ORA 0.4MM [S], ORA 1.0MM [S], ORA 2.0MM [S], ORA 3.0MM [S], ORA 4.0MM [S]
Applicant
Sterngold Dental, LLC
Product Code
NHA · Dental
Decision Date
Apr 16, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ORA Implant Abutment System is indicated for use with dental implants to support and/or retain removable dental prostheses in the treatment of partially edentulous patients to restore chewing function. The abutment screws directly into endosseous implants or they screw into SFI Abutments which are screwed into endosseous implants.
Device Story
ORA Implant Abutment System is a precision-machined, screw-retained ball-shaped abutment; connects dental implants to removable partial or complete overdentures. Device consists of a metal abutment screwed into an endosseous implant; retention provided by a rubber o-ring (red for processing, white for final use) housed in a metal cap. Metal housing is processed into the denture; white o-ring facilitates easier insertion/removal. Used in clinical dental settings by practitioners. Benefits include restoration of chewing function for edentulous patients. Design is compatible with various implant platforms; cuff area replicates previously cleared SFI Implant Abutments.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by tolerance analysis of platforms and comparison of design/materials to predicate devices.
Indicated for partially edentulous patients requiring support or retention of removable dental prostheses to restore chewing function. Used with endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
SFI Bar® Implant Abutments for 7 Platforms (K130183)
SFI Bar® Implant Abutments for 9 Platforms (K132814)
Submission Summary (Full Text)
{0}------------------------------------------------
April 14, 2014 ORA Implant Abutment System
# 510(k) Summary
APR 1 6 2014
| Sponsor: | Sterngold Dental, LLC<br>23 Frank Mossberg Drive<br>Attleboro, MA 02703 |
|----------|-------------------------------------------------------------------------|
|----------|-------------------------------------------------------------------------|
- Contact: Maria Rao, QA/RA Director Ph: 508-226-5660 ext 1206
- Trade Name: ORA Implant Abutments System
- Common Name: Implant Abutments
Classification Name: Endosseous Dental Implant Abutment
- Classification: According to Section 513 of the Federal Food, Drug, and Cosmetic Act, the device classification is Class II
Product Code: NHA (21CFR 872.3630)
Legally Marketed Device to which Equivalence is claimed (Predicate Devices): Predicate Device(s): K900099, K130183, K132814.
| K900099 | The O-Ring System - ORS |
|---------|--------------------------------------------|
| K130183 | SFI Bar® Implant Abutments for 7 Platforms |
| K132814 | SFI Bar® Implant Abutments for 9 Platforms |
#### Description of Device:
The ORA Implant Abutment is a precision machined ball shaped abutment that connects a compatible dental implant system with a removable partial or complete overdenture. The implant abutment is screwed into the dental implant. Connection to and retention of a denture is provided by a rubber o-ring, which may or may not be held within a metal housing. There are two color o-rings, a red processing o-ring and a white final o-ring.
The retaining ring (metal housing) comes with the red o-ring (which is a firm rubber) inside. The retaining ring (metal housing) with the red o-ring is pushed over the wide part of the ball until seated. The stiffness of the red o-ring holds the housing in position. Any exposed areas of the abutment are blocked out so that only the metal housing is exposed. The retaining ring (metal housing) is then processed into the denture. After the material has cured, the denture is removed. The red o-ring is pulled out of the metal housing and the white o-ring is inserted into its place. The white o-ring is more flexible making insertion and removal easier. The bottom portion of the abutment (cuff area to end of threads) is an exact replica of the SFI Implant Abutments previously cleared by K130183 and K132814.
ORA Implant Abutments are available in sixteen different platforms and each platform is compatible with one or more implant types. Table 1 demonstrates implant/abutment compatibility. The difference between each platform is the internal connection with the specific implant. This connection has been previously cleared by K130183 and K132814— SFI Implant Abutments.
{1}------------------------------------------------
The devices are supplied non-sterile, and there is no shelf life.
