← Product Code [JDR](/productcode/JDR) · K133780

# SPEED XL (K133780)

_Biomedical Ent., Inc. · JDR · Mar 19, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133780

## Device Facts

- **Applicant:** Biomedical Ent., Inc.
- **Product Code:** [JDR](/productcode/JDR.md)
- **Decision Date:** Mar 19, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Speed XLTM is indicated for: Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.

## Device Story

Speed XL is a metallic bone staple used for fracture/osteotomy fixation and joint arthrodesis in hand and foot procedures. Delivered to the OR in an 'open' martensitic state, the implant undergoes a phase transformation to a 'closed' austenitic state at room temperature upon insertion, providing compression without external heating. The device features a flattened bridge with a wider dimension compared to the predicate. It is intended for use by surgeons in a clinical setting. The staple provides mechanical fixation to stabilize bone segments, potentially improving healing outcomes for patients undergoing orthopedic procedures.

## Clinical Evidence

Bench testing only. Pull-out strength and bending stiffness compared to predicate using ASTM F564-10; corrosion resistance evaluated per ASTM F2129-08. Results confirm performance equivalent to predicate.

## Technological Characteristics

Nitinol (ASTM F2063-12); shape-memory alloy; room-temperature phase transformation (martensitic to austenitic); dimensions: 20-25mm bridge length, 20mm leg length; surface finish: mechanical tumbling, acid cleaning, chemical passivation; sterile packaged; non-software device.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Speed Shift ([K124022](/device/K124022.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# (510(k) Summary)

MAR 1 9 2014

#### Speed XLTM Product:

Submitter Information

BioMedical Enterprises, Inc. 14785 Omicron Drive, Ste. 205 San Antonio, Texas 78245 Telephone: (210) 677-0354 (210) 677-0355 Fax: Joe W. Soward Contact:

Date Prepared: December 3, 2013

Classification name: Single/multiple component metallic bone fixation appliances and accessories (21 CFR 888.3030)

| Classification:    | Class II    |
|--------------------|-------------|
| Product Code:      | JDR         |
| Common/Usual Name: | Bone Staple |
| Proprietary Name:  | Speed XLTM  |

#### Intended Use:

The Speed XLTM is indicated for: Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.

# Substantial Equivalence:

The Speed XL™ is substantially equivalent to the predicate Speed Shift™ cleared in K124022.

# Device Description

The Speed XLTM is for use in fracture and osteotomy fixation and joint arthrodesis of the hand and foot. The implant is designed so that it is delivered to the operating room in an "open" martensitic state. The implant is then transformed by room temperature after insertion, and contracts to a "closed" austenitic state. The implant does not require any external heating; it is completely transformed by room temperature.

This configuration change for the Speed XL™ consists of a change in device shape, including flattened bridge with wider dimension to be used in hand and foot procedures.

Technological Characteristics Comparison to the Predicates

{1}------------------------------------------------

| Product Name:                  | New Device<br>Speed XLTM                                         | Predicate Speed ShiftTM<br>(K124022)                                |
|--------------------------------|------------------------------------------------------------------|---------------------------------------------------------------------|
| Raw Material:                  | Nitinol, per ASTM<br>F2063-12                                    | Nitinol, per ASTM<br>F2063-12                                       |
| Bridge Lengths (mm):           | 20 and 25                                                        | 15 and 20                                                           |
| Leg Lengths (mm):              | 20                                                               | 20                                                                  |
| Cross-section Dimensions (mm): | Leg: 1.80 x 2.06<br>Bridge: 1.85 x 3.52                          | 1.80 x 2.00                                                         |
| Barbs:                         | Barbs on the legs                                                | Barbs on the legs                                                   |
| Pilot Drill Hole Sizes (mm):   | 2.65                                                             | 2.65                                                                |
| Heat Source:                   | Fully transformed<br>at room temperature                         | Fully transformed<br>at room temperature                            |
| Surface Finish:                | Mechanical tumbling, acid cleaning, and<br>chemical passivation. | Mechanical tumbling,<br>acid cleaning, and<br>chemical passivation. |
| Storage:                       | Sterile packaged and can be stored at<br>room temp until used.   | Sterile packaged and<br>can be stored at room<br>temp until used.   |

# Performance Bench Testing:

Standard ASTM F564-10 (2010) was used to compare the pull-out strength of the new Speed XLTM to the predicate Shift™. Two lots with three samples each (n=3) were compared to similarly sized Speed Shift™ predicates. The results for the Speed XLTM yielded substantially equivalent pull-out resistance than the comparable predicate Speed Shift™ for similar sizes. (TR-02-129-1316)

Standard ASTM F564-10 (2010) was used to compare the mechanical strength of the Speed XL™ to the Speed Shift™. The largest and smallest Speed XL™ sizes were tested and compared to similar Speed Shift™ predicates. The Speed XLTM devices yielded substantially equivalent bending stiffness when compared to the Speed Shift™ predicates. (TR-01-129-1315)

Standard ASTM F2129-08 was used to compare the corrosion resistance of representative samples of the new Speed XLIM to the predicate Speed Shift™. Test results demonstrate the corrosion resistance is equivalent to the predicate Speed Shift™. (FWM # 13-08-30946, FWM # 13-08-31279, & FWM # 13-10-31866)

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines that suggest wings and a tail.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 19, 2014

BioMedical Enterprises, Incorporated Mr. Joe W. Soward Director, Quality, Compliance and Regulatory Affairs 14785 Omicron Drive, Suite 205 San Antonio, Texas 78245

Re: K133780

Trade/Device Name: Speed XLTM Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR Dated: January 22, 2014 Received: January 23, 2014

Dear Mr. Soward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set

{3}------------------------------------------------

# Page 2 - Mr. Joe W. Soward

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Ronald P. Jean -S for

- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{4}------------------------------------------------

Device Name: Speed XLTM

# Indications for Use

The Speed XL™ is indicated for:

Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

Prescription Use V (Per 21 CFR 801.109)

ાર

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Casey L. Hanley, Ph.D.

---

Division of Orthopedic Devices

---

**Source:** [https://fda-staging.innolitics.com/device/K133780](https://fda-staging.innolitics.com/device/K133780)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
