The Fluid Level Monitor is a passive sensing device that alarms when the infusion fluid level is low. The device provides both visual and audible alarms when preset condition is met. The Fluid Level Monitor is intended for low-fluid-level alarm use in any healthcare setting where gravity flow infusion is utilized. The use of the device improves efficiency within a busy medical setting by providing healthcare personnel of timely notification of the low-fluid-level status. The available models are: LM771: 1-channel, DC battery power LM781 1-channel, DC plug-in power LM785 - 5-channel, DC plug-in power
Device Story
Passive sensing device; monitors infusion fluid levels in gravity flow systems. External sensor clips onto tubing; connects to main control unit via shielded cable. Control unit processes sensor input; triggers visual (LEDs) and audible (buzzer) alarms when fluid reaches preset low level. Operates on DC power (battery or plug-in). Used in clinical settings by healthcare personnel to improve monitoring efficiency. Provides timely notification of low-fluid status; assists in clinical decision-making regarding infusion management; benefits patient by preventing infusion interruption.
Clinical Evidence
Bench testing only. Performance testing confirmed compliance with design specifications and verified that all functions operate as designed. Testing demonstrated consistent power consumption across battery/plug-in power and 1-channel/5-channel configurations.
Technological Characteristics
Passive electronic monitor for gravity flow infusion. Components: main control unit, external sensor, shielded signal cable. Power: DC battery or DC plug-in. Monitoring capacity: 1-channel (LM771, LM781) or 5-channel (LM785). Output: Visual (LED) and audible (buzzer) alarms. Processor-based measurement.
Indications for Use
Indicated for use in any healthcare setting utilizing gravity flow infusion systems to provide low-fluid-level notification to healthcare personnel.
Regulatory Classification
Identification
An electronic monitor for gravity flow infusion systems is a device used to monitor the amount of fluid being infused into a patient. The device consists of an electronic transducer and equipment for signal amplification, conditioning, and display.
{0}------------------------------------------------
# 5. 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for Coleman Laboratories' Fluid Level Monitor 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
### 1) Submitter Information
| Company Name: | Coleman Laboratories |
|------------------------|----------------------------------------------------------------|
| Company Address: | 1150 First Avenue, Suite 501<br>King of Prussia, PA 19406, USA |
| Contact: | Le-Jun Yin, Vice President, Product Management |
| Email: | lejun.yin@colemanlabs.com |
| Phone: | (484) 727-8812 |
| Fax: | (484) 727-8828 |
| Date Summary Prepared: | December 18th, 2013 |
## 2) Device Identification
| Generic Device Name: | Electronic I.V. Monitor |
|-------------------------|------------------------------------------------------|
| Trade/Proprietary Name: | Fluid Level Monitor Model |
| Numbers: | LM771, LM781, LM785 |
| Classification: | Class II |
| Panel: | General Hospital (80) |
| Regulation Number: | 21 CFR 880.2420 |
| Regulation Name: | Electronic monitor for gravity flow infusion systems |
| Product Code: | FLN |
| 510(k) Number: | K133870 |
## 3) Predicate Device
| 510(k) Number: | K030136, K903193 |
|----------------|----------------------|
| Trade Name: | Drip Alert, Levelert |
| Product Code: | FLN |
## 4) Description of Device
Fluid Level Monitor is a passive sensing device that alarms when the infusion fluid level is low. Fluid Level Monitor provides both visual and audible alarms when preset condition is met. Fluid Level Monitor operates on DC power source.
Fluid Level Monitor includes both main control unit and external sensor. There are status LEDs on the main control unit to display run, alarm and battery-low status. There is a buzzer on the main control unit to provide audible alarm.
The external sensor is connected to the main control unit via shielded signal cable. The sensor can be clipped on the tubing to monitor fluid level.
### 5) Intended Use
The Fluid Level Monitor is a passive sensing device that alarms when the infysion fluid level is low. The device provides both visual and audible alarms when preset condition is met.