### Intended Use of the Device:
The ORA Implant Abutment System is indicated for use with dental implants to support and/or retain removable dental prostheses in the treatment of partially edentulous patients to restore chewing function. The abutment screws directly into endosseous implants or they screw into SFI Abutments which are screwed into endosseous implants.
| Implant Brand | Model | | |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Nobel Biocare Brånemark System | 3.3 Fixture, 3.75 Fixture, 4.0 Fixture, 5.0 Fixture (Old Version), 3.75 MkII Self-tapping Fixture, 4.0<br>MkII Self-tapping Fixture | | |
| Sterngold-ImpiaMed | 3.3 Hex Cylinder, 4.0 Hex Cylinder, 3.75 Standard Hex Screw, 3.75 Self-tapping Hex Screw, 3.75<br>Self-tapping "SST" Hex Screw, 4.0 Standard Hex Screw, 4.0 Self-tapping Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 5.0 RP "SST" Hex Screw, 3.75 RP<br>Acid Etched, 4.0 RP Acid Etched, 5.0 RP Acid Etched, 4.1 Stern IC (4.8 head), 3.3 Stern IC (4.8 head) | | |
| Nobel Biocare (Steri-Oss®) | 3.8 HL Cylinder, 3.8 HL Threaded, 4.5 HL Threaded, 3.5 NobelReplace™, Replace® Select (NP),<br>4.0 NobelReplace Straight, (RP), 4.3Replace®Select&NobelReplace™ (RP) | | |
| Keystone (Lifecore) | 3.75 Restore® Self-tapping Screw, 4.0 Restore® Self-tapping Screw, 3.75 Restore® External Hex<br>Screw, 4.0 Restore® External Hex Screw, 4.0 Restore® External Hex Cylinder, 4.2 Sustain®<br>External Hex Cylinder, 3.75 Sustain® External Hex Screw, 4.0 Sustain® External Hex Screw, 4.2<br>Sustain® External Hex MC Cylinder, Stage-1TM (3.3 and 4.0 fixtures) | | |
| 3i Implant Inovations | 3.25 External Hex Miniplant®, 3.25 ICE™ Miniplant®, 3.25 OSSEOTITE® Miniplant®, 3.3 Cylinder<br>Miniplant®, 3.3 External Hex Cylinder, 3.75 ICE™ Self-tapping, 3.75 OSSEOTITE®, 3.75 Self-tapping Threaded, 3.75 Standard Threaded, 4.0 External Hex Cylinder, 4.0 ICE™ Self-tapping, 4.0<br>OSSEOTITE®, 4.0 Standard Threaded, 4.25 External Hex Cylinder, TG OSSEOTITE® (4.8<br>Platform), 4.0 OSSEOTITE® Certain™, 4.0 OSSEOTITE® NTCertain™, 4.0<br>OSSEOTITE®CERTAIN PREVAIL, 5.0 Osseotite® Certain, 5.0 Osseotite® NT Certain, 5.0<br>Osseotite®Certain Prevail | | |
| IMTEC Corporation® | 3.3 Universal Flare Cylinder, 3.75 Universal Self-tapping, 3.75 Universal Self-tapping Coated, 4.0<br>Spike Cylinder, 4.0 Universal Cylinder | | |
| Interpore IMZ™ | 3.3 Hex Cylinder, 3.75 Self-tapping Threaded, 4.0 Hex Cylinder, 4.0 Self-tapping Threaded, 4.25 Hex<br>Cylinder | | |
