KLASSIC HD HIP SYSTEM, OFFSET FEMORAL STEM, NON-POROUS / POROUS
K133832 · Total Joint Orthopedics, Inc. · LPH · Mar 13, 2014 · Orthopedic
Device Facts
Record ID
K133832
Device Name
KLASSIC HD HIP SYSTEM, OFFSET FEMORAL STEM, NON-POROUS / POROUS
Applicant
Total Joint Orthopedics, Inc.
Product Code
LPH · Orthopedic
Decision Date
Mar 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Klassic HDTM Hip System is intended for prosthetic replacement without bone cement in treatment of the following: - Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli and painful hip dysplasia. - Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis. - Those patients with failed previous surgery where pain, deformity or dysfunction persists. - Revision of a previously failed hip arthroplasty. - Patients who require a total hip replacement.
Device Story
Klassic HD Offset Femoral Stem is a modular, uncemented femoral prosthesis used in total hip arthroplasty. Designed to restore hip joint biomechanics by achieving proper leg length and lateral offset. Monolithic titanium alloy (Ti6Al4V) Zweymuller blade-type stem; features 121° neck-shaft angle. Offered in porous and non-porous options. Used by orthopedic surgeons in clinical/OR settings. Implanted during surgery to replace the femoral head/neck; provides structural support for hip joint. Benefits patients by addressing pain, deformity, or dysfunction associated with degenerative or inflammatory joint disease.
Clinical Evidence
Bench testing only. Testing included neck and stem fatigue, axial and torsional disassembly evaluation, and range of motion determination per ASTM F1814-97a and FDA guidance for femoral stem prostheses. No clinical data presented.
Technological Characteristics
Monolithic titanium alloy (Ti6Al4V) Zweymuller blade-type stem. Features 121° neck-shaft angle. Porous and non-porous options available. Sterilized via gamma irradiation. Single-use. Modular design.
Indications for Use
Indicated for patients requiring total hip replacement due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory joint disease (rheumatoid arthritis), failed previous surgery, or revision of failed hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Klassic HD Hip System Total Joint Orthopedics (K100445)
{0}------------------------------------------------
#### 510(k) SUMMARY
# 510(k) Notification K133832 (pg 1/3)
#### GENERAL INFORMATION
# Applicant:
Total Joint Orthopedics, Inc. 1567 E Stratford Avenue Salt Lake City, UT 84106 United States Phone: 801-486-6070 Fax: 801-486-6117
#### Contact Person:
Ben Casey Product Development Engineer Total Joint Orthopedics l 567 E. Stratford Avenue Salt Lake City, UT 84106 United States Phone: 801-486-6070 FAX: 801-486-6117
Date Prepared: December 16, 2013
#### DEVICE INFORMATION
The Klassic HD™ Offset Femoral Stem is a modular femoral prosthesis to be used within the Klassic HD™ Hip System to help surgeons restore hip joint biomechanics intraoperatively by helping to achieve proper leg length and offset in total hip arthroplasty.
Trade Name: Klassic HD™ Offset Femoral Stem
Generic/Common Name: Femoral Hip Stem
# Classification:
21 CFR § 888.3358 - Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.
Product Code: LPH, MBL, LWJ
{1}------------------------------------------------
# 510(k) Summary
### PREDICATE DEVICES
The Klassic HD Offset Femoral Stem is substantially equivalent in intended use, design, function and performance testing to the following predicate devices:
- Klassic HD Hip System Total Joint Orthopedics K100445 .
- Alloclassic Zweymuller SL Offset Femoral Stem Centerpulse . Orthopedics K033664
# INTENDED USE
The Klassic HDTM Hip System is intended for prosthetic replacement without bone cement in treatment of the following:
- Patient conditions of non-inflammatory degenerative joint disease (NIDJD): . avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli and painful hip dysplasia.
- . Patient conditions of inflammatory joint disease (ID): rheumatoid arthritis.
