← Product Code [MUD](/productcode/MUD) · K133923

# MULTI-PARAMETER MOBILE CAREGUIDE 3100 OXIMETER WITH TABLET (K133923)

_Reflectance Medical, Inc. · MUD · Apr 11, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133923

## Device Facts

- **Applicant:** Reflectance Medical, Inc.
- **Product Code:** [MUD](/productcode/MUD.md)
- **Decision Date:** Apr 11, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended to allow for display of SmO2 and pHm data on an Android Tablet display, which would interface with the Multi-Parameter Mobile CareGuide 3100 Oximeter via USB connection. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is intended for prescriptive use (Rx only) by a trained healthcare professional in a hospital. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet provides display of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet has not been demonstrated in disease states.

## Device Story

Device uses Near Infrared (NIR) spectroscopy to measure muscle oxygen saturation (SmO2) and muscle pH (pHm). System consists of a reusable sensor, disposable sleeve, and an Android tablet display. Sensor contains optical/electronic elements to collect spectra from skin, fat, and muscle; communicates via proprietary protocol over USB to tablet. Tablet runs locked-down software displaying real-time values and historical trends. Used by healthcare professionals in hospitals as an adjunct monitoring tool. Does not provide diagnostic output; intended to support clinical decision-making by providing physiological trends. Benefits include non-invasive, real-time monitoring of microvascular status.

## Clinical Evidence

Bench testing only. No clinical data provided. Bench and software testing demonstrated that the device meets performance requirements for an oximeter and is equivalent to predicates.

## Technological Characteristics

NIR spectroscopy (700-900 nm wavelength). Components: reusable sensor (CPU, battery, optical board), disposable sleeve, Android tablet. Connectivity: USB interface. Software: proprietary algorithm for SmO2/pHm calculation; locked-down Android OS. Sterilization: not specified.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Multi-Parameter Mobile CareGuide Oximeter ([K130079](/device/K130079.md))
- CareGuide Oximeter ([K113656](/device/K113656.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K133923

Reflectance Medical, Inc.

510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet

APR 1 2014

# SECTION 5

-

#### 510(k) SUMMARY

# SUMMARY OF SAFETY AND EFFECTIVENESS FOR Multi-Parameter Mobile CareGuide™ 3100 Oximeter, with Tablet

# Submitter Information

| Name:                | Reflectance Medical, Inc. (RMI)                                                                                       |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|
| Address:             | 116 Flanders Road, Suite 1000 Westborough, MA 01581 USA                                                               |
| Telephone Number:    | 508.366.4700                                                                                                          |
| Registration Number: | NA (RMI will apply for registration number following 510(k) clearance, prior to commencement of commercial shipment.) |
| Contact Person:      | Dr. Babs Soller                                                                                                       |
| Telephone Number:    | 508.366.4700, Ext 223                                                                                                 |
| Fax Number:          | 508.366.4770                                                                                                          |
| Email:               | Babs.Soller@reflectancemedical.com                                                                                    |
| Date Prepared:       | 12/19/2013                                                                                                            |

### Device Name

| Trade name of New Device: | Multi-Parameter CareGuide™ Oximeter with Tablet |
|---------------------------|-------------------------------------------------|
| Model Number:             | 3100                                            |
| 510(k) Holder/Submitter:  | Reflectance Medical, Inc. (RMI)                 |
| 510(k) Number:            | N/A                                             |
| Product codes:            | 21 CFR § 870.2700, 21 CFR 868.1170              |
| Classification Panel:     | Cardiovascular                                  |

# Predicate Devices

| Predicate Device #1:     | Mobile CareGuide                                                             |
|--------------------------|------------------------------------------------------------------------------|
| Trade Name of Device:    | Multi-Parameter Mobile CareGuide Oximeter                                    |
| Model #:                 | 3100 :                                                                       |
| 510(k) Holder/Submitter: | Reflectance Medical Inc.                                                     |
| 510(k) Numbers:          | K130079, CareGuide 3100                                                      |
| Product codes:           | MUD, 21 CFR 870.2700, Cardiovascular<br>CBZ, 21 CFR 868.1170, Anesthesiology |

