The Traxcess 14 Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in coronary arteries.
Device Story
Intravascular guidewire; 0.014" stainless steel shaft; tapered nitinol tip; platinum/stainless steel coils; hydrophilic coating. Used in neuro and peripheral vasculature; operated by physicians to steer and place diagnostic or therapeutic catheters. Provides mechanical support and navigation through vasculature; facilitates catheter access. Benefits patient by enabling minimally invasive access to target vessels for diagnosis or therapy.
Clinical Evidence
Bench testing only. Testing included physical/dimensional attributes, tensile strength, tip flexibility, torque strength, coating adherence, torqueability, flexing/fracture resistance, radiopacity, tip shapeability, in-vitro simulated use, attachment compatibility, and particulate testing. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, sensitization, irritation, hemocompatibility, systemic toxicity, pyrogenicity). All results met specifications and were equivalent or superior to the predicate.
Technological Characteristics
Materials: stainless steel shaft, nitinol tip, platinum/stainless steel coils. Coating: hydrophilic [SLIP-COAT]. Dimensions: 0.014" proximal diameter, 0.012" distal diameter, 200 cm length. Mechanical navigation device; no energy source or software.
Indications for Use
Indicated for general intravascular use in neuro and peripheral vasculature to facilitate selective placement of diagnostic or therapeutic catheters. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
510(k) Summary
JUL 3 1 2014
K133725
| Trade Name: | Traxcess 14 Guidewire |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Generic Name: | Guidewire |
| Classification:<br>Product Code:<br>Submitted By: | Class II, 21 CFR 870.1330<br>MOF, DQX<br>MicroVention, Inc<br>1311 Valencia Avenue<br>Tustin, California U.S.A.<br>Ph: 714-247-8000<br>Fax: 714-247-8005 |
| Contact: | Naomi Gong |
| Date: | December 4, 2013 |
## Predicate Device:
| Number | Description | Clearance Date |
|---------|------------------------------------------|----------------|
| K080863 | Traxcess 0.014" Hydrophilic<br>Guidewire | April 7, 2008 |
## Device Description:
The Traxcess 14 Guidewire consists of a 0.014" stainless steel shaft and a tapered nitinol tip contained within 0.012" platinum and stainless steel coils. The guidewire is coated with a lubricious hydrophilic coating.
### Indication For Use:
The Traxcess 14 Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in coronary arteries.
{1}------------------------------------------------
# Technological Comparison:
.
| | Predicate | Traxcess 14 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Intended Use | The Traxcess 14 Guidewire is intended<br>for general intravascular use, including<br>the neuro and peripheral vasculature. The<br>guidewire can be steered to facilitate the<br>selective placement of diagnostic or<br>therapeutic catheters. This device is not<br>intended for use in coronary arteries. | Same |
| Diameter | Proximal = 0.014"<br>Distal = 0.012" | Same |
| Overall Length | 200 cm | Same |
| Distal Shaft Length<br>(Shapeable Length) | 1.4 cm | Same |
| Material | Core wire (proximal): Stainless steel<br>Core wire (distal): Nickel titanium alloy<br>Coil: Stainless steel and Platinum nickel<br>alloy<br>Other: Brazing material and solder | Same |
| Coating | Shaft: PTFE/silicone<br>Coil: Hydrophilic Coating<br>(dimethylacramide-glycidyl methacrylate<br>copolymer) | Shaft and Coil: Hydrophilic Coating<br>[SLIP-COAT] |
| Coating Length | Hydrophilic coating (distal) = 400 mm<br>PTFE/silicone (proximal) ~1600 mm | Hydrophilic Coating = 1450 mm |
| Coil Length | 40 cm | Same |
| Radiopaque Length | 3 cm | Same |
| Distal tip thickness<br>(core wire) | 0.037 mm | Same |
| Proximal end<br>configuration | Tapered to be compatible with Traxcess<br>docking wire | Same |
・
:
# Verification and Test Summary Table:
| Bench Testing | Result | Conclusion |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Physical and dimensional attributes | Test articles met specified<br>dimensional requirements for<br>OD, overall length, length of<br>coil section, length of SS<br>section, length of docking<br>section. | Device met established<br>dimensional and physical<br>specifications. |
| Tensile strength (distal and proximal) | Tensile strength of distal ≥ 3N<br>and of proximal ≥ 70N | Device met established tensile<br>strength |
| Tip flexibility | Demonstrated force to deflect<br>the distal tip is ≤ predicate | Device tip flexibility is ≤<br>predicate |
{2}------------------------------------------------
| Bench Testing | Result | Conclusion |
|--------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Torque strength | Equivalent or better than<br>predicate (turns to failure) | Equivalent to predicate |
