The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (Hydrophilic) Contact lenses are indicated for the correction of ametropia (myopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from 0.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Eye Care professionals may prescribe the lens for daily disposable or daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.
Device Story
Soft (hydrophilic) contact lens; manufactured via cast molding; material methafilcon A (HEMA/MAA copolymer crosslinked with EGDMA). Incorporates UV-absorbing monomer and C.I. Reactive Blue No. 4 tint. Designed for daily wear; prescribed by eye care professionals for ametropia correction. Patient wears lens directly on eye; requires chemical or hydrogen peroxide disinfection for planned replacement. Provides vision correction; UV protection; visibility tint.
Clinical Evidence
No clinical data. Safety and effectiveness established via bench testing including acute systemic injection, ocular irritation, cytotoxicity (ISO 10993-5), extractables, light transmittance, refractive index, water content, oxygen permeability, pH/osmolality, and mechanical comparative testing.
Technological Characteristics
Material: Methafilcon A (HEMA/MAA copolymer crosslinked with EGDMA). Manufacturing: Cast molded. Water content: 55%. Dk: 22.0 x 10^-11. UV blocking: <5% UVB, <50% UVA. Tint: C.I. Reactive Blue No. 4. Aspheric design.
Indications for Use
Indicated for correction of myopia in aphakic and non-aphakic persons with non-diseased eyes; powers 0.00 to -12.00D; astigmatism 2.00D or less. For daily wear in a frequent replacement program.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
SAUFLON 55 UV (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear (K013649)
Submission Summary (Full Text)
{0}------------------------------------------------
## MAR 2 1 2014
## 510(k) Summary
| 1 | Type of Submission: | Traditional |
|---|------------------------------------|------------------------------------------------------------------------|
| 2 | Orginal Submission<br>Date: | 2013/11/29 |
| 3 | Submitter: | Yung Sheng Optical Co., Ltd. |
| | Address: | 3F-1, No.6, Jhongke Road, Daya District, Taichung City<br>42881 Taiwan |
| | Manufacturer Address | No.8, Keya 2nd Road, Daya District, Taichung City 42881,<br>Taiwan |
| | Phone: | +886-4-25658384 #156 |
| | Fax: | +886-4-25658387 |
| | Contact: | Wen-Han Chen |
| | Establishment Registration Number: | N/A |
### 4 Identification of the Device:
| Proprietary/Trade name: | Eye Secret 55 UV Aspheric (methafilcon A) Soft |
|--------------------------|------------------------------------------------|
| | (hydrophilic) Contact Lens |
| Common Name: | Contact Lens |
| Classification Name: | Soft (Hydrophilic) Contact Lens (daily wear) |
| Device Classification: | II |
| Regulation Number: | 886.5925 |
| Panel: | Ophthalmic |
| Product Code: | LPL |
| Subsequent Product Code: | MVN |
### 5 Identification of the Predicate Device:
| Predicate Device Name: | SAUFLON 55 UV (methafilcon A) Soft |
|------------------------|-----------------------------------------|
| | (Hydrophilic) Visibility Tinted Contact |
| | Lens for Daily Wear |
| Manufacturer: | Sauflon Pharmaceuticals Ltd. |
| 510(k) Number: | K013649 |
{1}------------------------------------------------
#### Intended Use and Indications for Use of the subject device. 6
The Eye Secret 55 UV Aspheric (Methafilcon A) Soft (Hydrophilic) Contact lenses are indicated for the correction of ametropia (myopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from 0.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Eye Care professionals may prescribe the lens for daily disposable or daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.
## 7 Device Description
The Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) contact lens is manufactured by using advanced cast molding method. The lens material, methafilcon A, is a random copolymer of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA) crosslinked with ethylene glycol dimethacrylate (EGDMA). A UV absorbing monomer is used to block UV radiation. The average transmittance characteristics are less than 5% in the UVB range of 280~315 nm and less than 50% in the UVA range of 316 to 380 nm. The lenses are tinted blue for visibility purposes with the color additives, C.I. Reactive Blue No.4. The Lenses are available as aspheric lenses.
### 8 Summary of Clinical Study
Methafilcon A lenses have been used widely. Its safety and effectiveness have been well documented. Their safety and effectiveness can be further exemplified by the lenses cleared by FDA.
SAUFLON 55 UV (Methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear. K013649 submitted by Sauflon Pharmaceuticals Ltd.
Clinical study for Eye Secret 55 UV Aspheric (Methafilcon A) Soft (Hydrophilic) Contact Lens is not required for the premarket notification as the USAN name and process are the same as above mentioned predicated device.
{2}------------------------------------------------
## 9 Non-clinical Testing
A series of preclinical testing were performed to demonstrate the safety and effectiveness of the Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) contact lens. The results of all testing demonstrated that the safety and effectiveness of the Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) contact lens are equivalent to the SAUFLON 55 UV (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear (K013649). The following tests were conducted as recommended by the FDA Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, February 27, 1997:
- . Toxicity
- 1. Acute Systemic Injection Study: The lens material meets the requirements of the systemic injection test and is considered non-toxic.
- 2. White Rabbit Ocular Irritation Test: Ocular irritation test was performed and produced no ocular irritation.