The Fluid Level Monitor is intended for low-fluid-level alarm use in any healthcare setting where gravity flow infusion is utilized. The use of the device improves efficiency within a busy medical setting by providing healthcare personnel of timely notification of the low-fluid-level status. The available models are:
LM771: 1-channel, DC battery power LM781 1-channel, DC plug-in power
{1}------------------------------------------------
LM785 - 5-channel, DC plug-in power
Coleman Laboratories' Fluid Level Monitors (3 Models) are substantially equivalent to the predicate devices:
K030136 - Drip Alert, Drip Alert, Inc. K903193 - Smith & Nephew, Inc.
They have the same design principles, material, indications for use, and intended use.
| Device Name | Fluid Level Monitor | | | Drip Alert | Levelert |
|-----------------------------------------------------------|---------------------|--------------|--------------|---------------|--------------|
| Device Model | LM771 | LM781 | LM785 | - | - |
| 510(K) Number | | K133870 | | K030136 | K903193 |
| Passive device | Yes | Yes | Yes | Yes | Yes |
| Uses a processor to<br>perform<br>measurements | Yes | Yes | Yes | Yes | Yes |
| Use sensor to detect<br>fluid-level-low status | Yes | Yes | Yes | Yes | Yes |
| Visual and audible<br>alarm on fluid-level-<br>low status | Yes | Yes | Yes | Yes | Yes |
| Patient has no contact<br>with device | Yes | Yes | Yes | Yes | Yes |
| Class II device | Yes | Yes | Yes | Yes | Yes |
| Power source | DC<br>battery | DC<br>plugin | DC<br>plugin | DC<br>battery | DC<br>plugin |
| Monitoring Point | 1 | 1 | 5 | 1 | 1 |
# 7) Summary Performance Data
Based on the risk analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 18. Performance testing was conducted to confirm compliance to design specifications; all functions were verified to operate as designed.
{2}------------------------------------------------
The performance test data demonstrated that the system power consumption remains same when using:
- a) DC battery or DC plug-in power
- b) 1-channel or 5-channel sensing
We concluded that the design difference on number of monitoring points does not raise safety or effectiveness concerns.
### 8) Conclusion
Based on the analysis on nonclinical test data, and comparison on design principle, operation procedure and indications for use, we demonstrated that the device is as safe, as effective, and performs as well as the predicate devices. The design difference does not raise questions on safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2014
Coleman Laboratories Le-Jun Yin Vice President, Product Management 1150 First Avenue, Suite 501 King of Prussia, PA 19406, USA
Re: K133870
Trade/Device Name: Fluid Level Monitor, Models LM771, LM781, and LM785 Regulation Number: 21 CFR 880.2420 Regulation Name: Electronic monitor for gravity flow infusion systems Regulatory Class: II Dated: March 13, 2014 Received: March 28, 2014
Dear Mr. Yin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Yin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan R. DODDS, MA
Erin I. Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
#### 510(k) Number (if known) K133870
Device Name
Fluid Level Monitor, Models LM771, LM781, and LM785
#### Indications for Use (Describe)
The Fluid Level Monitor is a passive sensing device that alarms when the infusion fluid is low. The device provides both visual and audible alarms when preset condition is met.
The Fluid Level Monitor is intended for low-fluid-level alarm use in any healthcare setting where gravity flow infirsed. The use of the device improves efficiency within a busy medical setting by providing healthcare personnel of the low-fluid-level status.
LM771: 1-channel, DC battery power LM781: 1-channel, DC plug-in power LM785: 5-channel DC plug-in power
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Digitally signed by Sajjad H. Syed -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Sajjad H. Syed -S, 0.9.2342.19200300.100.1.1=2000601742 Date: 2014.04.24 12:11:28 -04'00' PSC Publishing Services (30) ) 443-6740 Page 1 of 2 FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.