical Evidence
No clinical data. Evidence is based on bench testing, including static and dynamic compression, compression-shear testing per ASTM F2077, load-induced subsidence testing per ASTM F2267, and expulsion testing. Plasma coating validation included material performance characterization and wear debris particulate analysis.
Technological Characteristics
Materials: Ti6Al4V (ASTM F136), PEEK-Optima LT1 (ASTM F2026), and plasma-sprayed CP Ti (ASTM F1580). Design: Box-shaped intervertebral spacer with integrated screw fixation. Energy source: None (mechanical). Sterilization: Steam autoclave for non-sterile components; sterile packaging for coated components.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD defined as discogenic back pain with radiographic confirmation. Patients must have failed at least 6 months of non-operative treatment. May include patients with prior non-fusion spinal surgery or Grade I spondylolisthesis/retrolisthesis.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Surgicraft Limited STALIF-TT Intervertebral Body Fusion System (K073109)
Globus Medical, Inc. INDEPENDENCE Spacer (K082252)
Spinal Elements, Inc. LUCENT OR LUCENT MAGNUM (K110632)
Submission Summary (Full Text)
{0}------------------------------------------------
K133947 Page 1 of 2
#### 510(k) Summary
JUL 2 9 2014
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name: | X-spine Systems, Inc.<br>452 Alexandersville Rd.<br>Miamisburg, OH 45342 |
|-------------------------|--------------------------------------------------------------------------|
| | Telephone (937) 847-8400<br>FAX (937) 847-8410 |
| Official Contact: | David Kirschman, M.D.<br>Chief Medical Officer |
| Date Prepared: | July 29, 2014 |
| DEVICE NAME | |
| Trade/Proprietary Name: | Irix-A™ Lumbar Integrated Fusion System |
| Classification Name: | Intervertebral Fusion Device with Integrated Fixation, Lumbar |
| Regulation Description: | Intervertebral body fusion device |
| Device Class: | Class II |
| Product Code: | OVD |
| Regulation Number: | §888.3080 |
## ESTABLISHMENT REGISTRATION NUMBER
The X-spine Systems, Inc. establishment registration number is 3005031160. The owner/operator number for X-spine Systems, Inc. is 9063903.
## INDICATIONS FOR USE
The Irix-A Lumbar Integrated Fusion System is a stand-alone intervertebral body fusion device intended for use in patients with degenerative disc disease (DDD) at one (1) contiguous levels of the lumbosacral spine (L2-S1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have had a previous non-fusion spinal surgery at the involved level(s) and may have had up to a Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Irix-A System is intended to be used with autogenous bone graft material, and is to be used with three titanium alloy screws included as part of the system.
#### DEVICE DESCRIPTION
The Irix-A Lumbar Integrated Fusion System is a stand-alone intervertebral fusion device to restore biomechanical height and act as an aid in fusion of the lumbar spine in anterior discectomy procedures. The device is generally box-shaped with teeth on the superior and inferior faces of the drix-A implant is manufactured from both titanium alloy (Ti6Al4V) in accordance with ASTM F136 and Invibio PEEK-Optima LT1 in accordance with ASTM F2026, or from Ti6AMV titanium alloy alone. The device will be supplied with the option of having the superior and inferior surfaces of the device plasma coated with
{1}------------------------------------------------
medical-grade commercially pure titanium (CP Ti) per ASTM F1580. The device is secured in location through the use of bone screws, also manufactured from titanium alloy (Ti6Al4V) per ASTM F136. The devices are provided in various sizes and screws are offered in multiple lengths to adjust for variations in patient anatomy.
The CP-Ti coated devices will be sterile packed with the balance of system components being provided clean and non-sterile. The non-sterile components are to be sterilized by a healthcare professional using a steam autoclave in accordance with the instructions for use provided by X-spine Systems Inc., as well as the instructions provided by the manufacturer of the autoclave.
## PREDICATE DEVICES
- Surgicraft Limited STALIF-TT Intervertebral Body Fusion System (K073109) .
- Globus Medical, Inc. INDEPENDENCE Spacer (K082252) .
- Spinal Elements, Inc. LUCENT OR LUCENT MAGNUM (K110632) .
## EQUIVALENCE TO MARKETED PRODUCT
X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the Irix-A Lumbar Integrated Fusion System is substantially equivalent to predicate devices based on a comparison including the following characteristics:
- . FDA Product Code
- . Indications for Use
- Surgical Approach .
.
- Anatomical Region ●
- . Implant Materials
- Product Dimensions .
- Mechanical Performance .
#### PERFORMANCE DATA
The implant components were tested using the following standards:
ASTM F2077 – Test Methods for Intervertebral Body Fusion Devices
- Static and dynamic compression ●
- Static and dynamic compression-shear .
ASTM F2267 – Standard Test Method for Measuring Load Induced Subsidence of an Intervertebral Body Fusion Device under Static Axial Compression
In addition to the above standard testing, expulsion testing was performed as part of this submission. There is no cited standard for this test.
For validation of the plasma coating, a full material performance qualification and characterization was performed per published FDA guidance, as well as wear debris particulate analysis from dynamic compression-shear testing.
In summary, the mechanical testing results indicate that the Irix-A Lumbar Integrated Fusion System is substantially equivalent to predicate device performance and is capable of performing in accordance with its intended use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 29, 2014
X-spine Systems. Incorporated David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342
Re: K133947
Trade/Device Name: Irix-ATM Lumbar Integrated Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: July 3. 2014 Received: July 3. 2014
Dear Dr. Kirschman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - David Kirschman, M.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Ronald P. Jean -S for
- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K133947
## Device Name
Irix-ATM Lumbar Integrated Fusion System
## Indications for Use (Describe)
The Irix-A Lumbar Integrated Fusion System is a stand-alone intervertebral body fusion device intended for use in patients with degenerative disc disease (DDD) at one (1) or two (2) contiguous levels of the lumbosacral spine (L2-S1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. In addition, these patients may have had a previous non-fusion spinal surgery at the involved level(s) and may have had up to a Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Irix-A System is intended to be used with autogenous bone graft material, and is to be used with three titanium alloy bone screws included as part of the system.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Katherine D. Kavlock, PhD
## Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the I he beroen time for the concess, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.