| Osstem | 4.1 US II, III, II Plus, III Plus, SS II, III (4.8 head) | | |
| Zimmer Dental | 3.5 Bio-Vent® X™, 3.75 Swede-Vent™ Conical Neck CST, 3.75 Swede-Vent™ Standard, 4.0<br>Swede-Vent™ Standard, 4.0 Bio-Vent® X™, 3.25 Micro-Vent® (3.5 head), 3.3 Screw-Vent® (3.5<br>head), 3.5 Bio-Vent® (3.5 head), 3.7 Screw-Vent® (3.5 head), 3.75 Screw-Vent® (3.5 head), 4.3<br>Core-Vent® (3.5 head), 4.25 Micro-Vent® (4.5 head), 4.5 Bio-Vent® (4.5 head), 4.7 Screw-Vent® (4.5<br>head), 5.3 Core-Vent® (4.5 head), 3.7 Tapered Swiss Plus™ (4.8 platform), 4.8 Tapered Swiss<br>Plus™ 4.1 Straight Swiss Plus™, 4.8 Straight Swiss Plus™ | | |
| Zimmer (Calcitek®, Centerpulse) | 3.75 ThreadLoc™ | | |
| Straumann | ITI TE™ 3.3 (4.8 head), ITI 3.3 Std & Std Plus (4.8 head), ITI TE™ 4.1 (4.8 head), ITI 4.1 Std. & Std.<br>Plus (4.8 head), ITI, 4.8 Std. & Std. Plus (4.8 head) | | |
| Biolok International | 4.5 Silhouette Screw, 4.0 Micro-Lok Screw, 4.0 Micro-Lok Cylinder, 3.75 Micro-Lok Screw, 3.3<br>Micro-Lok Cylinder | | |
| Bud | 3.25 Bud Screwvent, 3.75 Bud Screwvent | | |
| INNOVA | 4.1 ENDOPORE® External Connection, 4.0 ENTEGRATM External Connection | | |
| OIC | 3.0 Osteo Standard ST, 3.25 Osteo Standard ST, 3.75 Osteo Standard ST | | |
| MIS IMPLANTS | 3.3mm Internal Hex, 3.75mm Internal Hex, 4.20mm Internal Hex, 5.0mm Internal Hex | | |
| BioHorizons® | 3.5 Internal, 4.0 Internal, 4.5 Internal, 3.5 Single Stage, 4.5 Single Stage | | |
| Implant Direct | Legacy 3.5mm, Legacy 4.5mm, RePlant™ 4.3mm, RePlant™ 3.5mm, 3.7mm ScrewPlant, 4.7mm<br>ScrewPlant | | |
| Minimatic/Stryker | 3.3 External Hex Cylinder, 3.75 External Hex Screw, 4.0 External Hex Cylinder, 4.0 External Hex<br>Screw, 4.75 External Hex Screw, 5.0 External Hex Cylinder | | |
| Straumann | Straumann Bone Level RC, Blue Sky Bio Square Taper RC | | |
| Straumann | Straumann Bone Level NC, Blue Sky Bio Square Taper NC | | |
| Ankylos | Ankylos | | |
| Nobel Biocare | Nobel Replace WP, Nobel Replace Select WP, NobelSpeedy Replace WP, Implant Direct 5.0<br>RePlant, BlueSky Bio 5.0 Trilobe | | |
| Nobel Biocare | Nobel Conical Connection RP | | |
| Nobel Biocare | Nobel Conical Connection NP, Blue Sky Bio Max | | |
| Astra Dental | Astra 4.5 / 5.0, Blue Sky Bio Conus 12 4.5 / 5.0 | | |
| Astra Dental | Astra 3.5 / 4.0, Blue Sky Bio Conus 12 3.5 / 4.0 | | |
2 of 4
{2}------------------------------------------------
## Technical Characteristics:
.