- 3. Cytotoxicity Test: The test article meets the requirements of ISO 10993-5.
- . Extractables (Leachability)
- Finished Lens Parameters .
- . Light Transmittance
- Refractive Index .
- Water Content .
- Oxygen Permeability .
- pH and Osmolality Test .
- . Shelf-life
- Mechanical Comparative Testing
In addition to biocompatible tests, shelf-life, physical compatibility test of contact lens care solution, mechanical comparative testing and the comparative tests of the performance tests mentioned above are provided.
The results of the non-clinical testing demonstrate that the Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) contact lens is substantially equivalent to the predicate devices.
#### 10 Substantial Equivalence Determination
The Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) contact lens submitted in
{3}------------------------------------------------
this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared SAUFLON 55 UV (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear which is the subject of K013649. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.
| Item | | |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Number | Eye Secret 55 UV Aspheric<br>(methafilcon A) Soft (hydrophilic)<br>contact lens | Predicate Device (K013649)<br>SAUFLON 55 UV (methafilcon A)<br>Soft (Hydrophilic) Visibility Tinted<br>Contact Lens for Daily Wear |
| Classification | 886.5925 | 886.5925 |
| Intended Use | II<br>The Eye Secret 55 UV Aspheric<br>(Methafilcon A) Soft (Hydrophilic)<br>Contact lenses are indicated for the<br>correction of ametropia (myopia) in<br>aphakic and non-aphakic persons<br>with non-diseased eyes in powers<br>from 0.00 to -12.00 diopters. The<br>lenses may be worn by persons who<br>exhibit astigmatism of 2.00 diopters<br>or less that does not interfere with<br>visual acuity.<br>Eye Care professionals may<br>prescribe the lens for daily<br>disposable or daily wear in a<br>Frequent Replacement Program. As<br>prescribed for planned replacement,<br>the lens should be disinfected using a<br>chemical or hydrogen peroxide<br>disinfecting systems. | II<br>The Sauflon 55 UV soft<br>(hydrophilic) contact lens is<br>indicated for daily wear for the<br>correction of the refractive<br>ametropia (myopia and hyperopia)<br>and astigmatism in aphakic and<br>not-aphakic persons with<br>non-diseased eyes that may exhibit<br>astigmatism up to 2.00 Dioptres that<br>does not interfere with visual<br>acuity. Eyecare Practitioners may<br>prescribe the lens for daily wear in a<br>Frequent Replacement Program.<br>The lenses may be disinfected using<br>chemical or hydrogen peroxide<br>disinfection systems. |
| Prescription Use | Yes | Yes |
| Material | Methafilcon A | Methafilcon A |
| Manufacturing<br>Method | Cast Molded | Cast Molded |
| Water Content | 55% | 55% |
| Dk | 22.0 x 10-11 (cm²/sec) (mlO₂/ml x<br>mmHg) | 22.0 x 10-11 (cm²/sec) (mlO₂/ml x<br>mmHg) |
| Powers | 0 ~ -12.00 D | ± 20.00D |
| Light Transmittance | 95% ± 5% | 94.61% |
| UV-A | < 50% | 36.00% |
{4}------------------------------------------------
| UV-B | < 5% | 9.41% |
|------------------|-----------------------|---------------------|
| Refractive Index | 1.409 | 1.402 |
| Base Curve | 8.4 mm ~ 8.8 mm | 8.40mm ~ 9.30mm |
| Diameter | 14.0 mm ~ 14.4 mm | 14.0mm ~ 15.0mm |
| Tint | C.I. Reactive Blue #4 | Reactive Blue No. 4 |
## 11 Conclusion
After analyzing bench tests, safety testing data, it can be concluded that Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) contact lens is substantially equivalent to the predicate device.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with outstretched arms, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the figure. The logo is black and white.
March 21, 2014
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Mr. Wen-Han Chen, Director of Regulatory Affairs Division Yung Sheng Optical Co., 3F-1, NO. 6, Jhongke Road, Daya District, Taichung City 42881, Taiwan
Re: K133735
> Trade/Device Name: Eye Secret 55 UV Aspheric (methafilcon A) Soft (hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: December 23, 2013 Received: December 27, 2013
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
Page 2 - Mr. Wen-Han Chen
You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled,"Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
## 510(k) Number (if known) K133735
## Device Name
1
he Eye Secret 55 UV Aspheric (Methafilcon A) Soft (Hydrophilic) Contact lenses
### Indications for Use (Describe)
The Eye Secret SS UV Aspheric (Methafilcon A) Soft (Hydrophilic) Contact lenses are indicated for the correction of ametropia (myopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from 0.00 to -12.00 diopters. The lenses may be wom by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
Eve Care professionals may prescribe the lens for daily wear in a Frequent Replacement Program. As prescribed for planned replacement, the lens should be disinfected using a chemical or hydrogen peroxide disinfecting systems.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
. . . . . . . . . . . . . . . . . . . . . . . . ' FOR FDA USE ONLY ... :
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Leonid Livshitz -S 2014.02.19 10:14:43 -05'00'
FORM FDA 3881 (1/14)
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.