GENTLECARE AR APPLICATOR

K133922 · Gc Medtech, LLC · JAQ · Apr 18, 2014 · Radiology

Device Facts

Record IDK133922
Device NameGENTLECARE AR APPLICATOR
ApplicantGc Medtech, LLC
Product CodeJAQ · Radiology
Decision DateApr 18, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GentleCare AR Applicator is Intended for use during brachytherapy procedures. The nultiple lumens of the GC MedTech device are intended to provide pathways from which a prescribed radiation dose is delivered to the treatment area.

Device Story

The GentleCare AR Applicator is a sterile, single-use, disposable brachytherapy applicator inserted into the vagina or rectum. It features multiple lumens that serve as pathways for radiation sources delivered by compatible remote afterloading equipment. The device includes two expandable baffles (inner and outer) to position the treatment lumens in a circumferential array within the target anatomy. The physician controls inflation of the baffles to match patient anatomy and determines the treatment plan based on the device's geometry. Visualization is achieved via X-ray. The device facilitates precise radiation delivery to carcinoma sites, benefiting patients by enabling physician-controlled, localized brachytherapy.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included dimensional measurements, visual inspection, balloon inflation/deflation/burst, bond strength, and mechanical/performance verification. Biocompatibility testing (cytotoxicity, sensitization, irritation) was performed per ISO 10993.

Technological Characteristics

Sterile, single-use, disposable brachytherapy applicator. Materials are biocompatible (ISO 10993). Features 8 treatment lumens in a concentric ring and 2 expandable baffles (air-inflated). Dimensions: 20cm length, 6cm inflated diameter, 2cm insertion diameter. Compatible with standard remote afterloaders (VariSource, GammaMedplus, microSelectron) via 6 Fr catheter interface. Sterilized via Ethylene Oxide.