The proposed implant abutments are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate devices.
| Attribute | Candidate | Predicate Device | Predicate Device |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The ORA Implant Abutment<br>Sterngold Dental, LLC | The O-Ring System - ORS<br>Attachments International,<br>Inc.<br>K900099 | SFI Implant Abutments<br>Sterngold Dental, LLC<br>K130183, K132814 |
| Design/Construction | Machined, screw-retained | Machined, screw-retained | Machined, screw-retained |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity |
| Device Material | Wrought Titanium-6AL-4<br>Vanadium ELI Alloy | Wrought Titanium-6AL-4<br>Vanadium ELI Alloy | Wrought Titanium-6AL-4<br>Vanadium ELI Alloy |
| Indications for Use | Indicated for use with dental<br>implants to support and/or retain<br>removable dental prostheses in<br>the treatment of partially or<br>totally edentulous patients to<br>restore chewing function. The<br>abutment screws directly into<br>endosseous implants. | The ORS Implant Abutments are<br>intended for use with dental<br>implants as a support or<br>attachment for prosthetic<br>restoration. The abutment screws<br>directly into the implant. | The SFI-Bar® Implant<br>Abutments are indicated for<br>use with dental implant<br>bodies/fixtures to support and<br>/or retain removable dental<br>prostheses for partially or<br>totally edentulous patients to<br>restore chewing function. |
| Operating Principle/<br>Basic Design | Abutment Implant connection:<br>Screw fixation | Abutment Implant connection:<br>Screw fixation | Abutment Implant connection:<br>Screw fixation |
| | Connecting principle to<br>overdenture: Retentive system | Connecting principle to<br>overdenture: Retentive system | Connecting principle to<br>overdenture: Retentive system |
| | Cleaning procedures for patient:<br>Common procedure for oral<br>hygiene | Cleaning procedures for patient:<br>Common procedure for oral<br>hygiene | Cleaning procedures for<br>patient: Common procedure<br>for oral hygiene |
| | Patient handling: Common<br>cleaning and insertion of denture | Patient handling: Common<br>cleaning and insertion of denture | Patient handling: Common<br>cleaning and insertion of<br>denture |
| Packaging,<br>materials and<br>processes | Produced in a controlled CNC<br>machine process, previously<br>validated<br>Packaging: Pouch<br>Non-sterile | Produced in a controlled CNC<br>machine process, previously<br>validated<br>Packaging: Pouch<br>Non-sterile | Produced in a controlled CNC<br>machine process, previously<br>validated<br>Packaging: Pouch<br>Non-sterile |
| Cuff Sizes | 0.4, 1.0, 1.25, 2.0, 3.0, 4.0,<br>5.0mm | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0mm | 1.0, 1.5,1.75, 2.0, 2.2, 2.5, 3.0,<br>3.5, 4.0, 4.5, 5.0, 5.5mm |
| Prosthetic<br>Connection | RP, Conical, NP, WP, 3.5 Head,<br>4.0 Head, 4.1 Head, 4.5 Head,<br>4.8 Head | RP, Conical, NP, WP, 3.5 Head,<br>4.0 Head, 4.1 Head, 4.5 Head,<br>4.8 Head | RP, Conical, NP, WP, 3.5<br>Head, 4.0 Head, 4.1 Head, 4.5<br>Head, 4.8 Head |
{3}------------------------------------------------
#### Performance Date:
Application and functional testing have been conducted to evaluate the performance characteristics of the ORA Implant Abutments. The test methods used were the same as in predicate devices. Testing has shown that the ORA Implant Abutments included in this application are equivalent in performance characteristics to its predicate devices.
#### Safety and Effectiveness / Conclusion:
Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. The non-clinical testing consisted of tolerance analysis of platforms to identify worst case test samples. Fatigue testing was not done as the basic design is the same as the predicate devices. The summary of technological characteristics as well as application and functional testing indicate that the device is substantially equivalent to the declared predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### April 16, 2014
Sterngold Dental, LLC Ms. Maria Rao Director, Quality and Regulatory Affairs 23 Frank Mossberg Drive Attleboro, MA 02703
Re: K133791
Trade/Device Name: ORA Implant Abutments System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: March 17, 2014 · Received: March 19, 2014
Dear Ms. Rao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Rao
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor YowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/7 description: The image shows the word "Erin" followed by a logo that is difficult to decipher. The logo appears to be a combination of letters and symbols, possibly representing a company or organization. The text and logo are in black and white, and the overall image has a simple, clean design. The last part of the image shows the letters "h-S".
