A-Link Z is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device system is designed for use with autograft to facilitate fusion. One device is used per intervertebral body space. A-Link Z is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. A-Link Z may be used as a stand alone device when all four (4) vertebral body bone screws are used. If the physician chooses to use fewer than the four (4) screws, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used.
Device Story
A-Link Z is an intervertebral body fusion device implanted in the lumbar spine to provide stability and support fusion; used with autograft. Device consists of an interbody spacer (PEEK or titanium alloy), a titanium press-fit plate, and titanium bone screws. Implanted by surgeons in clinical settings. Device provides stand-alone fixation when four screws are used; otherwise, requires supplemental fixation. Screws are secured via a cover plate and screw to prevent back-out. Benefits include spinal stabilization and facilitation of fusion in patients with degenerative disc disease.
Clinical Evidence
No clinical data submitted. Substantial equivalence demonstrated via bench testing including static and dynamic compression, torsion, and compression-shear testing per ASTM F2077, and subsidence testing per ASTM F2267.
Technological Characteristics
Materials: PEEK (ASTM D2026) or Ti-6Al-4V ELI (ASTM F136). Tantalum markers (ASTM F560). Design: Interbody spacer with press-fit plate and bone screws. Lordotic options: 7° and 12°. Dimensions: 36 mm width, heights 11-19 mm. Fixation: 4-screw system with anti-back-out cover plate. Sterilization: Gamma radiation.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis at one or two contiguous lumbar levels (L2-S1) after at least six months of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Cynch Spinal System – Visualif Interbody Fusion Implant System (K102090)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, flowing line style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2014
OsteoVasive, LLC % Kevin A. Thomas, Ph.D. PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K133827
Trade/Device Name: A-Link Z Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: October 20, 2014 Received: October 21, 2014
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Kevin A. Thomas, Ph.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Device Name
A-Link Z
Indications for Use (Describe)
A-Link Z is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device system is designed for use with autograft to facilitate fusion. One device is used per intervertebral body space.
A-Link Z is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
A-Link Z may be used as a stand alone device when all four (4) vertebral body bone screws are used. If the physician chooses to use fewer than the four (4) screws, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary OsteoVasive LLC A-Link Z
# K133827
November 13, 2014
# ADMINISTRATIVE INFORMATION
| Manufacturer Name | OsteoVasive LLC<br>14215 Proton Road<br>Dallas, TX 75244 | |
|---------------------------|----------------------------------------------------------|------------------------------------------|
| | Telephone: | +1 (214) 405-2870 |
| | Fax: | +1 (214) 960-4313 |
| Official Contact | Bryan Cowan | |
| | President | |
| Representative/Consultant | Kevin A. Thomas, Ph.D. | |
| | Floyd G. Larson | |
| | PaxMed International, LLC | |
| | 12264 El Camino Real, Suite 400 | |
| | San Diego, CA 92130 | |
| | Telephone: | +1 (858) 792-1235 |
| | Fax: | +1 (858) 792-1236 |
| | Email: | kthomas@paxmed.com<br>flarson@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | A-Link Z |
|----------------------------|-------------------------------------------------------|
| Common Name | Intervertebral body fusion device |
| Classification Name | Intervertebral fusion device with integrated fixation |
| Classification Regulations | 21 CFR 888.3080, Class II |
| Product Code | OVD |
| Classification Panel | Orthopedic |
| Reviewing Branch | Anterior Spine Devices Branch |
{4}------------------------------------------------
#### INTENDED USE
A-Link Z is indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device system is designed for use with autograft to facilitate fusion. One device is used per intervertebral body space.
A-Link Z is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
A-Link Z may be used as a stand alone device when all four (4) vertebral body bone screws are used. If the physician chooses to use fewer than the four (4) screws, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used.
#### DEVICE DESCRIPTION
The A-Link Z is a spinal device that is implanted in the intervertebral body space to improve stability of the spine while supporting fusion. Components are offered in different shapes and sizes to meet the requirements of the individual patient anatomy.
