K133855 · Sunrise Medical (Us), LLC · IOR · Jan 31, 2014 · Physical Medicine
Device Facts
Record ID
K133855
Device Name
BREEZY ELEGANCE
Applicant
Sunrise Medical (Us), LLC
Product Code
IOR · Physical Medicine
Decision Date
Jan 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
Breezy Elegance Folding Manual Wheelchairs are manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
Breezy Elegance is a manual, user-propelled folding wheelchair. Input consists of user physical force applied to handrims on 24-inch rear wheels. Device features aluminum cross-braced frame, quick-release axles, front casters, padded seat/back slings, and swing-away arm/leg rests. Attendant-operated push handles and optional drum brakes allow for assisted mobility. Used in clinical or home environments to provide mobility for individuals restricted to a sitting position. Device folds for transport and storage. Output is physical mobility for the rider. Benefits include improved independence and accessibility for users with mobility limitations.
Clinical Evidence
Bench testing only. Design verification demonstrated performance equivalent to recognized wheelchair standards and the predicate device. Third-party design validation confirmed fitness for intended use regarding performance, human factors, and usability.
Technological Characteristics
Aluminum frame; triple cross-brace folding mechanism; 24-inch wheels with handrims; quick-release axles; swing-away arm/leg rests; fabric-covered foam upholstery; tension-adjustable back straps. Optional drum brake system. Manual energy source. No software or electronic components.
Indications for Use
Indicated for persons restricted to a sitting position requiring mobility assistance. Suitable for both prescription and over-the-counter use.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
improving people's lives
Image /page/0/Picture/1 description: The image shows the logo for Sunrise Medical. The logo consists of a stylized image of a sunrise on the left, with the word "SUNRISE" in large, bold letters to the right. Below "SUNRISE" is the word "MEDICAL" in smaller, bold letters. The logo is simple and professional, and the use of the sunrise image suggests themes of hope, new beginnings, and health.
## K133855
## 510(k) Summary
JAN 3 1 2014
| Device | Breezy Elegance Manual Folding Wheelchair | |
|-------------------|-------------------------------------------|-----------------------------|
| Owner | Sunrise Medical (US) LLC | |
| | 2842 Business Park Avenue | |
| | Fresno, CA 93727 | |
| | Phone: 559-348-2572 | |
| | Fax: 559-294-2872 | |
| Contact | Laurie H. Roberts | |
| | Senior Manager-Regulatory Affairs | |
| Date | 17 December 2013 | |
| Subject<br>Device | Trade name | Breezy Elegance |
| | Common name | Manual Folding Wheelchair |
| | Regulation | 21 CFR 890.3850 |
| | Device name | Mechanical wheelchair |
| | Product code | IOR |
| | Device class | 1 |
| | Panel | Physical Medicine |
| Predicate | Trade name | Quickie Q2 Lite |
| | Manufacturer | Sunrise Medical (US) LLC |
| | Market Clearance | 510(k): K072153 (10 Sep 07) |
## Device Description
Device purpose
・・・・・・・・・・・・
The product offering consists of a standard folding aluminum wheelchair. (Standard in the sense that it is not built to a specific order, rather it will be built to inventory.)
Device components
- The wheelchairs are
- User-propelled, as such, chair has 24" wheels with attached handrims - Foldable (using a cross-brace design) for easier transport and stowage
Continued on next page
2842 Business Park Avenue · Fresno, CA 93727-1328 · USA
800.333.4000
www.SunriseMedical.com
{1}------------------------------------------------
## 510(k) Summary: Breezy Elegance Wheelchair
| Device<br>components<br>(continued) | · The device comprises<br>- Cross-braced folding frame<br>- Quick-release Axle<br>- Large rear wheels with handrims for self-propulsion<br>- Smaller front wheels on casters for stability<br>- Padded seat sling material designed to support a seat cushion and to fold<br>when the wheelchair frame is folded<br>- An upholstered backrest designed to fold when the wheelchair frame is<br>folded, pre-set angle to 8<br>- Push handles at top of backrest to allow attendant to aid mobility of rider<br>- Swing-away armrests and swing-in/swing-out legrests (with footplates)<br>· Available Options<br>Elevating legrest<br>Anti-tip Tubes | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Technological<br>character-<br>istics | Aluminum frame to reduce weight and provide durability<br>Triple cross-brace for easy folding<br>Swing-away arm and leg rests for easier entry, egress, or transfer<br>Seat and back slings to accommodate separate cushion system for<br>rider comfort and ease of folding<br>Tension adjustable back straps for user comfort<br>Optional on Models B & C<br>●<br>Drum brake system to aid attendant to control chair movement<br>-<br>Height adjustable push handles for attendant comfort<br>-<br>Height adjustable armrests for user comfort | |
## Intended Use
Breezy Elegance Folding Manual Wheelchairs are manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position.