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Sterngold Dental, LLC Abbreviated 510(k) Premarket Notification
April 14, 2014 ORA Implant Abutment System
| 510(k) Number (if known): | K133791 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: ORA Implant Abutment System
#### Indications for Use:
The ORA Implant Abutment System is indicated for use with dental implants to support and/or retain removable dental prostheses in the treatment of partially edentulous patients to restore chewing function. The abutment screws directly into endosseous implants or they screw into SFI Abutments which are screwed into endosseous implants.
| Implant Brand | Model |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nobel Biocare Brånemark System | 3.3 Fixture, 3.75 Fixture, 4.0 Fixture, 5.0 Fixture (Old Version), 3.75 MkII Self-tapping Fixture, 4.0 MkII Self-tapping Fixture |
| Sterngold-ImplaMed | 3.3 Hex Cylinder, 4.0 Hex Cylinder, 3.75 Standard Hex Screw, 3.75 Self-tapping Hex Screw, 3.75 Self-tapping "SST" Hex Screw, 4.0 Standard Hex Screw, 4.0 Self-tapping Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 4.0 Self-tapping "SST" Hex Screw, 5.0 RP "SST" Hex Screw, 3.75 RP Acid Etched, 4.0 RP Acid Etched, 5.0 RP Acid Etched, 4.1 Stern IC (4.8 head), 3.3 Stern IC (4.8 head) |
| Nobel Biocare (Steri-Oss®) | 3.8 HL Cylinder, 3.8 HL Threaded, 4.5 HL Threaded, 3.5 NobelReplace™, Replace® Select (NP), 4.0 NobelReplace Straight, (RP), 4.3Replace®Select&NobelReplace™ (RP) |
| Keystone (Lifecore) | 3.75 Restore® Self-tapping Screw, 4.0 Restore® Self-tapping Screw, 3.75 Restore® External Hex Screw, 4.0 Restore® External Hex Screw, 4.0 Restore® External Hex Cylinder, 4.2 Sustain® External Hex Cylinder, 3.75 Sustain® External Hex Screw, 4.0 Sustain® External Hex Screw. 4.2 Sustain® External Hex MC Cylinder. Stage-1TM (3.3 and 4.0 fixtures) |
| 3i Implant Inovations | 3.25 External Hex Miniplant®, 3.25 ICE™ Miniplant®, 3.25 OSSEOTITE® Miniplant®, 3.3 Cylinder Miniplant®, 3.3 External Hex Cylinder, 3.75 ICE™ Self-tapping, 3.75 OSSEOTITE®, 3.75 Self-tapping Threaded, 3.75 Standard Threaded, 4.0 External Hex Cylinder, 4.0 ICE™ Self-tapping, 4.0 OSSEOTITE®, 4.0 Standard Threaded, 4.25 External Hex Cylinder, TG OSSEOTITE® (4.8 Platform), 4.0 OSSEOTITE® Certain™, 4.0 OSSEOTITE® NTCertain™, 4.0 OSSEOTITE®CERTAIN PREVAIL, 5.0 Osseotite® Certain, 5.0 Osseotite® NT Certain, 5.0 Osseotite®Certain Prevail |
| IMTEC Corporation® | 3.3 Universal Flare Cylinder, 3.75 Universal Self-tapping, 3.75 Universal Self-tapping Coated, 4.0 Spike Cylinder, 4.0 Universal Cylinder |
| Interpore IMZ™ | 3.3 Hex Cylinder, 3.75 Self-tapping Threaded, 4.0 Hex Cylinder, 4.0 Self-tapping Threaded, 4.25 Hex Cylinder |
| Osstem | 4.1 US II, III, II Plus, III Plus, SS II, III (4.8 head) |
