← Product Code [IOR](/productcode/IOR) · K133855

# BREEZY ELEGANCE (K133855)

_Sunrise Medical (Us), LLC · IOR · Jan 31, 2014 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133855

## Device Facts

- **Applicant:** Sunrise Medical (Us), LLC
- **Product Code:** [IOR](/productcode/IOR.md)
- **Decision Date:** Jan 31, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine

## Indications for Use

Breezy Elegance Folding Manual Wheelchairs are manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Device Story

Breezy Elegance is a manual, user-propelled folding wheelchair. Input consists of user physical force applied to handrims on 24-inch rear wheels. Device features aluminum cross-braced frame, quick-release axles, front casters, padded seat/back slings, and swing-away arm/leg rests. Attendant-operated push handles and optional drum brakes allow for assisted mobility. Used in clinical or home environments to provide mobility for individuals restricted to a sitting position. Device folds for transport and storage. Output is physical mobility for the rider. Benefits include improved independence and accessibility for users with mobility limitations.

## Clinical Evidence

Bench testing only. Design verification demonstrated performance equivalent to recognized wheelchair standards and the predicate device. Third-party design validation confirmed fitness for intended use regarding performance, human factors, and usability.

## Technological Characteristics

Aluminum frame; triple cross-brace folding mechanism; 24-inch wheels with handrims; quick-release axles; swing-away arm/leg rests; fabric-covered foam upholstery; tension-adjustable back straps. Optional drum brake system. Manual energy source. No software or electronic components.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Quickie Q2 Lite ([K072153](/device/K072153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

improving people's lives

Image /page/0/Picture/1 description: The image shows the logo for Sunrise Medical. The logo consists of a stylized image of a sunrise on the left, with the word "SUNRISE" in large, bold letters to the right. Below "SUNRISE" is the word "MEDICAL" in smaller, bold letters. The logo is simple and professional, and the use of the sunrise image suggests themes of hope, new beginnings, and health.

## K133855

## 510(k) Summary

JAN 3 1 2014

| Device            | Breezy Elegance Manual Folding Wheelchair |                             |
|-------------------|-------------------------------------------|-----------------------------|
| Owner             | Sunrise Medical (US) LLC                  |                             |
|                   | 2842 Business Park Avenue                 |                             |
|                   | Fresno, CA 93727                          |                             |
|                   | Phone: 559-348-2572                       |                             |
|                   | Fax: 559-294-2872                         |                             |
| Contact           | Laurie H. Roberts                         |                             |
|                   | Senior Manager-Regulatory Affairs         |                             |
| Date              | 17 December 2013                          |                             |
| Subject<br>Device | Trade name                                | Breezy Elegance             |
|                   | Common name                               | Manual Folding Wheelchair   |
|                   | Regulation                                | 21 CFR 890.3850             |
|                   | Device name                               | Mechanical wheelchair       |
|                   | Product code                              | IOR                         |
|                   | Device class                              | 1                           |
|                   | Panel                                     | Physical Medicine           |
| Predicate         | Trade name                                | Quickie Q2 Lite             |
|                   | Manufacturer                              | Sunrise Medical (US) LLC    |
|                   | Market Clearance                          | 510(k): K072153 (10 Sep 07) |

## Device Description

Device purpose

・・・・・・・・・・・・

The product offering consists of a standard folding aluminum wheelchair. (Standard in the sense that it is not built to a specific order, rather it will be built to inventory.)

Device components

- The wheelchairs are
	- User-propelled, as such, chair has 24" wheels with attached handrims - Foldable (using a cross-brace design) for easier transport and stowage

Continued on next page

2842 Business Park Avenue · Fresno, CA 93727-1328 · USA

800.333.4000

www.SunriseMedical.com

{1}------------------------------------------------

## 510(k) Summary: Breezy Elegance Wheelchair

| Device<br>components<br>(continued)   | · The device comprises<br>- Cross-braced folding frame<br>- Quick-release Axle<br>- Large rear wheels with handrims for self-propulsion<br>- Smaller front wheels on casters for stability<br>- Padded seat sling material designed to support a seat cushion and to fold<br>when the wheelchair frame is folded<br>- An upholstered backrest designed to fold when the wheelchair frame is<br>folded, pre-set angle to 8<br>- Push handles at top of backrest to allow attendant to aid mobility of rider<br>- Swing-away armrests and swing-in/swing-out legrests (with footplates)<br>· Available Options<br>Elevating legrest<br>Anti-tip Tubes |  |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Technological<br>character-<br>istics | Aluminum frame to reduce weight and provide durability<br>Triple cross-brace for easy folding<br>Swing-away arm and leg rests for easier entry, egress, or transfer<br>Seat and back slings to accommodate separate cushion system for<br>rider comfort and ease of folding<br>Tension adjustable back straps for user comfort<br>Optional on Models B & C<br>●<br>Drum brake system to aid attendant to control chair movement<br>-<br>Height adjustable push handles for attendant comfort<br>-<br>Height adjustable armrests for user comfort                                                                                                    |  |

