SAUFLON CLARITI WITH TINT, SAUFLON CLARITI TORIC WITH TINT, SAUFLON CLARITI MULTIFOCAL WITH TINT
Device Facts
| Record ID | K133740 |
|---|---|
| Device Name | SAUFLON CLARITI WITH TINT, SAUFLON CLARITI TORIC WITH TINT, SAUFLON CLARITI MULTIFOCAL WITH TINT |
| Applicant | Sauflon Pharmaceuticals, Ltd. |
| Product Code | LPL · Ophthalmic |
| Decision Date | Mar 5, 2014 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.5925 |
| Device Class | Class 2 |
Indications for Use
The SAUFLON CLARITI (somofilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity. The SAUFLON CLARITI TORIC (somofile) Contact Leas with UV blocker and handling tint is indicated for daily woar for the optical corrective anteropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters. The SAUFLON CLARITI MULTIFOCAL (somofileon A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive anetropia (myopia and or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +4.00 Diopters or less. The SAUFLON CLARITI MULTIFOCAL TORIC (somofilion A) Soft (hydrophilic) Contact Lens with UV blocker and handling tint is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and that may require a reading addition of +4.00 Diopters or less. The lenses may be prescribed for daily wear for up to one month with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion as recommended by the eye care professional. Saullon Claili (somofilcon A) Soll (hydrophilic) Contact lens with UV blocker and handling tint helps protect against transmission of harmful UV radiation to the cornea and into the eye. The lens has a itint added to make the lens more visible for handling.
Device Story
Soft contact lens (somofilcon A) for daily wear; corrects refractive ametropia (myopia, hyperopia, astigmatism, presbyopia). Lens material is hydrophilic silicone hydrogel copolymer with benzophenone UV blocker and D&C Green 6 handling tint. Available as single vision, toric, and multifocal designs. Prescribed by eye care professionals; patient performs daily removal, cleaning, and chemical disinfection. Lens covers cornea and adjacent sclera. Provides vision correction and UV radiation protection. Substantial equivalence established via physical, chemical, and clinical comparison to predicate lenses.
Clinical Evidence
Clinical study compared subject device to Air Optix Aqua (Ciba Vision Inc.) using OptiFree Replenish solution. Evaluated safety and efficacy for daily wear. Results confirmed safety, acceptability, and substantial equivalence to predicate device.
Technological Characteristics
Material: somofilcon A (silicone hydrogel copolymer cross-linked with Alkyl bis methacrylate). Water content: 56%. Oxygen permeability (Dk): 60 (Fatt method). Refractive index: 1.4008. UV blocker: benzophenone. Tint: D&C Green 6. Dimensions: 13.0-18.0mm diameter, 7.5-9.5mm base curve. Power range: -30.00 to +30.00 DS. Manufacturing: cast molding. Packaging: borate buffered saline with 0.005% poloxamer 407 in hermetically sealed blister packs.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, and/or presbyopia in phakic or aphakic persons with non-diseased eyes. Supports astigmatism correction up to 10.00 Diopters and reading additions up to +4.00 Diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
- Sauflon Clariti (Somofilcon A) Soft (Hydrophilic) Lens for Daily Wear (K130342)
- Air Optix Aqua (Lotrafilcon B) Soft (Hydrophilic) Visibility Tinted Contact Lens (K033919/K073459)