THE KINEGRAPH VMA (VERTEBRAL MOTION ANALYZER) SOFTWARE VERSION 2.2, THE MOTION NORMALIZER PATIENT HANDLING AND DATA
K133875 · Ortho Kinematics, Inc. · LLZ · Jan 17, 2014 · Radiology
Device Facts
Record ID
K133875
Device Name
THE KINEGRAPH VMA (VERTEBRAL MOTION ANALYZER) SOFTWARE VERSION 2.2, THE MOTION NORMALIZER PATIENT HANDLING AND DATA
Applicant
Ortho Kinematics, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 17, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The KINEGRAPH VMA™ software is a quantitative imaging software application intended to be used to process digital image files. It is designed for physicians and clinical professionals who are interested in the analysis of motion in medical images, particularly in musculoskeletal images of the spine. KINEGRAPH VMA™ software permits users to review static and dynamic digital lumbar and cervical spine images acquired with the assistance of the MOTION NORMALIZER™ patient handling and data collection device, which is designed for use by imaging technicians and intended to assist with patient lumbar and cervical bending and data collection during imaging. KINEGRAPH VMA™ software also facilitates quantitative assessment of vertebral motion in digital medical images. Information about the motion of selected objects, such as bone structures, can be generated and presented in the form of a 'motion analysis' report containing graphics, charts, and text.
Device Story
System comprises KineGraph VMA software and Motion Normalizer hardware; used for quantitative analysis of spinal motion. Motion Normalizer (electromechanical device) assists patient lumbar/cervical bending during standard fluoroscopic imaging; captures radiographic and motion data. Data uploaded to online servers via internet-connected computer. Ortho Kinematics employees operate software to process images and facilitate template placement. Prescribers access results via online account; can configure alerts for user-defined measurement thresholds. Output includes motion analysis reports with graphics, charts, and text. Benefits include objective assessment of vertebral motion to aid clinical decision-making.
Clinical Evidence
Bench testing only. Repeatability and accuracy testing performed for the integrated system. Software validation and verification conducted per FDA recognized standards. No clinical data presented.
Technological Characteristics
System includes KineGraph VMA software and Motion Normalizer electromechanical hardware. Operates with standard fluoroscope and procedure table. Data output as DICOM files. Connectivity via internet-connected computer to online servers. Software runs on OTS workstations; utilizes custom processing algorithms.
Indications for Use
Indicated for physicians and clinical professionals requiring quantitative analysis of vertebral motion in digital lumbar and cervical spine images. Used in conjunction with the Motion Normalizer device for patient bending and data collection during fluoroscopic imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K133875
Page 1 of 5
JAN 1 7 2014
#### 510(k) SUMMARY
#### ORTHO KINEMATICS, INC.
# INTEGRATED SYSTEM COMPRISED OF THE KINEGRAPH VMA™ (VERTEBRAL MOTION ANALYZER) SOFTWARE AND ITS ACCESSORY DEVICE, THE MOTION NORMALIZER IMAGE™ PATIENT HANDLING AND DATA COLLECTION DEVICE (VERSION 2.2)
| SUBMITTED BY | John Smith, M.D, J.D.<br>Hogan Lovells US LLP<br>555 13th Street, NW<br>Washington, DC 20004 | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON | Primary:<br>John Smith, M.D, J.D.<br>Partner<br>Hogan Lovells US LLP<br>Phone: (202) 637-5600<br>Fax: (202) 637-5910 | Alternate:<br>Adam Deitz<br>Chief Technology Officer<br>Ortho Kinematics<br>Phone: (415) 699-1736<br>Fax: (512) 334-5500 |
| DATE PREPARED | December 20, 2013 | |
| DEVICE PROPRIETARY NAME | The integral system comprised of:<br>- the KINEGRAPH VMA™ (VERTEBRAL MOTION<br>ANALYZER) software (version 2.0), and its accessory<br>device,<br>- the MOTION NORMALIZER™ patient handling and data<br>collection device (version 2.0) | |
| DEVICE COMMON NAME | The VMA system; the MOTION NORMALIZER device | |
| CLASSIFICATION NAME /<br>PRODUCT CODE | System, Image Processing, Radiological / LLZ | |
| DEVICE CLASS | Class II | |
| REGULATION NUMBER | 21 C.F.R. 892.2050 | |
| PREDICATE DEVICE | Version 2.0 of the VMA system (K130743) | |
| | | |
# PURPOSE OF THE SPECIAL 510K NOTICE
Version 2.2 of the VMA system is a modification to version 2.0 of the system, which was the originally cleared version.