Indications for Use

Indicated for use during brachytherapy procedures for patients requiring radiation delivery to a target site in the treatment of carcinoma. Used with remote afterloading equipment.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. General Information | APPLICANT: | GC MedTech, LLC | |------------------------|-------------------------------------------------------------------| | TRADE NAME: | GentleCare AR Applicator | | COMMON NAME: | Brachytherapy Applicator | | CLASSIFICATION NAME: | Remote Controlled Radionuclide Applicator System, 21 CFR 892.5700 | | DEVICE CLASSIFICATION: | Class II | | PRODUCT CODE: | JAQ | | PREDICATE DEVICES: | Capri Applicator (K092822) | | CONTACT | Colin Carpenter<br>President | | DATE: | December 16, 2013 | #### Substantially Equivalent to: The GentleCare AR Applicator is equivalent in intended use, principal of operation and technological characteristics to the CAPRI Applicator (K092822). # Description of the device subject to premarket notification The GentleCare AR Applicator is a specialized applicator that is temporarily inserted into the vagina or rectum to facilitate the application of radiation to the target site in the treatment of carcinoma. The GentleCare AR Applicator is provided sterile for single use and is disposable. ## Indications for Use The Gentle Care AR Applicator is intended for use with commercially available remote afterloading equipment used during brachytherapy procedures. The multiple lumens of the GentleCare AR Applicator are intended to provide pathways from which a prescribed radiation dose is delivered to the treatment area. {1}------------------------------------------------ #### Materials All materials used in the manufacture of the GentleCare AR Applicator are suitable for this use and have been used in numerous previously cleared products. The GentleCare AR Applicator was tested per ISO10993 and found to be biocompatible. Testing included the following: - Cytotoxicity . - Sensitization . - . Irritation ## Non-Clinical Testing Product testing was completed and met all of the acceptance criteria. Testing included dimensional measurements, visual inspection, balloon inflation, balloon deflation, balloon burst, bond strength, mechanical and performance. #### Performance Data: All necessary verification and validation testing has been performed for the GentleCare AR Applicator to assure substantial equivalence to the predicate devices. ## Basis for Determination of Substantial Equivalence: Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the GentleCare AR Applicator is determined by GC MedTech, to be substantially equivalent to existing legally marketed devices. | Trade name | GC MedTech System | CAPRI Applicator | SE Discussion | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product code | JAQ | JAQ | Same product code - JAQ | | 510k number | K133922 | K092822 | | | Device Classification | II | II | Same - Class II | | Device description | The GentleCare AR Applicator is<br>a specialized applicator that is<br>temporarily inserted into the<br>vagina or rectum to facilitate the<br>application of radiation to the<br>target site in the treatment of<br>carcinoma. The GentleCare AR<br>Applicator is provided sterile for<br>single use and is disposable. | The CAPRI Applicator is a<br>specialized applicator that is<br>temporarily inserted into the<br>vagina or rectum to facilitate<br>the application of radiation to<br>the target site in the treatment<br>of carcinoma. The CAPRI<br>Applicator is provided sterile<br>for single use and is disposable | The device description is the same. | | Intended Use | Intended for use during<br>brachytherapy procedures. The<br>multiple lumens of the | Intended for use during<br>brachytherapy procedures. The<br>multiple lumens of the CAPRI | Same Indications for Use | | | | | | | | GentleCare AR Applicator are<br>intended to provide pathways<br>from which a prescribed radiation<br>dose is delivered to the treatment<br>area. | applicator are intended to<br>provide pathways from which a<br>prescribed radiation dose is<br>delivered to the treatment area. | | | Length | Balloon: 10cm; Device: 20cm | Balloon: 8.5cm; Device: 15cm | The small differences in the balloon<br>length does not affect the safety or<br>efficacy of the device's ability to<br>interface with the Afterloader or the<br>delivery of physician controlled<br>therapy | | Diameter | 6cm | Balloon: 4cm when inflated | The diameter difference does not<br>affect the safety of the device as the<br>inflation of both devices is controlled<br>by the physician with the diameter<br>based on the individual patient's<br>anatomy | | Deflated insertion<br>diameter | 2cm | 3.3cm | The diameters are equivalent for the<br>insertion of the device to the target<br>area | | Number of treatment<br>lumens | 8: 8 evenly spaced in a single<br>concentric ring | 13: 12 evenly spaced in two<br>concentric rings (with six<br>lumens each) surrounding 1<br>central lumen. | During the treatment planning, the<br>number of lumens are considered in<br>the determination of the treatment<br>plan. The main issue is the<br>concentricity of the treatment lumens<br>which allow the physician to develop<br>an appropriate treatment plan. | | Number of<br>expandable baffles | 2 | 2 | Same | | Location of lumens | Inner baffle | Inner baffle | Same | | Inner expandable<br>baffle | Air | Foam | Both inner baffles serve the same<br>purpose - to space the treatment<br>lumens in a circumferential array | | Outer expandable<br>baffle | Air | Air | Same | | Method of<br>visualization | X-ray | X-ray | Same | | Conforming<br>treatment length | 10cm | 8.5cm | The treatment length is slightly longer<br>for the subject device. As stated<br>previously, the treatment plan is pre-<br>determined by the physician based on<br>the device itself. Therefore, the safety<br>profile of both devices is equivalent. | {2}------------------------------------------------ · י {3}------------------------------------------------ | Compatible<br>Afterloaders | VariSource 200- AL13301000<br>(channels 1-10)<br>VariSource 200-AL13301001<br>(channels 11-20)<br>VariSource 200-AL07356001<br>(channel X)<br>Varian GammaMedplus<br>Nucletron microSelectron<br>Digital microSelectron-HDR | VariSource 200- AL13301000<br>(channels 1-10)<br>VariSource 200-AL13301001<br>(channels 11-20)<br>VariSource 200-AL07356001<br>(channel X)<br>Varian GammaMedplus<br>Nucletron microSelectron<br>Digital microSelectron-HDR | All Afterloaders connect to 6 Fr<br>catheters. The connection interface is<br>also the same. | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Method of<br>sterilization | Ethylene Oxide | E-beam | Both devices are supplied sterile.<br>Both methods for sterilization are well<br>known and acceptable for medical<br>devices | | Single use | Y | Y | Same | | Shelf life | 180 days after production | 360 days after production | This will be on the product label | | Packaging | Paperboard insert in sterile pouch;<br>IFU on 8"x11" paper; both placed<br>inside outer box | Paperboard insert in sterile<br>pouch; IFU on 8"x11" paper;<br>both placed inside outer box | Same | | Materials | Biocompatible | Biocompatible | The GC MedTech device was tested to<br>the ISO10993 standard and found to<br>be biocompatible. As the CAPRI<br>device was reviewed and cleared by<br>FDA, we assume the materials of<br>construction for the CAPRI are<br>biocompatible as well. | # Conclusion The substantial equivalence was determined based on the similarity of materials, design, indications for use, packaging and clinical application. {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 18, 2014 GC Medtech, LLC. % Gregory Mathison President Regulatory Strategies, INC. 3924 Cascade Beach Road LUTSEN MN 25612 Re: K133922 Trade/Device Name: Gentlecare AR Applicator Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAO Dated: March 17, 2014 Received: March 25, 2014 Dear Mr. Mathison: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or my 1 vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Crice-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2-Mr. Mathison If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Michael D. O'Hara for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): K133922 Device Name: GentleCare AR Applicator Indications For Use: > The GentleCare AR Applicator is Intended for use during brachytherapy procedures. The nultiple lumens of the GC MedTech device are intended to provide pathways from which a prescribed radiation dose is delivered to the treatment area. Prescription Use _ S (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) Mecha D'Hara Page 1 of _ 1 __
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...