## Intended Use

Breezy Elegance Folding Manual Wheelchairs are manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Equivalence

The Breezy Elegance Wheelchair is substantially equivalent to its predicate device, Quickie Q2 Lite, based on intended use, materials, operating principle, technology, safety, and performance.

## Breezy Elegance equivalence to Quickie Q2 Lite

Intended Use Identical: Provides mobility to persons restricted to a sitting position

{2}------------------------------------------------

## 510(k) Summary: Breezy Elegance Wheelchair

| Materials                 | Substantially equivalent:<br>Aluminum frame, support members, wheels, and components Steel fasteners and components Flexible polymeric tires Fabric covered foam upholstery Fabric straps and slings                                                                                                                                                                                                                                                                   |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating<br>principle    | Identical:<br>User propelled by hand pressure against handrims attached to large (24 inch) wheels Wheel locks to hold chair on slight slopes Quick release wheel hubs for easy removal of wheel for transport and storage Cross-braced, hinged frame for folding of chair for transport and storage Seat cushion and back upholstery for user comfort Adjustable, swing-away legrests with angle-adjustable footrests Push handles to allow assistance by an attendant |
| Energy<br>source          | Identical: User propelled                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Technology                | Identical: Standard mechanical wheelchair construction and operation.                                                                                                                                                                                                                                                                                                                                                                                                  |
| Safety and<br>performance | Substantially equivalent:<br>Accomplishes the same performance in the same way as the predicate with slightly more weight and weight carrying capacity Design raises no new issues of safety or effectiveness                                                                                                                                                                                                                                                          |
| Bench Tests               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Design<br>Verification    | Design Control requirements per 21 CFR 820.30; testing/demonstration that<br>all essential specifications show equivalent performance to recognized<br>wheelchair standards or to the predicate device (Quickie Q2 Lite).                                                                                                                                                                                                                                              |
| Design<br>Validation      | Third-party evaluation of chair performance, human factors, and usability<br>demonstrate the fitness of the design to meet the requirements and demands<br>of its intended use.                                                                                                                                                                                                                                                                                        |

.

{3}------------------------------------------------

510(k) Summary: Breezy Elegance Wheelchair

## Biocompatibility

| Type of<br>Contact | • Momentary contact with uncompromised skin of hands—various parts<br>• Transient contact with uncompromised skin of hands—handrims<br>• Transient contact with uncompromised skin of arms—armrests                                                               |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Evaluation         | • Momentary contact of intact skin of hands on aluminum, steel, nylon fabric,<br>hard/soft plastics do not raise biocompatibility issues<br>• Handrim and armrest materials are identical with the predicate device and,<br>thus, raise no biocompatibility issue |

## Packaging / Shelf life

Based on the durability, materials, and use of wheelchairs, packaging and shelf life are not relevant to the safe and effective use of these devices.

## Conclusion

Breezy Elegance Manual Folding Wheelchairs are substantially equivalent to their predicate devices in technology, performance, and intended use. There are no significant differences between Breezy Elegance Wheelchairs and the Quickie Q2 Lite Wheelchair predicates in design which would raise new issues of safety and effectiveness, performance, function or intended use of the device. In addition, the chair design has been shown to be safe and effective for its intended uses.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

### January 31, 2014

Sunrise Medical (US) LLC c/o Laurie H. Roberts 2842 Business Park Ave. Fresno, CA 93727

Re: K133855

Trade/Device Name: Breezy Elegance Folding Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: January 2, 2014 Received: January 2, 2014

Dear Ms. Roberts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{5}------------------------------------------------

Page 2 - Ms. Laurie H. Roberts

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda,gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDeyjces/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Carlos L. Pena -S

Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ... ... . ....

### Indications for Use

510(k) Number (if known) K133855

### Device Name

### Breezy Elegance Folding Manual Wheelchair

### Indications for Use (Describe)

Breezy Elegance Folding Manual Wheelchairs are manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Carlos L. Pena -S

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### ~~ ~ . ~ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·

---

**Source:** [https://fda-staging.innolitics.com/device/K133855](https://fda-staging.innolitics.com/device/K133855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).