7004 Bee Caves Road Building Three, Suite 315 Austin, Texas 78746 Fax: (512) 334-5500 Ortho Kinematics
Page 110 of 120 Phone: (512) 334-5490
{1}------------------------------------------------
#### INTENDED USE
The KINEGRAPH VMA™ software is a quantitative imaging software application intended to be used to process digital image files. It is designed for physicians and clinical professionals who are interested in the analysis of motion in medical images, particularly in musculoskeletal images of the spine. KINEGRAPH VMA™ software permits users to review static and dynamic digital lumbar and cervical spine images acquired with the assistance of the MOTION NORMALIZER™ patient handling and data collection device, which is designed for use by imaging technicians and intended to assist with patient lumbar and cervical bending and data collection during imaging. KINEGRAPH VMA™ software also facilitates quantitative assessment of vertebral motion in digital medical images. Information about the motion of selected objects, such as bone structures, can be generated and presented in the form of a 'motion analysis' report containing graphics, charts, and text.
## TECHNOLOGICAL CHARACTERISTICS
The subject device consists of the KINEGRAPH VMA™ software that analyzes images from · the MOTION NORMALIZER™ on an Off-the-Shelf ("OTS") imaging workstation. The MOTION NORMALIZER patient handling and data collection device is an accessory device to the KINEGRAPH VMA™ software that is used to assist with subject lumbar bending and data collection while images are captured with a standard fluoroscope. The MOTION NORMALIZER is comprised of two powered, electromechanical patient handling devices connected to and controlled by a console-mounted OTS computer running custom software connected to various OTS hardware accessories. The subject system is able to capture and record radiographic image data, as well as data from the patient handling devices, and to output this data into DICOM compatible digital image files for analysis using the KINEGRAPH VMA™ software. Once images have been created, they can be uploaded to online servers using running KINEGRAPH VMA™ software using an internet connected computer. All software operation and data storage that occurs subsequent to uploading occurs via online servers running KINEGARPH VMA software. Image processing operators, who are Ortho Kinematics, Inc. employees, operate the KINEGRAPH VMA™ software to process images and facilitate template placement for the prescriber. Prescriber users are able to access processed image and measurement results via an online account administered by Ortho Kinematics, Inc. Prescriber users are able to configure account settings that enable the prescriber to be alerted any time specific user-defined measurement thresholds are exceeded.
### PERFORMANCE DATA
The KINEGRAPH VMA™ software and its accessory device, the MOTION NORMALIZER™ , have been designed and developed in accordance with FDA regulations, including validation and verification testing per FDA recognized standards. In addition, repeatability and accuracy testing was performed for the integrated system. In all instances, the KINEGRAPH VMA ™ software and its accessory device, the MOTION NORMALIZER ™, functioned as intended.
7004 Bee Caves Road Building Three, Suite 315 Austin, Texas 78746 Fax: (512) 334-5500 Ortho Kinematics
Page 111 of 120 Phone: (512) 334-5490
{2}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE
The KINEGRAPH VMA™ software and its accessory device, the MOTION NORMALIZER™, has the same intended use and indications for use, technological characteristics, and principles of operation as the identified predicate device. The minor technological differences between the KINEGRAPH VMA™ software and its accessory device, the MOTION NORMALIZER™ , and the predicate device raise no new issues of safety or effectiveness. Validation and verification data (including software validation) demonstrate that the subject device functions as intended, and performs functions substantially equivalent to the predicate device.
| | KineGraph VMA™ System v2.2<br>(subject device) | KineGraph VMA™ System v2.0<br>(K130743, the cleared device) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | System, Image Processing, Radiological | System, Image Processing, Radiological |
| Product Code | LLZ | LLZ |
| | KineGraph VMA™ System v2.2<br>(subject device) | KineGraph VMA™ System v2.0<br>(K130743, the cleared device) |
| Indications for<br>Use | The KINEGRAPH VMA™ software is a<br>quantitative imaging software<br>application intended to be used to<br>process digital image files. It is<br>designed for physicians and clinical<br>professionals who are interested in the<br>analysis of motion in medical images,<br>particularly in musculoskeletal images<br>of the spine. KINEGRAPH VMA™<br>software permits users to review static<br>and dynamic digital lumbar and cervical<br>spine images acquired with the<br>assistance of the MOTION<br>NORMALIZER™ patient handling and<br>data collection device, which is<br>designed for use by imaging<br>technicians and intended to assist with<br>patient lumbar and cervical bending<br>and data collection during imaging.<br>KINEGRAPH VMA™ software also<br>facilitates quantitative assessment of<br>vertebral motion in digital medical<br>images. Information about the motion<br>of selected objects, such as bone<br>structures, can be generated and<br>presented in the form of a 'motion<br>analysis' report containing graphics,<br>charts, and text. | The KINEGRAPH VMA™ software is a<br>quantitative imaging software<br>application intended to be used to<br>process digital image files. It is<br>designed for physicians and clinical<br>professionals who are interested in the<br>analysis of motion in medical images,<br>particularly in musculoskeletal images<br>of the spine. KINEGRAPH VMA ™<br>software permits users to review static<br>and dynamic digital lumbar and cervical<br>spine images acquired with the<br>assistance of the MOTION<br>NORMALIZER™ patient handling and<br>data collection device, which is<br>designed for use by imaging technicians<br>and intended to assist with patient<br>lumbar and cervical bending and data<br>collection during imaging. KINEGRAPH<br>VMA™ software also facilitates<br>quantitative assessment of vertebral<br>motion in digital medical images.<br>Information about the motion of<br>selected objects, such as bone<br>structures, can be generated and<br>presented in the form of a 'motion<br>analysis' report containing graphics,<br>charts, and text. |