| Coating adherence | Coating adherence maintained<br>after advance/retract cycles | Durability and lubricity of<br>coating was comparable to<br>predicate |
| Torqueability | Torque response was equal to<br>or better than predicate | Torque response was equal to<br>or better than predicate |
| Flexing and fracture resistance test | No damage to test articles or<br>coating after flexing and<br>fracture test | Device able to withstand<br>forces and stress during<br>navigation in vasculature. |
| Radiopacity | Distal coil section visible<br>under fluoroscopy | Device radiopacity equivalent<br>or better than predicate |
| Tip shapeability | Test articles able to maintain<br>tip shape | Device tip shapeability<br>equivalent or better than<br>predicate |
| In-vitro simulated use testing | Test articles achieved rating ≥<br>3 for prep of device,<br>introduction, and tracking | Device performed as intended<br>under simulated use. |
| Attachment with docking wire | Pull force of attachment was<br>equivalent or better than<br>predicate | Device is compatible with<br>docking wire |
| Particulate Testing | Particle count of test articles ≤<br>25 particles (≥ 10 microns)<br>and ≤ 3 particles (≥ 25<br>microns) | Device does not generate<br>particles under use. |
| Biocompatibility | Result | Conclusion |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Cytotoxicity - MEM Elution Test (ISO<br>10993-5) | Cell culture exhibited slight<br>reactivity (Grade 1) | Non-cytotoxic |
| Sensitization/Irritation - Kligman<br>Maximization Test (ISO10993-10) | Extracts of test article elicited<br>no reaction at challenge (0%)<br>following induction phase.<br>(Grade 1) | Weak allergenic potential |
| Sensitization/Irritation - Intracutaneous<br>Injection Test (ISO 10993-10) | Extracts of test article did not<br>show a significantly greater<br>biological reaction than sites<br>injected with the control | Non-irritant |
| Hemocompatibility - Hemolysis (Direct<br>and Indirect) (ISO 10993-4) | Hemolysis index was 4.69%<br>(direct) and 0.0% (indirect) | Non-hemolytic |
| Hemocompatibility - Unactivated<br>Partial Thromboplastin Assay (ISO<br>10993-4) | No statistically significant<br>difference found between<br>plasma exposed to test article,<br>negative control, and untreated<br>control. | No effect on coagulation |
| Hemocompatibility - Complement<br>Activation (ISO 10993-4) | C3a and SC5b-9 in plasma<br>exposed to test article was not<br>statistically higher than the<br>plasma exposed to negative<br>and untreated controls. | No effect on complement<br>system |
| Hemocompatibility - Dog<br>Thrombogenicity (ISO 10993-4) | Minimal thrombosis for test<br>article and control sites (Grade<br>1-2) | No significant thrombosis |
and the contraction of the comments of the comments of
{3}------------------------------------------------
| Biocompatibility | Result | Conclusion |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Systemic toxicity - Systemic Injection<br>Test (ISO 10993-11) | Extracts of test article did not<br>induce a significantly greater<br>biological reaction than the<br>control extracts when injected<br>into albino mice. | No toxic effects |
| Systemic toxicity - Rabbit Pyrogen Test<br>(ISO 10993-11) | Temperature increase was 0.0°<br>C from baseline. | Non-pyrogenic |
## Summary of Substantial Equivalence
The data presented in this submission demonstrates the technological similarity and equivalency of the Traxcess 14 Guidewire when compared with the predicate device.
1 - 1 - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
The devices,
- . Have the same intended use,
- 트 Use the same operating principle,
- Incorporate the same basic design,
- . Use similar construction and material,
- Are packaged and sterilized using same material and processes. .
In summary, the Traxcess 14 Guidewire described in this submission is, in our opinion, substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the United States. The logo consists of a stylized eagle with three curved lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-9002
July 31, 2014
MicroVention, Inc. Ms. Naomi Gong Sr. Regulatory Affairs Project Manager 1311 Valencia Ave. Tustin, CA 92780
Rc: K133725
> Trade/Device Name: Traxcess 14 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Neurovasculature Catheter Guide Wire Regulatory Class: Class II Product Code: MOF, DQX Dated: June 27, 2014 Received: June 30, 2014
Dear Ms. Gong.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{5}------------------------------------------------
Page 2 - Ms. Naomi Gong
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K133725
Device Name Traxcess 14 Guidewire
Indications for Use (Describe)
The Traxcess 14 Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic catheters. This device is not intended for use in coronary arteries.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Carlos L. Pena -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.