K133896 · Trophy Skin, Inc. · OHS · Nov 13, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133896
Device Name
REJUVALITE MD
Applicant
Trophy Skin, Inc.
Product Code
OHS · General, Plastic Surgery
Decision Date
Nov 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K133896 · Nov 13, 2014
REJUVALITE MD
Trophy Skin, Inc.
MAUDE (Manufacturer and User Facility Device Experience) database
The sponsor performed a search of the MAUDE database to assess the postmarket safety profile of the predicate device (K120775) as part of the substantial equivalence argument.
Postmarket surveillance; MAUDE; Safety history
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
MAUDE Database Search; Postmarket surveillance review; Follow-up/Duration: Since 2005; Study Period: 2005-2014
Users of the predicate device (LightStim for Wrinkles); Number of Sites: Not applicable
Not applicable for this study
Adverse events
Indications for Use
The Rejuvalite MD is an Over the Counter device that is intended for the treatment of full face wrinkles.
Device Story
Tabletop LED light therapy device; 120 LEDs emitting four wavelengths (600, 622, 660, 860nm); total power density 62 mW/cm2 at 4-inch distance. User-operated via simple On/Off switch; includes timer and distance-limiting arm to ensure consistent 4-inch treatment distance. Used in home setting (OTC); intended to reduce facial wrinkles through light exposure. Benefits patient by providing non-invasive wrinkle treatment; healthcare provider role limited to labeling/instructions for layman use. No software or complex processing involved.
Clinical Evidence
Bench testing only. Performance testing confirmed compliance with IEC 60601-1 and IEC 60601-1-2 standards. A usability/label comprehension study with 36 participants showed 97% comprehension of labeling and 100% successful device operation.
Technological Characteristics
Tabletop LED device; 120 LEDs; 4 wavelengths (600, 622, 660, 860nm); 62 mW/cm2 power density at 4 inches. Materials: ABS plastic. Power: AC to DC internal adapter. Connectivity: None. Software: None (On/Off switch only). Standards: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for the treatment of full face wrinkles in individuals.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be made of flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2014
Trophy Skin Incorporated % Ms. Susan R. Anthoney-DeWet AEGIS Regulatory Incorporated 2424 Dempster Drive Coralville, Iowa 52241
Re: K133896 Trade/Device Name: Rejuvalite MD Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: October 6, 2014 Received: October 7, 2014
Dear Ms. Anthoney-DeWet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133896
Device Name Rejuvalite MD
Indications for Use (Describe)
The Rejuvalite MD is an Over the Counter device that is intended for the treatment of full face wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K133896
This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 878.4810.
Submission Date: December 11, 2013
| 1. Submitter Information: | AEGIS Regulatory, Inc. - Robert T. Wagner<br>2424 Dempster Drive<br>Coralville, IA 52241<br>Tel.: 865-982-5552<br>Email: bob@fdalistingconsultants.com |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| For Sponsor: | Trophy Skin, Inc.<br>Attn: Imran Karim<br>4372 Kenmare Trail<br>Frisco, TX 75034<br>Tel.: 469-698-6270<br>Fax: 972-692-7194<br>Email: trophyskin@gmail.com |
#### 2. General Information
- 2.1 Classification Name: Light Based Over-The-Counter Wrinkle Reduction Device
- 2.2 Common/Usual Name: Rejuvalite MD
- 2.3 Proprietary Names: Rejuvalite MD
- 2.4 Classification: Class II
- 2.5 Classification Number: 878.4810
- 2.6 Product Code: OHS
- 2.7 Review Panel: General & Plastic Surgery
{4}------------------------------------------------
#### 3. Device Description:
The Rejuvalite MD, is a tabletop device that contains 120 LEDs split evenly into 4 different wavelengths. The total power density for combined wavelengths is approximately 62 mW/cm2 at a distance of 4" from the illuminating surface of the device.
The labeling requires Users to maintain the distance between the illuminating surface and the treatment site at 4"during the treatment. There is a 4" distance limiting arm attached to the device to help users on determining and maintaining the 4" distance between the device's illuminating surface and the treatment site.
The Rejuvalite MD system components include the tabletop unit consisting of the base, arm, and swivel head containing the LED module , 4" distance limiting arm, timer, internal UL approved power adapter, and goggles.
The device incorporates a power adapter mounted inside of the lamp base that converts the AC to DC voltage. The power switch is operated to turn the device on or off.
The unit is designed to be placed on a table for easy exposure to the face to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components. The device is sold as Over the Counter (OTC).
### 4. Indications / Intended Use:
The RejuvaliteMD is an Over -the- Counter device that is intended for the use in the
treatment of full face wrinkles.
Rx or OTC:
The Rejuvalite MD device is an Over the Counter (OTC) device. The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate device is OTC.