## Equivalence
The Breezy Elegance Wheelchair is substantially equivalent to its predicate device, Quickie Q2 Lite, based on intended use, materials, operating principle, technology, safety, and performance.
## Breezy Elegance equivalence to Quickie Q2 Lite
Intended Use Identical: Provides mobility to persons restricted to a sitting position
{2}------------------------------------------------
## 510(k) Summary: Breezy Elegance Wheelchair
| Materials | Substantially equivalent:<br>Aluminum frame, support members, wheels, and components Steel fasteners and components Flexible polymeric tires Fabric covered foam upholstery Fabric straps and slings |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating<br>principle | Identical:<br>User propelled by hand pressure against handrims attached to large (24 inch) wheels Wheel locks to hold chair on slight slopes Quick release wheel hubs for easy removal of wheel for transport and storage Cross-braced, hinged frame for folding of chair for transport and storage Seat cushion and back upholstery for user comfort Adjustable, swing-away legrests with angle-adjustable footrests Push handles to allow assistance by an attendant |
| Energy<br>source | Identical: User propelled |
| Technology | Identical: Standard mechanical wheelchair construction and operation. |
| Safety and<br>performance | Substantially equivalent:<br>Accomplishes the same performance in the same way as the predicate with slightly more weight and weight carrying capacity Design raises no new issues of safety or effectiveness |
| Bench Tests | |
| Design<br>Verification | Design Control requirements per 21 CFR 820.30; testing/demonstration that<br>all essential specifications show equivalent performance to recognized<br>wheelchair standards or to the predicate device (Quickie Q2 Lite). |
| Design<br>Validation | Third-party evaluation of chair performance, human factors, and usability<br>demonstrate the fitness of the design to meet the requirements and demands<br>of its intended use. |
.
{3}------------------------------------------------
510(k) Summary: Breezy Elegance Wheelchair
## Biocompatibility
| Type of<br>Contact | • Momentary contact with uncompromised skin of hands—various parts<br>• Transient contact with uncompromised skin of hands—handrims<br>• Transient contact with uncompromised skin of arms—armrests |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Evaluation | • Momentary contact of intact skin of hands on aluminum, steel, nylon fabric,<br>hard/soft plastics do not raise biocompatibility issues<br>• Handrim and armrest materials are identical with the predicate device and,<br>thus, raise no biocompatibility issue |
## Packaging / Shelf life
Based on the durability, materials, and use of wheelchairs, packaging and shelf life are not relevant to the safe and effective use of these devices.
## Conclusion
Breezy Elegance Manual Folding Wheelchairs are substantially equivalent to their predicate devices in technology, performance, and intended use. There are no significant differences between Breezy Elegance Wheelchairs and the Quickie Q2 Lite Wheelchair predicates in design which would raise new issues of safety and effectiveness, performance, function or intended use of the device. In addition, the chair design has been shown to be safe and effective for its intended uses.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
### January 31, 2014
Sunrise Medical (US) LLC c/o Laurie H. Roberts 2842 Business Park Ave. Fresno, CA 93727
Re: K133855
Trade/Device Name: Breezy Elegance Folding Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: January 2, 2014 Received: January 2, 2014
Dear Ms. Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Ms. Laurie H. Roberts
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda,gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDeyjces/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Carlos L. Pena -S
Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ... ... . ....
### Indications for Use
510(k) Number (if known) K133855
### Device Name
### Breezy Elegance Folding Manual Wheelchair
### Indications for Use (Describe)
Breezy Elegance Folding Manual Wheelchairs are manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
区 Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Carlos L. Pena -S
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### ~~ ~ . ~ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.