# SEAL SINGLE-USE BIOPSY VALVE (K133734)

_Endochoice, Inc. · OCX · Mar 19, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133734

## Device Facts

- **Applicant:** Endochoice, Inc.
- **Product Code:** [OCX](/productcode/OCX.md)
- **Decision Date:** Mar 19, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure, and provide access for irrigation.

## Device Story

Single-use biopsy valve provides sealable access to endoscope working channel ports. Device consists of cylindrical base, connected cap, and flexible diaphragm seal. During endoscopic procedures, surgical instruments are inserted through cap and diaphragm; elastomer deforms around instrument to create leak-free seal. Cap allows pressure modification during insufflation by venting gases. Used in clinical settings by physicians/endoscopy staff. Benefits include maintained insufflation and reduced leakage of biomaterials during instrument exchange.

## Clinical Evidence

Bench testing only. Performance testing included 1-minute pressure tests at 10 PSI with device inserted and after device removal. Biocompatibility testing conducted for cytotoxicity, sensitization, intracutaneous injection, and systemic injection per 21 CFR Part 58.

## Technological Characteristics

Thermoplastic elastomer construction. Cylindrical form factor with pre-perforated cap and internal diaphragm seal. Dimensions: 7.1 mm inner diameter. Compatible with Olympus (160, 180, 190), Fujinon (530, 590, 600), and EndoChoice Fuse endoscopes. Non-sterile, single-use.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Biopsy Valve ([K111821](/device/K111821.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## K133734 Page1of4

#### 510(k) Summary

#### Seal Single-Use Biopsy Valve

#### December 5, 2013

### 1. Company Identification

EndoChoice, Inc. 11810 Wills Road Alpharetta, GA 30009 Telephone: 678-708-4773 Fax: 678-567-8218 Establishment Registration: 300759133

#### Contact Person 2.

Janna Babson Regulatory Affairs Associate

#### 3. Device Name

| Commercial Name:     | Seal Single-Use Biopsy Valve          |
|----------------------|---------------------------------------|
| Common/Usual Name:   | Biopsy Valve                          |
| Classification Name: | Endoscopic, irrigation/suction system |

#### 4. Device Classification

| Product Code:      | OCX                      |
|--------------------|--------------------------|
| Regulation Number: | 876.1500                 |
| Class:             | 11                       |
| Review Panel:      | Gastroenterology/Urology |

#### 5. Intended Use:

The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure, and provide access for irrigation.

{1}------------------------------------------------

#### 6. Device Description:

The Seal Single-Use Biopsy valve is provided non-sterile, and is composed of a cylindrical base and a connected cap. The Single-Use Biopsy Valve provides surgical instruments with sealable access to the working channel port of an endoscope. The valve consists of an attached cap and valve body with a circular opening. The cap is pre-perforated, and ensures instrument access and removal is tight and leak-free.

The valve is designed with a hollow body with a distal end that releasably attaches to the inlet port of the working channel. The valve body provides a flexible diaphragm seal that separates the body into a proximal chamber and distal chamber. The distal chamber secures onto the endoscope, while the proximal chamber secures the cap. Instruments are inserted through the cap and chambers, through the diaphragm seal, and into the endoscope. The diaphragm seal Is configured such that a seal is formed due to deformation of the elastomer around an instrument when it is inserted through the opening, preventing fluids from passing through the biopsy valve.

The cap serves two primary functions. One function of the cap is to provide an additional seal around surgical instruments to prevent leakage. The second function is to allow modification of pressure during insufflation by opening the cap, allowing gasses to flow from the endoscope, and out of the biopsy valve.

#### 7. Substantial Equivalence:

The device submitted for review is a modification of the Biopsy Valve (K111821). Changes to the device include a modification in material, and a change in dimensions. Both the modified and unmodified device is composed of thermoplastic elastomers. The modified device also has a reduction in overall height of the biopsy valve to provide an improved fit between the cap and body, as well as the valve to the instrument port.