### 5. Predicate Device (s):
This device is substantially equivalent to the following predicates, which are currently in safe
and effective commerce:
### 1. K120775 - LightStim for Wrinkles (LED Intellectual)
{5}------------------------------------------------
### Predicate Comparison Chart
| Device | Rejuvalite MD | LightStim for<br>Wrinkles |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| | Trophy Skin | LED Intellectual |
| | K133896 | K120775 |
| | This Submission | A Predicate Device |
| Indications | The RejuvaliteMD<br>is an Over the<br>Counter device that<br>is intended for the<br>use in the treatment<br>of full face wrinkles. | The Lightstim for<br>Wrinkles is an OTC<br>hand-held device<br>intended for the use<br>in the treatment of<br>full-face wrinkles |
| Handheld | Table top | Yes |
| Wavelengths | 600,622,660,860nm | 605,630,660,855nm |
| Modes | On/Off | On/Off |
| Energy Source | 120 LEDs. Over<br>90cm2 | 72 LEDs. Over<br>40cm2 |
| Total Power<br>Density | 62 mW/cm2 at a<br>distance of 4" from<br>the LED head | 62 mW/cm2 in<br>contact with LED<br>head |
| Power Supply | AC to DC | 9-volt DC power<br>transformer |
| Treatment<br>Regime | 3 minutes daily, 5<br>days per week for 8<br>weeks | 3 minutes daily, 5<br>days per week for 8<br>weeks |
| Target<br>Population | Individuals with<br>wrinkles on the face | Individuals with<br>wrinkles on the face |
| Location for<br>Use | OTC | OTC |
### Summary of the technological characteristics of the device compared to predicate device:
The predicate device K120775 (LightStim for Wrinkles) has a smaller surface area with fewer LEDs than the proposed Rejuvalite MD device, but number of wavelengths, treatment duration, treatment regime and amount of energy (irradiance per area) delivered to the skin are the same.
{6}------------------------------------------------
This device also differs from the predicate device K120775 (LightStim for Wrinkles) in that it is not a handheld device in contact with the skin. The Rejuvalite MD is used at a 4" distance from the illuminated surface, but the power density per wavelength (mW/cm2), total power density of combined wavelengths (62 mW/cm2), amount of energy (irradiance per area) delivered to the skin measured at 4" are identical to the predicate device . The method of operation of the Rejuvalite MD device is similar to the predicate device K120775 (LightStim for Wrinkles) in that both devices have no software and use an On/Off switch.
No adverse effects on any of the predicate devices were found in MAUDE since clearance in 2005.
After analysis of safety, indications, intended uses, power density per wavelength (mW/cm2) , total power density of combined wavelengths (62 mW/cm2), performance, features, design materials, , technological properties, treatment areas, treatment reqimes and methods of operation, the manufacturer asserts that no significant differences exist between the applicant device and predicates listed in the predicate chart, and no new issues arise for safety and effectiveness. Therefore, substantial equivalency is hereby requested.
# 6. Biocompatibility:
The only patient contact material in the Reiuvalite MD is the body of the device.
The body of the device is constructed of ABS plastic, the same material used in the predicate device. The biocompatibility of this material is well known and considered safe when in contact with healthy skin.
The conclusion is that the Rejuvalite MD does not raise any new safety issues.
# 7. Performance Testing and Standards:
This device has been tested under and are in compliance with performance standards that have
been established for such devices under Section 878 of the Federal Food, Drug, and Cosmetics Act.
Safety and functionality testing demonstrates that the Rejuvalite MD conforms to various international consensus standards:
IEC 60601-1: (2005) +Corr. 1 (2006) + Corr. + (2007): Medical Electrical Equipment Part 1:
General Requirements for Basic Safety and Essential Performance.
IEC 60601-1-2: (2007): Medical Electrical Equipment Part 1: General Requirements for Basic
Safety and Essential Performance: Collateral Standard: Electromagnetic Compatibility
{7}------------------------------------------------
Performance Testing on optical and electrical characteristics demonstrated that the Rejuvalite MD device is almost identical to that of the predicate devices.
A Usability/Label Comprehension Study was conducted with 36 participants.
The results of the study found that: 97% of the participants were able to comprehend the labeling. 100% of the participants were able to use the device successfully.
The conclusions drawn from nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices.
# 8. Statement of Safety and Effectiveness:
The information in this 510(k) submission was used to support the safety and effectiveness of this device with respect to its cited predicates.
# 9. Substantial Equivalence Discussion
After analysis of safety, indications, intended uses, power density (mW/cm2) per wavelength, total power density of combined wavelengths (62 mW/cm2), performance, features, design materials, technological properties, treatment areas, treatment regimes and methods of operation, the manufacturer asserts that no significant differences exist between the applicant device and predicates listed in the predicate chart, and no new issues arise for safety and effectiveness. Therefore, substantial equivalency is hereby requested.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.