{1}------------------------------------------------

510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet

| Predicate Device #2:     | CareGuide                            |
|--------------------------|--------------------------------------|
| Trade Name of Device:    | CareGuide Oximeter                   |
| Model #:                 | 1100                                 |
| 510(k) Holder/Submitter: | Reflectance Medical Inc.             |
| 510(k) Numbers:          | K113656, CareGuide 1100              |
| Product code:            | MUD, 21 CFR 870.2700, Cardiovascular |

#### Device Description

The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet uses Near Infrared (NIR) Spectroscopy to calculate muscle oxygen saturation (SmO2) and muscle pH (pHm) and displays those parameters as real-time values and historical trends on a tablet device.

| Characteristics                                                                |                                                                 |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Reflectance Medical Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet |                                                                 |
| Principle of Operation                                                         | NIR spectroscopy                                                |
| Components                                                                     | Monitor with reusable sensor, disposable pad and display tablet |
| Light Source                                                                   | LEDs                                                            |
| Parameters Measured                                                            | Tissue oxygen saturation (SmO2) and muscle pH (pHm)             |

The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is a multiple parameter oximeter. The sensor contains algorithms that calculate SmO2 and pHm from collected spectra and communicates the current SmO2 and pHm results to an Android tablet with display software through a proprietary protocol. The Android tablet (qualified models Acer A500 and Asus Google Nexus 7) contains 3d party software that locks down the tablet so that only the CareGuide software may run and no other application or operating software can be modified. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet reusable sensor contains the optical and electronic elements necessary to collect spectra from skin, fat and muscle. The sensor has a 3m long cord with a USB connection to the Android tablet. The sensor is identical to the predicate (K130079) Multi-parameter Mobile CareGuide 3100 Oximeter. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet uses the same disposable element as the Multi-Parameter Mobile CareGuide 3100 Oximeter, a disposable sleeve that isolates the sensor optical elements from the patient's skin.

{2}------------------------------------------------

510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet

### Indications for Use

The Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended to allow for display of SmO2 and pHm data on an Android Tablet display, which would interface with the Multi-Parameter Mobile CareGuide 3100 Oximeter via USB connection. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is intended for prescriptive use (Rx only) by a trained healthcare professional in a hospital. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet provides display of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet has not been demonstrated in disease states.

# Rationale for Substantial Equivalence

The CareGuide 3100 device's oximetry feature has been already cleared under classification regulation 21 C.F.R § 870.2700, Oximeter. The device's pHm feature has been already cleared under classification regulation 21 C.F.R § 868.1170, Anesthesiology. This 510(k) is seeking clearance for use of the CareGuide 3100 with a dedicated Android Display device.

There is no change in how the User would use the information generated by the Multi-Parameter Care Guide 3100 with Tablet relative to the predicate devices. They are all intended for monitoring of respective parameters. Neither the Multi-Parameter CareGuide 3100 with Tablet nor any of the predicate devices identified by RMI provides diagnostic output.

The Multi-Parameter CareGuide 3100 with Tablet has the same technological characteristics as the previously cleared RMI devices, the CareGuide 1100 (K113656) and the Multi-Parameter Mobile CareGuide 2100 (K130079).

- The principle of operation of the Multi-Parameter Mobile CareGuide 3100 Oximeter with . Tablet is identical to that of the predicate CareGuide devices. They use the exact same NIR spectroscopic platform to measure tissue oxygen saturation and muscle. The same software quantitative algorithm for SmO2 and pHm is used in both devices.
- The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is equivalent to the ● Multi-Parameter Mobile CareGuide 3100 Oximeter predicate in reusable components. Both devices use the exact same sensor hardware: main sensor CPU board, battery, optical board (light sources, spectrometer and microprocessor), USB interfaces, plastic housing and cables.
- . The Multi-Parameter Mobile CareGuide 3100 Oximeter with Table is equivalent to the Multi-Parameter Mobile CareGuide 3100 Oximeter predicate in disposable components. Both devices use the exact same disposable sheath ("Ray") and disposable sensor check device ("Cradle").