| Technological<br>Characteristics<br>/ Functionality | • Lumbar, cervical intervertebral motion<br>measurements<br>• Lumbar, cervical device assisted<br>bending | • Lumbar, cervical intervertebral motion<br>measurements<br>• Lumbar, cervical device assisted<br>bending |
| Principles of<br>Operation | • Produce digital imaging output files<br>with the use of the MOTION<br>NORMALIZER device in<br>combination with a standard,<br>commercially available fluoroscope<br>and standard radiological procedure<br>table.<br><br>• KineGraph VMA™ Image<br>Processing software is operated by<br>Ortho Kinematics, Inc. staff as a<br>service to facilitate template<br>placement for prescriber users.<br><br>• Transfer output files from the<br>MOTION NORMALIZER device to | • Produce digital imaging output files<br>with the use of the MOTION<br>NORMALIZER device in combination<br>with a standard, commercially<br>available fluoroscope and standard<br>radiological procedure table.<br><br>• KineGraph VMA™ Image Processing<br>software is operated by Ortho<br>Kinematics, Inc. staff as a service to<br>facilitate template placement for<br>prescriber users.<br><br>• Transfer output files from the<br>MOTION NORMALIZER device to |
| KineGraph VMA™ System v2.2<br>(subject device) | KineGraph VMA™ System v2.0<br>(K130743, the cleared device) | |
| the company by uploading to a<br>server using the KineGraph VMA™<br>DataLink Software. | the Company using removable<br>media that is shipped to the<br>Company. | |
| • KineGraph VMA™ Image<br>processing and Image Review<br>software, as well as patient results<br>which have been processed by<br>Ortho Kinematics, Inc., are available<br>via any internet connected computer<br>terminal running installed with<br>internet browsing software. | • KineGraph VMA™ Image processing<br>software runs on standard off-the-<br>shelf image review workstations<br>installed with certain specific off the<br>shelf software (Matlab runtime<br>compiler, Microsoft Excel, others) | |
| • Prescribers can use online user<br>accounts to store alert preferences,<br>such that they can be alerted upon<br>login whenever patient results<br>exceed the user-defined thresholds. | • Patient results transferred from the<br>company to the prescriber via<br>removable media. | |
| | • KineGraph VMA™ Image Review<br>software runs on standard off-the-<br>shelf image review workstations<br>installed with certain specific off the<br>shelf software (Adobe Flash,<br>ghostscript, others) | |
| | • No alerting features. | |
7004 Bee Caves Road Building Three, Suite 315 Austin, Texas 78746 Fax: (512) 334-5500 Ortho Kinematics
Page 112 of 120 Phone: (512) 334-5490
{3}------------------------------------------------
Ortho Kinematics
7004 Bee Caves Road
Building Three, Suite 315
Austin, Texas 78746 Fax: (512) 334-5500
.
.
·
Page 113 of 120 Phone: (512) 334-5490
・
-
、
:
{4}------------------------------------------------
K133875
Page 5 of 5
7004 Bee Caves Road
Building Three, Suite 315
Austin, Texas 78746 Fux: (512) 334-5500
Ortho Kinematics
・
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 17, 2014
Ortho Kinematics, Inc. % John Smith, M.D., J.D. Hogan Lovells US LLP 555 13th Street, NW WASHINGTON DC 20004
Re: K133875
Trade/Device Name: KineGraph VMA™ (Vertebral Motion Analyzer) Software, and the Motion Normalizer™ Patient Handling and Data Collection Device (Version 2.2) Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 20, 2013 Received: December 20, 2013
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Dr. Smith Section of the Research and Career of the Ca
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Smh.7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K133875
#### Device Name
KINEGRAPH VMA™ (VERTEBRAL MOTION ANALYZER) and MOTION NORMALIZER (version 2.2)
#### Indications for Use (Describe)
The KineGraph VMA™ sollware is a quantitative imaging software application intended to be used to process digital image files. It is designed for physicians and clinical professionals who are interested in the analysis of motion in in musculoskeletal images of the spine. KineGraph VMA™ software permits users to review static and dynamic digital lumbar and cervical spine images acquired with the assistance of the Motion Normalizer patient handling and data collection device, which is designed for use by imaging technicians and intented to assist with patient lumbar and cara collection during imaging. KineGraph VMA™ software also facilitates quantitative assessment of vertebral motion in digital images. Information about the motion of selected objects, such as bone structures, can be generated in the form of a `motion analysis' report containing graphics, charts. and text.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sm.h.f)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# ...... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.