The modified device is identical in terms of intended use, operating principle, performance, technology, energy used, and packaging.

| Characteristic      | Single-Use Biopsy Valve<br>(Unmodified)                                                                                                                                                                                                                             | Single-Use Biopsy Valve<br>(Modified)                                                                                                                                                                                                                               |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number       | K111821                                                                                                                                                                                                                                                             | Pending                                                                                                                                                                                                                                                             |
| Indications for Use | Biopsy valves are intended to<br>provide access for endoscopic<br>device passage and exchange,<br>helps maintain sufflation,<br>minimizes leakage of biomaterial<br>from biopsy port throughout the<br>endoscopic procedure, and<br>provides access for irrigation. | Biopsy valves are intended to<br>provide access for endoscopic<br>device passage and exchange,<br>helps maintain sufflation,<br>minimizes leakage of biomaterial<br>from biopsy port throughout the<br>endoscopic procedure, and<br>provides access for irrigation. |

{2}------------------------------------------------

| Characteristic                                                                            | Single-Use Biopsy Valve<br>(Unmodified) | Single-Use Biopsy Valve<br>(Modified)                                                                                             |
|-------------------------------------------------------------------------------------------|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Material                                                                                  | Thermoplastic elastomer                 | Thermoplastic elastomer                                                                                                           |
| Valve Inner Diameter                                                                      | 7 mm                                    | 7.1 mm                                                                                                                            |
| Endoscope<br>compatibility                                                                | Olympus<br>Fujinon<br>EndoChoice Fuse   | Olympus series 160, 180, and 190<br>endoscopes.<br>Fujinon series 530, 590, and 600<br>endoscopes.<br>EndoChoice Fuse endoscopes. |
| Removable cap<br>with slit                                                                | Yes                                     | Yes                                                                                                                               |
| Slit accommodate<br>devices                                                               | Up to 3.2 mm                            | Up to 3.2 mm                                                                                                                      |
| Packaging                                                                                 | Individually packaged plastic pouch     | Individually packaged plastic pouch                                                                                               |
| Performance testing:                                                                      |                                         |                                                                                                                                   |
| • 1 minute<br>pressure test at<br>10 PSI – with<br>device in biopsy<br>valve              | Pass                                    | Pass                                                                                                                              |
| • 1 minute<br>pressure test at<br>10 PSI – with<br>device in biopsy<br>valve              | Pass                                    | Pass                                                                                                                              |
| • 1 minute<br>pressure test at<br>10 PSI after<br>removing<br>device from<br>biopsy valve | Pass                                    | Pass                                                                                                                              |

### 8. Non-Clinical Testing:

The modified device has undergone both bench testing of performance and laboratory biocompatibility testing for cytotoxicity, sensitization, intracutaneous injection test, and system injection test, in accordance with 21 CFR, Part 58. The modified Single-Use Biopsy Valve results in no safety or efficacy concerns regarding biocompatibility or performance. Likewise, in conformance with 21 CFR 807.92(b)(3), the modified device performs as well as the predicate in all testing performed.

{3}------------------------------------------------

.

### 9. Conclusion:

The modified Single-Use Biopsy Valve is substantially equivalent to the unmodified predicate device listed above in performance, technical characteristics, biocompatibility, and intended use.

.

·

:

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or supporting another, with flowing lines suggesting movement or support.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 19, 2014

EndoChoice, Inc. Janna Babson Regulatory Affairs Associate 11810 Wills Road Alpharetta, GA 30009

Re: K133734

> Trade/Device Name: Biopsy Valve Regulation Number: 21 CFR\$ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCX Dated: February 21, 2014 Received: February 24, 2014

Dear Janna Babson,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed prodicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Janna Babson

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K133734

Device Name Single-Use Biopsy Valve

Indications for Use (Describe)

The Single-Use Biopsy Valve provides access for endoscopic device passage and exchange, helps maintain sufflation, and minimizes leakage of biomaterial from biopsy port throughout the endoscopic procedure, and provides access for irrigation.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

THE THE SEOR FOR FOR FOR FOR FOR MEDIA MODEL HOW THE CONTRACT ્યુટિક ﺔ ﺍﻟﻘﺎﻧﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/13 description: The image shows the name "Benjamin R. Fisher" in bold, black font. Below the name is the date "2014.03.19" followed by the time "09:00:05". There is a logo to the right of the name and time. The logo is illegible.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K133734](https://fda-staging.innolitics.com/device/K133734)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com