{3}------------------------------------------------

510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet

- . The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet has the identical underlying LED light source as the CareGuide predicates, with the exact same range of wavelength (700-900 nm).
- . The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet and the predicated CareGuide 1100 Oximeter are equivalent. The CareGuide 1100 System includes a Monitor display while the CareGuide 3100 {subject of this 510(k)} includes a dedicated Android display. Both displays are functionally equivalent. Both CareGuide 1100 and 3110 displays are tools that interface with the CareGuide oximeter and display real-time parameters and historical trends per their cleared indications for use.
- . The predicate Multi-Parameter Mobile CareGuide 3100 Oximeter is compatible with any USB-connected display device that supports the specified communications protocol. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet {subject of this 510(k)} now includes such a USB-connected display device (i.e. dedicated Android tablet), supporting that specified communications protocol.

# Summary of Safety and Effectiveness Data

Bench testing demonstrates that the Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is safe and effective, meeting all relevant consensuses and FDA recognized standards. The bench and software test results in this submission demonstrate that the Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet meets the expected performance requirements for an Oximeter. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is therefore equivalent to the predicates by indications for use and device features and functionality.

#### Conclusion

The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is equivalent to predicate device in terms of technology (NIR Spectroscopy) and intended use. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet does not raise new questions of safety or effectiveness, as compared to the predicate. Therefore, the Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is substantially equivalent to the predicate device.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2014

Reflectance Medical, Inc. Babs Soller Chief Executive Officer 116 Flanders Road, Suite 1000 Westborough, MA 01581 US

Re: K133923

> Trade/Device Name: Multi-parameter Mobile CareGuide 3100 Oximeter with Tablet Regulation Number: 21 CFR 870.8700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: March 7, 2014 Received: March 13, 2014

Dear Dr. Soller,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be actised may be notal your device complies with other requirements of the Act 

{5}------------------------------------------------

Page 2 - Dr. Babs Soller

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/8 description: The image shows a logo with the letters FDA. The letters are stylized and have a bold, geometric design. There are lines drawn through the logo, which may indicate that it is not valid or has been rejected. The logo is black and white and appears to be a scanned image.

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Premarket Notification Submission: Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet

# Indications for Use Form

Indications for Use

510(k) Number (if known): K133923_

Device Name: Multi-Parameter Mobile CareGuide™ 3100 Oximeter with Tablet

Indications for Use:

The Multi-Parameter Mobile CarcGuide™ 3100 Oximeter with Tablet is intended for use as an adjunct, non-invasive monitor of the hemoglobin oxygen saturation and pH of microvascular blood in a region of skeletal muscle tissue beneath the sensor may be positioned on pigmented skin. The Multi-Parameter Mobile CareGuide 3100 Oximeter is intended to allow for display of SmO2 and pHm data on an Android Tablet display, which would interface with the Multi-Parameter Mobile CareGuide 3100 Oximeter via USB connection. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet is intended for prescriptive use (Rx only) by a trained healthcare professional in a hospital. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet provides display of the most recent values of SmO2 and pHm in a trend, as well as operational device information. The Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet should not be used as the sole basis for diagnosis or therapy. Note: The prospective clinical value of data from the Multi-Parameter Mobile CareGuide 3100 Oximeter with Tablet has not been demonstrated in disease states.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

Page of of

Image /page/6/Picture/14 description: The image shows a logo with the letters "PDA" in a stylized font. Next to the logo, there is a date and time stamp that reads "Date: 2014.04.11 07:57:56 -04'00". There is also a signature over the logo.

Page 1 of 1

Page 21 of 361

---

**Source:** [https://fda-staging.innolitics.com/device/K133923](https://fda-staging.innolitics.com/device/K133923)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