The interbody components are manufactured using medical grade polyetheretherketone (PEEK) conforming to ASTM D2026 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications or titanium alloy, Ti-6Al-4V ELI conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). The PEEK version of the interbody comes with two tantalum markers conforming to ASTM F560 Standard Specification for Unalloyed Tantalum for Surgical Implant Applications to facilitate implant placement. The A-Link Z is available in one footprint (36 mm wide). It has an anatomic shape with serrations on the superior and inferior surfaces. The A-Link Z is available in two lordotic angle options (79 and 12°). The 7° lordotic option has heights ranging from 11 mm in 2 mm increments and the 12° lordotic option has heights ranging from 11 mm to 19 mm increments.
The interbody comes with a press fit plate manufactured from titanium alloy, Ti-6Al-4V ELI conforming to ASTM F136. The plate has four screw holes that allow the passage of bone screws.
The 5.0 mm and 5.5 mm diameter bone screws are available in three lengths (20 mm, 25 mm and 30 mm). The screws are manufactured from titanium alloy, Ti-6A1-4V ELI according to ASTM F136. The A-Link Z is fixed to the adjacent vertebral bodies through the use of four screws inserted through anterior screw holes of the implant.
The screws are prevented from backing out after insertion by the attachment of a cover plate and a cover plate screw to the anterior side of the plate. The cover plate and cover plate screw are manufactured from titanium alloy, Ti-6A1-4V ELI according to ASTM F136.
{5}------------------------------------------------
If the physician chooses to use fewer than the four screws, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine should be used.
# PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included: performance testing, engineering analysis and dimensional analysis. Performance testing to demonstrate substantial equivalence included methods described in the standards ASTM F2077 Test Methods for Intervertebral Body Fusion Devices and ASTM F2267 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression. The following testing was performed:
- . Static & Dynamic Compression testing per ASTM F2077
- Static & Dynamic Torsion testing per ASTM F2077
- Static & Dynamic Compression Shear testing per ASTM F2077 ●
- Expulsion testing
- Subsidence testing per ASTM F2267
Static tensile testing also was performed to measure the force required to separate (pull) the device face plate component from the interbody spacer component.
Clinical data were not submitted in this premarket notification.
### EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent to the following predicate devices:
- . SpineSmith Partners, LP, Cynch Spinal System – Visualif Interbody Fusion Implant System (K102090)
- RSB Spine, LLC, InterPlate® (K071922) ●
- LDR Spine USA, LDR Spine ROI-A Implant System (K110327)
- Spinal USA, Spinal USA Interbody Fusion Device (K080314)
- Lanx, Inc., Lanx Lateral-SA System (K123767)
- Nexxt Spine, LLC, Honour Spacer System (K120345)
- Surgicraft Ltd., Stalif™ TT Intervertebral Body Fusion System (K073109) ●
- Life Spine, Inc., Life Spine Stand-Alone Spacer System (K091301)
- . Titan Spine, LLC, Endoskeleton® TAS (K111626)
The primary predicate device is K102090.
The implants of the subject device, A-Link Z, have footprint, height, and lordotic angle similar to those of implants cleared in K102090, K110327, K080314, K123767, K073109, K091301, and K111626. The screws of the subject device have dimensions similar to those of screws cleared in
{6}------------------------------------------------
K102090, K123767, and K091301. The screw locking mechanism of the subject device is similar to that of devices cleared in K102090, K071922, K123767, and K091301. The materials of the subject device are identical to those of devices cleared in K080314, K123767, K120345, and K111626. The subject device is gamma sterilized similarly to the device cleared in K110327.
## CONCLUSIONS
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including length, width, height, and lordotic angle of the implant and length and diameter of the screws. The subject and predicate devices are packaged in similar materials and sterilized using similar methods. Any differences in the technological characteristics between the subject and predicate devices do not raise different questions of safety or effectiveness.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Overall, A-Link Z has the following similarities to the predicate devices:
- has the same intended use,
- . uses the same operating principle.
- . incorporates the same basic design,
- incorporates the same or very similar materials, and ●
- has similar packaging and is sterilized using the same materials and processes. ●
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.