| Zimmer Dental | 3.5 Bio-Vent® X™, 3.75 Swede-Vent™ Conical Neck CST, 3.75 Swede-Vent™ Standard, 4.0 Swede-Vent™ Standard, 4.0 Bio-Vent® X™, 3.25 Micro-Vent® (3.5 head), 3.3 Screw-Vent® (3.5 head), 3.5 Bio-Vent® (3.5 head), 3.7 Screw-Vent® (3.5 head), 3.75 Screw-Vent® (3.5 head), 4.3 Core-Vent® (3.5 head), 4.25 Micro-Vent® (4.5 head), 4.5 Bio-Vent® (4.5 head), 4.7 Screw-Vent® (4.5 head), 5.3 Core-Vent® (4.5 head), 3.7 Tapered Swiss Plus™ (4.8 platform), 4.8 Tapered Swiss Plus™ 4.1 Straight Swiss Plus™, 4.8 Straight Swiss Plus™ |
| Zimmer (Calcitek®, Centerpulse) | 3.75 ThreadLoc™ |
| Straumann | ITI TE™ 3.3 (4.8 head), ITI 3.3 Std & Std Plus (4.8 head), ITI TE™ 4.1 (4.8 head), ITI 4.1 Std. & Std. Plus (4.8 head), ITI, 4.8 Std. & Std. Plus (4.8 head) |
| Biolok International | 4.5 Silhouette Screw, 4.0 Micro-Lok Screw, 4.0 Micro-Lok Cylinder, 3.75 Micro-Lok Screw, 3.3 Micro-Lok Cylinder |
| Bud | 3.25 Bud Screwvent, 3.75 Bud Screwvent |
| INNOVA | 4.1 ENDOPORE® External Connection, 4.0 ENTEGRATM External Connection |
| OIC | 3.0 Osteo Standard ST, 3.25 Osteo Standard ST, 3.75 Osteo Standard ST |
| MIS IMPLANTS | 3.3mm Internal Hex, 3.75mm Internal Hex, 4.20mm Internal Hex, 5.0mm Internal Hex |
| BioHorizons® | 3.5 Internal, 4.0 Internal, 4.5 Internal, 3.5 Single Stage, 4.5 Single Stage |
| Implant Direct | Legacy 3.5mm, Legacy 4.5mm, RePlant™ 4.3mm, RePlant™ 3.5mm, 3.7mm ScrewPlant, 4.7mm |
{7}------------------------------------------------
Sterngold Dental, LLC Abbreviated 510(k) Premarket Notification
| | ScrewPlant |
|-------------------|-------------------------------------------------------------------------------------------------|
| Minimatic/Stryker | 3.3 External Hex Cylinder, 3.75 External Hex Screw, 4.0 External Hex Cylinder, 4.0 External Hex |
| Minimatic/Stryker | Screw, 4.75 External Hex Screw, 5.0 External Hex Cylinder |
| Straumann | Straumann Bone Level RC, Blue Sky Bio Square Taper RC |
| Straumann | Straumann Bone Level NC, Blue Sky Bio Square Taper NC |
| Ankylos | Ankylos |
| Nobel Biocare | Nobel Replace WP, Nobel Replace Select WP, NobelSpeedy Replace WP, Implant Direct 5.0 |
| Nobel Biocare | RePlant, BlueSky Bio 5.0 Trilobe |
| Nobel Biocare | Nobel Conical Connection RP |
| Nobel Biocare | Nobel Conical Connection NP, Blue Sky Bio Max |
| Astra Dental | Astra 4.5 / 5.0, Blue Sky Bio Conus 12 4.5 / 5.0 |
| Astra Dental | Astra 3.5 / 4.0, Blue Sky Bio Conus 12 3.5 / 4.0 |
| Zimmer Dental | Zimmer TSV 5.7mm, Implant Direct Legacy 5.7, BioHorizon 5.7 |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Part 21 CFR 801 Subparts D) AND/OR
· Over-the -Counter Use (21 CFR 807 Subpart D)
Sheena A. Green-S
2014.04.16 11:34:03 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.