- Those patients with failed previous surgery where pain, deformity or dysfunction . persists.
- ♥ Revision of a previously failed hip arthroplasty.
- Patients who require a total hip replacement. .
# PRODUCT DESCRIPTION
The Klassic HD Offset Femoral Stem is a femoral component to be used within the Klassic HD Hip System as a prosthesis designed to help surgeons restore hip joint biomechanics intraoperatively by helping to achieve proper leg length and offset. The Klassic HD Offset Femoral Stem is offered in both porous and non-porous options. Each configuration of the Klassic HD Offset Femoral Stem is offered in 9 sizes, is sterilized by gamma irradiation and is intended for single-use only.
# TECHNOLOGICAL CHARACTERISTICS
The Offset Femoral Stem is a monolithic, titanium alloy, Zweymuller blade type stem featuring a neck shaft angle of 121° and increased lateral offset compared to the 131° neck shaft angle from the Klassic HD Femoral Stem. The Klassic HD Offset Femoral Stem utilizes the identical body geometry as the Klassic HD Femoral Stem, distal to the line of resection; along with the identical Ti6Al4V material and porous coated options.
The neck shaft angle of 121° and increased neck length are similar to the Alloclassic Zweymuller SL Offset Femoral Stem featuring a neck angle of 121° and 6.25mm increased lateral offset versus its predicate device.
# TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
Non-clinical testing was conducted on the Klassic HD Offset Femoral Stem to support a determination of substantial equivalence to the predicate devices. Non-clinical testing was based off of guidance from
{2}------------------------------------------------
# 510(k) SUMMARY
- Guidance for Industry and FDA Staff Non-clinical Information for . Femoral Stem Prostheses, Document 1647 dated September 17, 2007
- . ASTM F1814-97a - Standard Guide for Evaluating Modular Hip and Knee Joint Components
The performance testing for the Klassic HD Offset Femoral Stem consisted of neck and stem fatigue testing, axial and torsional disassembly evaluation and range of motion determination. The porous coating properties for the Klassic HD Offset Femoral Stem Porous are identical to that of the Klassic HD Femoral Stem Porous. All bench testing and evaluation demonstrates the Klassic HD Offset Femoral Stem is equivalent in regards to safety and efficacy, is suitable for Total Hip Arthroplasty and is substantially equivalent to predicate devices.
# SUBSTANTIAL EQUIVALENCE
The indications for use for the predicate devices are substantially equivalent to the proposed indications for use for the Klassic HD Offset Femoral Stem. The Klassic HD Offset Femoral stem is similar to the predicate devices based on technological characteristics, design, non-clinical performance testing, material and intended use. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the Klassic HD Offset Femoral Stem is substantially equivalent to the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 13, 2014
Total Joint Orthopedics, Inc. Mr. Ben Casey Product Development Engineer 1567 E. Stratford Avenue Salt Lake City, Utah 84106
Re: K133832
Trade/Device Name: Klassic HD™ Offset Femoral Stem Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, MBL, LWJ Dated: December 16, 2013 Received: December 17, 2013
Dear Mr. Casey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
# Page 2 - Mr. Ben Casey
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wigqins
- for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
TOTAL JOINT ORTHOPEDICS, INC.
# KLASSIC HD™ OFFSET FEMORAL STEM 510(k) PREMARKET NOTIFICATION
# SECTION 4 INDICATIONS FOR USE STATEMENT
K133832 (pg 1/1) 510(k) Number (if known):
Device Name: Klassic HDTM Offset Femoral Stem
# Indications For Use:
The Klassic HDTM Hip System is intended for prosthetic replacement without bone cement in treatment of the following:
- Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular . necrosis, osteoarthritis, ankylosis, protrusio acetabuli and painful hip dysplasia.
- Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis. ♥
- Those patients with failed previous surgery where pain, deformity or dysfunction . persists.
- . Revision of a previously failed hip arthroplasty.
- Patients who require a total hip replacement. ●
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabet旗峻Frank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.